K Number
K023615
Date Cleared
2003-05-06

(190 days)

Product Code
Regulation Number
876.5860
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

POLYFLUX LR is intended for use in hemodialysis for the treatment of acute and chronic renal failure. The POLYFLUX LR may be reprocessed for reuse on the same patient.

Device Description

The Gambro POLYFLUX 6 LR, 8 LR, and 10 LR, Capillary Dialyzers labeled for multiple use / reuse are identical in design, materials, function and intended use to the Gambro POLYFLUX 6L, 8L, and 10L Capillary Dialyzers/ Filters labeled for single use which have been previously cleared by the FDA under a 510(k) Notification for single use (510(k) Notification K010985). These devices are intended for use in hemodialysis for the treatment of acute and chronic renal failure and for certain types of intoxications. They may also be used in cases of acute fluid overload for the removal of plasma water. The membrane used in this device is polyarylethersulfone (PES) which is identical to the membrane utilized in the Gambro POLYFLUX 6L, 8L, and 10L Capillary Dialyzers/ Filters labeled for single use which have been previously cleared for marketing in the United States under 510K Notifications (K010985). Blood enters a blood inlet port where it is distributed to the hollow fibers. Each hollow fiber has an inner diameter of approximately 215 microns (wet hollow fiber internal diameter) and a wall thickness of 50 microns. The number of hollow fibers in each hemodialyzer / filter is 10,00 for the POLYFLUX 6 LR, 10,000 for the POLYFLUX 8 LR, and 12,500 for the POLYFLUX 10 LR. This effective membrane length is 210 mm for the POLYFLUX 6 LR, and 250 mm for the POLYFLUX 8 LR and 10 LR. The effective membrane surface area is 1.4 square meters for the POLYFLUX 6 LR, 1.7 square meters for the 8 LR and 2.1 square meters for the 10 LR. The housing and end caps of this hemodialyzer / filter are made of polycarbonate. The fibers used in the Gambro POLYFLUX 6 LR, 8 LR, and 10 LR are of the same composition as those previously approved for the Gambro POLYFLUX L 6L, 8L, and 10L Capillary Dialyzers/ Filters labeled for single use and multiple use (K010985). The patient's blood traverses the inside of the hollow fibers and exits the device via a blood exit port. By means of a hydrostatic pressure or transmembrane pressure which is created by a combination of positive and negative pressures across the membrane, plasma water along with certain lower molecular weight solutes pass through the membrane and into the dialysate or filtrate compartment of the device. Removal of uremic toxins and waste products are removed from the patient's blood in this device by means of both diffusion and convection through the membrane and into the counter current flowing dialysis solution during hemodialysis. The dialysate exits the devices via a dialysate outlet port.

AI/ML Overview

The provided text does not contain detailed acceptance criteria or a specific study proving the device meets these criteria in the format requested for AI/software-based devices. Instead, this document is a 510(k) summary for a medical device (hemodialyzer/filter) under the traditional regulatory pathway (K023615).

The core of this submission is to demonstrate "substantial equivalence" to a predicate device, not necessarily to meet pre-defined performance acceptance criteria through independent studies as would be typical for novel AI/software devices.

However, I can extract information related to the device's performance based on the provided text, interpreting "acceptance criteria" in the context of substantial equivalence to the predicate device.

Here's an attempt to answer using the provided information, noting where the requested details are not applicable or unavailable:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (Implied by Substantial Equivalence to Predicate)Reported Device Performance (POLYFLUX 6LR, 8LR, 10LR)
Identical design, materials, function, and intended use as predicate device (POLYFLUX 6L, 8L, 10L single-use) and predicate device (POLYFLUX L 6L, 8L, 10L single and multiple use)Stated as "identical in design, materials, function and intended use" to K010985 (POLYFLUX 6L, 8L, 10L single use). Also stated as "identical in design, materials, intended use and construction" to the currently marketed POLYFLUX 6L, 8L, 10L (single-use). The only stated difference is the addition of "multiple use".
Performance characteristics substantially equivalent to predicate devices."Both the predicate and the proposed devices, incorporate identical membranes and other blood and non-blood contact materials and have identical performance characteristics." "With respect to performance, these hemodialyzer/filters perform in a manner substantially equivalent to each other."
Safety and effectiveness."Testing performed on the Gambro POLYFLUX LR Capillary Dialyzers indicates that they are safe, effective, and perform as well as the predicate devices, when used in accordance with the instructions for use."
Compliance with FDA Guidance for Hemodialyzer Reuse Labeling (for multi-use)."In vitro data was collected according to the FDA Guidance for Hemodialyzer Reuse Labeling." "Clinical data was collected according to the FDA Guidance for Hemodialyzer Reuse Labeling."

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document mentions "In vitro data" and "Clinical data" were collected, but it does not specify the sample size for either. The data provenance (country of origin, retrospective/prospective) is not mentioned.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This is not applicable to this type of device and submission. The "ground truth" for a hemodialyzer's performance is typically established through direct measurement of physical and chemical parameters (clearance, ultrafiltration, etc.) in mechanical or biological models, rather than expert interpretation of data points.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This is not applicable as the evaluation of a hemodialyzer does not involve subjective expert adjudication in the way AI model performance is often assessed.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not applicable. This submission is for a physical medical device (hemodialyzer/filter), not an AI algorithm. Therefore, no MRMC study involving human readers and AI assistance was conducted or would be relevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable. This is not an AI algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" for a hemodialyzer's performance typically comes from:

  • In vitro measurements: Direct quantitative measurements of device parameters such as urea clearance, creatinine clearance, ultrafiltration rate, sieving coefficients, etc., using standardized test solutions and conditions.
  • Clinical outcomes: Observations of patient physiological parameters (e.g., reduction in uremic toxins) during actual use, compared to expected performance or to the predicate device.
    The document states "In vitro data was collected" and "Clinical data was collected."

8. The sample size for the training set

This is not applicable as this is not an AI algorithm requiring a training set.

9. How the ground truth for the training set was established

This is not applicable for the same reason as above.

{0}------------------------------------------------

K 023615
PAGE 1 OF 3

510K Notification Modification Gambro POLYFLUX 6LR, 8 LR, & 10LR Capillary Dialyzers / Filters October 22ªd , 2002

510K(k) SUMMARY

SUBMITTER:Gambro Renal Products10810 West Collins AvenueLakewood, CO 80215(303) 231-5075
MAY 0 6 2003
DATE PREPARED:September 15th, 2002
DEVICE NAME:Gambro POLYFLUX 6 LR, 8 LR, and 10 LRCapillary Dialyzer/Filter Labeled for Single& Multiple Use
CLASSIFICATION NAMES:Hemodialyzer
PREDICATE DEVICE:Gambro POLYFLUX 6L, 8L, & 10LCapillary Dialyzers / Filters

Device Description:

The Gambro POLYFLUX 6 LR, 8 LR, and 10 LR, Capillary Dialyzers labeled for multiple use / reuse are identical in design, materials, function and intended use to the Gambro POLYFLUX 6L, 8L, and 10L Capillary Dialyzers/ Filters labeled for single use which have been previously cleared by the FDA under a 510(k) Notification for single use (510(k) Notification K010985).

These devices are intended for use in hemodialysis for the treatment of acute and chronic renal failure and for certain types of intoxications. They may also be used in cases of acute fluid overload for the removal of plasma water. The membrane used in this device is polyarylethersulfone (PES) which is identical to the membrane utilized in the Gambro POLYFLUX 6L, 8L, and 10L Capillary Dialyzers/ Filters labeled for single use which have been previously cleared for marketing in the United States under 510K Notifications (K010985).

Blood enters a blood inlet port where it is distributed to the hollow fibers. Each hollow fiber has an inner diameter of approximately 215 microns (wet hollow fiber internal diameter) and a wall thickness of 50 microns. The number of hollow fibers in each hemodialyzer / filter is 10,00 for the POLYFLUX 6 LR, 10,000 for the POLYFLUX 8 LR, and 12,500 for the POLYFLUX 10 LR. This effective membrane length is 210 mm for the POLYFLUX 6 LR, and 250 mm for the POLYFLUX 8 LR and 10 LR. The effective membrane surface area is 1.4 square meters for the POLYFLUX 6 LR, 1.7 square meters for the 8 LR and 2.1 square meters for the 10 LR. The housing and end caps of this hemodialyzer / filter are made of polycarbonate. The fibers used in the Gambro POLYFLUX 6 LR, 8 LR, and 10 LR are of the same composition as those

CONFIDENTIAL

000768

{1}------------------------------------------------

K023615
PAGE 2 CF 3

510K Notification Modification Gambro POLYFLUX 6LR, 8 LR, & 10LR Capillary Dialyzers / Filters October 22td , 2002

previously approved for the Gambro POLYFLUX L 6L, 8L, and 10L Capillary Dialyzers/ Filters labeled for single use and multiple use (K010985). The patient's blood traverses the inside of the hollow fibers and exits the device via a blood exit port.

By means of a hydrostatic pressure or transmembrane pressure which is created by a combination of positive and negative pressures across the membrane, plasma water along with certain lower molecular weight solutes pass through the membrane and into the dialysate or filtrate compartment of the device. Removal of uremic toxins and waste products are removed from the patient's blood in this device by means of both diffusion and convection through the membrane and into the counter current flowing dialysis solution during hemodialysis. The dialysate exits the devices via a dialysate outlet port.

Predicate Devices:

The Gambro POLYFLUX 6LR, 8LR, and 10LR, Capillary Dialyzers labeled for multiple use, are identical in design, materials, intended use and construction to the currently marketed Gambro POLYFLUX 6L, 8L, and 10L, Capillary Dialyzers / Filters, labeled for single use. These dialyzers (POLYFLUX 6L, 8L, and 10L, Capillary Dialyzers / Filters) have been cleared for marketing / sale in the United States under 510K Notification K010985 for single use. Both the predicate and the proposed devices, incorporate identical membranes and other blood and non-blood contact materials and have identical performance characteristics. They are substantially equivalent to the listed predicate The intended use for the proposed and predicate devices is also the same, devices. hemodialyzer/ filter with the exception of single versus multiple use which has been added to the POLYFLUX LR.

PREDICATE DEVICES

DEVICE NAMESGambro POLYFLUX 6L, 8L & 10L Capillary Dialyzer/Filter
INTENDED USEHemodialyzer/Filter
510K NUMBERK010885
APPROVAL DATE10/10/01

With respect to performance, these hemodialyzer/filters perform in a manner substantially equivalent to each other. We therefore consider the proposed devices substantially equivalent to existing predicate devices in commercial distribution in the United States.

{2}------------------------------------------------

K023615
PAGE 3 OF 3

510K Notification Modification Gambro POLYFLUX 6LR, 8 LR, & 10LR Capillary Dialyzers / Filters October 22ªd , 2002

Intended Use:

POLYFLUX LR Indications:

POLYFLUX LR is intended for use in hemodialysis for the treatment of acute and chronic renal failure.

The POLYFLUX LR may be reprocessed for reuse on the same patient.

This indication statement is essentially the same as the indication statement for the predicate devices.

Technological Characteristics:

Comparing the proposed devices to the predicate devices, they are substantially equivalent to the predicate devices. Both the proposed and predicate devices use the same hollow fiber membrane and other blood and non-blood contact materials. Both the proposed and predicate devices use polycarbonate for the housing and header material and polyurethane for the membrane potting material and are steam sterilized.

Summary of Non-Clinical Tests:

In vitro data was collected according to the FDA Guidance for Hemodialyzer Reuse Labeling.

Clinical Test Results:

Clinical data was collected according to the FDA Guidance for Hemodialyzer Reuse Labeling.

Conclusions:

Testing performed on the Gambro POLYFLUX LR Capillary Dialyzers indicates that they are safe, effective, and perform as well as the predicate devices, when used in accordance with the instructions for use. In vitro and clinical performance data and directions for reuse have been included in the labeling.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its body and wings. The eagle is facing right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circle around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 0 6 2003

Jeffrey R. Shideman, Ph.D. Director, Therapy Group Americas Gambro Corporate Research Gambro® Renal Products 10810 West Collins Avenue LAKEWOOD CO 80215

Re: K023615

Trade/Device Name: POLYFLUX 6LR, 8LR, & 10LR Capillary Dialyzers/Filters labeled for Multiple/Re-Use Regulation Number: 21 CFR §876.5860 Regulation Name: High permeability hemodialysis system Regulatory Class: II Product Code: 78 MSF Dated: February 4, 2003 Received: February 5, 2003

Dear Dr. Shideman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

8xx.1xxx(301) 594-4591
876.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4616
884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx(301) 594-4616
892.2xxx, 3xxx, 4xxx, 5xxx(301) 594-4654
Other(301) 594-4692

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Page / of /

510 (k) NUMBER (IF KNOWN): Not yet usstgned

DEVICE NAME: Gambro POLYFLUX 6 LR, 8 LR, and 10 LR Capillary Dialyzer/Filter Labeled for Single & Multiple Use

INDICATIONS FOR USE:

POLYFLUX LR Indications:

POLYFLUX LR is intended for use in hemodialysis for the treatment of acute and chronic renal failure.

The POLYFLUX LR may be reprocessed for reuse on the same patient.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use l
(Per 21 CFR 801.109)

OR

(Division Sign-Off) Division of Reproductive, Abdominal. and Radiological Device

510(k) Number K023615

Over-The-Counter-Use

(Optional Format 1-2-96)

000174

§ 876.5860 High permeability hemodialysis system.

(a)
Identification. A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K
uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.).
(3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
(b)
Classification. Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”