(196 days)
THE MEDCOMP SC4 HEMODIALYSIS CATHETER IS INDICATED FOR USE IN ATTAINING LONG-TERM VASCULAR ACCESS FOR HEMODIALYSIS AND APHERESIS. IT MAY BE INSERTED PERCUTANEOUSLY AND IS PRIMARILY PLACED IN THE INTERNAL JUGULAR VEIN. ALTERNATE INSERTION SITES INCLUDE THE SUBCLAVIAN VEIN. CATHETERS GREATER THAN 40CM ARE INTENDED FOR FEMORAL VEIN INSERTION. THE SC4 EXTENSION SET IS INTENDED TO REPAIR THE SC4 HEMODIALYSIS CATHETER.
The Medcomp SC4 is a 16F polyurethane, double lumen catheter used to remove and return blood through two-secregated lumen passages. Both lumens are "D" shaped, tapered at the distal tip, each with a series of side holes. The distal venous lumen extends beyond the arterial lumen to reduce recirculation. The fixed polyester cuff allows for tissue ingrowth for long term placement. A removable suture wing hub is provided for securing the catheter after initial placement. The arterial and venous lumens are designed to be split, or peeled apart, prior to insertion to provide two free-floating lumens within the vessel. After the catheter has been positioned the catheter lumens can be trimmed and an extension set is assembled on the proximal end of the catheter. Red and blue luer connectors and clamps identify the arterial and venous extensions set. Priming volume information is printed on the catheter lumen.
Here's an analysis of the provided text regarding the Medcomp SC4 device, focusing on acceptance criteria and study details:
Summary of Acceptance Criteria and Device Performance for Medcomp SC4 Hemodialysis Catheter (K022678)
Based on the provided 510(k) summary, the Medcomp SC4 Hemodialysis Catheter's performance was evaluated through in-vitro testing to demonstrate substantial equivalence to predicate devices. There was no clinical data deemed necessary.
1. Table of Acceptance Criteria and Reported Device Performance:
Acceptance Criteria Category | Specific Performance Parameter | Reported Device Performance | Study Type |
---|---|---|---|
Mechanical Properties | Tensile Strength | Demonstrated substantial equivalence to predicate devices | In-vitro |
Mechanical Properties | Joint Strength | Demonstrated substantial equivalence to predicate devices | In-vitro |
Integrity | Leakage | Demonstrated substantial equivalence to predicate devices | In-vitro |
Hemodynamic Performance | Recirculation | Demonstrated substantial equivalence to predicate devices | In-vitro |
Hemodynamic Performance | Flow Performance | Demonstrated substantial equivalence to predicate devices | In-vitro |
Material Properties | Lumen Peel | Demonstrated substantial equivalence to predicate devices | In-vitro |
2. Sample Size Used for the Test Set and Data Provenance:
- Sample Size for Test Set: The document does not specify the exact sample sizes (number of catheters or test specimens) used for each in-vitro test.
- Data Provenance: The data is described as "In-vitro performance data," indicating it was generated in a controlled laboratory setting. There is no information regarding country of origin, nor is it retrospective or prospective clinical data.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:
- Not applicable. The ground truth for this device was established via validated in-vitro testing methods against established performance benchmarks or comparisons to predicate devices, not through expert human evaluation of a dataset.
4. Adjudication Method for the Test Set:
- Not applicable. As noted above, the evaluation was based on in-vitro test results and comparison to predicate devices, not on human adjudication of a clinical dataset.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:
- No. An MRMC comparative effectiveness study was not performed. The submission explicitly states: "Clinical data was not deemed necessary since substantial equivalence is addressed by way of comparison to a legally marketed device."
6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:
- Not applicable. This device is a physical medical device (hemodialysis catheter), not an algorithm or AI system.
7. The Type of Ground Truth Used:
- The "ground truth" for this submission was based on:
- Validated In-vitro Testing Methods: Standard engineering and material science tests to assess the physical and functional characteristics of the catheter.
- Performance of Legally Marketed Predicate Devices: The results of the in-vitro tests were compared against the known performance characteristics of the Medcomp Ash Split-Cath™ II (K020465) and Medcomp Bio-Flex™ CS (K971925) to demonstrate "substantial equivalence."
8. The Sample Size for the Training Set:
- Not applicable. There was no "training set" as this is not an AI/ML device. The device's design and manufacturing likely involved internal R&D and testing, but these are not referred to as a "training set" in the context of this 510(k) summary.
9. How the Ground Truth for the Training Set Was Established:
- Not applicable, as there was no training set for an algorithm. The design and validation of the catheter were based on engineering principles and regulatory requirements for medical devices.
§ 876.5540 Blood access device and accessories.
(a)
Identification. A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device.
(2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle).
(3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).
(b)
Classification. (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided.
(ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated.
(B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided.
(C) Priming volumes must be established.
(D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(E) Air leakage testing and liquid leakage testing must be conducted.
(F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage.
(G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern.
(H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established.
(iii) Performance data must demonstrate the sterility of the device.
(iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing.
(v) Labeling of implanted blood access devices for hemodialysis must include the following:
(A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated.
(B) Labeling must specify the forward and reverse recirculation rates.
(C) Labeling must provide the arterial and venous priming volumes.
(D) Labeling must specify an expiration date.
(E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device.
(F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient.
(G) Labeling must include a patient implant card.
(H) The labeling must contain comprehensive instructions for the following:
(
1 ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;(
2 ) Proper care and maintenance of the device and device exit site;(
3 ) Removal of the device;(
4 ) Anticoagulation;(
5 ) Management of obstruction and thrombus formation; and(
6 ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following:
(A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port.
(B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following:
(A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied.
(B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties.
(C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens.
(D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness.
(viii) The following must be included for A-V shunt cannulae (with vessel tips):
(A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply.
(B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(2) Class II (performance standards) for the nonimplanted blood access device.
(3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section.
(4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.