K Number
K030346

Validate with FDA (Live)

Manufacturer
Date Cleared
2003-05-02

(88 days)

Product Code
Regulation Number
876.5990
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Twinheads® TH-101 Extracorporeal Shock Wave Lithotripter is intended to fragment urinary stones in the kidney (renal pelvis and renal calyces) and ureter (upper, middle, and lower ureter).

Device Description

The Twinheads® TH-101 ESWL is a spark gap dual head shock wave lithotripter with an integrated fluoroscopy system for the fragmentation of kidney and ureteral calculi. The Twinheads® TH-101 delivers a pair of shock waves, which are separated from each other by a certain delay time, with perpendicular trajectories and overlapping focal zones. The pulse pairs or twin shocks are aligned with the calculi or stone utilizing the fluoroscopy system via two orientations. A rapid current pulse discharges through a small gap in each of the two heads. The pulse discharge creates a shock wave, which radiates outward. Two identical ellipsoid reflectors serve to reflect and focus the shock waves, which are transmitted into the patient by a water cushion through a membrane and contact gel. There are two shockwave sources and two cushions in the Twinheads® TH-101 ESWL. Also included are an accurate motorized table and U-arm movements, high resolution fluoroscopy, and efficient shock wave energy sources.

AI/ML Overview

The provided document describes a 510(k) premarket notification for the Twinheads® TH-101 Extracorporeal Shock Wave Lithotripter. The acceptance criteria and the study proving the device meets them are outlined in section 7, "Clinical study."

1. Table of Acceptance Criteria and Reported Device Performance:

The document implicitly refers to the special controls guidance for ESWL devices as the acceptance criteria. The specific performance metrics assessed in human clinical trials for a new ESWL device would typically include:

Acceptance Criteria (Implied from FDA Guidance for ESWL)Reported Device Performance
Safety: No unanticipated adverse effects."...the device was able to fragment stones successfully and not pose any unanticipated adverse effects."
Effectiveness/Functionality: Successful stone fragmentation for indicated uses (urinary stones in kidney and ureter)."...the device was able to fragment stones successfully..." and "The results of the study confirmed the functionality of the device and the adequacy of the proposed labeling."
Adequacy of Labeling: Labels accurately reflect device capabilities and safe use."...the adequacy of the proposed labeling."

2. Sample Size and Data Provenance for Test Set:

  • Sample Size: 20 patients.
  • Data Provenance: The study was conducted at "two sites." The document does not specify the country of origin, but given the 510(k) submission to the FDA, it is highly probable the data was collected in the United States.
  • Retrospective or Prospective: Prospective, as it states "20 patients were treated" indicating a clinical trial was performed.

3. Number of Experts and Qualifications for Ground Truth:

The document does not explicitly state the number of experts used to establish the ground truth or their specific qualifications (e.g., radiologist with 10 years of experience). In a clinical study for a lithotripsy device, the success of "stone fragmentation" would typically be assessed by the treating urologists or other medical specialists involved in post-procedure imaging and patient follow-up.

4. Adjudication Method for the Test Set:

The document does not specify an adjudication method (e.g., 2+1, 3+1) for assessing the outcomes of the clinical study.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not explicitly mentioned or indicated in this summary. The study described is a confirmatory clinical study to demonstrate the device's functionality, safety, and effectiveness, not a comparison of human reader performance with and without AI assistance.

6. Standalone (Algorithm Only) Performance:

No, a standalone (algorithm only) performance study was not conducted. This device is a physical medical device (Extracorporeal Shock Wave Lithotripter), not an AI algorithm. The study described assesses the device's ability to fragment stones in patients.

7. Type of Ground Truth Used:

The ground truth used was based on clinical outcomes (successful stone fragmentation) and safety assessments (absence of unanticipated adverse effects) in treated patients. While not explicitly detailed, these assessments would involve post-treatment imaging and clinical evaluation by medical professionals.

8. Sample Size for the Training Set:

The document does not mention a "training set" in the context of an AI/ML algorithm. This device is a physical lithotripsy machine; therefore, the concept of a training set for an algorithm is not applicable here. The "clinical study" described serves as the primary evidence for its intended use, rather than a separate training and test set distinction typically associated with AI.

9. How Ground Truth for Training Set Was Established:

As there is no training set for an AI algorithm, this question is not applicable. The device's fundamental design and technological characteristics (Section 6) are based on established physics and engineering principles, and its performance is validated through the clinical study.

{0}------------------------------------------------

MAY - 2 2003

K030346
PG. 1 OF 3

510(k) Summary

This summary of safety and effectiveness is being submitted in accordance with the requirements of 21 CFR 807.92.

Date of preparation: April 30, 2003

1. Sponsor information

Name and address:FMD, LLC8382 C Terminal RoadLorton, VA 22079
Contact:Yousry Faragalla, MD
Phone:703-339-8881
Fax:703-339-2922

2. Device information

Trade name:Twinheads® TH-101 Extracorporeal Shock Wave Lithotripter
Common name:Extracorporeal shock wave lithotripter
CFR Number:21 CFR 876.5990 – Extracorporeal shock wave lithotripter
Product code:78 LNS
Regulatory Class:Class II (special controls)

3. Substantial Equivalence

The Twinheads® TH-101 ESWL is substantially equivalent to the following legally marketed devices:

  • EDAP Technomed, Inc. Sonolith Praktis (K003529) ■
  • I PCK Electronic Industry and Trade Co., Inc. - Stonelith V5 (K011106)

4. Device description

The Twinheads® TH-101 ESWL is a spark gap dual head shock wave lithotripter with an integrated fluoroscopy system for the fragmentation of kidney and ureteral calculi. The Twinheads® TH-101 delivers a pair of shock waves, which are separated from each other by a certain delay time, with perpendicular trajectories and overlapping focal zones. The pulse pairs or twin shocks are aligned with the calculi or stone utilizing the fluoroscopy system via two orientations.

A rapid current pulse discharges through a small gap in each of the two heads. The pulse discharge creates a shock wave, which radiates outward. Two identical ellipsoid reflectors serve to reflect and focus the shock waves, which are transmitted

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows a document with the text "K030346" at the top, which appears to be a document or reference number. Below that, it says "PG 2 OF 3", indicating it is page 2 of a 3-page document. The text "Twinheads® TH-101 ESWL" is present, likely referring to a product name or model. Finally, the phrase "510(k) Premarket Notification" suggests the document is related to a medical device and its regulatory approval process.

into the patient by a water cushion through a membrane and contact gel. There are two shockwave sources and two cushions in the Twinheads® TH-101 ESWL.

Also included are an accurate motorized table and U-arm movements, high resolution fluoroscopy, and efficient shock wave energy sources.

5. Intended use

The Twinheads® TH-101 Extracorporeal Shock Wave Lithotripter is intended to fragment urinary stones in the kidney (renal pelvis and renal calyces) and ureter (upper, middle, and lower ureter).

6. Technological characteristics

Shock wave characteristics are provided in the following table, which includes data for the minimum, typical, and maximum shock wave generator output settings.

ParameterMin7 kVTypical11 kVMAX14 kV
Peak-Positive Acoustic Pressure—(Mpa)41.586.5110
Peak-Negative Acoustic Pressure—(Mpa)12.526.232.6
Rise Time (ns)424752
Compressional Pulse Duration (ns)420470492
Maximum Focal Width (mm)10.412.815.6
Orthogonal Focal Width (mm)9.612.215.4
Focal Extent (mm)28.430.631.4
Focal Volume (cm³)156248296
Distance between Focus & TargetLocation (mm) on Z Axis156
Derived Acoustic Pulse Energy (mJ)8.118.218.9
Derived Acoustic Pulse Energy at R=4mm Radius (mJ)40.151.255.4
Derived Acoustic Pulse Energy at R=7mm Radius (mJ)53.279.180.3

7. Clinical study

A confirmatory study as described in the FDA 510(k) guidance for ESWL was performed at two sites during which 20 patients were treated. The results of the study confirmed the functionality of the device and the adequacy of the proposed labeling, The safety and effectiveness of the device was also assessed through this study which concluded that the device was able to fragment stones successfully and not pose any unanticipated adverse effects. Therefore, the device has met the requirements for a confirmatory study as specified in the FDA Guidance Document.

{2}------------------------------------------------

8. Summary of nonclinical studies

The Twinheads® TH-101 conforms to the following consensus standards:

IEC 60601-2-36Particular Requirements for Safety Equipment forExtracorporeally Induced Lithotripsy
IEC 61846Ultrasonics- Pressure Pulse Lithotripters-Characteristics of Fields
IEC 60601-1-1Medical Electrical Equipment Part 1-1,General Requirements for Safety
IEC 60601-1-2Medical Electrical Equipment Part 1-2,General Requirements for SafetyCollateral Standard: Electromagnetic CompatibilityRequirements and Tests
IEC 60601-1-3Medical Electrical Equipment Part 1-3,General Requirements for SafetyCollateral Standard: General Requirements forRadiation Protection in Diagnostic X-RayEquipment
IEC 60601-2-7Medical Electrical Equipment Part 2-7,Particular Requirements for the Safety Of High-Voltage Generators of Diagnostic X-Ray Generators
IEC 60601-2-32Medical Electrical Equipment Part 2-32,Particular Requirements for the Safety of AssociatedEquipment of X-Ray Equipment

9. Conclusion

The Twinheads® TH-101 Extracorporeal Shock Wave Lithotripter is substantially equivalent to legally marketed devices and conforms to the requirements of FDA's special controls document "Guidance for the Content of Premarket Notifications (510(k)s) for Extracorporeal Shock Wave Lithotripters Indicated for the Fragmentation of Kidney and Ureteral Calculi", August 9, 2000.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three lines representing the snakes and a central form representing the staff.

MAY - 2 2003

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Yousry Faragalla, M.D. President FMD, LLC 8382C Terminal Road LORTON VA 22079

Re: K030346

Trade/Device Name: Twinheads® TH-101 Extracorporeal Shock Wave Lithotripter Regulation Number: 21 CFR 8876.5990 Regulation Name: Extracorporeal shock wave lithotripter Regulatory Class: II Product Code: 78 LNS Dated: January 30, 2003 Received: February 3, 2003

Dear Dr. Faragalla:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

8xx.1xxx(301) 594-4591
876.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4616
884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx(301) 594-4616
892.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4654
Other(301) 594-4692

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use Statement

510(k) Number (if known): K030346

Device Name: Twinheads® TH-101 Extracorporeal Shock Wave Lithotripter

Indication for Use:

The Twinheads® TH-101 Extracorporeal Shock Wave Lithotripter is intended to fragment urinary stones in the kidney (renal pelvis and renal calyces) and ureter (upper, middle, and lower ureter).

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use _X (Per 21 CFR 801.109) OR

Over-the-counter Use __

Carolyn Y Newland (for Nancy Brogdon)
Division Side Off

Intiste Ahdon

Indications for use

§ 876.5990 Extracorporeal shock wave lithotripter.

(a)
Identification. An extracorporeal shock wave lithotripter is a device that focuses ultrasonic shock waves into the body to noninvasively fragment urinary calculi within the kidney or ureter. The primary components of the device are a shock wave generator, high voltage generator, control console, imaging/localization system, and patient table. Prior to treatment, the urinary stone is targeted using either an integral or stand-alone localization/imaging system. Shock waves are typically generated using electrostatic spark discharge (spark gap), electromagnetically repelled membranes, or piezoelectric crystal arrays, and focused onto the stone with either a specially designed reflector, dish, or acoustic lens. The shock waves are created under water within the shock wave generator, and are transferred to the patient's body using an appropriate acoustic interface. After the stone has been fragmented by the focused shock waves, the fragments pass out of the body with the patient's urine.(b)
Classification. Class II (special controls) (FDA guidance document: “Guidance for the Content of Premarket Notifications (510(k)'s) for Extracorporeal Shock Wave Lithotripters Indicated for the Fragmentation of Kidney and Ureteral Calculi.”)