(90 days)
Pourchez RetrO™ Repair Kit is designed to replace worn or damaged extension connector adapters on Spire Biomedical, Inc.'s Pourchez RetrO catheters.
Pourchez RetrO™ Repair Kit is designed to replace worn or damaged extension connector adapters on Spire Biomedical, Inc.'s Pourchez RetrO Twin Lumen Silicone Hemodialysis Catheters with Separated Tips.
Here's an analysis of the provided text regarding the Pourchez RetrO Repair Kit, focusing on acceptance criteria and the study:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Minimum tensile strength required by ISO 10555-1 | "exceed the minimum tensile strength required by ISO 10555-1" |
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify the exact sample size used for the tensile tests. It only states "A series of tensile tests were performed."
The data provenance (country of origin, retrospective/prospective) is not mentioned.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
Not applicable. The study involves tensile strength testing against an ISO standard, not expert review of medical images or conditions.
4. Adjudication Method for the Test Set
Not applicable, as it is a physical device performance test against a standard, not a subjective assessment requiring adjudication.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done
No. This type of study is not applicable to
a physical device performance test.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study was Done
Not applicable. This is not an AI/algorithm-based device.
7. The Type of Ground Truth Used
The ground truth used is the minimum tensile strength required by ISO 10555-1. This is a recognized international standard for sterile intravascular catheters.
8. The Sample Size for the Training Set
Not applicable. There is no mention of a training set as this is not an AI/machine learning device.
9. How the Ground Truth for the Training Set was Established
Not applicable, as there is no training set for this type of device.
Summary of the Study and Device Performance:
The Spire Biomedical, Inc. Pourchez RetrO Repair Kit was evaluated through a series of tensile tests. The acceptance criteria for the device's performance was its ability to meet or exceed the minimum tensile strength specified by the ISO 10555-1 standard for hemodialysis catheters. The study concluded that the repair kit's extension adapters successfully "exceed the minimum tensile strength required by ISO 10555-1," demonstrating its substantial equivalence or conformity to established standards. The exact number of units tested (sample size) for these tensile tests is not detailed in the provided K510 summary.
{0}------------------------------------------------
Spire Biomedical, Inc. • One Patriots Park • Bedford, MA 01730-2396 (781) 275-6001 · (781) 275-6010 fax
6 2007
NOV
Section 7 510(k) Summary Pourchez RetrO Repair Kit
Date: November 5, 2002
Submitter: Spire Biomedical, Inc. One Patriots Park Bedford, MA 01730-2396 Phone: (781) 275-6001 Fax: (781) 275-6010
- Contact Person: Donald Fickett Director of RAQA Spire Biomedical, Inc. Phone: (781) 275-6001 x221 Fax: (781) 275-6010 e-mail: dfickett@spirecorp.com
Device Names:
| Trade Name: | Pourchez RetrO Repair Kit |
|---|---|
| Common Name: | Repair Kit |
| Classification Name: Kit, Repair Catheter, Hemodialysis |
Legally Marketed Devices to Which Substantial Equivalence is Claimed:
-
- Spire Biomedical, Inc. Pourchez Retro Twin Lumen Silicone Chronic Hemodialysis Catheter with Separated Tips (extension adapters only) "K022000."
-
- Kendall Tandem-Cath™ (in design and intended use) "K002900"
-
- Bard Access Systems, Inc. Catheter Repair Kit with Replacement connector ( in intended use) "K011015"
Device Description: Pourchez RetrO™ Repair Kit is designed to replace worn or damaged extension connector adapters on Spire Biomedical, Inc.'s Pourchez RetrO Twin Lumen Silicone Hemodialysis Catheters with Separated Tips.
{1}------------------------------------------------
Ko 2644
Park 2022
Image /page/1/Picture/1 description: The image shows the logo for Spire Biomedical. The word "spire" is in a bold, sans-serif font, with the word "Biomedical" in a smaller, cursive font underneath. Above the word "spire" is a circular symbol with radiating lines, resembling a sun or a stylized flower.
Spire Biomedical, Inc. · One Patriots Park · Bedford, MA 01730-2396 (781) 275-6001 • (781) 275-6010 fax
510(k) Summary (Continued)
Pourchez RetrO Repair Kit
Intended Use: The Pourchez RetrO Repair Kit is designed to replace worn or damaged Pourchez RetrO extension adapters.
Technological Characteristics Comparison to Predicate Devices: The Pourchez RetrO Repair Kit uses the same materials of constructions as those of the original extension adapters supplied with the Pourchez RetrO catheters.
Performance Data: A series of tensile tests were performed to demonstrate substantial equivalence to predicate devices or conformation to established ISO standards for hemodialysis catheters. In all cases, the Pourchez RetrO Repair Kit extension adapters exceed the minimum tensile strength required by ISO 10555-1.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized design featuring three abstract human profiles facing to the right, with flowing lines representing hair or clothing.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Donald Fickett Director of RA/OA Spire Biomedical, Inc. Spire Corporation One Patriots Park BEDFORD MA 01730-2396
NOV 6 2002
Re: K022644
Trade/Device Name: Pourchez RetrO™ Repair Kit Regulation Number: 21 CFR §876.5540 Regulation Name: Blood access device and accessories Regulatory Class: II Product Code: 78 NFK Dated: August 7, 2002 Received: August 8, 2002
Dear Mr. Fickett:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Mr. Donald Fickett
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration (21 CFR Part 807); listing (21 CFR Part 807), labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if . applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on the labeling regulation (21 CFR Part 801 and additionally Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4616. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification''(21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely vours.
yr
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Spire Biomedical. The logo consists of the word "spire" in a bold, sans-serif font, with a stylized sun or starburst symbol above the "i". Below "spire" is the word "Biomedical" in a smaller, cursive font. The entire logo is in black and white.
APPENDIX B - Indications for Use Statement
Device Name: Pourchez RetrO Repair Kit
Indications for Use: Pourchez RetrO™ Repair Kit is designed to replace worn or damaged extension connector adapters on Spire Biomedical, Inc.'s Pourchez RetrO catheters.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Optional Format 3-10-98)
(Division Sign-Off)
Division of Reproduct and Radiological Devices 510(k) Number
Prescription Use . (Per 21 CFR 801.109)
§ 876.5540 Blood access device and accessories.
(a)
Identification. A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device.
(2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle).
(3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).
(b)
Classification. (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided.
(ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated.
(B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided.
(C) Priming volumes must be established.
(D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(E) Air leakage testing and liquid leakage testing must be conducted.
(F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage.
(G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern.
(H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established.
(iii) Performance data must demonstrate the sterility of the device.
(iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing.
(v) Labeling of implanted blood access devices for hemodialysis must include the following:
(A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated.
(B) Labeling must specify the forward and reverse recirculation rates.
(C) Labeling must provide the arterial and venous priming volumes.
(D) Labeling must specify an expiration date.
(E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device.
(F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient.
(G) Labeling must include a patient implant card.
(H) The labeling must contain comprehensive instructions for the following:
(
1 ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;(
2 ) Proper care and maintenance of the device and device exit site;(
3 ) Removal of the device;(
4 ) Anticoagulation;(
5 ) Management of obstruction and thrombus formation; and(
6 ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following:
(A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port.
(B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following:
(A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied.
(B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties.
(C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens.
(D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness.
(viii) The following must be included for A-V shunt cannulae (with vessel tips):
(A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply.
(B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(2) Class II (performance standards) for the nonimplanted blood access device.
(3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section.
(4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.