K Number
K031158
Manufacturer
Date Cleared
2003-08-07

(118 days)

Product Code
Regulation Number
892.5700
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Swift is a software application intended for use with Brachytherapy procedures, i.e. intercavitary, interstitial, intraluminal, involving afterloading radioactive sources.

Swift is a software application for Brachytherapy Treatment Planning, for the treatment of cancer, i.e. intercavitary, interstitial, intraluminal, involving radioactive sources.

Device Description

Swift is a "real time" treatment planning system for brachytherapy especially meant for treatment of cancer in the prostate. Direct 3D Ultrasound imaging of the treatment gives the physician the possibility to update the planning of the dwell positions of the source in the prostate of the patient. The program provides a variety of plan evaluation tools to assist in generating the most optimal dose distribution, i.e. dose volume histograms, dose verification points and dose profiles. The software program provides the physician with anatomical and dosimetric information, to determine the positioning and loading of the radioactive sources, prior to radiation treatment. The program also provides the treatment time and dose distribution for the specified loading. From this the information the patient can be treated with the dwell positions of the source. The modified device is an accessory to an afterloader. The program consists of two modules: Treatment Planning Software: Swift and Database: Smoothbase. The software runs on a Windows XP platform.

AI/ML Overview

It sounds like you're looking for a detailed breakdown of the acceptance criteria and study data for the Nucletron Swift 2.0 device based on the provided text. However, the provided text does not contain the specific information you're asking for regarding acceptance criteria, device performance, study details, and ground truth establishment.

The document is a 510(k) summary for a medical device (Nucletron Swift 2.0), which primarily focuses on:

  • Device Identification: Trade name, common name, classification, manufacturer.
  • Predicate Device: Identification of the legally marketed device it's substantially equivalent to.
  • Description of Modifications: New functionalities added to the previously cleared device (e.g., Encoder, 3D Ultrasound acquisition, 3D VOI contouring, 3D Catheter recognition, Inverse Plan).
  • Intended Use: The same as the predicate device, for brachytherapy treatment planning.
  • Substantial Equivalence Statement: A declaration that it's equivalent to the predicate.
  • FDA Clearance Letter: Formal approval from the FDA.
  • Indications for Use Statement: The approved medical use for the device.

There is no information in the provided text about:

  1. Acceptance criteria
  2. Reported device performance metrics
  3. Sample size for test sets or training sets
  4. Data provenance (country, retrospective/prospective)
  5. Number or qualifications of experts
  6. Adjudication method
  7. MRMC comparative effectiveness studies (or effect size)
  8. Standalone (algorithm only) performance studies
  9. Type of ground truth used (pathology, expert consensus, outcomes data)
  10. How ground truth was established for training or test sets

This type of detailed performance data is typically found in the full 510(k) submission, specifically in sections related to performance testing, clinical data, or software validation, which are not included in this summary document.

To answer your request, you would need access to the full 510(k) submission documents for K031158 (Nucletron Swift 2.0) and potentially K022739 (Real Time HDR), as the Swift 2.0 is an updated version of the latter.

{0}------------------------------------------------

Page 36

Special 510(k) Nucletron Swift 2.0 April 2003

K031158
Page 1 of 2

etron

NUCLETRON B.V.

Waardgelder 1 3905 TH Veenendaal P.O.Box 930 3900 AX Veenendaal The Netherlands Phone +31 318 557133 Fax +31 318 550485

Department of Health and Human Services Centre of Device and Radiological Health Office of Device Evaluation Special 510(k) section

510(K) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION as required by section 807.92(c)

Submitter of 510(k):

Nucletron Corporation Company name: 1121753 Registration number: 7080 Columbia Gateway Drive Address: Columbia, MD 21046-2133 410-312-4100 Phone: 410-312-4197 Fax: Lisa Dimmick Correspondent: Director Assurance & Regulatory Affairs

Modified Device Name:

Trade/Proprietary Name:Swift 2.0
Common/Usual Name:Radiation Therapy Planning System
Classification Name:Accessory to Radiotherapy Device
Classification:21Cfr892.5050 Class II

Legally Marketed Device(s)

Our device is substantially equivalent to the legally marketed predicate device cited in the table below:

ManufacturerDevice510(k) #
Nucletron BVReal Time HDRK022739

Description:

Swift is a "real time" treatment planning system for brachytherapy especially meant for treatment of cancer in the prostate. Direct 3D Ultrasound imaging of the treatment gives the physician the possibility to update the planning of the dwell positions of the source in the prostate of the patient.

{1}------------------------------------------------

Special 510(k) Nucletron Swift 2.0 April 2003

The program provides a variety of plan evaluation tools to assist in generating the most optimal dose distribution, i.e. dose volume histograms, dose verification points and dose profiles. The software program provides the physician with anatomical and dosimetric information, to determine the positioning and loading of the radioactive sources, prior to radiation treatment. The program also provides the treatment time and dose distribution for the specified loading. From this the information the patient can be treated with the dwell positions of the source.

Modifications to the previously cleared device K022739, have been made to add functionality for:

  • Encoder ●
  • 3D Ultrasound acquisition .
  • 3D VOI contouring ●
  • 3D Catheter recognition ●
  • Inverse Plan ●

The modified device is an accessory to an afterloader. The program consists of two modules:

  • Treatment Planning Software: Swift ●
  • Database: Smoothbase .

The software runs on a Windows XP platform.

Intended use:

The modified device has the same intended use as the legally marketed predicate device cited:

Swift is a software application intended for use with Brachytherapy procedures, i.e. intercavitary, interstitial, intraluminal, involving afterloading radioactive sources.

Summary of technological considerations:

Swift is substantially equivalent to the cleared predicate device. Real Time HDR. 510(k)#: K022739.

Krieken

2-4-2003


Date

Name: Frits van Krieken Title: Business Segment Manager Nucletron B.V. Veenendaal, The Netherlands

Page 37

KOB1158?
Page 2 of 2

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo. The logo is circular and contains an image of a bird in the center. The bird is stylized with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES" is arranged around the top and left side of the circle.

ood and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG - 7 2003

Ms. Lisa Cole Dimmick Director of Regulatory Affairs Nucletron Corporation 8671 Robert Fulton Drive COLUMBIA MD 21046

Re: K031158

Trade/Device Name: Swift 2.0 Regulation Number: 21 CFR 892.5700 Regulation Name: Remote controlled radionuclide applicator system Regulatory Class: II Product Code: 90 JAQ Dated: July 14, 2003 Received: July 16, 2003

Dear Ms. Dimmick:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

8xx.1xxx(301) 594-4591
876.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4616
884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx(301) 594-4616
892.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4654
Other(301) 594-4692

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours.

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Special 510(k) Nucletron Swift 2.0 April 2003

Indications for Use Statement

510(k) Number

Device Name

K031158

Swift 2.0

Indications for Use

Swift is a software application for Brachytherapy Treatment Planning, for the treatment of cancer, i.e. intercavitary, interstitial, intraluminal, involving radioactive sources.

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109) \

OR

Over-The-Counter Use_

Daniel A. Lezerm

(Division Sign-Off) (Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number .

§ 892.5700 Remote controlled radionuclide applicator system.

(a)
Identification. A remote controlled radionuclide applicator system is an electromechanical or pneumatic device intended to enable an operator to apply, by remote control, a radionuclide source into the body or to the surface of the body for radiation therapy. This generic type of device may include patient and equipment supports, component parts, treatment planning computer programs, and accessories.(b)
Classification. Class II.