(77 days)
SureStep™ hCG/Combo (II) Pregnancy Test Name
Not Found
No
The device description and performance studies indicate a standard qualitative immunoassay for detecting hCG, with visual interpretation of results. There is no mention of AI, ML, image processing, or any computational analysis of the results.
No
A therapeutic device is used to treat or alleviate a medical condition. This device is a diagnostic test used to detect pregnancy, not to treat any condition.
Yes
The device is described as a "qualitative test for the rapid detection of human chorionic gonadotropin (hCG)...in urine." It is used to "obtain a visual result" for "early detection of pregnancy," which are all characteristics of a diagnostic device.
No
The device description and performance studies clearly indicate this is a physical in vitro diagnostic test kit that uses a visual result, not a software-only device.
Yes, this device is an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The "Intended Use / Indications for Use" section explicitly states that the device is an "in vitro qualitative test for the rapid detection of human chorionic gonadotropin (hCG) at a cut-off concentration of 25 mlU/ml in urine." The term "in vitro" means "in glass" or "in the lab," referring to tests performed outside of a living organism.
- Device Description: The description further clarifies that it's a "Human Chorionic Gonadotropin (hCG) test system for the use in early detection of pregnancy." This describes a test performed on a biological sample (urine) to diagnose a condition (pregnancy).
- Sample Type: The test is performed on urine, which is a biological specimen.
- Purpose: The purpose is to detect a substance (hCG) in the biological sample to provide information about a physiological state (pregnancy).
All of these characteristics align with the definition of an In Vitro Diagnostic device.
N/A
Intended Use / Indications for Use
The Bayer Diagnostics Clinitest® hCG Pregnancy Test is an in vitro qualitative test for the rapid detection of human chorionic gonadotropin (hCG) at a cut-off concentration of 25 mlU/ml in urine. This test kit is used to obtain a visual result and is intended for professional and laboratory use.
Product codes
JHI
Device Description
Human Chorionic Gonadotropin (hCG) test system for the use in early detection of pregnancy.
The Bayer Diagnostics Clinitest® hCG Pregnancy Test is a one-step chromatographic immunoassav for the rapid qualitative determination of human chorionic gonadotropin (hCG) in urine. The test cut-off is 25 mlU/ml hCG in urine. The test device contains a membrane strip that was pre-coated with anti-hCG capture antibody on the test line region and goat anti-mouse IgG on the control band region. During testing, the patient specimen is allowed to react with the colloidal gold particles that have been coated with anti-hCG monoclonal antibody. The mixture then moves along or across the membrane chromatographically by capillary action. For a positive result, a colored band with a specific anti-hCG-antibody-colloidal gold particle complex will form on the membrane in the test region. A strong colored line will always appear in the control region and another light colored line will always appear in the reference region. The color intensity of the reference line has been adjusted to a level of approximately 25 mlU/ml hCG. Comparison of the test line intensity to the reference line intensity will allow the estimation of whether the positive results are less than, equal to or greater than 25 mlU/ml hCG in serum.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
professional and laboratory use.
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Sensitivity study:
A sensitivity study was performed by spiking negative urine. medium specific gravity with hCG standard to the concentrations of 0, 5, 10, 15, 20, 25 and 50 miU/mL. The urine pool was prepared from 4 donors. The samples were blindlabeled and tested with the Clinitest® hCG Pregnancy Test using two different validation lots of product. A total of 280 samples were tested. Samples were randomly distributed among four visual readers for each product lot.
Key results (Clinitest hCG Lot 1 N=140):
- hCG Concentration (mlU/mL): 0, Number of Positives: 0, Number of Negatives: 20
- hCG Concentration (mlU/mL): 5, Number of Positives: 5, Number of Negatives: 15
- hCG Concentration (mlU/mL): 10, Number of Positives: 9, Number of Negatives: 11
- hCG Concentration (mlU/mL): 15, Number of Positives: 13, Number of Negatives: 7
- hCG Concentration (mlU/mL): 20, Number of Positives: 15, Number of Negatives: 5
- hCG Concentration (mlU/mL): 25, Number of Positives: 18, Number of Negatives: 2
- hCG Concentration (mlU/mL): 50, Number of Positives: 20, Number of Negatives: 0
Key results (Clinitest hCG Lot 2 N=140):
- hCG Concentration (mlU/mL): 0, Number of Positives: 0, Number of Negatives: 20
- hCG Concentration (mlU/mL): 5, Number of Positives: 3, Number of Negatives: 17
- hCG Concentration (mlU/mL): 10, Number of Positives: 9, Number of Negatives: 11
- hCG Concentration (mlU/mL): 15, Number of Positives: 10, Number of Negatives: 10
- hCG Concentration (mlU/mL): 20, Number of Positives: 13, Number of Negatives: 7
- hCG Concentration (mlU/mL): 25, Number of Positives: 19, Number of Negatives: 1
- hCG Concentration (mlU/mL): 50, Number of Positives: 20, Number of Negatives: 0
Accuracy study:
Two external sites were contracted to perform testing on the Clinitest® hCG test (validation lot 59421). A total of 200 visual test results were obtained. Site one tested 98 samples and site 2 tested 102 samples. The samples were prepared by pooling four urine specimens of medium specific gravity, aliquoting and spiking with hCG. The samples were divided by site.
Key results:
- Clinitest® hCG Visual Results for 0 mlU/mL Lablot # 1: Site 1 = 33, Site 2 = 35, Total = 68
- Clinitest® hCG Visual Results for 25 mlU/mL Lablot # 2: Site 1 = 33, Site 2 = 33, Total = 66
- Clinitest® hCG Visual Results for 100 mlU/mL Lablot # 3: Site 1 = 32, Site 2 = 34, Total = 66
The results of the Clinitest® hCG external site data compared to a commercial hCG assay results The commercial assay was an immunoradiometric assay (Magnetic Solid Phase) made by ADALTIS Italia, purchased through Polymedco, Inc. There was 100% agreement of sample results with the commercial hCG assay.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
The Bayer Diagnostics Clinitest® hCG Pregnancy Test detects urinary hCG concentrations at 25 mlU/mL (Calibrated against WHO 310 IRP).
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
SureStep™ hCG/Combo (II) Pregnancy Test Name
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 862.1155 Human chorionic gonadotropin (HCG) test system.
(a)
Human chorionic gonadotropin (HCG) test system intended for the early detection of pregnancy —(1)Identification. A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine.(2)
Classification. Class II.(b)
Human chorionic gonadotropin (HCG) test system intended for any uses other than early detection of pregnancy —(1)Identification. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.(2)
Classification. Class III.(3)
Date PMA or notice of completion of a PDP is required. As of the enactment date of the amendments, May 28, 1976, an approval under section 515 of the act is required before the device described in paragraph (b)(1) may be commercially distributed. See § 862.3.
0
Summary of Safety and Effectiveness
FEB 1 1 2003
As required by 21 CFR 807.92, the following 510(k) Summary is provided:
-
- Submitter Information
Contact person: Thomas F. Flynn
- Submitter Information
Baver Diagnostics Address: 63 North Street Medfield, MA 02052
Phone: | (508) 359-3877 |
---|---|
FAX: | (508) 359-3356 |
e-mail: | thomas.flynn.b@bayer.com |
2. Device Information
Proprietary Name: | Bayer Diagnostics Clinitest® hCG Pregnancy Test |
---|---|
Common Name: | An hCG test system is a device intended for early |
detection of Pregnancy. It is intended to measure hCG, a | |
placental hormone in urine. |
Classification Name: Human Chorionic Gonadotropin (hCG) test system for the |
---|
use in early detection of pregnancy. |
3. Predicate Device Information
SureStep™ hCG/Combo (II) Pregnancy Test Name:
4. Device Description
Human Chorionic Gonadotropin (hCG) test system for the use in early detection of pregnancy.
ട. Statement of Intended Use
The Bayer Diagnostics Clinitest® hCG Pregnancy Test is an in vitro qualitative test for the rapid detection of human chorionic gonadotropin (hCG) at a cut-off concentration of 25 mlU/ml in urine. This test kit is used to obtain a visual result and is intended for professional and laboratory use.
1
. Summary of Technological Characteristics
The Bayer Diagnostics Clinitest® hCG Pregnancy Test is a one-step chromatographic immunoassav for the rapid qualitative determination of human chorionic gonadotropin (hCG) in urine. The test cut-off is 25 mlU/ml hCG in urine. The test device contains a membrane strip that was pre-coated with anti-hCG capture antibody on the test line region and goat anti-mouse IgG on the control band region. During testing, the patient specimen is allowed to react with the colloidal gold particles that have been coated with anti-hCG monoclonal antibody. The mixture then moves along or across the membrane chromatographically by capillary action. For a positive result, a colored band with a specific anti-hCG-antibody-colloidal gold particle complex will form on the membrane in the test region. A strong colored line will always appear in the control region and another light colored line will always appear in the reference region. The color intensity of the reference line has been adjusted to a level of approximately 25 mlU/ml hCG. Comparison of the test line intensity to the reference line intensity will allow the estimation of whether the positive results are less than, equal to or greater than 25 mlU/ml hCG in serum.
7. Performance Data
Sensitivity
The Bayer Diagnostics Clinitest® hCG Pregnancy Test detects urinary hCG concentrations at 25 mlU/mL (Calibrated against WHO 310 IRP).
A sensitivity study was performed by spiking negative urine. medium specific gravity with hCG standard to the concentrations of 0, 5, 10, 15, 20, 25 and 50 miU/mL. The urine pool was prepared from 4 donors. The samples were blindlabeled and tested with the Clinitest® hCG Pregnancy Test using two different validation lots of product. A total of 280 samples were tested. Samples were randomly distributed among four visual readers for each product lot.
| hCG Concentration
mlU/mL | 0 | 5 | 10 | 15 | 20 | 25 | 50 |
---|---|---|---|---|---|---|---|
N = | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
Number of | |||||||
Positives | 0 | 5 | 9 | 13 | 15 | 18 | 20 |
Number of | |||||||
Negatives | 20 | 15 | 11 | 7 | 5 | 2 | 0 |
Clinitest hCG Lot 1 N=140
Clinitest hCG Lot 2 N=140
| hCG Concentration
mIU/mL | 0 | 5 | 10 | 15 | 20 | 25 | 50 |
---|---|---|---|---|---|---|---|
N = | 20 | 20 | 20 | 20 | 20 | 20 | 20 |
Number of | |||||||
Positives | 0 | 3 | 9 | 10 | 13 | 19 | 20 |
Number of | |||||||
Negatives | 20 | 17 | 11 | 10 | 7 | 1 | 0 |
2
Accuracy
Two external sites were contracted to perform testing on the Clinitest® hCG test (validation lot 59421). A total of 200 visual test results were obtained. Site one tested 98 samples and site 2 tested 102 samples. The samples were prepared by pooling four urine specimens of medium specific gravity, aliquoting and spiking with hCG. The samples were divided by site.
Sample | Site 1 | Site 2 | Total |
---|---|---|---|
0 mlU/mL Lablot # 1 | 33 | 35 | 68 |
25 mlU/mL Lablot # 2 | 33 | 33 | 66 |
100 mlU/mL Lablot # 3 | 32 | 34 | 66 |
200 |
Clinitest® hCG Visual Results
The results of the Clinitest® hCG external site data compared to a commercial hCG assay results The commercial assay was an immunoradiometric assay (Magnetic Solid Phase) made by ADALTIS Italia, purchased through Polymedco, Inc. There was 100% agreement of sample results with the commercial hCG assay.
3
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three human profiles facing to the right, with flowing lines suggesting movement or connection.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
FEB 1 1 2003
Mr. Thomas F. Flynn Director of Regulatory Affairs Bayer Diagnostics Baver Corporation 63 North Street Medfield, MA 02052-1688
Re: K023944
Trade/Device Name: Bayer Diagnostics Clinitest® hCG Pregancy Test Regulation Number: 21 CFR 862.1155 Regulation Name: Human chorionic gonadotropin (HCG) test system Regulatory Class: Class II Product Code: JHI Dated: November 21, 2002 Received: November 26, 2002
Dear Mr. Flynn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
4
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
5
510(k) Number (if known): K.O. 2394
Device Name:
Indications for Use:
The Bayer Diagnostics Clinitest® hCG Pregnancy Test is an in vitro qualitative test for the rapid detection of human chorionic gonadotropin (hCG) at a cut-off concentration of 25 mlU/ml in urine. This test kit is used to obtain a visual result and is intended for professional and laboratory use.
Jlan Coogen
(Division Sign-Off)
Division of Clinical Labora 510(k) Number -
## (PLEASE DO NOT WRITE BELOW THIS LINE------------------------------------------------------------------------------------------------------------------------------------------NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
OR
Over-The-Counter
(Optional
Use (Per 21 CFR 801.109) Format 1-2-96)