(108 days)
Not Found
No
The description details the use of a microprocessor and software to analyze pressure oscillations using a standard oscillometric algorithm, which is a traditional method and does not indicate the use of AI or ML. There are no mentions of AI, ML, or related concepts like training/test sets.
No
The device is a blood pressure monitor intended for measuring blood pressure and heart rate, not for treating or rehabilitating a condition.
No
The device measures physiological parameters (blood pressure and heart rate) but does not provide a diagnosis or aid in diagnosing a disease or condition. It is a measurement device.
No
The device description explicitly mentions physical components like transducers, a pneumatic pump, a cuff, an LCD display, and a signal transmission cable, indicating it is a hardware device with embedded software.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- IVD Definition: In vitro diagnostics are tests performed on samples taken from the human body, such as blood, urine, or tissue, to detect diseases, conditions, or infections. They are used to examine these samples outside of the body.
- Device Function: The description clearly states that the M100 blood pressure monitor is a non-invasive electronic medical device that measures blood pressure and heart rate by using a cuff placed on the patient's wrist or arm. This is a measurement taken on the body, not a test performed on a sample taken from the body.
- Method: The device uses the oscillometric method, which involves measuring pressure and its oscillations during cuff deflation. This is a physical measurement, not a chemical or biological analysis of a sample.
Therefore, the M100 blood pressure monitor falls under the category of a non-invasive medical device for physiological measurement, not an in vitro diagnostic device.
N/A
Intended Use / Indications for Use
The M100 Blood Pressure Monitor is an electric device that provides a signal from which the systolic, diastolic, and rate of heart beats can be derived through the use of transducers placed on the surface of body. The signals of measurement can be stored in the memory of device, and then can be displayed on the screen of device and/or be transmitted out of device by using the standard signal transmission cable.
The M100 of blood pressure monitor is an electronic device that provides a signal from which the systolic, diastolic and rate of heart beats can be derived through the use of transducers placed on the surface of the body. The signals of measurement can be stored in the memory of device, and then can be displayed on the screen of device and/or be transmitted out of device by using the standard signal transmission cable.
Product codes
DXN
Device Description
The Meditek Blood Pressure Monitor, model M100, is the non-invasive electronic medical device intended for the measuring of blood pressure (including systolic pressure and diastolic pressure) and the number of heart beats of patient by using the oscillometric method. According to the general principle of the oscillometric method, the cuff was placed on patient's wrist or arm. The cuff was inflated by using pneumatic pump to the estimated pressure and then was deflated at a certain deflation rate. During the deflation of pressure, the pressure and it's oscillation was measured. The measuring pressure and its oscillation are calculated and analysed by the software of the microprocessor according to the algorithm of oscillometric method.
After completing the measuring cycle, the systolic pressure, diastolic pressure, and heart beat rate are displayed on the LCD of device. All the measuring information and the time of measuring may be stored in the memories of device. Those stored memories may be transmitted out of device through the signal transmission cable.
Basically the blood pressure monitor M100 is the device measuring blood pressure at wrist.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
wrist or arm
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Not Found
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
HL-168(K990807), HL-168 (K990807)
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 870.1130 Noninvasive blood pressure measurement system.
(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).
0
NOV 1 0 2003
510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: ____________________
-
- Submitter's Identifications: Meditek Technology Inc. 1F, No. 11, Lane 223, Sec. 3, Chung Ching N. Rd., Taipei, Taiwan, R.O.C.
Contact: Mr. Rodin Chang President
- Submitter's Identifications: Meditek Technology Inc. 1F, No. 11, Lane 223, Sec. 3, Chung Ching N. Rd., Taipei, Taiwan, R.O.C.
Date of Summary Preparation: September 25, 2002.
-
- Name of the Device: Meditek Blood pressure Monitor, model M100.
-
- Information of the 510(k) Cleared Device (Predicate Device): Health care: HL-168(K990807).
4. Device Description:
The Meditek Blood Pressure Monitor, model M100, is the non-invasive electronic medical device intended for the measuring of blood pressure (including systolic pressure and diastolic pressure) and the number of heart beats of patient by using the oscillometric method. According to the general principle of the oscillometric method, the cuff was placed on patient's wrist or arm. The cuff was inflated by using pneumatic pump to the estimated pressure and then was deflated at a certain deflation rate. During the deflation of pressure, the pressure and it's oscillation was measured. The measuring pressure and its oscillation are calculated and analysed by the software of the microprocessor according to the algorithm of oscillometric method.
After completing the measuring cycle, the systolic pressure, diastolic pressure, and heart beat rate are displayed on the LCD of device. All the measuring information and the time of measuring may be stored in the memories of device. Those stored memories may be transmitted out of device through the signal transmission cable.
Basically the blood pressure monitor M100 is the device measuring blood pressure at wrist.
5. Intended Use:
The M100 Blood Pressure Monitor is an electric device that provides a signal from which the systolic, diastolic, and rate of heart beats can be derived through the use of transducers placed on the surface of body. The signals of measurement can be stored in the memory of device, and then can be displayed on the screen of device and/or be transmitted out of device by using the standard signal transmission cable.
1
mediek technology mc.
-
- Comparison to the 510(k) Cleared Device (Predicate Device):
The cleared device for the specification comparison of M100 is HL-168 (K990807).
- Comparison to the 510(k) Cleared Device (Predicate Device):
-
- Discussion of Non-Clinical Tests Performed Determination of Substantial Equivalence are as follows:
Compliance to applicable voluntary standards includes ANSI/AAMI, SP 10-1992, as well as EN 60601-1, EN 60601-1-1, and EN 60601-1-2 requirement.
- Discussion of Non-Clinical Tests Performed Determination of Substantial Equivalence are as follows:
In addition to the compliance of voluntary standards, the software verification has been carried out according to the FDA software guidance.
-
- Conclusions
The Meditek Blood Pressure Monitor, including model M100, has the same intended use and technological characteristics as the cleared device of HL-168. Moreover, verification and validation tests contained in this submission demonstrate that the difference in the submitted demonstrate that the difference in the submitted models could maintain the same safety and effectiveness as that of cleared device.
- Conclusions
In the other words, those engineering difference do not: (1) affect the intended use or (2) alter the fundamental scientific technology of the device.
2
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of a stylized eagle.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 1 0 2003
Meditek Technology, Inc. c/o Mr. Tony C. S. Chang Official Correspondent 1F, No. 11, Lane 223, Sec. 3 Chung Ching N. Road Taipei Taiwan, R.O.C.
Re: K032308
Trade Name: Blood Pressure Monitor, Model M100 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Monitor Regulatory Class: Class II (two) Product Code: DXN Dated: November 3, 2003 Received: November 3, 2003
Dear Mr. Chang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
3
Page 2 – Mr. Tony C. S. Chang
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Uffaxmy
for Bram D. Zuckerman, M.D.
Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
4
510(k) Number (if known):
Device Name: Blood Pressure Monitor / M100 .
Indications For Use:
The M100 of blood pressure monitor is an electronic device that provides a signal from which the systolic, diastolic and rate of heart beats can be derived through the use of transducers placed on the surface of the body. The signals of measurement can be stored in the memory of device, and then can be displayed on the screen of device and/or be transmitted out of device by using the standard signal transmission cable.
PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
K032308
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number | |
---|---|
--------------- | -- |
| Prescription Use (Per 21 CFR 801.109) | | OR | Over-The-Counter Use |
✓ |
--------------------------------------- | -- | ---- | ---------------------- | ----------------------------------------- |
---|
K03230
And All and