K Number
K022561
Manufacturer
Date Cleared
2002-08-23

(21 days)

Product Code
Regulation Number
876.5540
Panel
GU
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

To replace: Cracked or broken female luer lock connectors or repair damaged extension leg where there is a minimum of 4.5 cm of viable extension tubing on the following Vas-Cath catheters: Soft-Cell® Long-Term Dual Lumen Catheter, Opti-Flow® Long-Term Dual Lumen Catheter, Slim-Cath™ Short-Term Dual Lumen Catheter, Vaccess® Short-Term Single Lumen Catheter, Flexxicon® Short-Term Dual Lumen Catheter, Niagara™ Short-Term Dual Lumen Catheter, Flexxicon® II Short-Term Dual Lumen Catheter, HemoGlide™ Long-Term Dual Lumen Catheter.

Device Description

The Catheter Repair Kit with Replacement Connector is exactly the same as the predicate device.

AI/ML Overview

The provided document is a 510(k) summary for a Catheter Repair Kit with Replacement Connector. This type of submission is for demonstrating substantial equivalence to a legally marketed predicate device, rather than proving the safety and effectiveness of a novel device through clinical performance studies with acceptance criteria.

Therefore, this document does not contain information regarding:

  • Acceptance criteria and reported device performance in the context of a clinical study.
  • Sample sizes for test sets or training sets.
  • Data provenance (country of origin, retrospective/prospective).
  • Number/qualifications of experts for ground truth establishment.
  • Adjudication methods.
  • Multi-Reader Multi-Case (MRMC) comparative effectiveness studies or effect sizes.
  • Standalone algorithm performance.
  • Type of ground truth used (pathology, outcomes data, expert consensus).
  • How ground truth for a training set was established.

Instead, the document focuses on demonstrating substantial equivalence based on technological characteristics and intended use.

Here's a breakdown of what the document does state in relation to the prompt:

1. A table of acceptance criteria and the reported device performance:

  • Not Applicable. This document does not describe performance in terms of clinical acceptance criteria for a device. It asserts that the new device has "exactly the same" technological characteristics as the predicate device (K011015) and that the intended use is "identical" or "the same." The only difference noted is the addition of one catheter trade name (HemoGlide™) to the list of repairable catheters and a minor labeling consistency change (temporary to short-term).
  • The "performance" implied is that it will perform identically to the predicate device because it is technologically the same.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

  • Not Applicable. No test set or clinical data is presented or referenced for device performance. The 510(k) relies on comparison to a predicate.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

  • Not Applicable. No "ground truth" establishment by experts for a test set is described.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not Applicable. No test set or adjudication is mentioned.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not Applicable. This is not an AI (Artificial Intelligence) device, and no MRMC study or AI assistance is discussed.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not Applicable. This is a physical medical device (catheter repair kit), not an algorithm or software. "Standalone performance" in this context would typically refer to bench testing, which is assumed to be part of the product development but not detailed as a comparative study in this 510(k) summary.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • Not Applicable. The concept of "ground truth" in the context of clinical data for performance assessment is not relevant to this 510(k) submission. The basis for approval is substantial equivalence to a predicate device.

8. The sample size for the training set:

  • Not Applicable. There is no mention of a training set as this is not an AI/machine learning device.

9. How the ground truth for the training set was established:

  • Not Applicable. No training set or ground truth establishment for a training set is discussed.

In summary: The provided document is a 510(k) premarket notification that demonstrates substantial equivalence of a physical medical device (Catheter Repair Kit) to a predicate device, rather than presenting a clinical study with acceptance criteria, ground truth, or statistical performance metrics.

§ 876.5540 Blood access device and accessories.

(a)
Identification. A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device.
(2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle).
(3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).
(b)
Classification. (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided.
(ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated.
(B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided.
(C) Priming volumes must be established.
(D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(E) Air leakage testing and liquid leakage testing must be conducted.
(F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage.
(G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern.
(H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established.
(iii) Performance data must demonstrate the sterility of the device.
(iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing.
(v) Labeling of implanted blood access devices for hemodialysis must include the following:
(A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated.
(B) Labeling must specify the forward and reverse recirculation rates.
(C) Labeling must provide the arterial and venous priming volumes.
(D) Labeling must specify an expiration date.
(E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device.
(F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient.
(G) Labeling must include a patient implant card.
(H) The labeling must contain comprehensive instructions for the following:
(
1 ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;(
2 ) Proper care and maintenance of the device and device exit site;(
3 ) Removal of the device;(
4 ) Anticoagulation;(
5 ) Management of obstruction and thrombus formation; and(
6 ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following:
(A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port.
(B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following:
(A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied.
(B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties.
(C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens.
(D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness.
(viii) The following must be included for A-V shunt cannulae (with vessel tips):
(A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply.
(B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(2) Class II (performance standards) for the nonimplanted blood access device.
(3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section.
(4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.