(27 days)
Skintact ECG electrodes are designed for use in general electrocardiographic procedures where ECG monitoring is deemed necessary and is ordered by a physician. Such procedures include in particular patient ECG surveillance and ECG diagnosis recording. Skintact ECG electrodes are non-sterile and are to be used on intact (uninjured) skin.
Skintact® ECG Electrodes (and as also to be offered for sale under various private label tradenames) will now also be offered with solid adhesive gel. Just like the liquid gel electrodes, solid adhesive gel electrodes are self-adhesive, non-sterile, single use disposable snap electrodes. The solid adhesive electrodes are identical in size, shape and configuration to the liquid gel Skintact ECG electrodes currently marketed by Leonhard Lang, GmbH. All electrode configurations include a stainless steel stud to guarantee an unimpaired performance during the shelf-life of the product. All electrodes include an ABS sensor element coated with silver. The silver layer is either completely or partially (in the areas in contact with the conductive gel) covered with a silver chloride layer. This is the same construction as the current Skintact® ECG electrode using liguid gel conducting media.
Here's a breakdown of the acceptance criteria and study information for the Skintact® ECG Electrode, based on the provided text:
Acceptance Criteria and Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Biocompatibility | Materials are biocompatible. Testing showed no adverse results for Cytotoxicity, Skin Irritation, and Sensitization. |
| Electrical Performance (ANSI/AAMI EC 12:2000) | - AC Impedance: Met requirements of ANSI/AAMI EC 12:2000. - DC Offset Voltage: Met requirements of ANSI/AAMI EC 12:2000. - Defibrillation Overload Recovery: Met requirements of ANSI/AAMI EC 12:2000. - Combined Offset Instability and Internal Noise: Met requirements of ANSI/AAMI EC 12:2000. - Bias Current Tolerance: Met requirements of ANSI/AAMI EC 12:2000. |
| Shelf Life (24 months) | The electrical performance of electrodes with solid adhesive gel stays within ANSI/AAMI EC12-2000 limits for 24 months, as confirmed by real-time and accelerated aging (3 months at 40°C). |
| Equivalence to Liquid Gel Electrodes (Electrical) | Electrical performance of solid adhesive gel electrodes is equivalent to liquid gel electrodes, with negligible differences compared to ANSI/AAMI EC12-2000 limits. |
| Clinical Performance (Signal Quality) | Clinical trace testing confirmed that the solid adhesive gel performs the same as liquid gel, demonstrating equivalence in ECG traces. |
| Clinical Performance (Wear Test - 48 hours) | - No deterioration of electrical performance. - Gel did not dry out. - Electrodes were not replaced and showed no displacement, even with normal activities and sports. - No skin problems/irritation for volunteers. |
Study Information
2. Sample Size Used for the Test Set and the Data Provenance:
- Test Set Sample Size:
- For biocompatibility: Not specified in detail, but implied to be sufficient for cytotoxicity, skin irritation, and sensitization tests.
- For electrical testing: Not specified, but a "certification to conformance EC12:2000 with this standard has been provided." This implies sufficient devices were tested to meet the standard.
- For shelf-life: Electrodes were used for real-time and accelerated aging, but the specific number is not provided.
- For clinical trace testing: Not specified, but implied to be sufficient to evaluate the potential effect of material change.
- For wear tests: "Three wear test reports" were provided. The number of volunteers per report is not explicitly stated.
- Data Provenance: The document implies the testing was conducted by Leonhard Lang GmbH (based in Austria). The data is retrospective in the sense that it's reported after the testing.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and their Qualifications:
- Wear Tests: "A physician reviewed the performance of the wear tests and examined for skin irritation." The number of physicians is not specified (it states "a physician," implying one, but doesn't explicitly rule out more if separate reports were reviewed). Qualifications are stated as "physician."
4. Adjudication Method for the Test Set:
- Not applicable/Not specified as human "reader" adjudication in the typical sense for image-based diagnostics. The assessments were based on meeting objective electrical standards and physician review for wear tests.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This device is an ECG electrode, not an AI-assisted diagnostic tool.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Not applicable. This device is an ECG electrode; its performance is standalone in the sense that it collects the ECG signal itself, without an AI algorithm.
7. The Type of Ground Truth Used:
- Electrical Performance: Ground truth was defined by the requirements of the ANSI/AAMI EC 12:2000 "Disposable ECG electrodes" standard.
- Biocompatibility: Ground truth was based on established biocompatibility testing standards (cytotoxicity, skin irritation, sensitization).
- Clinical Performance (Wear Test): Ground truth was based on direct observation and clinical assessment by a physician regarding electrode performance (electrical stability, gel integrity, adhesion) and patient outcomes (skin irritation).
- Equivalence: Comparisons were made against performance data from the predicate liquid gel electrodes and against the ANSI/AAMI EC 12:2000 limits.
8. The Sample Size for the Training Set:
- Not applicable. This is a medical device (electrode), not a machine learning algorithm that requires a training set.
9. How the Ground Truth for the Training Set was established:
- Not applicable (as above).
{0}------------------------------------------------
NOV 1 4 2002
K023503
p.1/2
510(k) Summary
SUBMITTED ON BEHALF OF:
| Company Name:Address: | Leonhard Lang GmbHArchenweg 566010 InnsbruckAustria | |||||
|---|---|---|---|---|---|---|
| Telephone:Fax: | ++ 43 / 512 / 33 4 25 7++ 43 / 512 / 39 22 10 | |||||
| by: | Elaine Duncan, MS.M.E., RACPresident, Paladin Medical, Inc.PO Box 560Stillwater, MN 55082 | |||||
| Telephone:Fax: | 715-549-6035715-549-5380 | |||||
| CONTACT PERSON: | Elaine Duncan | |||||
| DATE PREPARED: | October 16, 2002 | |||||
| Trade Name:Common Name:Classification Name: | Skintact® ECG ElectrodeDisposable ECG ElectrodesElectrocardiograph (ECG) electrode |
SUBSTANTIALLY EQUIVALENT TO: substantially equivalent to the Skintact® ECG Electrodes with liquid gel (the manufacturer's predicate device cleared under K982521). FDA previously cleared predicate solid adhesive gel electrodes in K960968.
DESCRIPTION of the DEVICE: Skintact® ECG Electrodes (and as also to be offered for sale under various private label tradenames) will now also be offered with solid adhesive gel. Just like the liquid gel electrodes, solid adhesive gel electrodes are self-adhesive, non-sterile, single use disposable snap electrodes. The solid adhesive electrodes are identical in size, shape and configuration to the liquid gel Skintact ECG electrodes currently marketed by Leonhard Lang, GmbH.
All electrode configurations include a stainless steel stud to guarantee an unimpaired performance during the shelf-life of the product. All electrodes include an ABS sensor element coated with silver. The silver layer is either completely or partially (in the areas in contact with the conductive gel) covered with a silver chloride layer. This is the same construction as the current Skintact® ECG electrode using liguid gel conducting media.
INDICATIONS FOR USE:
Skintact ECG electrodes are designed for use in general electrocardiographic procedures where ECG monitoring is deemed necessary and is ordered by a physician. Such procedures include in particular patient ECG surveillance and ECG diagnosis recording. Skintact ECG electrodes are non-sterile and are to be used on intact (uninjured) skin. (NO CHANGE to ORIGINAL INDICATION for USE)
{1}------------------------------------------------
510(k) Summary-Continued
SUMMARY of TESTING:
Biocompatibility testing confirms the materials are biocompatible and the change does not introduce new risks. The following testing showed no adverse results: Cytotoxicity; Skin Irritation; Sensitization.
The ANSI/AAMI EC 12:2000 "Disposable ECG electrodes" was used to define the requirements for Skintact ECG Electrodes with solid adhesive gel. All electrical tests are according to ANSI/AAMI EC 12:2000. A certification to conformance EC12:2000 with this standard has been provided. The testing conducted was: AC impedance; DC offset voltage; Defibrillation overload recovery; Combined offset instability and internal noise; Bias current tolerance.
The shelf life of the electrodes with solid adhesive gel was tested in real-time aging and in accelerated aging. For accelerated aging the electrodes were in an incubator for a time of 3 months with an increased temperature of 40°C. In accelerated shelf life testing the electrodes are subjected to a controlled environment in which one or more extrinsic factors (e.g., temperature, humidity, gas atmosphere, light) is maintained at a higher than normal level. Leonhard Lang has experience for about 20 years of using the current packaging and this ensures all requirements for the 24 months shelf-life of the electrodes. No differences were required for packaging of the solid adhesive electrodes compared to the predicate electrode.
The results of these tests confirm that the shelf-life of Leonhard Lang Skintact ECG Gel Electrodes with solid adhesive gel is well inside the limits defined in ANSUAAMI EC12-2000, both for the first test and with real-time and accelerated aged electrodes. Thus the conclusion that the electrical performance of the solid adhesive gel electrodes will stay within the limits during their shelf-life of 24 months. The comparison with the predicate device and the data from the solid adhesive gel electrodes shows similar results. The difference is negligible in the limits defined in ANSI/AAMI EC12-2000. Therefore electrical performance of the predicate device and solid adhesive gel electrodes is equivalent.
Clinical data: The potential effect of material change to the conducting signal was evaluated by repeating the clinical trace testing per the FDA guidance on the solid adhesive gel and determined that the solid adhesive gel performs the same. Comparing the ECG traces between solid adhesive gel and liquid gel electrodes demonstrate that solid adhesive gel electrodes are equivalent to the liquid gel electrodes. Three wear test reports for Holter monitoring electrodes used for 48 hours were provided. A physician reviewed the performance of the wear tests and examined for skin irritation. The tests confirm that the performance of electrodes is quite good and that wearing of the electrodes does not cause any skin problems for the volunteers. During a time of 48 hours there was no deterioration of electrical performance and the gel did not dry out. The electrodes were not replaced during this time and did not show any displacement; although the volunteers worked and participated in sports as usual demonstrating the adhesion was also very good. (No change to adhesive formulation was made.)
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 1 4 2002
Leonhard Lang GmbH c/o Ms. Elaine Duncan, M.S.M.E., RAC President Paladin Medical, Inc. P.O. Box 560 Stillwater, MN 55082-0560
Re: K023503
Trade Name: Skintact ECG Electrodes Regulation Number: 21 CFR 870.2360 Regulation Name: Electrocardiograph Electrode Regulatory Class: Class II (two) Product Code: DRX Dated: October 16, 2002 Received: October 18, 2002
Dear Ms. Duncan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Elaine Duncan, M.S.M.E., RAC
Please be advised that FDA's issuance of a substantial equivalence determination does not mean. that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Elia Mallis
Por
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known)
KD23503
. . . .
Device Name:
Skintact ECG Electrodes
Indications for Use:
Skintact ECG electrodes are designed for use in general electrocardiographic procedures where ECG monitoring is deemed necessary and is ordered by a physician. Such procedures include in particular patient ECG surveillance and ECG diagnosis recording.
Skintact ECG electrodes are non-sterile and are to be used on intact (uninjured) skin.
| (Please Do Not Write Below This Line-Continue On Another Page If Needed) |
|---|
| Concurrence of CDRH, Office of Device Evaluation (ODE) |
Prescription Use __ OR . Over-The-Counter Use __
(Optional Format 1-2-96)
Ko 23503 Division of Cardiovascular & Responsion / Devices
510(k) Number ______________________________________________________________________________________________________________
§ 870.2360 Electrocardiograph electrode.
(a)
Identification. An electrocardiograph electrode is the electrical conductor which is applied to the surface of the body to transmit the electrical signal at the body surface to a processor that produces an electrocardiogram or vectorcardiogram.(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9. The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Electrocardiograph Electrodes.” See § 870.1(e) for availability information of guidance documents.