(36 days)
Fracture reduction and external fixation for treatment of distal radius fractures.
The WristJack ORIF System is an external fixation system used for reduction and fixation of distal radius fractures. The system includes an adjustable reduction/fixation frame (fixator), application instrumentation and skeletal fixation pins.
The fixator element has multiple adjustments to aid in fracture reduction and stabilization of distal radius fractures. The device and instrumentation are constructed of polyetherimide resin, carbon fiber, stainless steel, titanium and aluminum alloy. The fixation pins are constructed of implant grade 316 stainless steel per ASTM F138.
The provided text is a 510(k) summary for the WristJack ORIF System, an external fixation system for distal radius fractures. This document focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed study proving the device meets specific acceptance criteria in the way a clinical trial for an AI/CADe device would.
Therefore, the information requested about acceptance criteria, study details, sample sizes, expert involvement, and ground truth establishment, which are typical for studies validating AI or diagnostic devices, is largely not present in this type of regulatory submission.
Here's an attempt to answer based on the provided text, highlighting what is (and isn't) available:
1. Table of acceptance criteria and the reported device performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Mechanical/Functional Equivalence to Predicate Device | The device is deemed "substantially equivalent" to the predicate device (Agee WristJack Fracture Reduction System, K984442) with respect to application technique and adjustment function. The fixator element has multiple adjustments to aid in fracture reduction and stabilization. |
| Material Equivalence (where applicable) | The fixator is comprised of a beam element with two carbon fiber rods (compared to an injection molded Ultem component in the predicate). The fixation pins and instrumentation are "essentially identical" to the predicate. Materials include polyetherimide resin, carbon fiber, stainless steel, titanium, and aluminum alloy. Fixation pins are implant grade 316 stainless steel per ASTM F138. |
| Intended Use Equivalence | Intended use for "Fracture reduction and external fixation for treatment of distal radius fractures" is consistent with the predicate. |
| Safety and Effectiveness (implied via substantial equivalence) | The FDA's substantial equivalence determination implies that the device is as safe and effective as the predicate device for its intended use, based on the information provided in the 510(k) submission. |
2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Not applicable / Not provided. This document is a premarket notification (510(k)) for a medical device (external fixator), not an AI/CADe device. The assessment of substantial equivalence relies on comparing the new device's design, materials, and intended use to a legally marketed predicate device, rather than a clinical study with a "test set" of patient data in the context of an AI algorithm. There is no mention of patient data, clinical trial, or geographical origin of data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable / Not provided. The concept of "ground truth" established by experts for a test set is relevant for AI or diagnostic performance studies. For this external fixator, the "ground truth" for its safety and effectiveness is largely established by its similarity to a previously cleared predicate device, and the engineering and material specifications provided in the submission. The FDA reviewers (Celia M. Witten, Ph.D., M.D., Director, Division of General, Restorative and Neurological Devices, Office of Device Evaluation) are the experts who reviewed the submission.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable / Not provided. This concept is for resolving discrepancies in expert interpretations within an AI/diagnostic study.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. An MRMC study is specific to evaluating diagnostic performance, often with AI assistance. This device is an external fixator, not an AI/CADe system.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- No. This question is also specific to AI algorithms.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- The "ground truth" for this 510(k) submission is the demonstration of substantial equivalence to a predicate device. This involves comparing technological characteristics, materials, and intended use to established standards of safety and effectiveness, rather than a clinical "ground truth" from patient data. The basis for this is engineering comparison, material specifications, and regulatory precedent.
8. The sample size for the training set
- Not applicable / Not provided. This concept applies to AI algorithm development.
9. How the ground truth for the training set was established
- Not applicable / Not provided. This concept applies to AI algorithm development.
In summary, the provided document is a regulatory submission for a physical medical device (an external fixator). It demonstrates compliance through substantial equivalence to a predicate device, rather than through a clinical study designed to meet specific performance metrics with patient data, which would be typical for an AI algorithm or diagnostic device.
{0}------------------------------------------------
OV - 5 2003
Image /page/0/Picture/1 description: The image shows the logo for the Hand Biomechanics Lab. On the left side of the logo is a black silhouette of a hand with a curved line above it. To the right of the hand is the text "Hand Biomechanics Lab" stacked on top of each other. The text is in a simple, sans-serif font.
510(k) Summary
033112
page 1 of 1
Hand Biomechanics Lab. Inc. 77 Scripps Drive, Suite 104 Sacramento, CA 95825-6209 Contact: Timothy R. Stallings (916) 923-5073 Phone: Fax: (916) 920-2215 Email: tstallings@handbiolab.com
Prepared September 29, 2003
Name of Device:
| Regulatory Classification: | Class II |
|---|---|
| Classification Name: | Component, Traction, Invasive [888.3040] |
| Common Name: | External Fixator System |
| Trade Name/Proprietary Name: | WristJack ORIF System, Item No. CFD-347WristJack ORIF System, non-sterile, Item No. CFD-347-NS |
| Performance Standards: | No performance standards exist for this device. |
Predicate Device:
Agee WristJack Fracture Reduction System (sterile), Item No. CFD-147, K984442
Description of Device:
The WristJack ORIF System is an external fixation system used for reduction and fixation of distal radius fractures. The system includes an adjustable reduction/fixation frame (fixator), application instrumentation and skeletal fixation pins.
The fixator element has multiple adjustments to aid in fracture reduction and stabilization of distal radius fractures. The device and instrumentation are constructed of polyetherimide resin, carbon fiber, stainless steel, titanium and aluminum alloy. The fixation pins are constructed of implant grade 316 stainless steel per ASTM F138.
Intended Use:
Fracture reduction and external fixation for treatment of distal radius fractures.
Technological Characteristics Compared to Predicate Device:
The WristJack ORIF System is comparable to the predicate device with respect to application technique and adjustment function. The WristJack ORIF fixator has a beam element comprised of two carbon fiber rods compared to an injection molded Ultem component in the predicate device. The fixation pins and instrumentation are essentially identical to those included in the predicate device. The subject device is delivered to the customer in either sterile form. In the non-sterile model, the customer is responsible for sterilization before use. The predicate device is supplied sterile.
{1}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/1/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV - 5 2003
Mr. Timothy R. Stallings Manufacturing Manager Hand Biomechanics Lab, Inc. 77 Scripps Drive, Suite 104 Sacramento, California 95825
Re: K033112
Trade/Device Name: WristJack ORIF System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: II Product Code: KTT Dated: September 29, 2003 Received: September 30, 2003
Dear Mr. Stallings:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Page 2 - Mr. Timothy R. Stallings
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Sincerely yours,
Mark N Millman
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
| 510(k) Number (if known): | K033112 |
|---|---|
| Device Name: | WristJack ORIF System |
Indications For Use:
Fracture reduction and external fixation for treatment of distal radius fractures.
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mark N Mullener
(Division Sign-Off) Division of General, Restorative and Neurological Devices
510(k) Number K033112
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.
(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.