K Number
K031378

Validate with FDA (Live)

Date Cleared
2003-05-30

(29 days)

Product Code
Regulation Number
876.5860
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Fresenius Ultraflux dialyzer is designed for single use acute and chronic hemodialysis.

Device Description

Not Found

AI/ML Overview

The presented document is a 510(k) Premarket Notification for the Fresenius Ultraflux Single Use Hemodialyzer. This type of submission focuses on demonstrating substantial equivalence to a predicate device, rather than proving a new device's efficacy through a clinical study with detailed acceptance criteria as one might find in a PMA application or for novel software devices.

Here's a breakdown based on the provided text, addressing your questions where possible:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not present a formal table of acceptance criteria with specific thresholds for performance metrics. Instead, it states that "In vitro testing was performed to determine the following: priming volume, bloodside and dialysate-side pressure drops, ultrafiltration coefficient, urea clearance, creatinine clearance, Vitamin B12 clearance, phosphate and lysozyme clearance." The acceptance criteria for these tests are implicitly that the Ultraflux dialyzer performs "substantially equivalent" to the predicate devices (Fresenius Optiflux 200NR and Gambro Polyflux 17S) for its intended use.

Therefore, for this device, a direct table of numerical "acceptance criteria" and "reported device performance" in the way you might expect for a diagnostic or AI device is not available in the provided text. The "acceptance criteria" here is met by demonstrating performance equivalent to legally marketed predicate devices through in vitro testing.

2. Sample Size for the Test Set and Data Provenance

  • Sample Size for the Test Set: Not explicitly stated in the document. The testing described is "in vitro testing," which generally refers to laboratory bench tests.
  • Data Provenance: The data is generated from in vitro testing of the Fresenius Ultraflux hemodialyzers. The document does not specify the country of origin for these specific tests, but it's for a device from "Fresenius Medical Care North America." The testing is analogous to prospective data collection in a lab setting, as the tests are performed on the device itself to assess its characteristics.

3. Number of Experts Used to Establish Ground Truth and Qualifications

This information is not applicable to this 510(k) submission. The ground truth for this medical device is established by its physical and functional characteristics measured through standardized in vitro tests, not by expert consensus on interpretations of data (like medical images).

4. Adjudication Method for the Test Set

Not applicable. There is no "adjudication method" in the context of expert consensus as described for an AI/diagnostic device, since the testing is in vitro and objective.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. An MRMC comparative effectiveness study is not relevant for this type of medical device (a hemodialyzer). This type of study investigates the performance of human readers, sometimes with AI assistance, in interpreting medical data. The hemodialyzer is a therapeutic device, not a diagnostic one involving human interpretation.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Not applicable. This is not a software algorithm or AI device; it is a physical medical device (hemodialyzer). Its performance is assessed directly through in vitro measurements.

7. Type of Ground Truth Used

The ground truth is based on objective measurements from in vitro testing of the device's physical and functional properties (e.g., priming volume, pressure drops, clearance rates, ultrafiltration coefficient). These measurements are then compared to the known performance characteristics of the predicate devices to demonstrate substantial equivalence.

8. Sample Size for the Training Set

Not applicable. There is no "training set" as this is not a machine learning or AI device that requires training data.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows a black and white illustration of a funnel. The funnel is composed of four sections that are stacked on top of each other. Each section is a trapezoid shape, with the widest part at the top and the narrowest part at the bottom.

K031378

Ray 1/3

Fresenius Medical Care

Fresenius Ultraflux Single Use Hemodialyzer "Special" 510(k) Premarket Notification

Summary of Safety and Effectiveness

This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of SMDA 1990.

A. Submitter's Information:

Name:Fresenius Medical Care North America
Address:95 Hayden AveLexington, MA 02420
Phone:1-781-402-9068
Fax:(781) 402-9635
Contact Person:Arthur Eilinsfeld, Director of Regulatory Affairs
Date of Preparation:29 April 2003

B. Device Name:

Common Name:Dialyzer, High Permeability with or withoutSealed Dialysate System
Product Code/Classification Panel:78KDI/Gastroenterology-Urology
Classification:Class II per §876.5860

C. Predicate Device

The predicate devices for the Fresenius Ultraflux Hemodialyzers are:

  • Fresenius Optiflux 200NR - #K002277 (8/25/00);
  • . Gambro Polyflux 17S - #K982414 (3/26/99).

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the text "Page 2/3" written in black ink on a white background. The word "Page" is written in cursive, while the numbers "2/3" are written in a more standard font. The slash between the 2 and 3 is underlined.

Image /page/1/Picture/1 description: The image shows a geometric figure that resembles a downward-pointing pyramid or funnel. The figure is composed of four stacked, truncated triangles, each smaller than the one above it. The triangles are outlined with thick, black lines, creating a layered effect.

K031378

Fresenius Medical Care

Fresenius Ultraflux Single Use Hemodialyzer "Special" 510(k) Premarket Notification

Summary of Safety and Effectiveness

D. Indications for Use:

The Fresenius Ultraflux dialyzer is designed for single use acute and chronic hemodialysis.

E. Substantial Equivalence:

1. Is the product a device?

YES - The Ultraflux dialyzers are devices pursuant to 21 CFR §201 [321] (h).

2. Does the new device have the same intended use?

YES - The intended use for the Ultraflux is identical to that for the Optiflux 200NR and is as follows:

Intended Use for Ultraflux

The Fresenius Ultraflux dialyzer is designed for single use acute and chronic hemodialysis.

Intended Use for Optiflux 200NR

Optiflux 200NR dialyzers are designed for single use acute and chronic hemodialysis.

3. Does the device have technological characteristics that raise new types of safety or effectiveness questions?

NO - The Ultraflux are equivalent in materials, sterilization and indications for use to the currently manufactured Optiflux 200NR. The technological characteristics of the Ultraflux are equivalent to those of the Optiflux 200NR and Gambro Polyflux 17S and raise no new types of safety or effectiveness questions.

4. Does descriptive or performance information demonstrate equivalence?

YES - Fresenius Medical Care North America believes that the information provided in this submission clearly describes the Ultraflux and demonstrates that it is substantially equivalent to the Optiflux 200NR.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the text "Page 3/3" in a handwritten style. The word "Page" is written in cursive, followed by the numbers "3/3". This indicates that it is the last page of a document. The text is in black ink on a white background.

Image /page/2/Picture/1 description: The image shows the logo for Fresenius Medical Care. The logo consists of a stylized triangle shape with horizontal lines inside, positioned above the text "Fresenius Medical Care". The text is in a simple, sans-serif font and is left-aligned.

Ko 3 1378

Fresenius Ultraflux Single Use Hemodialyzer "Special" 510(k) Premarket Notification

Summary of Safety and Effectiveness

F. Safety Summary

The Ultraflux hemodialyzers are substantially equivalent in construction, design, materials, and intended use to the commercially available Fresenius Optiflux hemodialyzers. In addition, testing of Ultraflux hemodialyzers indicates that the dialyzers are safe and effective for their intended use.

In vitro testing was performed to determine the following: priming volume, bloodside and dialysate-side pressure drops, ultrafiltration coefficient, urea clearance, creatinine clearance, Vitamin B12 clearance, phosphate and lysozyme clearance. No clinical testing was performed.

G. General Safety and Effectiveness Concerns

The device labeling contains a Package Insert, which includes indications for use, cautions and warnings, as well as the general operating instructions required br proper use of the device. This information promotes safe and effective use of the dialyzer.

Arthur Eilinfeld

ur Eilinsfeld Director of Regulatory Affairs

4/29/03
Date

Date

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three curved lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle image.

MAY 3 0 2003

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Nichole Riek Regulatory Affairs Supervisor Fresenius Medical Care North America Two Ledgemont Center 95 Hayden Avenue LEXINGTON MA 02173

Re: K031378

Trade/Device Name: Fresenius Ultraflux Single Use Hemodialyzer Regulation Number: 21 CFR §876.5860 Regulation Name: High permeability hemodialysis system Regulatory Class: II Product Code: 78 KDI Dated: April 29, 2003 Received: May 1, 2003

Dear Ms. Riek:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If vour device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

8xx.1xxx(301) 594-4591
876.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4616
884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx(301) 594-4616
892.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4654
Other(301) 594-4692

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy Crogdon

Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Fresenius Medical Care. The logo consists of an inverted pyramid shape made up of several horizontal lines, with the company name written below it. The text "Fresenius Medical Care" is written in a simple, sans-serif font. The logo is black and white.

Fresenius Ultraflux Single Use Hemodialyzer "Special" 510(k) Premarket Notification

Indications for Use Statement

Device Name:

Fresenius Ultraflux Hemodialyzers

Indications for Use:

The Fresenius Ultraflux dialyzer is designed for single use acute and chronic hemodialysis.

PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter Use

David W. Sezmore

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devi 510(k) Number

0000001

§ 876.5860 High permeability hemodialysis system.

(a)
Identification. A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K
uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.).
(3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
(b)
Classification. Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”