K Number
K022905
Date Cleared
2002-12-13

(101 days)

Product Code
Regulation Number
876.5540
Panel
GU
Reference & Predicate Devices
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The LifeJet™ F16 Hemodialysis/Apheresis Chronic Catheter is indicated for use in attaining short and long term vascular access for hemodialysis, hemoperfusion or apheresis therapy via the jugular or subclavian vein. The catheter is intended for implantation dwell time of greater than 30 days.

Device Description

The LifeJet™ F16 Hemodialysis/Apheresis Chronic Catheter is a radiopaque dual lumen polyurethane catheter intended to remove and return blood. The two lumens are designed in a Circle Crescent (Circle "C") configuration. The distal venous lumen extends beyond the arterial lumen to reduce recirculation.

The lumens are connected to the extensions via a molded hub with suture wing. The arterial and venous extensions are identified by red and blue luer connectors and clamps. Priming volume information is printed on identification rings on the clamps for ease in identification. The fixed retention cuff on the shaft provides an anchoring site for tissue ingrowth during long-term placement.

The LifeJet™ F16 Hemodialysis/Apheresis Chronic Catheter is available in varied implantable lengths, with straight or J-Cannula versions.

AI/ML Overview

This 510(k) summary is for the LifeJet™ F16 Hemodialysis/Apheresis Chronic Catheter.
Here's an analysis of the provided text for acceptance criteria and supporting studies:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria / Performance MetricReported Device Performance
Tensile StrengthMet (Implied by "sufficient to demonstrate safety and effectiveness by way of comparison to the legally marketed predicate device")
Catheter Pressurization/Leak TestingMet (Implied by "sufficient to demonstrate safety and effectiveness by way of comparison to the legally marketed predicate device")
Flow RateMet (Implied by "sufficient to demonstrate safety and effectiveness by way of comparison to the legally marketed predicate device")
Catheter StiffnessMet (Implied by "sufficient to demonstrate safety and effectiveness by way of comparison to the legally marketed predicate device")
Substantial Equivalence to PredicateAchieved (FDA determination)

Notes:

  • The document states that "in-vitro testing was sufficient to demonstrate safety and effectiveness by way of comparison to the legally marketed predicate device." This implies that the predicate device's performance metrics served as the de facto acceptance criteria for these in-vitro tests, and the LifeJet™ F16 met or performed comparably to these.
  • Specific numerical acceptance thresholds or performance values are not provided in this document. The FDA's substantial equivalence determination supports that the device met the necessary criteria.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not explicitly stated. The performance data is based on "in-vitro performance data," which refers to laboratory testing, not human subjects. The number of samples/units tested for each in-vitro test (tensile strength, leak testing, flow rate, stiffness) is not provided.
  • Data Provenance: The data is in-vitro (laboratory testing). It is not derived from human subjects, so concepts like country of origin or retrospective/prospective do not directly apply in the same way they would for clinical data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • Not Applicable. This study did not involve human expert interpretation of data to establish a ground truth. Performance was determined through physical and mechanical testing comparing the device's characteristics to a predicate device.

4. Adjudication Method for the Test Set

  • Not Applicable. There was no human interpretation or adjudication process for the in-vitro tests described.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

  • No. An MRMC study was not conducted as this study focused on in-vitro performance testing of a physical medical device, not on AI or diagnostic imaging interpretation.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

  • Not Applicable. This pertains to a physical medical device (catheter), not an algorithm or AI system.

7. The Type of Ground Truth Used

  • For the in-vitro performance tests: The ground truth was based on the established performance characteristics and safety profile of the legally marketed predicate device (Super C™ (LifeJet™) Hemodialysis/Apheresis Chronic Catheter). The new device's performance was compared to this predicate to demonstrate substantial equivalence.

8. The Sample Size for the Training Set

  • Not Applicable. This is a physical medical device without a "training set" in the context of machine learning or algorithms.

9. How the Ground Truth for the Training Set Was Established

  • Not Applicable. As there is no training set for this type of device submission, this question is not relevant.

§ 876.5540 Blood access device and accessories.

(a)
Identification. A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device.
(2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle).
(3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).
(b)
Classification. (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided.
(ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated.
(B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided.
(C) Priming volumes must be established.
(D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(E) Air leakage testing and liquid leakage testing must be conducted.
(F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage.
(G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern.
(H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established.
(iii) Performance data must demonstrate the sterility of the device.
(iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing.
(v) Labeling of implanted blood access devices for hemodialysis must include the following:
(A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated.
(B) Labeling must specify the forward and reverse recirculation rates.
(C) Labeling must provide the arterial and venous priming volumes.
(D) Labeling must specify an expiration date.
(E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device.
(F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient.
(G) Labeling must include a patient implant card.
(H) The labeling must contain comprehensive instructions for the following:
(
1 ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;(
2 ) Proper care and maintenance of the device and device exit site;(
3 ) Removal of the device;(
4 ) Anticoagulation;(
5 ) Management of obstruction and thrombus formation; and(
6 ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following:
(A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port.
(B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following:
(A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied.
(B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties.
(C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens.
(D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness.
(viii) The following must be included for A-V shunt cannulae (with vessel tips):
(A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply.
(B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(2) Class II (performance standards) for the nonimplanted blood access device.
(3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section.
(4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.