K Number
K022769
Manufacturer
Date Cleared
2002-11-13

(84 days)

Product Code
Regulation Number
862.1660
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

LifeScan OneTouch Ultra control solution is intended for use to verify the performance of LifeScan BGM systems using the OneTouch Ultra or OneTouch FastTake test strips at glucose levels within the reportable range. The Glucose Control Solution is intended for use by healthcare professionals and people with diabetes mellitus at home. For In Vitro Diagnostic Use

Device Description

LifeScan OneTouch Ultra control solution is a single-level, viscosity-adjusted, aqueous liquid glucose control solution. LifeScan OneTouch Ultra control solution provides a convenient method of performing periodic QC checks for laboratories selecting to measure liquid QC material as a part of their quality assurance program. The product is packaged in plastic bottles with dropper tips for application of the solution to test strips. The control has a red color to help users see the solution while dispensing onto a test strip. LifeScan OneTouch Ultra control solution contains glucose representing typical, normal-range glucose values in blood to verify performance of LifeScan BGM systems using the OneTouch Ultra or OneTouch FastTake test strips. LifeScan OneTouch Ultra control solution is a non-hazardous aqueous solution containing no biological materials.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the LifeScan OneTouch Ultra control solution, based on the provided text:

Acceptance Criteria and Reported Device Performance

The submission is for an extension of an existing control solution, stating "There has been no change in formulation, labeling or intended use of the control solution since the original 510(k) submission, K000318." Therefore, the acceptance criteria are implicitly tied to the performance requirements of the control solution's ability to verify the various LifeScan BGM systems. The reported performance is that the control solution successfully addresses these requirements.

Acceptance Criteria CategoryReported Device Performance
Closed bottle stabilityVerified
Stability after openingVerified
Test precision and rangeVerified

The control solution's intended use is "to monitor and evaluate the analytical performance of LifeScan BGM family systems using the OneTouch Ultra or OneTouch FastTake test strips." The device performance is deemed adequate because it provides a convenient method of performing periodic QC checks and contains glucose representing typical, normal-range glucose values in blood to verify performance.

1. Sample Sized Used for the Test Set and Data Provenance

The provided 510(k) summary does not explicitly state the sample sizes (number of tests, number of control solution lots, etc.) used for the non-clinical tests (closed bottle stability, stability after opening, test precision and range). No detailed testing protocols or results are included in this summary.

The data provenance is not specified beyond "Tests were conducted to verify specific performance requirements." It does not mention the country of origin of the data or whether it was retrospective or prospective.

2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. This device is a control solution for glucose meters, not a diagnostic device requiring expert interpretation of results. The "ground truth" for a control solution would be its known glucose concentration, established through a traceable reference method, not through expert consensus on diagnostic images or pathology.

3. Adjudication Method for the Test Set

Not applicable. As this is not a diagnostic device involving subjective interpretation, there is no adjudication process for the test set.

4. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a control solution for blood glucose meters, not an AI-powered diagnostic device or a system for human readers.

5. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is not directly applicable in the context of a control solution. A control solution itself operates "standalone" in the sense that its chemical properties and stability are assessed intrinsically. It doesn't involve a separate algorithm for interpretation in the way an AI diagnostic tool would. Its performance is measured by how well it verifies the performance of the BGM system.

6. The Type of Ground Truth Used

The ground truth for a glucose control solution's performance stability, precision, and range would be:

  • Reference Glucose Concentration: The known, accurately measured glucose concentration of the control solution itself, established through highly precise and traceable chemical analytical methods.
  • Performance of Predicate Device/System: The expected and validated performance (accuracy, precision) of the OneTouch Ultra and OneTouch FastTake glucose meters when measuring known glucose concentrations. The goal of the control solution is to ensure these meters perform within their established specifications.

7. The Sample Size for the Training Set

Not applicable. This is a chemical control solution, not a machine learning or AI model that requires a training set.

8. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for a chemical control solution.

{0}------------------------------------------------

KC022769

BIONOSTI

NOV 13 2002

510(k) Summary1

  • (a) (1) Submitter's name, address Bionostics, Inc. 7 Jackson Road Devens, MA 01432
    Contact Person Kathleen Storro Director, QA & Regulatory Affairs (978) 772-7070 x 220

Date of preparation of this summary: 15 August 2002

  • (2) Device trade or proprietary name: LifeScan OneTouch Ultra control solution

Device common or usual name or classification name:

Single Analyte Control, All Types (Assayed and Unassayed)

CLASSIFICATION
PRODUCT NOMENCLATURENUMBERCLASSPANEL
SINGLE ANALYTE CONTROL SOLUTION862.1660 75 JJXICHEMISTRY

(3) Substantial Equivalence

There has been no change in formulation, labeling or intended use of the control solution since the original 510(k) submission, K000318. The purpose of this 510(k) submission is to extend the use of this control to all LifeScan family of meters using the OneTouch Ultra or OneTouch FastTake glucose test strips, namely:

OneTouch FastTake Blood Glucose Meter, K011479 InDuo Blood Glucose Meter, K011616 UltraSmart Blood Glucose Meter, K021819

(4) Description of the new device

LifeScan OneTouch Ultra control solution is a single-level, viscosity-adjusted, aqueous liquid glucose control solution. LifeScan OneTouch Ultra control solution provides a convenient method of performing periodic QC checks for laboratories selecting to measure liquid QC material as a part of their quality assurance program. The product is packaged in plastic bottles with dropper tips for application of the solution to test strips. The control has a red color to help users see the solution while dispensing onto a test strip.

LifeScan OneTouch Ultra control solution contains glucose representing typical, normal-range glucose values in blood to verify performance of LifeScan BGM systems using the OneTouch Ultra or OneTouch FastTake test strips.

4 This summary of safety and effectiveness is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

{1}------------------------------------------------

LifeScan OneTouch Ultra control solution is a non-hazardous aqueous solution containing no biological materials.

(5) Intended use of the device

LifeScan OneTouch Ultra control solution is intended to be used to monitor and evaluate the analytical performance of LifeScan BGM family systems using the OneTouch Ultra or OneTouch FastTake test strips.

(6) Technological characteristics of the device.

This material consists of viscosity-adjusted, aqueous glucose control solution prepared in to represent normal, whole blood glucose values. The solution has been optimized to simulate the response of whole blood on LifeScan BGM family systems using the OneTouch Ultra or OneTouch FastTake test strips.

(b) (1) Summary of non-clinical tests submitted with the premarket notification for the device.

Tests were conducted to verify specific performance requirements:

  • a) Closed bottle stability
  • b) Stability after opening
  • c) Test precision and range

Summary of clinical tests submitted with the premarket notification for the (b) (2) device.

N/A

(b) (3) Conclusions drawn from the clinical and non-clinical trials.

Comparison of technological characteristics, formulation and intended use to predicate devices listed in this summary support the claim of substantial equivalence.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three wavy lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

NOV 1 3 2002

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Ms. Kathleen Storro Director, QA and Regulatory Affairs Bionostics, Inc. 7 Jackson Road Devens, MA 01432

Re: K022769

Trade/Device Name: LifeScan OneTouch Ultra Control Solution Regulation Number: 21 CFR 862.1660 Regulation Name: Quality Control Material (Assayed and Unassayed) Regulatory Class: Class I Product Code: JJX Dated: August 19, 2002 Received: August 21, 2002

Dear Ms. Storro:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

. Page 2 -

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number

LifeScan OneTouch Ultra control solution Device Name:

Indications for Use:

LifeScan OneTouch Ultra control solution is intended for use to verify the performance of LifeScan BGM systems using the OneTouch Ultra or OneTouch FastTake test strips at glucose levels within the reportable range. The Glucose Control Solution is intended for use by healthcare professionals and people with diabetes mellitus at home.

For In Vitro Diagnostic Use

ttan Cooper

Division Sign-Off) Division of Clinical Laboratory D 510(k) Number .

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109) OR

Over-The-Counter Use

(Optional Format 1-2-96)

OneTouch Ultra 510(k)

§ 862.1660 Quality control material (assayed and unassayed).

(a)
Identification. A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.(b)
Classification. Class I (general controls). Except when intended for use in donor screening tests, quality control materials (assayed and unassayed) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.