K Number
K032947
Date Cleared
2003-10-17

(25 days)

Product Code
Regulation Number
870.2700
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Dolphin 2000 Oximetry Sensors are indicated for use in continuous monitoring of arterial oxygen saturation and pulse rate.

Device Description

The Dolphin 2000 Y Oximetry Sensor is a fully compatible re-usable replacement sensor for use with Nellcor pulse oximeter monitors. The Re-usable Y sensor is for use on the ear, finger, hand, or neonatal foot and held in place with a disposable bandage. The sensor can also be used on the adult ear with the ear clip accessory. The emitter and detector are mounted in a sealed pouch (same material as in the re-usable clip sensor above) constructed in a Y shape. The sensor is provided non-sterile.

AI/ML Overview

Here's a summary of the acceptance criteria and study details for the Dolphin 2000™ Pulse Oximetry Y Sensor, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

ParameterAcceptance Criteria (Implied)Reported Device Performance (Accuracy)
SpO2 Accuracy (Adults, Pediatrics, Infants)Less than 2.0% ARMs (Absolute Root Mean Square difference)< 2.0% ARMs in the range of 70-100% SaO2
SpO2 Accuracy (Neonates)Less than 3% ARMs< 3% ARMs in the range of 70-100% SaO2
SpO2 Accuracy (Ear Clip Applications)Less than 3.5% ARMs< 3.5% ARMs
BiocompatibilityNon-toxic, non-irritant, non-sensitizingTest results demonstrated materials to be non-toxic, non-irritant, and non-sensitizing.
Electrical SafetyCompliance with specific standards (EN 60601-1, EN 60601-1-1, EN 60601-1-2, ASTM F1415-92)Tested and found to comply with applicable clauses of the listed standards.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not explicitly stated as a number of volunteers. The document mentions "Volunteers participated in the breathe-down protocol."
  • Data Provenance: Prospective clinical study conducted in the USA. (VA Hospital of Wisconsin - Milwaukee).

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • Number of Experts: Not explicitly stated. The study was conducted at the "VA Hospital of Wisconsin - Milwaukee, (Dr. Phillip Clifford, MD.)." Dr. Clifford, as an MD, would be considered an expert. It's unclear if other experts were involved in establishing the ground truth.
  • Qualifications of Experts: Dr. Phillip Clifford, MD.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not explicitly stated. The document refers to "statistically comparing Dolphin 2000 SpO2 values to functional SaO2 values." This suggests a direct comparison method rather than an adjudication process involving multiple human interpretations of the ground truth.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • MRMC Study Done? No, a MRMC comparative effectiveness study was not done. The study focuses on the device's accuracy against a physiological reference, not on human reader performance with or without AI assistance.

6. Standalone (Algorithm Only) Performance Study

  • Standalone Study Done? Yes, the clinical validation described focused on the direct performance of the Dolphin 2000 sensors (the "algorithm/device only") compared to functional SaO2 values. There's no mention of a human-in-the-loop component in this specific accuracy testing.

7. Type of Ground Truth Used

  • Type of Ground Truth: Functional SaO2 values. This is a direct physiological measurement, likely obtained through co-oximetry, considered the gold standard for arterial oxygen saturation.

8. Sample Size for the Training Set

  • Sample Size for Training Set: Not applicable/not stated. The document describes a medical device, not an AI/Machine Learning algorithm that would typically have a separate training set. The device's underlying principles are based on established oximetry technology, not on machine learning trained on large datasets.

9. How the Ground Truth for the Training Set was Established

  • How Ground Truth for Training Set was Established: Not applicable. As mentioned above, this is a hardware medical device using established photoplethysmography principles, not an AI system requiring a 'training set' with associated ground truth in the typical machine learning sense. The device's calibration and design would be based on fundamental scientific principles and existing validated technologies.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows a series of handwritten alphanumeric characters, seemingly a code or identifier. The characters are 'K032947' and are written in a simple, slightly tilted manner. The handwriting appears somewhat casual, with variations in stroke thickness and character alignment.

OCT 1 7 2003

14. 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

Dolphin 2000™ Pulse Oximetry Y Sensor 9/09/03

Submitter ( Consultant name and Address)

Bill Curnan 9433 S. Morning Glory Lane Highlands Ranch, CO 80130

Phone: 720-939-6482 Fax: 786-551-8221

Sponsor Company Name and Address and Contact Person

Dolphin Medical Inc. 12525 Chadron Avenue Hawthorne, CA 90250

Tammy Conway, QA Manager (310) 978-0516 phone:

(310) 978-1816 fax:

Manufacturing Facility Name and Address

Opto Sensors (M) Sdn. Bhd. No. 6 Jalan Angkasa Mas 1 Tabrau Industrial Estate II 81100 Johor Bahru, Malaysia

Common, Classification & Proprietary Names

Common Name: oximetry sensor Classification Name: oximeter Dolphin™ 2000 Oximetry Sensors Proprietary Name:

Predicate Devices

SensorDolphinModelDolphin 2000Predicate Modelfound in K030952 #Nellcor Predicatefound inK991823
Dolphin 2000 NellcorCompatable ReusableOximetry Y Sensor22102010D-YSD-YSE

{1}------------------------------------------------

Device Description

The Dolphin 2000 Y Oximetry Sensor is a fully compatible re-usable replacement sensor for use with Nellcor pulse oximeter monitors.

The Re-usable Y sensor is for use on the ear, finger, hand, or neonatal foot and held in place with a disposable bandage. The sensor can also be used on the adult ear with the ear clip accessory. The emitter and detector are mounted in a sealed pouch (same material as in the re-usable clip sensor above) constructed in a Y shape. The sensor is provided non-sterile.

Intended Use

The Dolphin 2000 Oximetry Sensors are indicated for use in continuous monitoring of arterial oxygen saturation and pulse rate.

Technological Characteristics Comparison

The Dolphin 2000 Oximetry Sensors are substantially equivalent in intended use, design, principles of operation, materials, and performance to commercially available oximetry sensors.

All of the Dolphin 2000 oximetry sensors and the predicate devices operate on the identical principles of non-invasive optical assessment of tissue oxygenation using emitters (LEDs) and detectors (photodiode).

The Dolphin 2000 oximetry sensors are designed, and manufactured for full compatibility for use with the labeled, commercially-available oximetry monitors. They are constructed of similar materials and components of equivalent specifications as used in the predicate devices.

The Dolphin 2000 oximetry sensors, like the predicate devices are available in both disposable and re-usable styles, labeled for use in adult, pediatric, infant and neonatal populations.

The labeled accuracy of the Dolphin 2000 sensors is equivalent to those of the predicate devices.

Performance Testing

프 Biocompatibility

Biocompatibility tests, appropriate for skin-contacting devices for prolonged exposure, were performed on the each of the device components used in the assembly of the Dolphin 2000™ pulse oximetry sensors. Test results demonstrated the materials to be non-toxic, non-imitant, and non-sensitizing.

{2}------------------------------------------------

t Electrical Safety

The Dolphin 2000 Oximetry Sensors have been tested and found to comply with the applicable clauses of the following standards:

  • Medical electrical equipment part 1: General EN 60601-1 (1990) . requirements for safety
  • EN 60601-1-1 (1993) Medical electrical equipment part 1: General requirements for safety - 1. Collateral standard: Safety requirements for medical electrical systems
  • EN 60601-1-2 (1993) Medical electrical equipment part 1: General requirements for safety - 2. Collateral standard: Electromagnetic compatibility requirements and tests
  • ASTM F1415-92 Standard Specification for Pulse Oximeters -

Clinical Testing

The sensors were validated in breathe-down protocols at the VA Hospital of Wisconsin - Milwaukee, (Dr. Phillip Clifford, MD.). Scientific accuracy was demonstrated by statistically comparing Dolphin 2000 SpO2 values to functional SaO2 values. Volunteers participated in the breathe-down protocol at rest (i.e. no motion) while fully conscious at SaO2 values ranging from 70-100%. Data was analyzed to determine the ARMS for each probe. Clinical Validation for the Dolphin 2000 Reusable, Adult disposable, and Neonatal disposable probes resulted in an accuracy determination of less then 2.0% ARMs in the range of 70-100% Sa02 for adults, pediatrics, and infants, less than 3% Arms in the range of 70-100 for Neonates, and less than 3.5 % for ear clip applications..

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with three curved lines above them that resemble wings or feathers.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 17 2003

Mr. Bill Curnan Regulatory Specialist Dolphin Medical Incorporated 9433 S. Morning Glory Lane Littleton, Colorado 80130

Re: K032947

Trade/Device Name: Nellcor Compatible Dolphin 2000 Y Oximetry Sensor (Model 2210) Regulation Number: 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA

Dated: September 19, 2003 Received: September 22, 2003

Dear Mr. Curnan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Curnan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Patricia Cucinotta

Chiu Lin. Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Statement of Indications for Use (FDA Form)

510(k): ҚОЗΖ947

Device:

DOLPHIN 2000 Oximetry Sensors

Indications for Use:

The Dolphin 2000 Oximetry Sensors are indicated for use in continuous monitoring of arterial oxygen saturation and pulse rate.

(Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Device

510(k) Number:

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).