(46 days)
The VALIDATE Chem 9 Calibration Verification Test Set is intended for in vitro diagnostic use for quantitatively verifying calibration, validating reportable ranges, and determining linearity for automated and manual chemistry systems for the following eighteen analytes: albumin (ALB), calcium (CA), cholesterol (CHOL), chloride (CL), carbon dioxide (CO2), creatinine (CRE), glucose (GLU), iron (FE), lactate (LAC), lithium (LI), magnesium (MG), phosphorus (PQ4), potassium (K), sodium (NA), total protein (TP), triglyceride (TRIG), urea nitrogen (BUN), and uric acid (UA).
The VALIDATE Chem 9 Calibration Verification Test Set is used by trained laboratory professionals for the quantitative determination of linearity, calibration verification and verification of reportable range in automated, semi-automated, and manual clinical chemistry systems for the following eighteen analytes. albumin, calcium, cholesterol, chloride, carbon dioxide, creatinine, glucose, iron, lactate, lithium, magnesium, phosphorus, potassium, sodium, total protein, triglyceride, urea nitrogen, and uric acid.
VALIDATE Chem 9 Calibration Verification Test Set contains purified chemicals in a protein matrix. Multiple levels are provided to establish the relationship between theoretical operation and actual performance of each of the included analytes. Each set contains one bottle each of five (5) levels. Each bottle contains 5 milliliters.
Here's a breakdown of the acceptance criteria and study information for the VALIDATE Chem 9 Calibration Verification Test Set, based on the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria for the VALIDATE Chem 9 device are not explicitly stated as numerical thresholds (e.g., "correlation coefficient must be > 0.99"). Instead, the primary acceptance criterion is "functional equivalence for calibration verification and linearity assessment" to the predicate devices. This functional equivalence is demonstrated through linear regression analysis and correlation coefficients.
The reported device performance, represented by the correlation coefficients (r) and regression equations, establishes this equivalence.
| Analyte | Acceptance Criteria (Implied) | VALIDATE Chem 9 Performance (Correlation Coefficient (r)) | VALIDATE Chem 9 Regression Equation (Y = slope(X) + intercept) | Predicate 1 (DOCUMENT Serum) Performance (Correlation Coefficient (r)) | Predicate 1 (DOCUMENT Serum) Regression Equation (Y = slope(X) + intercept) | Predicate 2 (Ortho-Clinical VITROS) Performance (Correlation Coefficient (r)) | Predicate 2 (Ortho-Clinical VITROS) Regression Equation (Y = slope(X) + intercept) |
|---|---|---|---|---|---|---|---|
| ALB | Functional equivalence to predicate devices (demonstrated by high correlation and similar regression). | 0.9942 | x - 0.1 | 0.998 | 0.997x + 0.022 | (Not applicable for this analyte) | (Not applicable for this analyte) |
| BUN | Functional equivalence to predicate devices (demonstrated by high correlation and similar regression). | 0.9999 | 1.0138x - 0.4276 | 0.999 | 1.005x + 0.618 | (Not applicable for this analyte) | (Not applicable for this analyte) |
| CA | Functional equivalence to predicate devices (demonstrated by high correlation and similar regression). | 0.9974 | 1.01x + 0.288 | 0.999 | 0.995x + 0.035 | (Not applicable for this analyte) | (Not applicable for this analyte) |
| CHOL | Functional equivalence to predicate devices (demonstrated by high correlation and similar regression). | 0.9999 | 0.9935x + 1.8102 | 0.995 | 1.058x - 6.192 | (Not applicable for this analyte) | (Not applicable for this analyte) |
| CL | Functional equivalence to predicate devices (demonstrated by high correlation and similar regression). | 0.9996 | 0.9581x + 3.6226 | 0.998 | 1.010x - 1.389 | (Not applicable for this analyte) | (Not applicable for this analyte) |
| CO2 | Functional equivalence to predicate devices (demonstrated by high correlation and similar regression). | 0.9971 | 0.875x + 1.675 | 0.998 | 1.009x - 0.176 | (Not applicable for this analyte) | (Not applicable for this analyte) |
| CRE | Functional equivalence to predicate devices (demonstrated by high correlation and similar regression). | 0.9986 | 0.9364 + 0.1791 | 0.999 | 0.974x + 0.073 | (Not applicable for this analyte) | (Not applicable for this analyte) |
| FE | Functional equivalence to predicate devices (demonstrated by high correlation and similar regression). | 0.9988 | 0.9205x + 14.617 | 0.998 | 1.119x - 14.492 | (Not applicable for this analyte) | (Not applicable for this analyte) |
| GLU | Functional equivalence to predicate devices (demonstrated by high correlation and similar regression). | 0.9999 | 1.0132x - 0.7707 | 0.999 | 1.050x - 9.370 | (Not applicable for this analyte) | (Not applicable for this analyte) |
| K | Functional equivalence to predicate devices (demonstrated by high correlation and similar regression). | 0.9987 | 0.9984x - 0.2576 | 0.999 | 0.983 + 0.170 | (Not applicable for this analyte) | (Not applicable for this analyte) |
| LITH | Functional equivalence to predicate devices (demonstrated by high correlation and similar regression). | 0.9997 | 0.9733x + 0.0187 | 0.994 | 0.924x + 0.015 | (Not applicable for this analyte) | (Not applicable for this analyte) |
| MG | Functional equivalence to predicate devices (demonstrated by high correlation and similar regression). | 0.9993 | 0.9832 + 0.1053 | 0.998 | 1.051x - 0.082 | (Not applicable for this analyte) | (Not applicable for this analyte) |
| NA | Functional equivalence to predicate devices (demonstrated by high correlation and similar regression). | 0.9991 | 1.0042x + 1.8979 | 0.998 | 1.076x - 7.561 | (Not applicable for this analyte) | (Not applicable for this analyte) |
| PO4 | Functional equivalence to predicate devices (demonstrated by high correlation and similar regression). | 0.9998 | 0.975x + 0.035 | 0.999 | 0.992x + 0.093 | (Not applicable for this analyte) | (Not applicable for this analyte) |
| TP | Functional equivalence to predicate devices (demonstrated by high correlation and similar regression). | 0.9986 | 1.097x - 0.5097 | 0.998 | 1.086x - 0.261 | (Not applicable for this analyte) | (Not applicable for this analyte) |
| TRIG | Functional equivalence to predicate devices (demonstrated by high correlation and similar regression). | 0.9999 | 1.0226x - 3.6259 | 0.999 | 1.018x + 2.292 | (Not applicable for this analyte) | (Not applicable for this analyte) |
| URIC ACID | Functional equivalence to predicate devices (demonstrated by high correlation and similar regression). | 0.9997 | 0.97x + 0.161 | 0.999 | 0.992x + 0.040 | (Not applicable for this analyte) | (Not applicable for this analyte) |
| LAC | Functional equivalence to predicate devices (demonstrated by high correlation and similar regression). | 1.0000 | 1.012x - 0.0272 | (Not applicable for this analyte) | (Not applicable for this analyte) | 0.9995 | 0.9398x + 0.0692 |
Note: For the LAC analyte, the comparison is specifically against Ortho-Clinical Diagnostics VITROS Calibrator Kit 1, which primarily targets lactate.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: The document states that performance was evaluated on "pre-production lots of VALIDATE Chem 9 Calibration Verification Test Set." It also mentions, "Each analyte was tested in triplicate." However, the specific number of distinct samples or cases in the test set is not explicitly stated. The core of the test involves comparing values obtained from the VALIDATE Chem 9 kit to those from predicate devices across multiple concentration levels (5 levels for VALIDATE Chem 9 and DOCUMENT Serum, 3 levels for Ortho-Clinical VITROS).
- Data Provenance: The document does not specify the country of origin of the data. The study appears to be prospective in nature, as it uses "pre-production lots" of the VALIDATE Chem 9 device for comparison.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
This type of study does not involve "experts" in the human diagnostic sense (e.g., radiologists). Instead, the "ground truth" or reference values are established by the well-characterized and legally marketed predicate devices (DOCUMENT Serum Multi-Analyte CAL·VER and Ortho-Clinical Diagnostics VITROS Calibrator Kit 1). The performance of these predicate devices on the VITROS instrument system serves as the benchmark against which the new device is compared.
4. Adjudication Method for the Test Set
Not applicable. This study involves a direct measurement comparison between calibration verification materials on an instrument system, not human interpretation requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This is a study comparing in vitro diagnostic calibration materials, not a human-in-the-loop diagnostic imaging study involving human readers or AI.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
This study is inherently a "standalone" performance evaluation of the calibration verification test set, as it assesses its behavior on an instrument system (specifically the VITROS instrument system). There is no human intervention in the performance of the calibration verification process itself, beyond the initial setup by laboratory professionals. The comparison is between the new device and existing devices, all performing their function as "algorithms" (or chemical/physical standards) on the instrument.
7. The Type of Ground Truth Used
The "ground truth" is established by the performance of the legally marketed predicate devices (DOCUMENT Serum Multi-Analyte CAL·VER and Ortho-Clinical Diagnostics VITROS Calibrator Kit 1) on the VITROS instrument system. This can be considered a form of reference standard comparison against established and accepted laboratory calibration materials.
8. The Sample Size for the Training Set
Not applicable. This device is a chemical calibration verification test set, not an AI/ML algorithm that requires a training set. Its "performance" is inherent to its chemical composition and how it interacts with analytical instruments.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as this is not an AI/ML device requiring a training set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Maine Standards Company LLC. The word "maine" is on the top left, and the word "standards" is on the bottom right. A diagonal line separates the two words, and the words "COMPANY LLC" are below the word "standards".
1023352
NOV 22 2002
510(k) SUMMARY
"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."
"The assigned 510(k) number is: _
Maine Standards Company Submitter: Address: 765 Roosevelt Trail Windham, ME 04062 207-892-1300 Telephone: Fax: 207-892-2266 Christine Beach, Mgr. RA/QA Contact:
Summary prepared on: September 30, 2002
| Proprietary Name: | VALIDATE Chem 9 Calibration Verification Test Set |
|---|---|
| Common Name: | Calibration Verification |
| Classification Name: | Calibrator, Multi-Analyte |
Predicate Devices:
-
- DOCUMENT Serum Multi-Analyte CAL-VER, K950469, manufactured by CASCO NERL Diagnostics.
-
- Ortho-Clinical Diagnostics VITROS Calibrator Kit 1
Device description: VALIDATE Chem 9 Calibration Verification Test Set contains purified chemicals in a protein matrix. Multiple levels are provided to establish the relationship between theoretical operation and actual performance of each of the included analytes. Each set contains one bottle each of five (5) levels. Each bottle contains 5 milliliters.
Intended use: VALIDATE Chem 9 Calibration Verification Test Set is intended for in vitro diagnostic use for quantitatively verifying calibration, validating reportable ranges, and determining linearity for automated and manual chemistry systems for the following eighteen analytes: albumin (ALB), calcium (CA), cholesterol (CHOL), chloride (CL), carbon dioxide (CO2), creatinine (CRE), glucose (GLU), iron (FE), lactate (LAC), lithium (LI), magnesium (MG), phosphorus (PQ4), potassium (K), sodium (NA), total protein (TP), triglyceride (TRIG), urea nitrogen (BUN), and uric acid (UA).
{1}------------------------------------------------
Comparison of VALIDATE Chem 9 Calibration Verification Test Set to the predicate devices:
Table 1 compares characteristics of the VALIDATE Chem 9 Calibration Verification Test Set with those of the DOCUMENT Serum Multi-Analyte CAL.•VER and Ortho-Clinical Diagnostics VITROS Calibrator Kit 1.
| VALIDATE CHEM 9CalibrationVerification Test Set | DOCUMENT SerumMulti-AnalyteCAL·VER | Ortho-ClinicalDiagnostics VITROSCalibrator Kit 1 | |
|---|---|---|---|
| Catalog # | 109 | M-115 | 1882208 |
| Intended Use | For in vitro diagnosticuse in quantitativelyverifying calibration,validating reportableranges, anddetermining linearity inautomated, semi-automated and manualchemistry systems | For in vitro diagnosticuse in the quantitativedetermination oflinearity in manual,automated and semi-automated chemistrysystems. | For use in thecalibration of VITROSchemistry systems forthe quantitativemeasurementof lactate. |
| Analytes | ALB, CA, CHOL, CL,CO2, CREAT, GLU, FE,LAC, LITH, MG, PO4,K, NA, TPROT, TRIG,BUN, URIC ACID | ALB, CA, CHOL, CL,CO2, CREAT, GLU, FE,LITH, MG, PO4, K, NA,TPROT, TRIG, BUN,URIC ACID | LAC |
| Matrix | Bovine serum albumin | Human serum | Bovine serum albumin |
| Number ofLevels | 5 | 5 | 3 |
| Preparation | Liquid,ready to use | Liquid,ready to use | Lypholized |
| Packaging | 5.0 mL each level | 2 x 3.0 mL each level | 3.0 mL each level |
| Stability | 9 months | 9 months | 24 hours afterreconstitution |
| Storage | -10 to -20°C | -10 to -20°C | -10 to -20°C |
Comparison of Products TABLE 1.
The performance of VALIDATE Chem 9 Calibration Verification Test Set solutions on the VITROS instrument system as compared to DOCUMENT Serum Multi-Analyte CAL.•VER and Ortho-Clinical Diagnostics VITROS Calibrator Kit 1 been shown to be substantially equivalent using pre-production lots of VALIDATE Chem 9 Calibration Verification Test Set. The results of correlation comparisons between the VALIDATE Chem 9 Calibration Verification Test Set and the predicate devices are presented in Table 2.
{2}------------------------------------------------
| VALIDATE | DOCUMENT | |||
|---|---|---|---|---|
| Chem 9 Calibration Verification Test Set | Serum Multi-Analyte CAL•VER | |||
| Analyte | Correlation Coefficient(r) | Regression EquationY = slope(X) + intercept | Correlation Coefficient(r) | Regression EquationY = slope(X) + intercept |
| ALB | 0.9942 | x - 0.1 | 0.998 | 0.997x + 0.022 |
| BUN | 0.9999 | 1.0138x - 0.4276 | 0.999 | 1.005x + 0.618 |
| CA | 0.9974 | 1.01x + 0.288 | 0.999 | 0.995x + 0.035 |
| CHOL | 0.9999 | 0.9935x + 1.8102 | 0.995 | 1.058x - 6.192 |
| CL | 0.9996 | 0.9581x + 3.6226 | 0.998 | 1.010x - 1.389 |
| CO2 | 0.9971 | 0.875x + 1.675 | 0.998 | 1.009x - 0.176 |
| CRE | 0.9986 | 0.9364 + 0.1791 | 0.999 | 0.974x + 0.073 |
| FE | 0.9988 | 0.9205x + 14.617 | 0.998 | 1.119x - 14.492 |
| GLU | 0.9999 | 1.0132x - 0.7707 | 0.999 | 1.050x - 9.370 |
| K | 0.9987 | 0.9984x - 0.2576 | 0.999 | 0.983 + 0.170 |
| LITH | 0.9997 | 0.9733x + 0.0187 | 0.994 | 0.924x + 0.015 |
| MG | 0.9993 | 0.9832 + 0.1053 | 0.998 | 1.051x - 0.082 |
| NA | 0.9991 | 1.0042x + 1.8979 | 0.998 | 1.076x - 7.561 |
| PO4 | 0.9998 | 0.975x + 0.035 | 0.999 | 0.992x + 0.093 |
| TP | 0.9986 | 1.097x - 0.5097 | 0.998 | 1.086x - 0.261 |
| TRIG | 0.9999 | 1.0226x - 3.6259 | 0.999 | 1.018x + 2.292 |
| URICACID | 0.9997 | 0.97x + 0.161 | 0.999 | 0.992x + 0.040 |
| TABLE 2. Linear Regression Statistical Comparison of VALIDATE Chem 9 | ||||
|---|---|---|---|---|
| Calibration Verification Test Set to the predicate devices. |
| VALIDATEChem 9Calibration Verification Test Set | Ortho-Clinical Diagnostics VITROSCalibrator Kit 1 | ||||
|---|---|---|---|---|---|
| Analyte | Correlation Coefficient(r) | Regression EquationY = slope(X) + intercept | Correlation Coefficient(r) | Regression EquationY = slope(X) + intercept | |
| LAC | 1.0000 | 1.012x - 0.0272 | 0.9995 | 0.9398x + 0.0692 |
Summary:
Linear regression analysis was carried out on recovered values for each analyte. Each analyte was tested in triplicate. The VALIDATE Chem 9 Calibration Verification Test Set has been shown to be functionally equivalent for calibration verification and linearity assessment to DOCUMENT Serum Multi-Analyte CAL.�VER and Ortho-Clinical Diagnostics VITROS Calibrator Kit 1.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the edge. Inside the circle is a stylized image of three birds in flight, one behind the other. The birds are black and have a simple, abstract design.
od and Drug Administra 98 Gaither Road ckville MD 20850
OV 22 2002
Ms. Christine Beach Manager, RA/QA Maine Standards Company 765 Roosevelt Trail - Suite 9A Windham, ME .04062
Re: K023352
Trade/Device Name: Validate Chem 9 Calibration Verification Test Set Regulation Number: 21 CFR 862.1660 Regulation Name: Quality control material (assayed and unassayed) Regulatory Class: Class I Product Code: JJY Dated: September 30, 2002 Received: October 7, 2002
Dear Ms. Beach:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{4}------------------------------------------------
Page 2 – 2017 - 12:00 PM - 12 - 11:00 PM - 12 - 11 - 12 - 11 - 11 - 11 - 11 - 1
- 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 - 11 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours.
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
INDICATIONS FOR USE STATEMENT
510(k) Number: ________________
Device Name: Validate Chem 9 Calibration Verification Test Set
Indications for Use:
The VALIDATE Chem 9 Calibration Verification Test Set is used by trained laboratory professionals for the quantitative determination of linearity, calibration verification and verification of reportable range in automated, semi-automated, and manual clinical chemistry systems for the following eighteen analytes. albumin, calcium, cholesterol, chloride, carbon dioxide, creatinine, glucose, iron, lactate, lithium, magnesium, phosphorus, potassium, sodium, total protein, triglyceride, urea nitrogen, and uric acid.
Gran Crogn
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K023352
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use
§ 862.1660 Quality control material (assayed and unassayed).
(a)
Identification. A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.(b)
Classification. Class I (general controls). Except when intended for use in donor screening tests, quality control materials (assayed and unassayed) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.