K Number
K021793
Device Name
EBV EA-D IGG ELISA KIT, MODEL EBG-100
Manufacturer
Date Cleared
2002-09-27

(119 days)

Product Code
Regulation Number
866.3235
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
The Epstein Barr Virus Early Diffuse Antigen (EBV EA-D) IgG ELISA Test is for the qualitative detection of IgG antibodies to EBV EA-D as an aid in the diagnosis of EBV infection in patients with clinical symptoms of Infectious Mononucleosis (IM). The PANBIO EBV EA-D IgG ELISA should be used in conjunction with other EBV serologies.
Device Description
The EBV EA-D IgG ELISA Kit is an enzyme-linked immunosorbent assay (ELISA) for the detection of IgG antibodies to EBV EA-D antigen in human serum.
More Information

Trinity Biotech Captia™ EBV EA-D IgG ELISA

Not Found

No
The summary describes a standard ELISA assay for detecting antibodies and does not mention any AI or ML components.

No.
This device is an in-vitro diagnostic (IVD) test used for the qualitative detection of antibodies to aid in the diagnosis of EBV infection. It does not provide any direct therapeutic benefit or treatment to the patient.

Yes

The device aids in the diagnosis of EBV infection by detecting IgG antibodies, which is a diagnostic purpose.

No

The device is described as an ELISA kit, which is a laboratory-based assay involving physical reagents and procedures, not solely software.

Yes, this device is an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use explicitly states it is for the "qualitative detection of IgG antibodies to EBV EA-D as an aid in the diagnosis of EBV infection in patients with clinical symptoms of Infectious Mononucleosis (IM)." This clearly indicates it is used to test samples taken from the human body to provide information for diagnostic purposes.
  • Device Description: The device description states it is an "enzyme-linked immunosorbent assay (ELISA) for the detection of IgG antibodies to EBV EA-D antigen in human serum." This confirms it is a test performed on a biological sample (human serum) outside of the body.
  • Performance Studies: The document details performance studies conducted on human serum samples to evaluate the device's sensitivity, specificity, and agreement compared to other methods and the EBV serological status. This is typical for IVD devices.

The definition of an In Vitro Diagnostic (IVD) is a medical device that is used to perform tests on samples such as blood, urine, or tissues, taken from the human body, to detect diseases, conditions, or infections. This device fits that definition perfectly.

N/A

Intended Use / Indications for Use

The Epstein Barr Virus Early Diffuse Antigen (EBV EA-D) IgG ELISA Test is for the qualitative detection of IgG antibodies to EBV EA-D as an aid in the diagnosis of EBV infection in patients with clinical symptoms of Infectious Mononucleosis (IM). The PANBIO EBV EA-D IgG ELISA should be used in conjunction with other EBV serologies.

Product codes

LSE

Device Description

The EBV EA-D IgG ELISA Kit is an enzyme-linked immunosorbent assay (ELISA) for the detection of IgG antibodies to EBV EA-D antigen in human serum.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Study Site 2:
346 frozen retrospective sera of various ages and genders were submitted to a state health laboratory in Maryland USA for EBV testing. The sera include samples from the following groups: 52 seronegative samples, 51 samples from patients with acute Infectious Mononucleosis, and 243 samples from patients with past exposure to EBV.
These sera were tested on the PANBIO EBV EA-D IgG ELISA and the Trinity Biotech EBV EA-D IgG ELISA. The PANBIO results were compared to the EBV status of the sera to determine the sensitivity, specificity, and agreement of the assay relative to the EBV serological status (table 1). Additionally, the Trinity Biotech results were compared to the EBV serological status (table 2), and the PANBIO versus the Trinity Biotech results (table 3) and are summarised below.

Study Site 3:
330 prospective sera of various ages and genders were tested at a private pathology laboratory in Queensland Australia for EBV testing. The sera include the following groups: 47 seronegative, 35 with acute infectious mononucleosis and 248 with past exposure to EBV.
These sera were tested on the PANBIO EBV EA-D IgG ELISA and the Trinity Biotech EBV EA-D IgG ELISA. The PANBIO results were compared to the Sera to determine the sensitivity, specificity, and agreement of the assay relative to the EBV serological status (table 4). Additionally, the Trinity Biotech results were compared to the EBV serological status (table 5), and the PANBIO versus the Trinity Biotech results (table 6) and are summarised below.

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Study Type: Diagnostic performance comparison with a predicate device and EBV serological status.
Sample Size: 346 retrospective sera in Study Site 2, 330 prospective sera in Study Site 3.
Standalone Performance:
Study Site 2 (PANBIO ELISA vs. EBV Status):
Relative Sensitivity (Acute) = 16/51 = 31.4 % (95% CI: 19.1 – 45.9 %)
Relative Sensitivity (Past) = 60/243 = 24.7% (95% CI: 19.3 – 30.1%)
Relative Specificity (Past) = 161/243 = 66.3 % (95% CI: 60.3 – 72.2 %)
Relative Specificity (Negative) = 51/52 = 98.1 % (95% CI: 89.7 – 100.0 %)
Relative Agreement = 228/346 = 65.9 % (95% CI: 60.9 – 70.9 %)
Study Site 3 (PANBIO ELISA vs. EBV Status):
Relative Sensitivity (Acute) = 13/35 = 37.1% (95% CI: 21.5 - 55.1%)
Relative Sensitivity (Past) = 33/248 = 13.3% (95% CI: 9.1 - 17.5%)
Relative Specificity (Past) = 214/248 = 86.3% (95% CI: 82.0 - 90.6%)
Relative Specificity (Negative) = 45/47 = 95.7% (95% CI: 85.5 - 99.5%)
Relative Agreement = 272/330 = 82.4% (95% CI: 78.3 - 86.5%)

Reproducibility Study:
Study Sites 3, 4 & 5 (Australia, including PANBIO, 2 private labs).
Sample size: 8 sera tested 3 times each on three different days.
Results presented as Within-run, Between Day, Between Site, and Total precision (Standard Deviation and Coefficient of Variation).
Key results: For a positive sample (mean ratio 1.94), Total CV was 20.1%. For a cut-off sample (mean ratio 1.00), Total CV was 10.1%. For a negative sample (mean ratio 0.11), Total CV was 19.4%.

Potential Cross-Reactivity Study:
Study Site 5.
Sample size: Panel of 50 specimens with IgG antibodies to diseases other than EBV (Cytomegalovirus, Varicella zoster, Herpes simplex virus 1, Herpes simplex virus 2, Anti-Nuclear Antibody, Rheumatoid Factor).
Key Result: 0/50 specimens were positive when analyzed with the EBV EA-D IgG ELISA Kit, indicating 100% true negative result for the panel and good analytical specificity.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Relative Sensitivity (Acute)
Relative Sensitivity (Past)
Relative Specificity (Past)
Relative Specificity (Negative)
Relative Agreement

Predicate Device(s)

Trinity Biotech Captia™ EBV EA-D IgG ELISA

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 866.3235 Epstein-Barr virus serological reagents.

(a)
Identification. Epstein-Barr virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to Epstein-Barr virus in serum. The identification aids in the diagnosis of Epstein-Barr virus infections and provides epidemiological information on diseases caused by these viruses. Epstein-Barr viruses are thought to cause infectious mononucleosis and have been associated with Burkitt's lymphoma (a tumor of the jaw in African children and young adults) and postnasal carcinoma (cancer).(b)
Classification. Class I (general controls).

0

SEP 2 7 2002

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS 1.10

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: K021793

Applicant Information:

Date Prepared:September 26, 2002
Name:PANBIO Limited
Address:116 Lutwyche Road
Windsor 4030 Australia
Contact Person:Helen Jennings
Phone Number:61-(0)7-3357-1177

Device Information:

Fax Number:

Trade Name:EBV EA-D IgG ELISA Kit
Common Name:EBV EA-D IgG EIA Test
Classification Name:EBV EA-D IgG Serological Reagent

60-(0)7-3357-1222

Equivalent Device:

Trinity Biotech Captia™ EBV EA-D IgG ELISA

Device Description: The EBV EA-D IgG ELISA Kit is an enzyme-linked immunosorbent assay (ELISA) for the detection of IgG antibodies to EBV EA-D antigen in human serum.

Intended Use: The Epstein Barr Virus Early Diffuse Antigen (EBV EA-D) IgG ELISA Test is for the qualitative detection of IgG antibodies to EBV EA-D as an aid in the diagnosis of EBV infection in patients with clinical symptoms of Infectious Mononucleosis (IM). The PANBIO EBV EA-D IgG ELISA should be used in conjunction with other EBV serologies.

Principle of Procedure:

Serum antibodies of the IgG class, when present, combine with EBV Early Antigen, which is purified by immunoaffinity chromatography, and attached to the polystyrene surface of the microwells. Residual serum is removed by washing and peroxidase conjugated anti-human IgG is added. The microwells are washed and a colourless substrate system, tetramethylbenzidine/ hydrogen peroxide (TMB/H2/22) is added. The substrate is hydrolysed by the enzyme and the chromogen changes to a blue colour. After stopping the reaction with acid, the TMB becomes yellow. Colour development is indicative of the presence of EBV EA-D IgG antibodies in the test sample.

1

PERFORMANCE CHARACTERISTICS

Study Site 2:

346 frozen retrospective sera of various ages and genders were submitted to a state health laboratory in Maryland USA for EBV testing. The sera include samples from the following groups: 52 seronegative samples, 51 samples from patients with acute Infectious Mononucleosis, and 243 samples from patients with past exposure to EBV.

These sera were tested on the PANBIO EBV EA-D IgG ELISA and the Trinity Biotech EBV EA-D IgG ELISA. The PANBIO results were compared to the EBV status of the sera to determine the sensitivity, specificity, and agreement of the assay relative to the EBV serological status (table 1). Additionally, the Trinity Biotech results were compared to the EBV serological status (table 2), and the PANBIO versus the Trinity Biotech results (table 3) and are summarised below.

PANBIO ELISA
EBV StatusPositiveEquivocalNegativeTotal
Seronegative
VCA IgG (-)
VCA IgM (-)
EBNA IgG (-)015152
Acute
VCA IgM (+)
EBNA IgG (-)1662951
Past Infection
VCA IgG (+)
VCA IgM (-)
EBNA IgG (+)6022161243
Total7629241346

TABLE 1 EBV STATUS VERSUS PANBIO ELISA

95% Confidence Interval

Relative Sensitivity (Acute)= 16/51 = 31.4 %19.1 – 45.9 %
Relative Sensitivity (Past)= 60/243 = 24.7%19.3 – 30.1%
Relative Specificity (Past)= 161/243 = 66.3 %60.3 – 72.2 %
Relative Specificity (Negative)= 51/52 = 98.1 %89.7 – 100.0 %
Relative Agreement= 228/346 = 65.9 %60.9 – 70.9 %

*Retesting of equivocal samples was not conducted, as the samples were unavailable.

2

PANBIO ELISA
EBV StatusPositiveEquivocalNegativeTotal
Seronegative
VCA IgG (-)
VCA IgM (-)
EBNA IgG (-)424652
Acute
VCA IgM (+)
EBNA IgG (-)2961651
Past Infection
VCA IgG (+)
VCA IgM (-)
EBNA IgG (+)9221130243
Total12529192346

TABLE 2 EBV STATUS VERSUS TRINITY BIOTECH ELISA

PANBIO ELISA

95% Confidence Interval

Relative Sensitivity (Acute)= 29/51= 56.9%42.2 – 70.7%
Relative Sensitivity (Past)= 92/243= 37.9%31.8 – 44.0%
Relative Specificity (Past)= 130/243= 53.5%47.2 – 59.8%
Relative Specificity (Negative)= 46/52= 88.5%76.6 – 95.6%
Relative Agreement= 205/346= 59.2%54.1 – 64.4%

*Retesting of equivocal samples was not conducted, as the samples were unavailable.

TABLE 3 TRINITY BIOTECH VERSUS PANBIO ELISA

PANBIO ELISA
Trinity ResultPositiveEquivocalNegativeTotal
Positive721637125
Equivocal242329
Negative29181192
Total7629241346

95% Confidence Interval

Relative Sensitivity= 72/125= 57.6%48.9 – 66.3%
Relative Specificity= 181/192= 94.3 %90.0 – 97.1%
Relative Agreement= 253/346= 73.1%68.5 – 77.8%

*Retesting of equivocal samples was not conducted, as the samples were unavailable.

3

Study Site 3:

330 prospective sera of various ages and genders were tested at a private pathology laboratory in Queensland Australia for EBV testing. The sera include the following groups: 47 seronegative, 35 with acute infectious mononucleosis and 248 with past exposure to EBV.

These sera were tested on the PANBIO EBV EA-D IgG ELISA and the Trinity Biotech EBV EA-D IgG ELISA. The PANBIO results were compared to the Sera to determine the sensitivity, specificity, and agreement of the assay relative to the EBV serological status (table 4). Additionally, the Trinity Biotech results were compared to the EBV serological status (table 5), and the PANBIO versus the Trinity Biotech results (table 6) and are summarised below.

PANBIO ELISA
EBV StatusPositiveEquivocalNegativeTotal
Seronegative
VCA IgG (-)
VCA IgM (-)
EBNA IgG (-)204547
Acute
VCA IgM (+)
EBNA IgG (-)1302235
Past Infection
VCA IgG (+)
VCA IgM (-)
EBNA IgG (+)331*214248
Total481281330
Relative Sensitivity (Acute) = 13/35 = 37.1% 95% Confidence In 21.5 - 55.1

TABLE 4 EBV STATUS VERSUS PANBIO ELISA

95% Confidence Interval
Relative Sensitivity (Acute)= 13/35= 37.1%21.5 - 55.1%
Relative Sensitivity (Past)= 33/248= 13.3%9.1 - 17.5%
Relative Specificity (Past)= 214/248= 86.3%82.0 - 90.6%
Relative Specificity (Negative)= 45/47= 95.7%85.5 - 99.5%
Relative Agreement= 272/330= 82.4%78.3 - 86.5%

*Equivocal samples were inconclusive following repeat testing on the alternative method (IFA).

4

TABLE 5
EBV STATUS VERSUS TRINITY BIOTECH ELISA
PANBIO ELISA
EBV StatusPositiveEquivocalNegativeTotal
Seronegative
VCA IgG (-)
VCA IgM (-)
EBNA IgG (-)104647
Acute
VCA IgM (+)
EBNA IgG (-)2101435
Past Infection
VCA IgG (+)
VCA IgM (-)
EBNA IgG (+)381*209248
Total601269330
95% Confidence Interval
Relative Sensitivity (Acute)= 21/35= 60.0%42.1 – 76.1%
Relative Sensitivity (Past)= 38/248= 15.3%10.8 – 19.8%
Relative Specificity (Past)= 209/248= 84.3%79.7 – 88.8%
Relative Specificity (Negative)= 46/47= 97.9%88.7 – 99.9%
Relative Agreement= 276/330= 83.7%79.8 – 87.7%

*Equivocal samples were inconclusive following repeat testing on the alternative method (IFA).

TABLE 6
TRINITY BIOTECH VERSUS PANBIO ELISA
PANBIO ELISA
PANBIO ELISA
Trinity BiotechPositiveEquivocalNegativeTotal
Positive3512460
Equivocal0011
Negative130256269
Total481281330
95% Confidence Interval
Relative Sensitivity= 35/60= 58.3%44.9 - 70.9%
Relative Specificity= 256/269= 95.2%91.9 - 97.4%
Relative Agreement= 291/330= 88.2%84.7 - 91.7%

*Equivocal samples were inconclusive following repeat testing on the alternative method (IFA).

5

REPRODUCIBILITY

Study Sites 3, 4 & 5:

The reproducibility of the PANBIO EBV EA-D IgG ELISA kit was determined by testing 8 sera 3 times each on three different days at three Australian study sites. Two sites were private pathology laboratories and the third site was PANBIO. Within-run, between site and total precision were estimated by analysis of variance (ANOVA Type II). The results are presented in table 7 below.

TABLE 7 -- Reproducibility Data PANBIO EBV EA-D IgG Study Sites 1, 2 & 3

Samplen*MeanWithinBetween DayBetween SiteTotal
*S.DCV*S.DCV*S.DCV*S.DCV
Positive271.940.3920.3%0.000.0%0.084.3%0.3920.1%
Cut-off271.000.1110.9%0.000.0%0.000.0%0.1010.1%
Negative270.110.0218.9%0.000.0%0.015.8%0.0219.4%
1273.780.349.0%0.000.0%0.164.1%0.369.4%
2270.940.1111.7%0.022.4%0.044.7%0.1212.5%
3271.610.1710.8%0.000.0%0.000.0%0.1710.4%
4270.870.0910.9%0.000.0%0.000.0%0.0910.2%
5271.260.1814.0%0.000.0%0.043.3%0.1713.9%
6272.510.3012.2%0.072.7%0.000.0%0.3112.2%
7271.160.119.5%0.119.1%0.065.5%0.1513.0%
8271.120.1614.0%0.119.9%0.043.7%0.1916.5%

Precision Measures (Using Ratio)

SD = Standard Deviation

CV = Coefficient of Variation (%) * Values calculated from ratios

Site 1: Three days of triplicates Site 2: Three days of triplicates Site 3: Three days of triplicates

Note:

Standard Deviation results have been rounded to two decimal places for tabulation purposes.

6

POTENTIAL CROSS-REACTIVITY

Study Site 5:

This study consisted of a panel of 50 specimens screened for IgG antibodies detectable by ELISA to disease types other than Epstein Barr Virus. The purpose of this study is to establish the analytical specificity of the EBV EA-D IgG ELISA Test, through the analysis of specimens with diseases that have the potential for cross-reactivity. Each of the specimens included in the study was characterized with respect to disease diagnosis prior to analysis of the specimens with the EBV EA-D IgG ELISA Test. Table 9 on the following page lists the cross-reactivity results for each type of specimen included in the disease panel. Table 8 below provides a summary of the cross-reactivity data.

TABLE 8 - PANBIO EBV EA-D IgG

Disease (IgG Antibodies)Total SpecimensPositive Result
Cytomegalovirus10(0/10)
Varicella zoster13(0/13)
Herpes simplex virus 18(0/8)
Herpes simplex virus 22(0/2)
Anti-Nuclear Antibody8(0/8)
Rheumatoid Factor9(0/9)
Total Antibody50(0/50)

CROSS-REACTIVITY SPECIMEN PANEL SUMMAY

Results indicate that no specimens (0/50) were positive when analysed with the EBV EA-D IgG ELISA Kit.

The true negative result of 100% for the above disease panel is consistent with good analytical specificity for the EBV EA-D IgG ELISA Test.

7

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three wing-like shapes extending from its head. The eagle faces right and is positioned within a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".

Food and Drug Administration 098 Gaither Road Rockville MD 20850

SEP 27 2002

Ms. Kate Wersin Regulatory Affairs Officer PANBIO Limited 116 Lutwyche Road Windsor Brisbane, Queensland Australia 4030

Re: K021793

Trade/Device Name: EBV-EA-D IgG ELISA Regulation Number: 21 CFR 866.3235 Regulation Name: Epstein-Barr Virus Serological Reagents Regulatory Class: Class I Product Code: LSE Dated: August 12, 2002 Received: August 15, 2002

Dear Ms. Wersin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050,

8

Page 2 -

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

9

Page 1 of 1

510(k) Number (if known): K021793

Device Name:

Indications For Use:

The Epstein Barr Virus Early Diffuse Antigen (EBV EA-D) IgG ELISA Test is for the qualitative detection of IgG antibodies to EBV EA-D as an aid in the diagnosis of EBV infection in patients with clinical symptoms of Infectious Mononucleosis (IM). The PANBIO EBV EA-D IgG ELISA should be used in conjunction with other EBV serologies.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Worley Dubose


(Division Sign Off)

sion Sign-Laboratory Dev Division of Clink 510(k) Number

(Optional Format 3-10-98)