K Number
K033094
Date Cleared
2003-12-19

(81 days)

Product Code
Regulation Number
862.3100
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Ecstasy Enzyme Immunoassay is a homogeneous enzyme immunoassay with a 500 ng/mL cutoff. The assay is intended for use in the qualitative and semi-quantitative analyses of Ecstasy (MDMA) in human urine. The assay is designed for professional use with a number of automated clinical chemistry analyzers.

The Ecstasy Enzyme Immunoassay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgement should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used.

The Ecstasy Drug of Abuse Calibrators are intended for in vitro diagnostic use for the calibration of the Ecstasy enzyme immunoassay to detect ecstasy in human urine.

The Ecstasy Drug of Abuse Controls are intended for in vitro diagnostic use for the validation of the Ecstasy enzyme immunoassay to detect ecstasy in human urine.

Device Description

LZI's Ecstasy Enzyme Immunoassay is a ready-to-use, liquid reagent, homogeneous enzyme immunoassay. The assay uses specific antibody that can detect Ecstasy (MDMA) in human urine with minimal cross-reactivity to various, common prescription drugs and abused drugs. The assay is based on competition between ecstasy labeled with glucose-6-phosphate dehydrogenase (G6PDH) enzyme and free drug from the urine sample for a fixed amount of specific antibody. In the absence of free drug from the urine sample the specific antibody binds to the drug labeled with G6PDH enzyme causing a decrease in enzyme activity. The G6PDH enzyme activity is determined spectrophotometrically at 340 nm by measuring its ability to covert nicotinamide adenine dinucleotide (NAD) to NADH.

All of the Single Analyte Urine DAU Calibrators and Controls are human urine-based liquid, and ready to use. These Calibrators and Controls do not have any especially unique technical characteristics. Each contains a known concentration of a specific drug analyte. The Negative DAU calibrator is a processed, drug-free human urine matrix, which has also been used with all assays. The calibrators and controls are prepared by spiking known concentrations of drug analyte into the Negative DAU Calibrator matrix.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Lin-Zhi International, Inc. Ecstasy Enzyme Immunoassay, based on the provided text:

Acceptance Criteria and Device Performance

The acceptance criteria are implicitly defined by comparing the performance of the LZI Ecstasy EIA to the performance of the predicate device, DRI Ecstasy Assay, and demonstrating "acceptable results." Specific quantitative acceptance criteria are not explicitly stated in the document as numerical thresholds, but rather through the reported performance relative to the predicate or confirmed methods.

FeatureImplicit Acceptance Criteria / Predicate PerformanceLZI's Ecstasy EIA Reported Performance
Within Run Precision (Qualitative)Comparable Mean Rate, SD, and %CV to DRI's Ecstasy EIA for 375, 500, and 625 ng/mLNegative: Mean Rate 254.4, SD 2.28, %CV 0.90 375 ng/mL: Mean Rate 327.9, SD 2.50, %CV 0.76 500 ng/mL: Mean Rate 384.2, SD 3.13, %CV 0.81 625 ng/mL: Mean Rate 420.8, SD 3.54, %CV 0.84 1000 ng/mL: Mean Rate 453.5, SD 3.27, %CV 0.72
Within Run Precision (Semi-quantitative)Comparable Mean Conc., SD, and %CV to DRI's Ecstasy EIA for 375, 500, and 625 ng/mL375 ng/mL: Mean Conc. 381.4, SD 5.24, %CV 1.37 500 ng/mL: Mean Conc. 517.4, SD 8.05, %CV 1.56 625 ng/mL: Mean Conc. 649.0, SD 11.38, %CV 1.75
Run-To-Run Precision (Qualitative)Comparable Mean Rate, SD, and %CV to DRI's Ecstasy EIA for 375, 500, and 625 ng/mLNegative: Mean Rate 252.4, SD 2.37, %CV 0.94 375 ng/mL: Mean Rate 326.9, SD 2.55, %CV 0.78 500 ng/mL: Mean Rate 383.2, SD 1.20, %CV 0.31 625 ng/mL: Mean Rate 420.3, SD 3.31, %CV 0.79 1000 ng/mL: Mean Rate 453.6, SD 3.39, %CV 0.75
Run-To-Run Precision (Semi-quantitative)Comparable Mean Conc., SD, and %CV to DRI's Ecstasy EIA for 375, 500, and 625 ng/mL375 ng/mL: Mean Conc. 367.6, SD 8.73, %CV 2.37 500 ng/mL: Mean Conc. 502.6, SD 9.46, %CV 1.88 625 ng/mL: Mean Conc. 637.1, SD 8.21, %CV 1.29
SensitivityDRI's Ecstasy EIA: 22 ng/mL50 ng/mL (This is less sensitive than the predicate, but the device is cleared for a 500 ng/mL cutoff, making this a secondary metric).
Accuracy (Positive Samples)DRI's Ecstasy EIA: 100% agreement (GC/MS confirmed)97% agreement (Vs. GC/MS, n=127)
Accuracy (Negative Samples)DRI's Ecstasy EIA: 100% agreement100% agreement (Vs. GC/MS, n=127)
Analytical Recovery (Qualitative)DRI's Ecstasy EIA: No data available100% accuracy on positive vs. negative tests
Analytical Recovery (Semi-quantitative)DRI's Ecstasy EIA: No data availableQuantitates within ±10% of the nominal concentration between 50 ng/mL and 920 ng/mL.
SpecificityComparable to the predicate device (see attached DRI's Ecstasy EIA package insert)Comparable to the predicate device. (Specific data not provided in this summary, but implies the LZI device meets the cross-reactivity profiles demonstrated by the predicate.)

Study Details

  1. Sample Size used for the test set and the data provenance:

    • Test Set Sample Size: 127 samples for the Accuracy study (comparison against GC/MS).
    • Data Provenance: Not explicitly stated but clinical performance data typically comes from laboratory testing of human urine samples. The document does not specify country of origin or whether it's retrospective or prospective. Given the nature of a 510(k) for an in vitro diagnostic, these are typically laboratory-based studies.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Number of experts: Not applicable for this type of device.
    • Qualifications of experts: The ground truth was established by Gas Chromatography/Mass Spectrometry (GC/MS), which is an analytical chemical method, not human expert consensus.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Adjudication Method: Not applicable. The ground truth was established by GC/MS, which is considered a definitive analytical method, not requiring human adjudication for interpretation in this context.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • MRMC Study: No. This device is an enzyme immunoassay for drug detection in urine, not an imaging or diagnostic device requiring human reader interpretation in the same way as AI in radiology, for example. The output is a quantitative or qualitative chemical result.
  5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Standalone Performance: Yes, the performance described (precision, sensitivity, accuracy, analytical recovery, specificity) is "standalone" performance of the assay itself, without human intervention in the result determination beyond instrument operation and result recording. The "human-in-the-loop" aspect for such a device is typically clinical interpretation of the analytically confirmed result, not the initial assay performance.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • Ground Truth Type: GC/MS (Gas Chromatography/Mass Spectrometry). This is a highly specific and sensitive analytical chemical method considered the gold standard for drug confirmation.
  7. The sample size for the training set:

    • Training Set Sample Size: Not explicitly mentioned. For an immunoassay, "training" typically refers to the development and optimization of the assay reagents and parameters, which would involve numerous experiments and titrations, rather than a distinct "training set" of patient samples in the way an AI algorithm uses for machine learning. The stability, precision, and accuracy studies would be considered part of the validation/testing after the assay is developed.
  8. How the ground truth for the training set was established:

    • Ground Truth for Training Set: Not applicable in the context of a machine learning "training set". During the development of the immunoassay, known concentrations of MDMA in urine (spiked samples) would be used to develop the assay's dose-response curve and optimize its components, with the "ground truth" being the known, spiked concentrations verified by reference methods if necessary.

{0}------------------------------------------------

DEC 1 9 2003

510(k) Summary of Safety and Effectiveness

This summarv of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Introduction

According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.

Submitter name, Address, and Contact

Lin-Zhi International, Inc. 687 North Pastoria Avenue Sunnyvale, CA 94085 Phone: (408) 732-3856 Fax: (408) 732-3849

Cheng-I Lin, Ph.D. Contact: President, R&D Director

Device Name and Classification

Classification Name:Enzyme Immunoassay, Amphetamine,Class II, DKZ(91 Toxicology),21CFR 862.3100
Common Name:Homogeneous enzyme immunoassay for the determination ofEcstasy (MDMA) levels in urine.
Proprietary Name:None

Legally Marketed Predicate Device(s)

Lin-Zhi International, Inc.' Ecstasy Enzyme Immunoassay is substantially equivalent to the DRI Ecstasy Assay (By Microgenics Corp.), cleared under premarket notification K012110.

LZI's Ecstasy Enzyme Immunoassay is identical or similar to its predicate in terms of intended use, method principle, device components, and clinical performance.

Device Description

LZI's Ecstasy Enzyme Immunoassay is a ready-to-use, liquid reagent, homogeneous enzyme immunoassay. The assay uses specific antibody that can detect Ecstasy (MDMA) in human urine with minimal cross-reactivity to various, common prescription drugs and abused drugs.

{1}------------------------------------------------

The assay is based on competition between ecstasy labeled with glucose-6-phosphate dehydrogenase (G6PDH) enzyme and free drug from the urine sample for a fixed amount of specific antibody. In the absence of free drug from the urine sample the specific antibody binds to the drug labeled with G6PDH enzyme causing a decrease in enzyme activity. The G6PDH enzyme activity is determined spectrophotometrically at 340 nm by measuring its ability to covert nicotinamide adenine dinucleotide (NAD) to NADH.

Intended Use

The Ecstasy Enzyme Immunoassay is a homogeneous enzyme immunoassay with a 500 ng/mL cutoff. The assay is intended for use in the qualitative and semi-quantitative analyses of Ecstasy (MDMA) in human urine.

Comparison to Predicate Device

LZI's Ecstasy Enzyme Immunoassay is substantially equivalent to other products in commercially distribution intended for similar use. Most notably it is substantially equivalent to the currently, commercially marketed DRI Ecstasy Assay (K012110) by Microgenics Corporation.

The following table compares LZI's Ecstasy Enzyme Immunoassay with the predicate device, DRI Ecstasy Enzyme Immunoassay by Microgenics Corp.

Similarities:

  • Both assays are for qualitative and semi-quantitative determination of Ecstasy in . human urine.
  • Both assays use 500 ng/mL cutoff, and 5 points calibration for semi-quantitative . determination.
  • Both assays use the same method principle, and device components. .

Differences:

  • Microgenics assay uses 5 points calibration (0, 250, 500, 750, 1000 ng/ml) for . semi-quantitative determination. LZI assay uses 5 calibrator set (0, 100, 500, 750, and 1000 ng/ml) for semi-quantitative determination.

{2}------------------------------------------------

(Comparison to Predicate Device, continued)

Performance Characteristics

FeatureDRI's Ecstasy EIALZI's Ecstasy EIA
Within Run Precision:
QualitativeMean RateSD% CVMean RateSD% CV
375 ng/mL2542.30.9Negative254.42.280.90
500 ng/mL3323.41.0375 ng/mL327.92.500.76
625 ng/mL3953.81.0500 ng/mL384.23.130.81
625 ng/mL420.83.540.84
1000 ng/mL453.53.270.72
Semi-quantitative:375 ng/mL3595.71.6Mean Conc.SD% CV
500 ng/mL5006.91.4375 ng/mL381.45.241.37
625 ng/mL6309.51.5500 ng/mL517.48.051.56
625 ng/mL649.011.381.75
Run-To-Run Precision:
Qualitative:Mean RateSD% CVMean RateSD% CV
375 ng/mL2547.02.8Negative252.42.370.94
500 ng/mL3328.92.7375 ng/mL326.92.550.78
625 ng/mL3957.21.8500 ng/mL383.21.200.31
625 ng/mL420.33.310.79
1000 ng/mL453.63.390.75
Semi-quantitative:Mean Conc.SD% CVMean Conc.SD% CV
375 ng/mL3599.12.5375 ng/mL367.68.732.37
500 ng/mL50010.72.1500 ng/mL502.69.461.88
625 ng/mL6302.12.2625 ng/mL637.18.211.29
Sensitivity:22 ng/mL50 ng/mL
Accuracy:Vs. GC/MS (n=127)
Positive Samples100 % agreement (GC/MS confirmed)97% agreement
Negative Samples100 % agreement100 % agreement
Analytical Recovery:
Qualitative: No data available100 % accuracy on positive vs. negative tests
Semi-quantitative: No data availableQuantitates within ±10% of the nominal
concentration between 50 ng/mL and 920
ng/mL.
Specificity:See attached DRI's Ecstasy EIA packageComparable to the predicate device.
insert

{3}------------------------------------------------

Conclusion

LZI's Ecstasy Enzyme Immunoassay was evaluated for several performance characteristics including precision, sensitivity, accuracy, analytical recovery, and specificity. All the studies showed acceptable results when compared to the predicate device.

We trust the information provided in this Premarket Notification [510(k)] submission will support a determination of substantial equivalence of the LZI's Ecstasy Enzyme Immunoassay to other Ecstasy test systems currently marketed in the United States.

{4}------------------------------------------------

510(k) Summary of Safety and Effectiveness

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Introduction

According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.

Submitter name, Address, and Contact

Lin-Zhi International, Inc. 687 North Pastoria Avenue Sunnyvale, CA 94085 Phone: (408) 732-3856 Fax: --(408) 732-3849

Contact:Cheng-I Lin, Ph.D.
President

Device Name and Classification

(a) Classification Name:Calibrators, Drug Specific;Class II, DLJ (91 Toxicology), 21 CFR 862.3200
Common/Usual Name:Ecstasy Calibrators
Proprietary Name:None
(b) Classification Name:Single (Specified) Analyte Controls (Assayed and Unassayed);Class I, LAS (91 Toxicology), 21 CFR 862.3280
Common/Usual Name:Ecstasy Controls
Proprietary Name:None

Legally Marketed Predicate Device(s)

Lin-Zhi International, Inc.'s Ecstasy Drug of Abuse Calibrators and Controls are substantially equivalent to the DRI Ecstasy Urine Calibrators and controls, by Microgenics Corporation, cleared under premarket notifications (K012109).

{5}------------------------------------------------

Device Description

All of the Single Analyte Urine DAU Calibrators and Controls are human urine-based liquid, and ready to use. These Calibrators and Controls do not have any especially unique technical characteristics. Each contains a known concentration of a specific drug analyte.

The Negative DAU calibrator is a processed, drug-free human urine matrix, which has also been used with all assays. The calibrators and controls are prepared by spiking known concentrations of drug analyte into the Negative DAU Calibrator matrix. The concentrations of drug analyte in the calibrators and controls are summarized as follows:

Reference MaterialEcstasy EIA
Negative Calibrator0 ng/mL
Low Calibrator100 ng/mL
Cutoff Calibrator500 ng/mL
Intermediate Calibrator750 ng/mL
High Calibrator1000 ng/mL
Control Level I375 ng/mL
Control Level II625 ng/mL

Intended Use

The Ecstasy DAU Calibrators are intended for in vitro diagnostic use for the calibration of the Ecstasy enzyme immunoassay to detect ecstasy in human urine.

The Ecstasy DAU Controls are intended for in vitro diagnostic use for the validation of the Ecstasy enzyme immunoassay to detect ecstasy in human urine.

Comparison to Predicate Device

LZI's Ecstasy DAU Calibrators and Controls are similar in intended use, matrix, and performance to the DRI's ecstasy calibrators and controls included in the DRI Ecstasy Calibrators and Controls.

Similarities:

  • Both are for the calibration and validation of Ecstasy Enzyme Immunoassay to detect . drug of abuse in human urine.
  • A total of 5 levels of calibrators including the negative calibrator for each analyte. .
  • The nominal concentrations of the analyte in the calibrators and controls are determined . and confirmed by GC/MS.
  • Both are urine-based liquids. ●
  • Storage condition is the same, at 2℃ to 8℃. .
  • Performance characteristics on precision, accuracy and stability are similar. .

{6}------------------------------------------------

Differences:

  • For semi-quantitative assay, DRI uses 0, 250, 500, 750, and 1000 ng/mL as . calibrators. LZI uses 0, 100, 500, 750, and 1000 ng/mL as calibrators.

Conclusion

The information provided in the premarket notification demonstrates that the LZI's Ecstasy Urine Drugs of Abuse Calibrators and Controls are substantially equivalent to previously approved predicate devices, notably the DRI Ecstasy Urine Calibrators and Controls, and safe and effective for its intended use.

{7}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three wavy lines representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

DEC 1 9 2003

Cheng-I Lin, Ph.D. President, R & D Director Lin-Zhi International, Inc. 687 North Pastoria Avenue Sunnyvale, CA 94085

K033094 Trade/Device Name: Ecstasy Enzyme Immunoassay Regulation Number: 21 CFR 862.3610 Regulation Name: Methamphetamine test system Regulatory Class: Class II Product Code: DJC; DLJ; LAS Dated: September 25, 2003 Received: September 29, 2003

Dear Dr. Lin:

Re:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{8}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device. or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97), You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Steven Putman

Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{9}------------------------------------------------

Premarket Notification Supplement

Indications for Use Statement

Ko37094

510(k) Number (if known): _

Device Name: Ecstasy Drug of Abuse Calibrators and Controls

Indications for Use:

The Ecstasy Drug of Abuse Calibrators are intended for in vitro diagnostic use for the The Lestasy Drug of Nouse Canolation as a to detect ecstasy drugs in human urine.

The Ecstasy Drug of Abuse Controls are intended for in vitro diagnostic use for the The Ecstasy Drug of Abase Controls none of to detect ecstasy drugs in human urine.

Concurrence of CDRH, Office of Device Evaluation (ODE) Over-The-Counter Use OR Prescription Use (Per 21 CFR 801.109)

(Optional Format 1-2-96)

Alberts
Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K033094

Page 2

{10}------------------------------------------------

Premarket Notification

Indications for Use Statement

510(k) Number (if known): K03 3094

Device Name: Ecstasy Enzyme Immunoassay

Indications for Use:

The Ecstasy Enzyme Immunoassay is a homogeneous enzyme immunoassay with a 500 ng/mL cutoff. The assay is intended for use in the qualitative and semi-quantitative analyses of Ecstasy (MDMA) in human urine. The assay is designed for professional use with a number of automated clinical chemistry analyzers.

The Ecstasy Enzyme Immunoassay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical consideration and professional judgement should be applied to any drug-ofabuse test result, particularly when preliminary positive results are used.

Albato Curtis

Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K033094

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109) OR

Over-The-Counter Use

(Optional Format 1-2-96)

§ 862.3100 Amphetamine test system.

(a)
Identification. An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.(b)
Classification. Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).