K Number
K022724

Validate with FDA (Live)

Manufacturer
Date Cleared
2002-09-12

(27 days)

Product Code
Regulation Number
862.1345
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SURESTEP System is for quantitative measurement of glucose in whole blood. It may be used by persons with diabetes in the home and by healthcare professionals in clinical settings as an aid to monitor effectiveness of diabetes control.

Device Description

The LifeScan SURESTEP Blood Glucose Monitoring System consists of a glucose reagent test strip, a reflectance photometer, and a quality control solution. A lancing device and lancets are also included with the system. The user obtains a blood sample and applies it to the test strip before inserting the strip into the meter. The user obtains a blood sample and applies it to the test strip before inserting the strip into the meter. After inserting the strip into the meter, the glucose value displays on the meter's liquid crystal display (LCD). The test strip color can be compared visually to a color chart provided with the system to estimate if the glucose concentration is very high or very low as a backup to the meter.

When blood is placed on the test strip, glucose in the blood sample reacts with oxygen to yield gluconic acid and hydrogen peroxide. This reaction is catalyzed by the enzyme glucose oxidase present in the test strip. Hydrogen peroxide then reacts with the indicator dye, catalyzed by the presence of peroxidase in the test strip, creating a blue color proportional to the concentration of glucose present in the blood sample. An absorbent layer in the test strip acts as a reservoir for excess blood.

The blood glucose meter is a reflectance photometer that measures the light reflectance of the blue color that develops on the test strip. The meter electronically converts the reflectance data to a digital result that is displayed on a liquid crystal display (LCD). The meter is powered by two 1.5 V AAA alkaline batteries. The user calibrates the meter for each lot number of test strips in use by entering a calibration code specific to that lot number. A quality control solution is included in the system so that the user can check system performance.

AI/ML Overview

This document only provides a 510(k) Summary for the SURESTEP® Blood Glucose Monitoring System, which primarily focuses on establishing "substantial equivalence" to a predicate device. It does not present a detailed study with acceptance criteria and performance data in the way typically expected for a rigorous clinical or performance evaluation study of a new medical device.

Therefore, most of the requested information cannot be extracted from the provided text. The document states that the existing labeling was "reformatted and simplified to increase understanding and provide clear explanations of the performance capabilities and the performance limitations of the system," but it does not provide those performance capabilities or the study details behind them.

Here's an attempt to answer the questions based only on the provided text, highlighting what is not available:


1. A table of acceptance criteria and the reported device performance

This information is not provided in the 510(k) Summary. The summary focuses on comparing the modified device to a predicate device and stating substantial equivalence, not on presenting specific performance data against defined acceptance criteria.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not provided in the 510(k) Summary. There is no mention of a specific test set or data provenance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not provided in the 510(k) Summary. Blood glucose monitoring systems typically use laboratory reference methods (e.g., YSI analyzer) as ground truth, not expert consensus in the same way imaging devices might. However, even the type of ground truth used is not specified.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not provided in the 510(k) Summary. Adjudication methods are typically relevant for subjective assessments, which are not usually part of blood glucose meter performance testing.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

An MRMC study is not applicable to this type of device (blood glucose meter). This document describes a standalone diagnostic device where the result is read by the user, not an AI-assisted diagnostic intended to improve human interpretation of complex data.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, the device described is a standalone device that performs a quantitative measurement of glucose. The user obtains a blood sample, applies it to the test strip, inserts it into the meter, and the glucose value is displayed. This is an "algorithm only" (or rather, a measurement system only) performance in the sense that the meter provides a direct numerical reading without human interpretation of complex images or data that an AI might assist with. However, no specific standalone study details or performance metrics are provided in this summary.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

This information is not explicitly stated in the 510(k) Summary. For blood glucose meters, the typical ground truth in performance studies is a highly accurate laboratory reference method (e.g., YSI glucose analyzer), but this document does not specify.

8. The sample size for the training set

This information is not provided in the 510(k) Summary. This device is a reflectance photometer based on a chemical reaction, not a machine learning or AI algorithm that would typically have a "training set" in the common sense.

9. How the ground truth for the training set was established

This information is not provided as the concept of a "training set" does not directly apply in the context of this device described in the summary.


Summary of what is available from the document:

  • Device Name: SURESTEP® Blood Glucose Monitoring System
  • Intended Use: Quantitative measurement of glucose in whole blood by persons with diabetes in the home and by healthcare professionals in clinical settings as an aid to monitor effectiveness of diabetes control.
  • Predicate Device: SURESTEP Blood Glucose Monitoring System (implying the submission is for a modified version of an existing device).
  • Nature of Modification: Labeling reformatted and simplified, and the visual backup system modified by eliminating comparison colors equivalent to 100 mg glucose/dL and 180 mg glucose/dL.
  • Conclusion: The modified device is substantially equivalent to the predicate device.

The provided text simply serves as a 510(k) summary for FDA clearance based on substantial equivalence, not a detailed scientific paper outlining a performance study.

{0}------------------------------------------------

16022724

510(k) Summarv

SEP 1 2 2002

SubmitterLifeScan, Inc.1000 Gibraltar DriveMilpitas, CA 95035Contact: John E. HughesDate Prepared: August 12, 2002
Device NameSURESTEP® Blood Glucose Monitoring SystemCommon name: Glucose test system
Predicate DeviceSURESTEP Blood Glucose Monitoring System

Device Description

The LifeScan SURESTEP Blood Glucose Monitoring System consists of a glucose reagent test strip, a reflectance photometer, and a quality control solution. A lancing device and lancets are also included with the system. The user obtains a blood sample and applies it to the test strip before inserting the strip into the meter. The user obtains a blood sample and applies it to the test strip before inserting the strip into the meter. After inserting the strip into the meter, the glucose value displays on the meter's liquid crystal display (LCD). The test strip color can be compared visually to a color chart provided with the system to estimate if the glucose concentration is very high or very low as a backup to the meter.

When blood is placed on the test strip, glucose in the blood sample reacts with oxygen to yield gluconic acid and hydrogen peroxide. This reaction is catalyzed by the enzyme glucose oxidase present in the test strip. Hydrogen peroxide then reacts with the indicator dye, catalyzed by the presence of peroxidase in the test strip, creating a blue color proportional to the concentration of glucose present in the blood sample. An absorbent layer in the test strip acts as a reservoir for excess blood.

The blood glucose meter is a reflectance photometer that measures the light reflectance of the blue color that develops on the test strip. The meter electronically converts the reflectance data to a digital result that is displayed on a liquid crystal display (LCD). The meter is powered by two 1.5 V AAA alkaline batteries. The user calibrates the meter for each lot number of test strips in use by entering a calibration code specific to that lot number. A quality control solution is included in the system so that the user can check system performance.

Intended Use

The SURESTEP System is for quantitative measurement of glucose in whole blood. It may be used by persons with diabetes in the home and by healthcare professionals in clinical settings as an aid to monitor effectiveness of diabetes control.

{1}------------------------------------------------

Comparison to Predicate Device

The existing labeling of SURESTEP® Blood Glucose Monitoring System has been reformatted and simplified to increase understanding and provide clear explanations of the performance capabilities and the performance limitations of the system. In addition, the visual backup system has been modified by eliminating comparison colors equivalent to 100 mg glucose/dL and 180 mg glucose/dL.

Conclusion

The modified SURESTEP Blood Glucose Monitoring System is substantially equivalent to the predicate device.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol. The logo is in black and white.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

SEP 1 2 2002

Mr. John E. Hughes Manager, Regulatory Submissions LifeScan, Inc. 1000Gibraltar Drive Milpitas, CA 95035-6312

Re: K022724

Trade/Device Name: SureStep Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: Class II Product Code: NBW Dated: August 15, 2002 Received: August 16, 2002

Dear Mr. Hughes:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device. please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmaldsmamain.html".

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use Statement

510(k) Number:

K022724

Device Name:

SURESTEP Blood Glucose Monitoring System

Indications for Use:

The SURESTEP System is for quantitative measurement of glucose in whole blood. It may be used by persons with diabetes in the home and by healthcare professionals in clinical settings as an aid to monitor effectiveness of diabetes control.

for Dr. Cooper
(Division Sign-Off)

Division of Clinical Laboratory Devi 510(k) Number .

Concurrence of CDRH, Office of Device Evaluation

Prescription Use (Per 21 CFR 801.109) OR

Over-the-Counter Use

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.