(92 days)
The BD Phoenix™ Automated Microbiology System is intended for the rapid identification and in vitro antimicrobial susceptibility testing of isolates from pure culture of most aerobic and facultative anaerobic gram-negative and gram-positive bacteria of human origin.
The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus and Enterococcus.
This premarket notification is for the addition of the antimicrobial agent Meropenem at concentrations of 0.5-16 ug/mL to Gram Positive ID/AST or AST only Phoenix panels. Meropenem has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.
Active In Vitro Against:
Staphylococcus aureus (B-lactamase and non-ß-lactamase producing)
Staphylococcus epidermidis (B-lactamase and non-B-lactamase producing)
The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:
- . BD Phoenix instrument and software.
- . BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents or AST determinations.
- BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum. ●
- BD Phoenix AST Broth used for performing AST tests only. .
- BD Phoenix AST Indicator solution added to the AST broth to aid in bacterial growth . determination.
The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a gram-negative or gram-positive isolate. For each isolate, an inoculation equivalent to a 0.5 McFarland standard is prepared in Phoenix ID broth.
The Phoenix AST method is a broth based microdilution test. The Phoenix system utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.
The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, or R (sensitive, intermediate, or resistant).
Acceptance Criteria and Device Performance Study for BD Phoenix™ Automated Microbiology System – Meropenem
This document summarizes the acceptance criteria and the study demonstrating the performance of the BD Phoenix™ Automated Microbiology System for Meropenem, as presented in the provided 510(k) summary.
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for the BD Phoenix™ Automated Microbiology System when compared to the NCCLS reference broth microdilution method are based on Essential Agreement (EA) and Category Agreement (CA). While specific numerical thresholds for acceptance criteria are not explicitly stated as "acceptance criteria" in the provided text, the study results are presented against the FDA's guidance document for substantial equivalence. Historically, for AST devices, typical acceptance criteria for EA and CA are often >90%.
| Performance Metric | Acceptance Criteria (Implied by FDA Guidance/Historical Norms) | Reported Device Performance (Meropenem, Gram-Positive Organisms) | Result |
|---|---|---|---|
| Essential Agreement (EA) | >90% (often >95%) | 98.4% | Met |
| Category Agreement (CA) | >90% (often >95%) | 96.0% | Met |
| Intra-site Reproducibility | >90% | >90% | Met |
| Inter-site Reproducibility | >95% | >95% | Met |
Note: The reported "concentration" and other fields in the provided table (Table 1) were largely unreadable due to formatting issues in the input. The focus here is on the EA and CA percentages.
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size: The exact number of isolates for the clinical and stock/challenge sets is not explicitly stated in numerical form. However, the study involved a "panel of gram-positive isolates" for reproducibility and "Clinical, stock and challenge isolates" for performance assessment. Table 1 indicates an 'n' value for EA (though obscured by text, it appears to be around "OZL" which is not a standard numerical representation, but likely a placeholder for a count of isolates).
- Data Provenance: The clinical studies were conducted "across multiple geographically diverse sites across the United States." The data includes "stock and challenge isolates" and "clinical isolates." This indicates a mix of prospectively collected clinical samples and curated challenge strains, originating from the United States.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
The document does not mention the use of "experts" in the traditional sense for establishing ground truth. Instead, the ground truth for the test set was established by the NCCLS reference broth microdilution method. This is a standardized laboratory method, not dependent on individual expert interpretation for each case. Therefore, the concept of "number of experts" and their "qualifications" is not directly applicable in this context.
4. Adjudication Method for the Test Set
The concept of an "adjudication method" (like 2+1 or 3+1) is not applicable here because the ground truth was established by a reference laboratory method (NCCLS broth microdilution), not by a panel of human experts whose disagreements would require adjudication. The Phoenix System results were directly compared to these reference results.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, an MRMC comparative effectiveness study was not conducted. This device is an automated system for antimicrobial susceptibility testing, not an imaging or diagnostic aid that humans interpret. Therefore, the effect size of how much human readers improve with AI vs. without AI assistance is not relevant to this type of device. The comparison is between the automated system's output and a reference laboratory method.
6. Standalone Performance Study
Yes, a standalone performance study was done. The entire evaluation focuses on the performance of the "BD Phoenix™ Automated Microbiology System" (the algorithm/device only) in comparison to the NCCLS reference method. There is no human-in-the-loop component described for the system's operation or interpretation during the performance comparison. The instrument takes readings and interprets them to provide MIC values and categorical interpretations (S, I, R).
7. Type of Ground Truth Used
The type of ground truth used was a reference laboratory method: the NCCLS (National Committee for Clinical Laboratory Standards) reference broth microdilution method.
8. Sample Size for the Training Set
The document does not explicitly state the sample size used for the training set. The descriptions provided relate to the performance evaluation (test set) rather than the development or training of the system.
9. How the Ground Truth for the Training Set Was Established
The document does not explicitly describe how the ground truth for the training set was established. Given the nature of AST devices, it is highly likely that similar reference laboratory methods (like NCCLS broth microdilution) would have been used during the development and calibration of the Phoenix system to establish the expected antimicrobial susceptibility patterns for various organisms against different agents. However, this is inferred, not explicitly stated in the provided text.
{0}------------------------------------------------
510(k) SUMMARY
| SUBMITTED BY: | Becton, Dickinson and Company7 Loveton CircleSparks, MD 21152Phone: 410-316-4287Fax: 410-316-4499 |
|---|---|
| CONTACT NAME: | Monica E. GiguereRegulatory Affairs Specialist |
| DATE PREPARED: | January 10, 2003 |
| DEVICE TRADE NAME: | BD Phoenix™ Automated Microbiology System –Meropenem 0.5-16.0 µg/mL |
| DEVICE COMMON NAME: | Antimicrobial susceptibility test system-short incubation |
| DEVICE CLASSIFICATION: | Fully Automated Short-Term Incubation CycleAntimicrobial Susceptibility Device, 21 CFR 866.1645 |
| PREDICATE DEVICES: | VITEK® System (PMA No. N50510) and BD Phoenix™Automated Microbiology System with Gatifloxacin(K020321, May 23, 2002), Ofloxacin (K020323, April 14,2002), and Levofloxacin (K020322, March 27, 2002). |
| INTENDED USE: | The BD Phoenix™ Automated Microbiology System isintended for the rapid identification and in vitroantimicrobial susceptibility testing of isolates from pureculture of most aerobic and facultative anaerobic gram-negative and gram-positive bacteria of human origin. |
DEVICE DESCRIPTION:
The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. The system includes the following components:
- . BD Phoenix instrument and software.
- . BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents or AST determinations.
- BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum. ●
- BD Phoenix AST Broth used for performing AST tests only. .
- BD Phoenix AST Indicator solution added to the AST broth to aid in bacterial growth . determination.
{1}------------------------------------------------
The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells containing dried reagents. Organisms for susceptibility testing must be a pure culture and preliminarily identified as a gram-negative or gram-positive isolate. For each isolate, an inoculation equivalent to a 0.5 McFarland standard is prepared in Phoenix ID broth.
The Phoenix AST method is a broth based microdilution test. The Phoenix system utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. Measurements of changes to the indicator as well as bacterial turbidity are used in the determination of bacterial growth. Each AST panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.
The instrument houses the panels where they are continuously incubated at a nominal temperature of 35°C. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, or R (sensitive, intermediate, or resistant).
DEVICE COMPARISON:
The BD Phoenix™ Automated Microbiology System demonstrated substantially equivalent performance when compared with the NCCLS reference broth microdilution method. This premarket notification provides data supporting the use of the BD Phoenix™ Automated Microbiology System Gram positive ID/AST or AST only Phoenix panels with this antimicrobial agent.
SUMMARY OF SUBSTANTIAL EQUIVALENCE TESTING:
The BD Phoenix™ Automated Microbiology System has demonstrated substantially equivalent performance when compared to the NCCLS reference broth microdilution method (AST panels prepared according to NCCLS M7). The system has been evaluated as defined in the FDA Draft guidance document, "Guidance on Review Criteria for Assessment of Antimicrobial Susceptibility Devices", March 8, 2000.
Site Reproducibility
Intra- and inter-site reproducibility of this antimicrobial agent in the BD Phoenix System was evaluated at three sites using a panel of gram-positive isolates. Each site tested the isolates in triplicate on three different days using one lot of Gram Positive Phoenix panels containing this antimicrobial agent and associated reagents.
The results of the study demonstrate for the this antimicrobial agent there was an overall intrasite reproducibility of greater than 90% and an overall inter-site reproducibility greater than 95% for the gram-positive isolates tested.
{2}------------------------------------------------
Clinical Studies
Clinical. stock and challenge isolates were tested across multiple geographically diverse sites across the United States to demonstrate the performance of the Phoenix antimicrobial susceptibility test with the Gram Positive Phoenix Panel format containing this antimicrobial agent. Phoenix System results for Challenge set isolates were compared to the expected results. Phoenix System results for clinical isolates were compared to the results obtained from the NCCLS reference broth microdilution method.
The performance of the Phoenix System was assessed by calculating Essential Agreement (EA) and Category Agreement (CA) to expected/reference results for all isolates tested. Essential Agreement (EA) occurs when the BD Phoenix™ Automated Microbiology System agrees exactly or within + one two-fold dilution to the reference result. Category Agreement (CA) occurs when the BD Phoenix™ Automated Microbiology System agrees with the reference method with respect to the FDA categorical interpretive criteria (susceptible, intermediate, and resistant).
Table 1 summarizes the performance for the isolates tested in this study.
Table 1: Performance of BD Phoenix System for Gram-Positive Organisms by Drug
| 11 11 11 11 11 11 11 19 11Antimicrobial | ConcentrationSTER LEASE OF STATE AND STATES AND A BEACH ANDComments assess to 4 W Pages of | EA (nor other control and property and for a local be and | EA (%) | ( DCA | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------CA (%)THE CONTRACT THE THE THE THE THE THE THE THE THE THE THE THE THE THE THE THE THE THE THE THE THE THE THE THE THE THE THE THE THE THE THE THE THE THE THE THE THE THE THE THE T |
|---|---|---|---|---|---|
| and and the computer of the production of the concession of the comments of theMeropenem | '6 µg/mLJ- J = I' | rarOZL | 98. | 1198 | 96.0 |
Conclusions Drawn from Substantial Equivalence Studies
The data collected from the substantial equivalence studies demonstrate that testing on the BD Phoenix™ Automated Microbiology System with this antimicrobial agent is substantially equivalent as outlined in the FDA draft guidance document, "Guidance on Review Criteria for Assessment of Antimicrobial Susceptibility Devices", March 8, 2000. Technological characteristics of this system are substantially equivalent to those used in the VITEK® system, which received approval by the FDA under N50510 and BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321. May 23, 2002), Offloxacin (K020323. April 14, 2002), and Levofloxacin (K020322, March 27, 2002).
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is a stylized image of a bird or eagle, with three overlapping profiles facing to the right.
JAN 2 4 2003
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Monica E. Giguere, RAC Regulatory Affairs Specialist BD Diagnostics Systems Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152
Re: K023568 Trade/Device Name: BD Phoenix TM Automated Microbiology Systems Meropenem (0.5-16 ug/ml) Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Short-Term Incubation Cycle Antimicrobial Susceptibility Devices Regulatory Class: Class II Product Code: LON Dated: January 10, 2003 Received: January 13, 2003
Dear Ms. Giguere:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{4}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page I of 1
510(k) Number: K023568
Device Name: BD Phoenix™ Automated Microbiology System for use with the antimicrobial agent Meropenem (0.5-16 ug/mL) - Gram positive ID/AST or AST only Phoenix panels.
Indications for Use:
The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus and Enterococcus.
This premarket notification is for the addition of the antimicrobial agent Meropenem at concentrations of 0.5-16 ug/mL to Gram Positive ID/AST or AST only Phoenix panels. Meropenem has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.
Active In Vitro Against:
Staphylococcus aureus (B-lactamase and non-ß-lactamase producing) Staphylococcus epidermidis (B-lactamase and non-B-lactamase producing)
Neile Sephyloggen with he resistent a meinterling saghin mist it vonstitlered essisem Start Ste williat
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Freddie Poole
(Division Sign-Off)
Division of Clinical Laborato
Over-the-Counter Use
510(k) Number K02 556
Prescription Use (Per 21 CFR 801.109)
BD Diagnostic Systems Becton, Dickinson and Company
Page 9
§ 866.1645 Fully automated short-term incubation cycle antimicrobial susceptibility system.
(a)
Identification. A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.(b)
Classification. Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”