K Number
K032309

Validate with FDA (Live)

Manufacturer
Date Cleared
2003-12-04

(132 days)

Product Code
Regulation Number
886.5928
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The NoveLens® Contact Lens Case is indicated for storage of hard, rigid gas permeable (silicone acrylate and fluro silicone acrylate) and soft (hydrophilic) contact lenses during chemical disinfection.

Device Description

The NoveLens® Contact Lens Case is a contact Iens storage case comprised of a plastic case that houses two chambers. The upper chamber which holds the lenses separately in individually attached concave reservoirs and a lower chamber that provides space for rinsed solution as it passes through holes from the upper chamber and through a drainage toube into the lower portion of the device. The NoveLens® Contact Lens Case is constructed to take solution introduced into the upper chamber in a rinse fashion and force, by gravity, solution through the drainage holes through a drainage tube and into a lower collecting chamber. Additional solution is added until both chambers are completely filled to the top. Screw caps when closed hermetically seal the system to hold disinfection solution while the lenses are stored and disinfected in accordance with directions consistent with labeled multipurpose contact lens solutions. The design provides for one-way ingress of solution and expulsion of spent solution through external drainage holes from the lower chamber upon tipping the unit on its side.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the NoveLens® Contact Lens Case, extracted from the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Biocompatibility: CytotoxicityPassed
Biocompatibility: Ocular IrritationPassed
Biocompatibility: Systemic ToxicityPassed
Effectiveness (for storage during chemical disinfection)Equivalence to predicate devices established without further testing.

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: Not explicitly stated for the toxicology tests.
  • Data Provenance: Not specified, but implied to be from laboratory testing conducted on the plastic components of the NoveLens® Contact Lens Case. It is retrospective in the sense that the materials were tested prior to submission for clearance.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • Not applicable in the context of toxicology testing; the "ground truth" is determined by established scientific methods and regulatory guidance for chemical safety. Expertise would be in toxicology and materials science, but specific numbers and qualifications are not provided for reviewers of the toxicology test results beyond FDA.

4. Adjudication Method for the Test Set

  • Not applicable. Toxicology testing results are typically evaluated against pre-defined thresholds and regulatory guidelines rather than through expert adjudication of individual cases.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

  • No, an MRMC study was not done. This device is a contact lens case, not an imaging or diagnostic AI device that would typically involve human readers.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This is a physical medical device (contact lens case), not an algorithm or AI system.

7. The Type of Ground Truth Used

  • For the toxicity testing, the "ground truth" was established by standardized laboratory testing procedures and accepted regulatory guidance (FDA Guidance for contact lens case products; Guidance for Industry-Premarket Notification 510(k) Guidance Document for Contact Lens Care Products- dated May 1, 1997, and ISO-10993-1).
  • For effectiveness, the ground truth was demonstrated by substantial equivalence to predicate devices without additional effectiveness testing.

8. The Sample Size for the Training Set

  • Not applicable. This is a physical medical device, not an AI or machine learning model that requires a training set.

9. How the Ground Truth for the Training Set was Established

  • Not applicable as there is no training set for this type of device.

{0}------------------------------------------------

K032309 510(k) Summary Statement

SUBMITTER:

Submitted on behalf of:

Company Name:NoveLens®, Inc.
Address:17-25 Hunter Place
Fair Lawn, New Jersey 07410
Phone:(201) 794-7999
Fax:(201) 794-7958
CONTACT PERSON:Richard E. Lippman, O.D., F.A.A.O.
Official Representative and Correspondent
R.P. Chiacchierini & Associates, LLC
15825 Shady Grove Road
Suite 30
Rockville, Maryland 20850
Phone:(240) 683-3738
FAX:(240) 683-9236
DATE SUMMARY PREPARED:November 10, 2003
TRADE NAME:NoveLens® Contact Lens Case

COMMON NAME: CLASSIFICATION:

contact lens case Class II: LRX

ESTABLISHMENT REGISTRATION: PENDING

SUBSTANTIALLY EQUIVALENT TO:

The NoveLens® Contact Lens Case is substantially equivalent to the following devices:

K013232Bausch & Lomb CyberCase Contact Lens Case
K852384Bausch & Lomb CyberCase Contact Lens Case
K971618Alcon Contact Lens Case
K974281Alcon Contact Lens Case
K021248iCase: IKeeps, Inc,

DESCRIPTION of the DEVICE:

The NoveLens® Contact Lens Case is a contact Iens storage case comprised of a plastic case that houses two chambers. The upper chamber which holds the lenses separately in individually attached concave reservoirs and a lower chamber that provides space for rinsed solution as it passes through holes from the upper chamber and through a drainage toube into

. .

{1}------------------------------------------------

the lower portion of the device. The NoveLens® Contact Lens Case is constructed to take solution introduced into the upper chamber in a rinse fashion and force, by gravity, solution through the drainage holes through a drainage tube and into a lower collecting chamber. Additional solution is added until both chambers are completely filled to the top. Screw caps when closed hermetically seal the system to hold disinfection solution while the lenses are stored and disinfected in accordance with directions consistent with labeled multipurpose contact lens solutions. The design provides for one-way ingress of solution and expulsion of spent solution through external drainage holes from the lower chamber upon tipping the unit on its side.

INDICATIONS FOR USE:

The NoveLens® Contact Lens Case is indicated for storage of hard, rigid gas permeable (silicone acrylate and fluro silicone acrylate) and soft (hydrophilic) contact lenses during chemical disinfection.

ACTIONS:

The NoveLens® Contact Lens Case provides for lens storage during chemical disinfection as well as storage up to the labeled time listed on the user's multipurpose disinfection solution. Directions for use accompany the NoveLens® Contact Lens Case for continued safe use of the device.

TOXICITY TESTING

A battery of toxicology tests were conducted on the plastic components to verify the safety of the materials used in the device. The tests included Cytotoxicity, Ocular Irritation, and Systemic Toxicity in conformance with FDA Guidance for contact lens case products; Guidance for Industry-Premarket Notification 510(k) Guidance Document for Contact Lens Care Products- dated May 1, 1997, and ISO-10993-1.

Results of the testing demonstrate that the safety and effectiveness of the NoveLens® Contact Lens Case is equivalent to the predicate devices listed above.

EFFECTIVENESS TESTING for the NoveLens® Contact Lens Case .

No further testing for case effectiveness was warranted for demonstration of substantial equivalence.

CLINICAL INFORMATION

Clinical studies were deemed unnecessary as the solutions used in the NoveLens® Contact Lens Case are already cleared for marketing as cleaning, rinsing, disinfection, and storage solution for soft, RGP, and hard contact lenses.

{2}------------------------------------------------

LABELING

、 . . . . . .

.

The NoveLens® Contact Lens Case is provided to the user with an outer carton as well as User Instructions.

.

:

. 1

: :

..

1

: 199

1 11:44

: : 上一篇:

.

:

,在线上的

NoveLens® Inc. 17-25 Hunter Place Fair Lawn, New Jersey 07410 (201) 794-7999 FAX: (201) 794-7958

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of a stylized eagle with three lines representing its wings.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

$$\mathtt{2EC} \text{ - 4 } \mathtt{2003}$$

NoveLens®, Inc. C/O Richard Lippman, O.D., F.A.A.O. Vice President, Ophthalmic Regulatory Medical Products R.P. Chiacchierini & Associates, LLC 15825 Shady Grove Rd., Suite 30 Rockville, MD 20850

Re: K032309

Trade/Device Name: NoveLens® Contact Lens Case Regulation Number: 21 CFR 886.5928, 21 CFR 886.5918 Regulation Name: Soft (hydrophilic) contact lens care products, Rigid gas permeable contact lens care products Regulatory Class: Class II

Product Code: LRX Dated: November 10, 2003 Received: November 10, 2003

Dear Dr. Lippman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Richard Lippman, O.D., F.A.A.O.

This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4613. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Sincerely yours,

A. Ralph Rosenthal

A. Ralph Rosenthal, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indication Statement

510(k) Number (if known): K032309

Device Name: NoveLens® Contact Lens Case

Indications for Use:

The NoveLens® Contact Lens Case is indicated for storage of hard, rigid gas permeable (silicone acrylate and fluro silicone acrylate) and soft (hydrophilic) contact lenses during chemical disinfection.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Presription Use _

Over -The-Counter Use

(Optional Format 1.26)

(Oprtional Format 1-2-96)

Karen Wabont

ion Sion-Off Division of Ophthalmic Ear Nose and Throat Devi

510(k) Number K032309

11:11:11 : 11:11:11

N/A