(49 days)
The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic Gram-negative and Gram-positive bacteria of human origin.
The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus and Enterococcus.
This premarket notification is for the addition of the antimicrobial agent trimethoprim at concentrations of 0.5-16 ug/mL to Gram-negative ID/AST or AST only Phoenix panels. trimethoprim has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.
Active In Vitro and in Clinical Infections Against:
Escherichia coli
Proteus mirabilis
Klebsiella pneumoniae
Enterobacter species
Active In Vitro Against:
Common urinary tract pathogens with the exception of Pseudomonas aeruginosa
The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:
- BD Phoenix instrument and software. ●
- . BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents for AST determinations.
- BD Phoenix ID Broth used for performing ID.tests and preparing AST Broth inoculum. .
- BD Phoenix AST Broth used for performing AST tests only.
- . BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth determination.
The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. For each isolate, an inoculation equivalent to a 0.5 McFarland standard is prepared in Phoenix ID Broth.
The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.
The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, or R (sensitive, intermediate, or resistant).
The BD Phoenix™ Automated Microbiology System is designed for rapid identification and antimicrobial susceptibility testing (AST) of bacterial isolates. This summary focuses on its performance with the antimicrobial agent Trimethoprim (0.5-16 µg/mL) for Gram-negative organisms.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for the BD Phoenix System, specifically for Trimethoprim against Gram-negative organisms, are not explicitly stated as numerical targets within the provided text (e.g., "EA must be >90%"). However, the study aims to demonstrate substantial equivalence to the NCCLS reference broth microdilution method. The performance metrics used are Essential Agreement (EA) and Category Agreement (CA). Based on the results provided, the de facto acceptance criteria appear to be:
- Essential Agreement (EA) for Gram-negative Organisms: High percentage (demonstrated as 90.5%).
- Category Agreement (CA) for Gram-negative Organisms: High percentage (demonstrated as 98.7%).
- Site Reproducibility: Overall intra-site reproducibility >90% and inter-site reproducibility >95% for Gram-negative isolates.
| Metric | Acceptance Criteria (Implied) | Reported Device Performance (Trimethoprim, Gram-negative) |
|---|---|---|
| Essential Agreement (EA) | High Agreement | 90.5% (n=1856) |
| Category Agreement (CA) | High Agreement | 98.7% (n=1856) |
| Overall Intra-site Reproducibility | >90% | >90% |
| Overall Inter-site Reproducibility | >95% | >95% |
2. Sample Size and Data Provenance
- Test Set Sample Size:
- Clinical Study (Performance): 1856 isolates for the performance metrics (EA and CA).
- Site Reproducibility: Not explicitly stated as a single number, but a "panel of Gram-negative isolates" was tested in triplicate on three different days at three sites, using one lot of panels.
- Data Provenance: The clinical studies were conducted across multiple geographically diverse sites across the United States. The data is prospective in nature, comparing the Phoenix System results to a reference method.
3. Number of Experts and Qualifications for Ground Truth
The document does not explicitly state the number or specific qualifications of experts used to establish the ground truth. However, the ground truth for the clinical isolates was established using the NCCLS reference broth microdilution method. This method is a standardized laboratory procedure, performed by trained laboratory personnel, rather than relying on individual expert interpretation of images or observations.
4. Adjudication Method for the Test Set
The document does not describe an adjudication method in the context of multiple readers or experts reviewing results. The comparison is made between the BD Phoenix System's automated results and the results from the NCCLS reference method. Discrepancies would likely be investigated through re-testing or confirmation by standard laboratory protocols rather than an adjudication panel.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No multi-reader multi-case (MRMC) comparative effectiveness study was mentioned. The device is an automated system, and the study focuses on the agreement of its automated output with a reference standard, not on how human readers' performance might change with or without AI assistance.
6. Standalone Performance
Yes, a standalone performance study was conducted. The "Clinical Studies" section directly assesses the BD Phoenix System's (algorithm only, with its automated readings) performance by comparing its Essential Agreement (EA) and Category Agreement (CA) to the NCCLS reference broth microdilution method. This evaluates the algorithm's performance without direct human intervention in interpreting the final susceptibility results after the automated reading.
7. Type of Ground Truth Used
The ground truth used for the clinical isolates was the NCCLS reference broth microdilution method. For the "Challenge set isolates," the Phoenix System results were compared to "expected results," which are typically pre-determined and validated results for known strains.
8. Sample Size for the Training Set
The document does not explicitly mention a separate training set or its sample size. The description focuses on the evaluation of the device against reference methods. It's plausible that the underlying algorithms were developed and trained using various data, but this information is not provided in this specific submission summary.
9. How Ground Truth for the Training Set Was Established
Since a training set is not explicitly mentioned, the method for establishing its ground truth is also not provided. The performance evaluation relies on comparing the device's output to established reference methods for clinical and challenge isolates.
{0}------------------------------------------------
510(K) SUMMARY
| SUBMITTED BY: | Becton, Dickinson and Company7 Loveton CircleSparks, MD 21152Phone: (410) 316-4287Fax: 410-316-4499MAY 1 6 2003 | ||
|---|---|---|---|
| CONTACT NAME: | Monica Evelyn GiguereRegulatory Affairs Specialist | ||
| DATE PREPARED: | March 26, 2003 | ||
| DEVICE TRADE NAME: | BD Phoenix™ Automated Microbiology System –Trimethoprim 0.5-16 µg/mL | ||
| DEVICE COMMON NAME: | Antimicrobial susceptibility test system-short incubation | ||
| DEVICE CLASSIFICATION: | Fully Automated Short-Term Incubation Cycle AntimicrobialSusceptibility Device, 21 CFR 866.1645 | ||
| PREDICATE DEVICES: | VITEK® System (PMA No. N50510) and BD Phoenix™Automated Microbiology System with Gatifloxacin (K020321,May 23, 2002), Ofloxacin (K020323, April 14, 2002), andLevofloxacin (K020322, March 27, 2002). | ||
| INTENDED USE: | The BD Phoenix™ Automated Microbiology System isintended for the rapid identification and in vitro antimicrobialsusceptibility testing of isolates from pure culture of mostaerobic and facultative anaerobic Gram-negative and Gram-positive bacteria of human origin. |
DEVICE DESCRIPTION:
The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:
- BD Phoenix instrument and software. ●
- . BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents for AST determinations.
- BD Phoenix ID Broth used for performing ID.tests and preparing AST Broth inoculum. .
- BD Phoenix AST Broth used for performing AST tests only.
- . BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth determination.
{1}------------------------------------------------
The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. For each isolate, an inoculation equivalent to a 0.5 McFarland standard is prepared in Phoenix ID Broth.
The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.
The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, or R (sensitive, intermediate, or resistant).
DEVICE COMPARISON:
The BD Phoenix™ Automated Microbiology System demonstrated substantially equivalent performance when compared with the NCCLS reference broth microdilution method. This premarket notification provides data supporting the use of the BD Phoenix™ Automated Microbiology System Gram-negative ID/AST or AST only Phoenix panels with this antimicrobial agent.
SUMMARY OF SUBSTANTIAL EQUIVALENCE TESTING:
The BD Phoenix™ Automated Microbiology System has demonstrated substantially equivalent performance when compared to the NCCLS reference broth microdilution method (AST panels prepared according to NCCLS M7). The system has been evaluated as defined in the FDA Draft guidance document, "Guidance on Review Criteria for Assessment of Antimicrobial Susceptibility Devices", March 8, 2000.
Site Reproducibility
Intra- and inter-site reproducibility of this antimicrobial agent in the BD Phoenix System was evaluated at three sites using a panel of Gram-negative isolates. Each site tested the isolates in triplicate on three different days using one lot of Gram-negative Phoenix panels containing this antimicrobial agent and associated reagents.
The results of the study demonstrate for the this antimicrobial agent there was an overall intra-site reproducibility of greater than 90% and an overall inter-site reproducibility greater than 95% for the Gram-negative isolates tested
{2}------------------------------------------------
Clinical Studies
Clinical, stock and challenge isolates were tested across multiple geographically diverse sites across the United States to demonstrate the performance of the Phoenix antimicrobial susceptibility test with the Gram-negative Phoenix panel format containing this antimicrobial agent. Phoenix System results for Challenge set isolates were compared to the expected results. Phoenix System results for clinical isolates were compared to the results obtained from the NCCLS reference broth microdilution method.
The performance of the Phoenix System was assessed by calculating Essential Agreement (EA) and Category Agreement (CA) to expected/reference results for all isolates tested. Essential Agreement (EA) occurs when the BD Phoenix™ Automated Microbiology System agrees exactly or within ± one two-fold dilution to the reference result. Category Agreement (CA) occurs when the BD Phoenix™ Automated Microbiology System agrees with the reference method with respect to the FDA categorical interpretive criteria (susceptible, intermediate, and resistant).
Table 1 summarizes the performance for the isolates tested in this study.
Table 1: Performance of BD Phoenix System for Gram-negative Organisms by Drug
| Carreland Control Concession of ChildrenThe first of the county of the county of the first of the first of the first of the first of the first of the first of the first and Antimicrobialand the many of the many of the same of the same of the same of the distance of the summer of the summer of the summer of the summer of the summer of the summer of the summer | Carrely ControllerConcentration | Carrely Concession of Carder of Canadian Company of' EA (n)Acres of the later in and in the count of | EA (%)------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | CA (n)Classic and C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. C. | " CA (%) |
|---|---|---|---|---|---|
| Trimethoprim | 0.5-16 ug/mL | 1856 | 1 05 €1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 1 | 1856 | , 98.7 |
Conclusions Drawn from Substantial Equivalence Studies
The data collected from the substantial equivalence studies demonstrate that testing on the BD Phoenix™ Automated Microbiology System with this antimicrobial agent is substantially equivalent as outlined in the FDA draft guidance document, "Guidance on Review Criteria for Assessment of Antimicrobial Susceptibility Devices", March 8, 2000. Technological characteristics of this system are substantially equivalent to those used in the VITEK® system, which received approval by the FDA under PMA number N50510 and BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, May 23, 2002), Ofloxacin (K020323, April 14, 2002), and Levofloxacin (K020322, March 27, 2002).
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines. The overall design is simple and conveys a sense of government authority and public service.
Public Health Service
MAY 1 6 2003
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Monica Evelyn Giguere Regulatory Affairs Specialist BD Diagnostics Systems Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152
K030986 Trade/Device Name: BD Phoenix™ Automated Microbiology System Trimethoprim (0.5-16 ug/ml) Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Devices Regulatory Class: Class II Product Code: LON Dated: March 26, 2003 Received: March 28, 2003
Dear Ms. Giguere:
Re:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{4}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page 1 of 1
510(k) Number: _____________
Device Name: BD Phoenix™ Automated Microbiology System for use with the antimicrobial agent trimethoprim (0.5-16 ug/mL) - Gram-negative ID/AST or AST only Phoenix panels.
Indications for Use:
The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus and Enterococcus.
This premarket notification is for the addition of the antimicrobial agent trimethoprim at concentrations of 0.5-16 ug/mL to Gram-negative ID/AST or AST only Phoenix panels. trimethoprim has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.
Active In Vitro and in Clinical Infections Against:
Escherichia coli Proteus mirabilis Klebsiella pneumoniae Enterobacter species
Active In Vitro Against:
Common urinary tract pathogens with the exception of Pseudomonas aeruginosa
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Division Sign-Off | |
|---|---|
| ------------------- | -- |
| Prescription Use (Per 21 CFR 801.109) | Office of In Vitro Diagnostic Device Evaluation and Safety |
|---|---|
| --------------------------------------- | ------------------------------------------------------------ |
| 510(k) | K030986 |
|---|---|
| -------- | --------- |
Over-the-Counter UseBD Diagnostic Systems Becton, Dickinson and Company
§ 866.1645 Fully automated short-term incubation cycle antimicrobial susceptibility system.
(a)
Identification. A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.(b)
Classification. Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”