K Number
K030728
Device Name
FYRELINK VL
Manufacturer
Date Cleared
2003-05-23

(77 days)

Product Code
Regulation Number
892.2050
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP Authorized
Intended Use
This device is intended for use when there is a need to convert endoscopic and microscopic surgical video to a digital video format for subsequent review and archiving. This device will also transfer the digital video over digital networks and/or digital communication lines.
Device Description
fyreLINK VL (formerly echoLINK) now provides the endoscopy and surgical market a real-time solution for dealing with video data in a complete image management system. There are three major components that make up the fyreLINK VL product line; an image acquisition unit, a review station (either the fyreLINK station or a third party product), and a DVD library system or other archive device. The telemedicine system uses the same components except for the DVD library system. Appendix D shows an example of how an image acquisition unit can be arranged to provide a particular customer solution. The fyreLINK VL Image Acquisition Unit (IAU) is the key component of the fyreLINK VL device. It uses a hardware video converter to convert analog video coming from a surgical video system to streaming digital video in real time. This streaming digital video is then compressed using MPEG 2 technology and stored on its local hard drive as a MPEG2 file. The IAU can store up to 1000 minutes of digital video data on its internal hard disk. The data is available for immediate review on the IAU, or can be exported to another application or another workstation or archive. Each acquired recording is an individual file. Exams can consist of a single continuous recording of the entire procedure, or a collection of shorter recordings of areas of interest. Still images may also be captured and saved as bitmap or JPEG files. The IAU is built on a Windows 2000 platform, and is network-ready for transferring patient data to a local or central archive system and/or to a physician's review station. The IAU's compact size allows it to be a direct replacement for existing VCR's used in endoscopic and microscopic surgical procedures. The IAU platform can consist of either a laptop or desktop type computer. The device has an electronic report generator that will allow the user to enter exam information as part of a paperless reporting system. The IAU, when configured as part of a WAN environment, can be a part of a remote access telemedicine system. The IAU is controlled by MPACS proprietary software. fyreLINK VL offers two review station solutions. Acquired images can be reviewed at the IAU, or the image files can be exported for use with another application. THE REVIEW FUNCTION OF FYRELINK VL SOFTWARE LABELS ALL RECALLED IMAGES WITH THE TYPE OF COMPRESSION USED AND THE COMPRESSION RATIO. fyreLINK VL used in a network environment can integrate with a mass archive solution already in place. As an alternative, the IAU includes a DVD recorder built in, to permit immediate local archival of endoscopic or microscopic surgical procedures. MPACS can supply network solutions for integrating the fyreLINK VL components. In a network configuration, the image acquisition unit moves patient studies directly to a server. The networked review station is able to review both on-line and near-line studies. Network solutions can also include telecommunication links such as T1, DSL, and cable modem connections for telemedicine.
More Information

Not Found

No
The description focuses on video conversion, compression, storage, and network transfer. There is no mention of AI or ML algorithms for analysis, interpretation, or decision support.

No
The device is described as converting video data for review and archiving, not for directly treating or diagnosing patients.

No

The device is described as a system for converting and managing digital video from endoscopic and microscopic surgical procedures for review and archiving purposes. Its primary function is data acquisition, conversion, storage, and networking, not the diagnosis of a medical condition.

No

The device description explicitly states that the fyreLINK VL product line includes an "image acquisition unit" which is described as a "hardware video converter" and can be built on a "laptop or desktop type computer." It also mentions a "DVD library system or other archive device" and a "DVD recorder built in." These are all hardware components, indicating the device is not software-only.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use is clearly stated as converting endoscopic and microscopic surgical video to a digital format for review and archiving, and transferring this digital video over networks. This is related to image management and data handling, not the analysis of biological samples or the diagnosis of disease based on in vitro testing.
  • Device Description: The description focuses on the hardware and software components for video acquisition, conversion, compression, storage, and transfer. It describes a system for managing surgical video data.
  • Lack of IVD Characteristics: There is no mention of analyzing biological samples, detecting analytes, or providing diagnostic information based on in vitro tests. The device handles video data from surgical procedures, which is distinct from in vitro diagnostics.
  • Predicate Device: The listed predicate device (K011944; Dyonics Vision 635 Digital Capture System) is also a digital capture system for surgical video, further supporting the classification of this device as a surgical video management system, not an IVD.

Therefore, the fyreLINK VL device, as described, falls under the category of medical image management and archiving systems, not In Vitro Diagnostics.

N/A

Intended Use / Indications for Use

This device is intended for use when there is a need to convert endoscopic and microscopic surgical video to a digital video format for subsequent review and archiving. This device will also transfer the digital video over digital networks and/or digital communication lines.

Product codes

LLZ

Device Description

fyreLINK VL (formerly echoLINK) now provides the endoscopy and surgical market a real-time solution for dealing with video data in a complete image management system. There are three major components that make up the fyreLINK VL product line; an image acquisition unit, a review station (either the fyreLINK station or a third party product), and a DVD library system or other archive device. The telemedicine system uses the same components except for the DVD library system. Appendix D shows an example of how an image acquisition unit can be arranged to provide a particular customer solution.

The fyreLINK VL Image Acquisition Unit (IAU) is the key component of the fyreLINK VL device. It uses a hardware video converter to convert analog video coming from a surgical video system to streaming digital video in real time. This streaming digital video is then compressed using MPEG 2 technology and stored on its local hard drive as a MPEG2 file. The IAU can store up to 1000 minutes of digital video data on its internal hard disk. The data is available for immediate review on the IAU, or can be exported to another application or another workstation or archive. Each acquired recording is an individual file. Exams can consist of a single continuous recording of the entire procedure, or a collection of shorter recordings of areas of interest. Still images may also be captured and saved as bitmap or JPEG files. The IAU is built on a Windows 2000 platform, and is network-ready for transferring patient data to a local or central archive system and/or to a physician's review station. The IAU's compact size allows it to be a direct replacement for existing VCR's used in endoscopic and microscopic surgical procedures. The IAU platform can consist of either a laptop or desktop type computer. The device has an electronic report generator that will allow the user to enter exam information as part of a paperless reporting system. The IAU, when configured as part of a WAN environment, can be a part of a remote access telemedicine system. The IAU is controlled by MPACS proprietary software.

fyreLINK VL offers two review station solutions. Acquired images can be reviewed at the IAU, or the image files can be exported for use with another application. THE REVIEW FUNCTION OF FYRELINK VL SOFTWARE LABELS ALL RECALLED IMAGES WITH THE TYPE OF COMPRESSION USED AND THE COMPRESSION RATIO.

fyreLINK VL used in a network environment can integrate with a mass archive solution already in place. As an alternative, the IAU includes a DVD recorder built in, to permit immediate local archival of endoscopic or microscopic surgical procedures.

MPACS can supply network solutions for integrating the fyreLINK VL components. In a network configuration, the image acquisition unit moves patient studies directly to a server. The networked review station is able to review both on-line and near-line studies. Network solutions can also include telecommunication links such as T1, DSL, and cable modem connections for telemedicine.

Mentions image processing

Yes

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Video

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Not Found

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K011944

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

0

MPACS

MAY 2 3 2003

Ko30728

7601 Ganser Way Madison, WI 53719 (608) 827-7111

510(K) SUMMARY (As Required by 21 CFR 807.92(c))

A. Submitter's Name and Address

MPACS, LLC 7601 Ganser Way Madison, WI 53719

Phone: (608) 827-7111 Fax:

B. Contact Person

Greg Sopotnick Phone: (608) 827-7111 Fax: (608) 827-0162

C. Date of Submission: March 6, 2003

D. Device Name

D.1. Device Trade or Proprietary name: fyreLINK VL™

D.2. Device Common or Usual Name: Picture Archiving and Communications Systems (PACS)

D.3. Classifications: Image Processing System

D.4. Product Code: LLZ

D.5. Class: Class II

D.6. Classification Panel: Radiology

1

E. Equivalent Device

The equivalent legally marketed device is the Dyonics Vision 635 Digital Capture System (K011944) from Smith & Nephew, Inc., Endoscopy Division.

F. Device Description

fyreLINK VL (formerly echoLINK) now provides the endoscopy and surgical market a real-time solution for dealing with video data in a complete image management system. There are three major components that make up the fyreLINK VL product line; an image acquisition unit, a review station (either the fyreLINK station or a third party product), and a DVD library system or other archive device. The telemedicine system uses the same components except for the DVD library system. Appendix D shows an example of how an image acquisition unit can be arranged to provide a particular customer solution.

The fyreLINK VL Image Acquisition Unit (IAU) is the key component of the fyreLINK VL device. It uses a hardware video converter to convert analog video coming from a surgical video system to streaming digital video in real time. This streaming digital video is then compressed using MPEG 2 technology and stored on its local hard drive as a MPEG2 file. The IAU can store up to 1000 minutes of digital video data on its internal hard disk. The data is available for immediate review on the IAU, or can be exported to another application or another workstation or archive. Each acquired recording is an individual file. Exams can consist of a single continuous recording of the entire procedure, or a collection of shorter recordings of areas of interest. Still images may also be captured and saved as bitmap or JPEG files. The IAU is built on a Windows 2000 platform, and is network-ready for transferring patient data to a local or central archive system and/or to a physician's review station. The IAU's compact size allows it to be a direct replacement for existing VCR's used in endoscopic and microscopic surgical procedures. The IAU platform can consist of either a laptop or desktop type computer. The device has an electronic report generator that will allow the user to enter exam information as part of a paperless reporting system. The IAU, when configured as part of a WAN environment, can be a part of a remote access telemedicine system. The IAU is controlled by MPACS proprietary software.

fyreLINK VL offers two review station solutions. Acquired images can be reviewed at the IAU, or the image files can be exported for use with another application. THE REVIEW FUNCTION OF FYRELINK VL SOFTWARE LABELS ALL RECALLED IMAGES WITH THE TYPE OF COMPRESSION USED AND THE COMPRESSION RATIO.

fyreLINK VL used in a network environment can integrate with a mass archive solution already in place. As an alternative, the IAU includes a DVD recorder built in, to permit immediate local archival of endoscopic or microscopic surgical procedures.

2

MPACS can supply network solutions for integrating the fyreLINK VL components. In a network configuration, the image acquisition unit moves patient studies directly to a server. The networked review station is able to review both on-line and near-line studies. Network solutions can also include telecommunication links such as T1, DSL, and cable modem connections for telemedicine.

G. Intended Use

This device is intended for use when there is a need to convert endoscopic and microscopic surgical video to a digital video format for subsequent review and archiving. This device will also transfer the digital video over digital networks and/or digital communication lines.

H. Substantial Equivalency Comparison

fyreLINK VL and Dyonics Vision 635 Digital Capture System are technologically equivalent in their capture of video images. Both convert an analog video signal to digital data, and then use the MPEG compression standard to create a file for storage and playback. In addition, both devices are capable of exporting data files in DICOM format, and both devices also capture still images in bitmap and JPEG format.

Data files on both devices are initially stored to a large internal hard drive. Both devices also can export to a network storage device via an Ethernet connection.

I. Conclusions

fyreLINK VL's performance is the same as the equivalent marketed device. Both devices are required to acquire or digitize full-motion (30 frames per second) analog video in real time. Likewise, both devices can playback the stored digital video in realtime.

Based on the intended use and the comparisons between the fyreLINK VL device and the legally marketed device, there are all the indications that the fyreLINK VL device is substantially equivalent to the Dyonics Vision 635 Digital Capture System device.

(End of 510(k) Summary)

3

Image /page/3/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the department's name around the perimeter. In the center of the seal is an abstract image of an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 2 3 2003

Mr. Greg Sopotnick OA Manager MPACS, LLC 7601 Ganser Way MADISON WI 53719 Re: K030728

Trade/Device Name: fyreLINK VL Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II

Product Code: 90 LLZ Dated: March 6, 2003 Received: March 7, 2003

Dear Mr. Sopotnick:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

4

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

8xx. 1xxx(301) 594-4591
876.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4616
884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx(301) 594-4616
892.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4654
Other(301) 594-4692

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

5

Page 1 of

ڊيسٽمبر

510(k) Number (if known):

fyreLINK VL Device Name:

Indications for Use:

This device is intended for use when there is a need to convert endoscopic and microscopic surgical video to a digital video format for subsequent review and archiving. This device will also transfer the digital video over digital networks and/or digital communication lines.