(137 days)
ImmunoCard STAT! HpSA is a rapid in vitro qualitative assay for the detection of Helicobacter pylori antigen (HpSA) in human stool. The stool antigen detection is intended to aid in the diagnosis of H. pylori infection and to demonstrate loss of H. pylori stool antigen following treatment. Conventional medical practice recommends that testing by any method to confirm the loss of antigen be done at least four weeks following completion of therapy.
ImmunoCard STAT! HpSA is a qualitative horizontal flow in vitro diagnostic device used to detect the presence of H. pylori antigen in human stool specimens. The intended use of the device is identical to that of Premier Platinum HpSA (Meridian Bioscience, Inc., Cincinnati, OH) an enzyme-linked immunoassay previously cleared to market under 510(k) K983255. While assay methods differ, both are designed to detect Helicobacter pylori antigen in the stools of patients. The results of both tests are intended to aid in the diagnosis of H. pylori infection and to monitor bacterial reduction in response to anti-bacterial therapy.
Acceptance Criteria and Study for ImmunoCard STAT!® HpSA®
This report describes the acceptance criteria and the study that proves the ImmunoCard STAT! HpSA device meets these criteria, based on the provided 510(k) summary.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implicitly derived from the comparative and clinical study results that demonstrate substantial equivalence to the predicate device and acceptable diagnostic accuracy.
| Performance Metric | Acceptance Criteria (Implicit) | Reported Device Performance (ImmunoCard STAT! HpSA) |
|---|---|---|
| Comparative Study | High correlation with predicate device (Premier Platinum HpSA) | Correlation with Premier Platinum HpSA: 95% initial correlation, improving after referee laboratory analysis. After referee analysis: 436 concordant (105 positive, 331 negative), 20 discordant. |
| Clinical Sensitivity (Diagnosis) | High sensitivity to detect H. pylori infection in dyspeptic patients. | 90.6% (95% CI: 84.9% to 97.1%) in dyspeptic patients (before treatment). |
| Clinical Specificity (Diagnosis) | High specificity to correctly identify patients without H. pylori infection in dyspeptic patients. | 91.5% (95% CI: 87.5% to 96.5%) in dyspeptic patients (before treatment). |
| Predictive Value Positive (Diagnosis) | High positive predictive value. | 86.5% (95% CI: 79.9% to 94.1%) in dyspeptic patients (before treatment). |
| Predictive Value Negative (Diagnosis) | High negative predictive value. | 94.2% (95% CI: 90.1% to 97.9%) in dyspeptic patients (before treatment). |
| Overall Correlation (Diagnosis) | High overall correlation with gold standard. | 91.2% (95% CI: 87.3% to 94.7%) in dyspeptic patients (before treatment). |
| Clinical Sensitivity (Eradication) | High sensitivity to confirm eradication of H. pylori post-treatment. | 95.4% (95% CI: 86.0% to 100%) in post-treatment patients. |
| Clinical Specificity (Eradication) | High specificity to confirm eradication of H. pylori post-treatment. | 100% in post-treatment patients. |
| Predictive Value Positive (Eradication) | High positive predictive value post-treatment. | 100% in post-treatment patients. |
| Predictive Value Negative (Eradication) | High negative predictive value post-treatment. | 98.4% (95% CI: 94.5% to 100%) in post-treatment patients. |
| Overall Correlation (Eradication) | High overall correlation with gold standard post-treatment. | 98.8% (95% CI: 96.8% to 100%) in post-treatment patients. |
| Reproducibility | Consistent results across different tests and sites for known positive and negative samples. | Intra-assay and inter-assay reproducibility was 100%. Tested with 5 negative and 5 positive samples (2 near limit of detection). Conducted at four independent test sites over three days. |
| Lower Limit of Detection | Ability to detect H. pylori antigen at a specified concentration. | 64 ng/mL in tests with sonicated antigen prepared from H. pylori strain TV1970. |
| Assay Specificity | No positive results from common interfering microorganisms, and no interference with detection of H. pylori in positive stools. | None of the tested bacterial, viral, or yeast strains (≥ 1 X 106 bacteria or yeast spiked into stools) yielded a positive result in negative stool or interfered with positive stool detection. Both negative and positive stools were positive when spiked with H. pylori strain 43504. |
| Interfering Substances | No interference with results from common medications or stool components at specified concentrations. | Tums® Antiacid (5 mg/mL), Tagamet® (5 mg/mL), Prilosec® (5 mg/mL), Mylanta® Antacid (1:20), Pepto-Bismol® (1:20), Barium sulfate (5%), Whole Blood (50%), Leukocytes (50%), Mucin (3.4%), Stearic acid/palmitic acid (fecal fat) (4%), Hemoglobin (tarry stool) (12.5%) were found to have no effect. |
2. Sample Sizes Used for the Test Set and Data Provenance
-
Comparative Study (with predicate device):
- Sample Size: 457 total samples.
- Data Provenance: Not explicitly stated, but implies "in-house" due to the involvement of "four independent laboratories" for initial testing and "a referee laboratory (Meridian Bioscience, Inc., Cincinnati)" for discordant samples. The nature of this study suggests it might be a retrospective assessment of banked samples or a controlled, prospective comparison within the specified labs. Given the device was marketed outside the US since 2002, some of this data may originate from those regions.
-
Clinical Studies:
- Sample Size:
- Diagnosis (Dyspeptic Patients): 227 consecutive dyspeptic patients.
- Eradication (Post-Treatment): 85 patients from the initial 227 were followed up post-treatment (22 True Positive, 63 True Negative for H. pylori post-treatment).
- Data Provenance: Not explicitly stated, but typical for clinical trials involving patient samples. Given the device's prior marketing outside the US since 2002 and references to studies conducted outside the United States (5-9), it is highly probable that this data originates from outside the United States. The text mentions "clinical studies" and "studies conducted outside of the United States."
- Sample Size:
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
-
Comparative Study (with predicate device):
- Number of Experts: Not specified.
- Qualifications: The "referee laboratory" (Meridian Bioscience, Inc.) evaluated discordant samples, but the number and specific qualifications of individuals are not detailed. It can be inferred that they were expert laboratory personnel capable of performing thorough investigations and potentially additional confirmatory tests beyond the two assays being compared.
-
Clinical Studies:
- Number of Experts: Not specified.
- Qualifications: The ground truth for H. pylori infection was established by "conventional in vitro diagnostic methods." These methods involve histology, rapid urease test, and culture, typically interpreted by trained pathologists and microbiologists. While the specific number of experts isn't given, these methods inherently require specialized medical professionals (e.g., pathologists, microbiologists) to establish a diagnosis. The description "histology and urease tests were positive, or if culture was positive" indicates a consensus or composite reference standard approach using widely accepted diagnostic techniques.
4. Adjudication Method for the Test Set
-
Comparative Study: An adjudication by a referee laboratory was used for discordant results between the ImmunoCard STAT! HpSA and the Premier Platinum HpSA assays. "Samples giving discordant results...were sent to and evaluated by a referee laboratory" for "further investigation." This suggests a mechanism to resolve discrepancies and arrive at a corrected ground truth or understanding of the discrepancy.
-
Clinical Studies: For establishing the H. pylori status of patients, the ground truth was defined as a composite reference standard: "Patients were defined as infected with H. pylori if histology and urease tests were positive, or if culture was positive." This is a form of consensus or composite ground truth rather than a direct adjudication between expert readers of the device. If all three (histology, urease, culture) were performed, two positive results might trigger the positive ground truth, or merely one positive result from any of the three, as implied by the "or if culture was positive" statement.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
There is no indication of a Multi-Reader Multi-Case (MRMC) comparative effectiveness study being performed with human readers and AI assistance. The ImmunoCard STAT! HpSA is a qualitative lateral flow immunoassay with a visual endpoint (pink-red line or no line), interpreted by a laboratory technician. It is not an imaging-based AI device where human readers are typically evaluated with and without AI assistance.
6. Standalone Performance Study
Yes, a standalone performance study was done for ImmunoCard STAT! HpSA. The "Clinical Studies" section provides the standalone diagnostic accuracy (sensitivity, specificity, predictive values, and correlation) of the device against a composite gold standard for H. pylori infection, both for initial diagnosis and for confirming eradication post-treatment. The "Reproducibility," "Assay Specificity," and "Tests for Interfering Substances" sections also represent standalone performance assessments of the device itself.
7. Type of Ground Truth Used
-
Comparative Study: The "corrected results" for the comparative study were based on referee laboratory evaluation of discordant samples, which implies further testing or expert review to determine the true status of those samples against the two devices being compared.
-
Clinical Studies: The ground truth for H. pylori status was a composite reference standard employing established clinical and laboratory methods:
- Histology (e.g., biopsy for H&E staining)
- Rapid Urease Test
- Bacterial Culture
Patients were defined as H. pylori positive if histology and urease tests were positive, or if culture was positive. This combines multiple evidence types, often considered a strong form of ground truth in clinical diagnostics.
8. Sample Size for the Training Set
The document does not explicitly state the sample size for a training set. ImmunoCard STAT! HpSA is a qualitative immunoassay, not a machine learning or AI-driven diagnostic device that typically undergoes a distinct "training phase" comparable to algorithms. Its development and validation would involve optimization of reagents and protocols rather than training on a separate dataset in the AI sense.
9. How Ground Truth for the Training Set Was Established
As mentioned above, since this is not an AI/machine learning device, the concept of a "training set" ground truth in the conventional sense does not apply. The development of the assay's capture and detector antibodies and optimizing the assay parameters would be guided by established biological principles and analytical validation using characterized positive and negative controls/samples, rather than a data-driven training set with established ground truth labels for an algorithm.
{0}------------------------------------------------
DEC - 5 2003
Image /page/0/Picture/2 description: The image contains the text "K032222", "510(k) Notification", and "ImmunoCard STAT! HpSA". The text appears to be part of a document or label. The text "K032222" is at the top, followed by "510(k) Notification", and then "ImmunoCard STAT! HpSA".
AMENDED 510(k) SUMMARY OF SAFETY & EFFECTIVENESS (Amended 11/25/03)
IDENTIFICATION INFORMATION
SUBMITTER'S INFORMATION
This summary of 510(k) safety and effectiveness is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: K032222
SUBMITTER'S NAME AND ADDRESS: Meridian Bioscience, Inc. 3471 River Hills Drive Cincinnati, OH 45244
PHONE NUMBER: (513) 271-3700
FAX NUMBER: (513) 272-5213
CONTACT PERSON: Susan Rolih Vice President, Regulatory Affairs and Quality Assurance Official Correspondent
DATE SUMMARY PREPARED: July 18, 2003
NAME OF DEVICE: ImmunoCard STAT!® HpSA® (ImmunoCard STAT! and HpSA are registered trademarks of Meridian Bioscience, Inc.)
COMMON NAME: Lateral flow immunoassay for H. pvlori stool antigen
CLASSIFICATION NAME: Campylobacter pvlori [83LYR]
REGULATION: 866,3110
PREDICATE EQUIVALENT DEVICES: Premier Platinum HpSA® (K983255)
INTENDED USE:
ImmunoCard STAT! HpSA is a rapid in vitro qualitative assay for the detection of Helicobacter pylori antigen (HpSA) in human stool. The stool antigen detection is intended to aid in the diagnosis of H. pylori infection and to demonstrate loss of H. pylori stool antigen following treatment. Conventional medical practice recommends that testing by any method to confirm the loss of antigen be done at least four weeks following completion of therapy. (1)
BACKGROUND:
pylori is a spiral gram-negative bacterium that invades the mucosal membrane of the H. gastrointestinal tract. It causes chronic gastritis, predisposes some infected patients to gastric and peptic duodenal ulcers. (1-3) Noninvasive in vitro diagnostic assays, such as ImmunoCard STAT! HpSA, have been shown to be effective in differentiating infected patients. Such noninvasive assays are also recommended to
{1}------------------------------------------------
monitor the success and failure of treatment regimens to eradicate the organism. (3)
H. pylor is found in the stomachs of humans. Infections with the organism are distributed world -wide, however the preponderance appears in developing countries, where the incidence of infection is 70-80%. In developed or more industrialized countries, the incidence of infection is only 25-50%. The incidence continues to decrease in persons in higher socioeconomic levels. Infections in all groups appear to occur in childhood and many before the age of 10 years. Males and females appear to be infected at the same rates. (4) Transmission of the organism between humans is not well understood. particularly since the harbor for the infection is the human stomach. It is most likely that all infections have occurred through ingestion of fecally contaminated materials.
All infected patients develop chronic gastric inflammation but the condition is usually asymptomatic. H. pylor is the direct cause of most gastric and duodenal ulcers. Eradication of the orqanism leads to oure of the ulcers. Infection due to H. pylor is strongly associated with atrophic gastritis (which is a precursor to gastric cancer) and with adenocarcinoma of the distal stomach.
A variety invasive and noninvasive tests are used to detect and isolate H pylori. Invasive testing includes histological biopsy for hematoxylin and eosin (H and E) staining, bacterial culture, urease testing and PCR analysis. Invasive tests present some slight degree of risk for the patient due to complications. Noninvasive tests include those to monitor breath, serum, gastric juice and urine for the direct or indirect presence of organisms.
DEVICE DESCRIPTION:
ImmunoCard STAT! HpSA is a qualitative horizontal flow in vitro diagnostic device used to detect the presence of H. pylori antigen in human stool specimens. The intended use of the device is identical to that of Premier Platinum HpSA (Meridian Bioscience, Inc., Cincinnati, OH) an enzyme-linked immunoassay previously cleared to market under 510(k) K983255. While assay methods differ, both are designed to detect Helicobacter pylori antigen in the stools of patients. The results of both tests are intended to aid in the diagnosis of H. pylori infection and to monitor bacterial reduction in response to anti-bacterial therapy.
A. Technological characteristics compared to predicate device:
| Characteristics | IC STAT! HpSA | Premier Platinum HpSA |
|---|---|---|
| Device Type | ||
| In vitro diagnostic device | Yes | Yes |
| Control | No | No |
| Calibrator | No | No |
| Intended Use | ||
| Detection of H. pylori antigens in human stool | Yes | Yes |
| Acceptable Sample | ||
| Formed stool | Yes | Yes |
| Semi-solid stool | Yes | Yes |
| Liquid stool | Yes | Yes |
| Watery stool samples | No | No |
| Stool collected in transport media | No | No |
{2}------------------------------------------------
Comparison of Assay Methods
| Characteristic | IC STAT! HpSA | Premier Platinum HpSA |
|---|---|---|
| Intended use | Detection of H. pylori antigen in stool | Detection of H. pylori antigen in stool |
| Results | Qualitative | Qualitative |
| Specimen Required | 1. Stool | 1. Stool |
| Technology | Lateral flow chromatography | Enzyme-linked immunoassay |
| Level of skill required | Laboratory Technician | Laboratory Technician |
| Assay steps | 1. Dilute specimen in Sample Diluent2. Add diluted specimen to test port3. Incubate at 20-26 C for 5 minutes4. Read results visually | 1. Dilute specimen in Sample Diluent2. Add diluted specimen to test well3. Incubate at 22-27 C for 60 minutes4. Wash test well5. Add Conjugate Reagent6. Add Substrate Reagent7. Incubate at 22-27 C for 10 minutes8. Add Stop Solution9. Read results using spectrophotometer |
| End point | Visual color line | Color change, change in optical density of solution |
| Interpretation of testresult | Positive = pink-red lineNegative = no line | Positive = OD ≥ 0.120 at A450/630 nm or> 0.160 at A450 nmNegative = OD ≤ 0.100 at A450/630 nm or< 0.140 at A450 nm |
B. Device Components:
- Test Devices: lateral flow membrane strips impregnated with monoclonal anti-H. 1. pylori as the capture antibody, red latex-conjugated detector antibody. The strips are enclosed in a plastic case with a window.
- Positive Control: a dilute suspension of inactivated H. pylori in a buffered solution 2. containing <0.1% sodium azide as a preservative.
- Specimen Diluent: a buffered salt solution containing <0.1% sodium azide as a റ്റ് preservative.
C. Principle of the Test:
ImmunoCard STAT! HpSA uses capture solid phase technology to detect the presence of antigen in test specimens.
To perform the test, patient stool is added to the Sample Diluent using the applicator stick that is ro perform the Sample Diluent Vial. The diluted stool sample (approximately a 1 in 10 dilution) is dispensed through the tip of the Sample Diluent Vial into the round sample window of the device. H. pylori antigen, if present in the diluted sample, binds to the detector antibody-latex conjugate as the sample moves through the device. The capture monoclonal antibody, which is bound to as the sample throught through, binds the antigen-antibody-latex complex and yields a visible pink-red line. When no antigen is present, no complex is formed and no pink-red line will appear at the test position of the central window.
A control line, appearing at the control position in the test window, shows whether adequate flow has occurred through the device during a test run. The control line is a protein of nonmammalian origin.
{3}------------------------------------------------
Blue latex particles conjugated with a monoclonal antibody to this protein co-migrate with the latex Blue latex particles conjugated with a monocolar antibody only be control on the device in the device bound detector antibody during the incubation step. "A black our of the some of the test is considered invalid.
D. Contraindications, precautions, Warnings:
There are no known contraindications for ImmunoCard STAT! HpSA. See product labeling for There are no Khowr Contraindications for immans sala in vitro diagnostic device.
MARKETING HISTORY:
ImmunoCard STAT! HpSA has been marketed since 2002 outside the United States. It is currently ImmunoCard STAT! HDSA has been marketed shiec 2002 outcle with and and studies conducted outside of the United States. (5-9)
ADVERSE EFFECTS OF THE DEVICE ON HEALTH:
There are no potential adverse effects of health associated with the use of this in with and There are no potential adverse ellects of health associated withings symptoms of the patient and device. The diagnosis of H. pylon mieculor is news of the or stool speciments. Conventional in vitro diagnostic methods for confirming H pylori infection include:
- iagnostic methods for confilming in pylor mecaon molder.
13c or 4C-labeled urea breath test to detect 33c or 14C-labeled CO2 expired in air as a result of 1) H. pylori urease activity - H. pylon drease activity
Serology to detect circulating H. pylori IgG antibody in serum or whole blood -
- Stool assay for the detection of bacterial antigen 3)
SUBSTANTIAL EQUIVALENCE:
Four independent laboraories tested specimens in parallel with Comparative studies: Comparative Studies: Four intepence ELSA in in ImmunoCard STAT! HpSA and a leleleine Lawner Samples giving discordant results of the HpSA (Meridian Broscience, The, Ciffelinan) Only between the two assays were sent to and evaluated by a rolerone and more investigation of discordant samples by the referee laboratory.
| Initial Trial Results | Corrected Results | |
|---|---|---|
| Total samples tested | 457 | 457* |
| Concordant test results | 433 | 436 |
| Positive samples | 102 | 105 |
| Negative samples | 331 | 331 |
| Discordant test results | 21 | 20 |
| Premier +, ImmunoCard - | 6 | 6 |
| Premier -, ImmunoCard + | 15 | 14 |
| Indeterminant | 3 | 1 |
| Premier Equivocal, ImmunoCard + | 2 | 1 |
| Premier Equivocal, ImmunoCard - | 1 | 0 |
| % correlation | 95% | N/A |
- Two discordant samples were QNS for follow up analysis.
{4}------------------------------------------------
The lower limit of detection of this assay is 64 ng/mL in tests with sonicaled antigen prepared The lower limit of delection of this assay is 64 igine in took and to semi-solid stool.
from H. pylori strain TV1970. This limit does not vary from formed (solid) to semi-so
Clinical studies: Stool samples from 227 consecutive dyspeptic patients, who were not usings Clinical studies: Stool Samples from 227 Golloodi.ro oferred for endoscopy were tested with acid suppressant therapy of antibiolos, and who were taken for histology, rapid urease test and ImmunoCard STAT! HpSA. Diopsy Specifiched with H. pylor if histology and urease tests were culture. Patlents were denned as linected with 1. pyron in keeregy of
positive, or if culture was positive. Eighty five of the 227 patients were found H. pylori positive. The results are summarized in the following table.
Diagnostic accuracy of ImmunoCard STAT! HpSA before and after H. pylori eradication treatment.
| H. pylori status byendoscopy/biopsy/gold standard | True Positive | True Negative | Total | |
|---|---|---|---|---|
| IC STAT! HpSA + | 77 | 12 | 89 | |
| IC STAT! HpSA - | 8 | 130 | 138 | |
| Total | 85 | 142 | 227 | |
| Estimated clinical sensitivity (95% CI) | 90.6% (84.9 to 97.1%) | |||
| Estimated clinical specificity (95% CI) | 91.5% (87.5 to 96.5%) | |||
| Predictive value, positive test (95% CI) | 86.5% (79.9 to 94.1%) | |||
| Predictive value, negative test (95% CI) | 94.2% (90.1 to 97.9%) | |||
| Correlation (CI 95%) | 91.2% (87.3 to 94.7%) |
Correlation of ImmunoCard STAT! HpSA test results with eradication treatment
| H. pylori status byendoscopy/biopsy/gold standard | |||
|---|---|---|---|
| True Positive | True Negative | Total | |
| IC STAT! HpSA + | 21 | 0 | 21 |
| IC STAT! HpSA - | 1 | 63 | 64 |
| Total | 22 | 63 | 85 |
| Estimated clinical sensitivity (95% CI) | 95.4% (86.0 to 100%) | ||
| Estimated clinical specificity (95% CI) | 100% | ||
| Predictive value, positive test (95% CI) | 100% | ||
| Predictive value, negative test (95% CI) | 98.4% (94.5 to 100%) | ||
| Correlation (CI 95%) | 98.8% (96.8 to 100%) |
REPRODUCIBILITY
The reproducibility of ImmunoCard STAT! HpSA was determined with known negative (n = 5) and The reproduciblity of Inn fullocald OTAT: riper wo works sorted to prevent their identities. Two of positive (n = 5) samples), that were couce and randemly on the assay The reproducibility
the five positive samples were near the limit of detection Intro-ass the five positive samples were near the links of assession of thee independent test sites. Intra-assay and interassay reproducibility was 100%.
{5}------------------------------------------------
| Sample Status | Premier OD reading | IC STAT! graded reading * | Referee (MBI) Day 1 | Referee (MBI) Day 2 | Referee (MBI) Day 3 | Clinical Site # 1 Day 1 | Clinical Site # 1 Day 2 | Clinical Site # 1 Day 3 | Clinical Site # 2 Day 1 | Clinical Site # 2 Day 2 | Clinical Site # 2 Day 3 | Clinical Site # 3 Day 1 | Clinical Site # 3 Day 2 | Clinical Site # 3 Day 3 |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Neg | 0.017 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 |
| Pos | 0.737 | 4/5 | + | + | + | + | + | + | + | + | + | + | + | + |
| Neg | 0.016 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 |
| Pos | 1.140 | 7 | + | + | + | + | + | + | + | + | + | + | + | + |
| Neg | 0.028 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 |
| Low Pos | 1.442 | 1 | + | + | + | w | w | w | + | w | + | + | + | + |
| Neg | 0.042 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 |
| Low Pos | 1.041 | 2 | + | + | + | + | w | + | + | + | + | + | + | + |
| Pos | 1.493 | 5 | + | + | + | + | + | + | + | + | + | + | + | + |
| Neg | 0.058 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 |
| Pos Cont | 2.309 | N/A | + | + | + | + | + | + | + | + | + | + | + | + |
| Neg Cont | 0.034 | N/A | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 |
- The signal intensity (strength) of a positive reaction in ImmunoCard STAT! will not necessarily correlate with the OD value obtained in Premier Platinum HpSA EIA.
Legend: 0 = negative, 1-10 = semiquanitiative scoring scale used in the interpretation of ImmunoCard STAT! positive test results. (A value was assigned to the intensity of color in the Test Line, where 1 is the weakest visible positive reaction and 10 is the strongest. A 4/5 means the reaction fell between a grade of 4 and a gradicor (s) sub = weak (correlates with a semiquantitative reaction grade of +/-, 1 or 2)
ASSAY SPECIFICITY
The specificity of ImmunoCard STAT! HpSA was tested utilizing the following bacterial, viral and yeast strains. Positive and negative stools were spiked with ≥ 1 X 10° bacteria or yeast. None of the microorganisms tested yielded a positive result in the negative stool or interfered with detection of the positive stool. Both the negative and positive stool was positive when spiked with Helicobacter pylori strain 43504.
Adenovirus Type 2 Adenovirus Type 40 Coxsackie Type B1 Coxsackie Type B6 Echovirus Type 22 Feline calicivirus Rotavirus
Aeromonas hydrophila Campylobacter coli
રન્ડ
{6}------------------------------------------------
Meridian Bioscience, Inc. Cincinnati, OH
Campylobacter jejuni Candida albicans Citrobacter freundii Clostridium perfringens Clostridium difficile (2) Enterobacter cloacae Enterococcus faecalis (2) E. coli (2) E. coli 0157:H7 (2) E. ferqusonii Helicobacter felis Klebsiella pneumoniae Proteus vulgaris Pseudomonas aeruginosa Salmonella dublin Salmonella (Group B) Salmonella hilversum Salmonella minnesota Salmonella typhimurium Staphylococcus aureus Staphylococcus aureus (Cowan I) Staphylococcus epidermidis Serratia liquifaciens Shigella boydii Shigella dysenteriae Shigella flexneri Shigella sonnei Yersinia enterocolitica
Borrelia burgdorferi (Stool inoculated with antigen protein to a final conc. of 32 ug/mL)
TESTS FOR INTERFERING SUBSTANCES
The following substances were found to have no effect on results when present in stool at the concentrations indicated. Tums® Antiacid (5 mg/mL) Tagamet® (5 mq/mL) Prilosec® (5 mg/mL) Mylanta® Antacid (1:20) Pepto-Bismol® (1:20) Barium sulfate (5%) Whole Blood (50%) Leukocytes (50%) Mucin (3.4%) Stearic acid/palmitic acid (fecal fat) (4%) Hemoglobin (tarry stool) (12.5%)
BIBLIOGRAPHY:
{7}------------------------------------------------
- Marshall BJ, Warren JR. Unidentified curved bacilli in the stomach of patients with gastritis (1) and peptic ulceration Lancet 1984:1:1311-14.
- Dunn BE, Cohen H, Blaser MJ. Helicobacter pylori. Clin Miocobiol Reviews, 1997;10:720-41. (2)
- Vaira D, Malferthainer P, Megraaud F, et al. Diagnosis of Helicobacter pylori infection with a (4) valla D, Mallerthallier F , Mogralassay. Lancet 1999:345:30-3.
- Graham D. Helicobacter pylori: Its epidemiology and its role in duodenal disease. J (5) Gastroenterol, 1991:4:105-13.
- Calvet K, Quesada M, Rosello M, et al. Stool antigen for the ddiagnosis of Helicobacter pylori (6) Calver K, Quesada M, Rosello M, Erai. Ottor anigen for and the one of the macol Ther 2003:12:727-31.
- 2003;12:727-31.
Calvet X, Salceda F, Sanfeliu I, et al. Testing a new in-office test for determination of faecal
Calvet X, Salceda F, Sanfeliu I, et al. Testing a new in-of (6) Calvel A, Salecual P, Samillan, Med Clin (Barc) 202; 118:126-9. - Antos D, Konstantopoulos N, Crone J, Koletzko S. Evaluation of a novel rapid one-step (7) Antos D, Konstantopoulos N, Cruite J, Koletion of H. pylori antigen in stool in children.
monoclonal enzyme immunoassay for detection of H. pylori antigen in the Podiatic monocional enzyme immanodosay for desiring European Society for Pediatric Abstract presented at the 20 "Annual Mocally Laropour Co., Progue, Czech Republic, June 2003. - Perna F, Tampieri A, Rici C et al. Evaluation of a new rapid one step stool antigen test for (8) Perna F, Tamplen A, Rici C et al. Evandalion of d'hother eneshalt the XV International Helicobacter pylori (HP) intection ulagnosis. Abstracted prosunted and Creece, September 11-14, 2002.
- Lebdolter A, Wolle K, Wex T et al. Evaluation of a novel rapid H. pylori stool antigen test: Is a
rease (છ) Lebdolter A, VVolle K, VVEX T et al. Evaluation of a novel rapia villa in pyromation
reliable now possible in the doctor's office? Abstract presented at the Digestive Dise Week, Orlando Florida, May 2003.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for the Department of Health & Human Services (USA). The logo consists of a stylized caduceus symbol, which features a staff with a serpent entwined around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" arranged in a circular fashion around the symbol. The text is in all caps and appears to be in a sans-serif font. The logo is black and white.
Public Health Service
DEC - 5 2003
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Susan Rolih Vice President, Regulatory Affairs and Quality Assurance Official Correspondent Meridian Bioscience, Inc. 3471 River Hills Drive Cincinnati, OH 45244
K032222 Re:
Trade/Device Name: ImmunoCard STAT! HpSA Regulation Number: 21 CFR 866.3110 Regulation Name: Campylobacter Fetus Serological Reagents Regulatory Class: Class I Product Code: LYR Dated: October 24, 2003 Received: October 27, 2003
Dear Ms. Rolih:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{9}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Putman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
Indications for Use
510(k) Number (if known): K032222
Device Name: ImmunoCard STAT! HpSA
Indications For Use: ImmunoCard STAT! HpSA is a rapid in vitro qualitative assay for the detection of Helicobacter pylori antigen (HpSA) in human stool. The stool antigen detection is intended to aid in the diagnosis of H. pylori infection and to demonstrate loss of H. pylori stool antigen following treatment. Conventional medical practice recommends that testing by any method to confirm the loss of antigen be done at least four weeks following completion of therapy.
Prescription Use V No (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use No (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Fidelity Poolo
Division Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
1/ 1/32222
§ 866.3110
Campylobacter fetus serological reagents.(a)
Identification. Campylobacter fetus serological reagents are devices that consist of antisera conjugated with a fluorescent dye used to identifyCampylobacter fetus from clinical specimens or cultured isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by this bacterium and provides epidemiological information on these diseases.Campylobacter fetus is a frequent cause of abortion in sheep and cattle and is sometimes responsible for endocarditis (inflammation of certain membranes of the heart) and enteritis (inflammation of the intestines) in humans.(b)
Classification. Class I (general controls).