K Number
K030677
Date Cleared
2003-04-01

(28 days)

Product Code
Regulation Number
866.1645
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic Gram-negative and Grampositive bacteria of human origin.

The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus and Enterococcus.

This premarket notification is for the addition of the antimicrobial agent vancomycin at concentrations of 0.5-32 µg/mL to Gram Positive ID/AST or AST only Phoenix panels. Vancomycin has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.

Active In Vitro and in Clinical Infections Against:

Staphylococcus species
Enterococcus species

Device Description

The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:

  • BD Phoenix instrument and software.
  • BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents or AST determinations.
  • BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum.
  • BD Phoenix AST Broth used for performing AST tests only.
  • BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth determination.

The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. For each isolation equivalent to a 0.5 McFarland standard is prepared in Phoenix ID broth.

The Phoenix AST method is a broth based microdilution test. The Phoenix system utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.

The instrument houses the panels where they are continuously incubated at a nominal temperature of 35℃. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, or R (sensitive, intermediate, or resistant).

AI/ML Overview

Here's an analysis of the provided text, focusing on the acceptance criteria and study details for the BD Phoenix™ Automated Microbiology System for Vancomycin:

The document describes the performance evaluation of the BD Phoenix™ Automated Microbiology System when testing the antimicrobial agent Vancomycin against Gram-positive organisms.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implied by the comparison to the NCCLS reference broth microdilution method and the FDA's "Guidance on Review Criteria for Assessment of Antimicrobial Susceptibility Devices." Specifically, Essential Agreement (EA) and Category Agreement (CA) are the key metrics. While explicit numeric targets for "acceptance criteria" are not given in the provided text as "acceptance criteria," the study results demonstrate performance relative to these criteria.

MetricAcceptance Criteria (Implied/Standard for AST Devices)Reported Device Performance (BD Phoenix™ for Vancomycin)
Overall Essential Agreement (EA)Typically > 90% (Industry Standard/FDA Guidance Expectation)97.7% (with n=1981)
Overall Category Agreement (CA)Typically > 90% (Industry Standard/FDA Guidance Expectation)99.4% (with n=1981)
Intra-site Reproducibility> 90%> 90%
Inter-site Reproducibility> 95%> 95%

Note: The exact numerical acceptance thresholds for EA and CA are not explicitly stated as "acceptance criteria" in the text but are values against which the device performance is measured in substantial equivalence studies for antimicrobial susceptibility testing. The achieved values (97.7% EA, 99.4% CA) are generally considered very good and acceptable for such devices.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: 1981 isolates (as indicated by 'n=1981' for both EA and CA).
  • Data Provenance: Clinical, stock, and challenge isolates were tested. The testing was conducted across multiple geographically diverse sites across the United States. The study is prospective in nature as it involved collecting and testing isolates specifically for this evaluation.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document does not specify the number of experts or their qualifications for establishing the ground truth. It states that:

  • "Phoenix System results for Challenge set isolates were compared to the expected results."
  • "Phoenix System results for clinical isolates were compared to the results obtained from the NCCLS reference broth microdilution method."

The "NCCLS reference broth microdilution method" is the gold standard for antimicrobial susceptibility testing, and its results are considered the ground truth. While performing this method requires expert technicians following standardized protocols, the document doesn't detail external expert review or consensus for the ground truth itself, as the method is inherently the "expert" ground truth.

4. Adjudication Method for the Test Set

The document does not describe an explicit adjudication method for discrepancies between the Phoenix System and the NCCLS reference method. The comparison focused on calculating Essential Agreement and Category Agreement based on the direct results. It's common in AST studies to have a predefined method for resolving major discrepancies (e.g., very major errors), but this specific detail is not provided.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, an MRMC comparative effectiveness study was not done. This study focuses on a standalone device performance comparison against a reference method, not on human reader performance with or without AI assistance. The device in question is an automated system for susceptibility testing, where the output is directly interpreted (S, I, R) by the system itself based on observed growth, rather than by human readers interpreting images or complex data fields with AI assistance.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Yes, a standalone study was performed. The entire evaluation described is a standalone performance assessment. The BD Phoenix™ Automated Microbiology System is an automated device designed to autonomously perform susceptibility testing and report results (MIC values and categories of S, I, R) without direct human intervention in the interpretation process once the panel is loaded. Its results were directly compared to the NCCLS reference method.

7. Type of Ground Truth Used

The ground truth used was the NCCLS reference broth microdilution method for clinical isolates and "expected results" for challenge isolates. The NCCLS method is widely recognized as the gold standard for antimicrobial susceptibility testing.

8. Sample Size for the Training Set

The document does not specify a sample size for the training set. Antimicrobial susceptibility testing devices like the BD Phoenix™ are often developed using a large historical dataset for algorithm training and refinement during product development. The current document focuses on the validation study for a specific antimicrobial agent (Vancomycin) addition to an existing system, not the initial development or training phase of the core algorithm.

9. How the Ground Truth for the Training Set Was Established

As the document does not mention the training set size, it also does not explicitly describe how the ground truth for the training set was established. However, based on industry standards for such devices, it is highly probable that the ground truth for any training data (if new training was done for Vancomycin, or for the core system development) would have been established using similar reference methods like the NCCLS broth microdilution method, or other accepted standard methods appropriate for the specific organisms and antimicrobials.

{0}------------------------------------------------

RO 3061
CONFIDENTIAL AND PROPRIETARY

510(k) SUMMARY

410-316-4499

Regulatory Affairs Specialist

Becton, Dickinson and Company

SUBMITTED BY:

CONTACT NAME:

DATE PREPARED:

DEVICE TRADE NAME:

DEVICE COMMON NAME:

DEVICE CLASSIFICATION:

PREDICATE DEVICES:

March 3, 2003

Michelle B. Bandy,

7 Loveton Circle Sparks, MD 21152 Phone: (410) 316-4778

Fax:

BD Phoenix™ Automated Microbiology System -Vancomycin 0.5-32 ug/mL

Antimicrobial susceptibility test system-short incubation

Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Device, 21 CFR 866.1645

VITEK® System (PMA No. N50510) and BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, May 23, 2002), Ofloxacin (K020323, April 14, 2002), and Levofloxacin (K020322, March 27, 2002).

The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic Gram-negative and Grampositive bacteria of human origin.

DEVICE DESCRIPTION:

The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:

  • BD Phoenix instrument and software.
  • BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents or AST determinations.
  • BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum.
  • BD Phoenix AST Broth used for performing AST tests only.
  • BD Phoenix AST Indicator solution added to the AST Broth to aid in bacterial growth determination.

APR 0 1 2003

INTENDED USE:

INTENDED USE:

{1}------------------------------------------------

The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. For each isolation equivalent to a 0.5 McFarland standard is prepared in Phoenix ID broth.

The Phoenix AST method is a broth based microdilution test. The Phoenix system utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.

The instrument houses the panels where they are continuously incubated at a nominal temperature of 35℃. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, or R (sensitive, intermediate, or resistant).

DEVICE COMPARISON:

The BD Phoenix™ Automated Microbiology System demonstrated substantially equivalent performance when compared with the NCCLS reference broth microdilution method. This premarket notification provides data supporting the use of the BD Phoenix™ Automated Microbiology System Gram positive ID/AST or AST only Phoenix panels with this antimicrobial agent.

SUMMARY OF SUBSTANTIAL EQUIVALENCE TESTING:

The BD Phoenix™ Automated Microbiology System has demonstrated substantially equivalent performance when compared to the NCCLS reference broth microdilution method (AST panels prepared according to NCCLS M7). The system has been evaluated as defined in the FDA Draft guidance document, "Guidance on Review Criteria for Assessment of Antimicrobial Susceptibility Devices", March 8, 2000.

Site Reproducibility

Intra- and inter-site reproducibility of this antimicrobial agent in the BD Phoenix System was evaluated at three sites using a panel of Gram-positive isolates. Each site tested the isolates in triplicate on three different days using one lot of Gram Positive Phoenix panels containing this antimicrobial agent and associated reagents.

The results of the study demonstrate for the this antimicrobial agent there was an overall intra-site reproducibility of greater than 90% and an overall inter-site reproducibility greater than 95% for the Gram-positive isolates tested.

{2}------------------------------------------------

Clinical Studies

Clinical, stock and challenge isolates were tested across multiple geographically diverse sites across the United States to demonstrate the performance of the Phoenix antimicrobial susceptibility test with the Gram Positive Phoenix Panel format containing this antimicrobial agent. Phoenix System results for Challenge set isolates were compared to the expected results. Phoenix System results for clinical isolates were compared to the results obtained from the NCCLS reference broth microdilution method.

The performance of the Phoenix System was assessed by calculating Essential Agreement (EA) and Category Agreement (CA) to expected/reference results for all isolates tested. Essential Agreement (EA) occurs when the BD Phoenix™ Automated Microbiology System agrees exactly or within + one two-fold dilution to the reference result. Category Agreement (CA) occurs when the BD PhoenixTM Automated Microbiology System agrees with the reference method with respect to the FDA categorical interpretive criteria (susceptible, intermediate, and resistant).

Table 1 summarizes the performance for the isolates tested in this study.

Performance of BD Phoenix System for Gram-Positive Organisms by Drug Table 1:

Annual Property of Children Comments of Children Comments of Children Comments of Children Comments of ChildrenAntimicrobialConcentration- Research Property of Concession Comments of Children Comments of Children Comments of ChildrenEA (n)And Company of Children Company ofEA (%)and and the result of the commended to the control of theCA (n)And And Children Comments of the ChildrenCA (%)
the first of the Research and the state of the state of the state of the states of the states of the states of the status of the status of the states of the states of the staVancomvcin0.5-32 µg/mL198107 7198199.4

Conclusions Drawn from Substantial Equivalence Studies

The data collected from the substantial equivalence studies demonstrate that testing on the BD Phoenix™ Automated Microbiology System with this antimicrobial agent is substantially equivalent as outlined in the FDA draft guidance document, "Guidance on Review Criteria for Assessment of Antimicrobial Susceptibility Devices", March 8, 2000. Technological characteristics of this system
are substantially equivalent to those used in the VITEK® system, which re FDA under PMA number N50510 and BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, May 23, 2002), Ofloxacin (K020323, April 14, 2002), and Levofloxacin (K020322, March 27, 2002).

Page 12

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)" is arranged in a circular fashion around the image.

Public Health Service

APR 0 1 2003

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Michelle B. Bandy Regulatory Affairs Specialist BD Diagnostics Systems Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152-0999

Re: K030677

Trade/Device Name: BD Phoenix™ Automated Microbiology System Vancomycin (0.5-32 ug/ml) Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Devices Regulatory Class: Class II Product Code: LON Dated: March 3, 2003 Received: March 4, 2003

Dear Ms. Bandy:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{4}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Page 1 of 1

510(k) Number: __

Device Name: BD Phoenix™ Automated Microbiology System for use with the antimicrobial agent vancomycin (0.5-32 µg/mL) - Gram positive ID/AST or AST only Phoenix panels.

Indications for Use:

The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus and Enterococcus.

This premarket notification is for the addition of the antimicrobial agent vancomycin at concentrations of 0.5-32 µg/mL to Gram Positive ID/AST or AST only Phoenix panels. Vancomycin has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.

Active In Vitro and in Clinical Infections Against:

Staphylococcus species Enterococcus species

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Freddie. Poole

Division Sign-Off) Division of Clinical Laboratory Devices

Prescription Use (Per 21 CFR 801.109)

0(k) Number KD30677

Over-the-Counter Use

Page 9

§ 866.1645 Fully automated short-term incubation cycle antimicrobial susceptibility system.

(a)
Identification. A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.(b)
Classification. Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”