K Number
K021784
Date Cleared
2002-07-24

(55 days)

Product Code
Regulation Number
878.4460
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The surgeon's glove is a device made of natural rubber latex intended to be worn by surgeons and/or operating room personnel to protect a surgical wound from contamination.

Device Description

The Powder Free Polymer Coated Brown Latex Surgical Gloves, Sterile (Protein Labeling Claim) meets all the requirements of ASTM standard D 3577-0151 and FDA 21 CFR 800.20.

AI/ML Overview

Here's an analysis of the provided 510(k) summary regarding the acceptance criteria and the study proving the device meets them:

1. Table of Acceptance Criteria and Reported Device Performance:

CharacteristicAcceptance Criteria (Standards)Reported Device Performance
DimensionsASTM D 3577 - 01E1Meets
Physical PropertiesASTM D 3577 - 01E1Meets
Freedom from pinholesASTM D 3577 - 01E1; FDA 21 CFR 800.20Meets
Powder-FreeASTM D 6124 - 01Meets (2 mg/glove maximum)
Protein LevelASTM D 5712 – 95< 50 µg/g
BiocompatibilityPrimary Skin Irritation in RabbitsPasses (Not a primary skin irritant)
BiocompatibilityDermal SensitizationPasses (Not a contact sensitizer)

2. Sample Size Used for the Test Set and the Data Provenance:

The document does not explicitly state the sample sizes used for each specific test (e.g., how many gloves were tested for pinholes, dimensions, etc.). It refers to meeting the requirements of the listed ASTM and FDA standards, which would imply that the sample sizes and methodologies prescribed by those standards were followed.

Data provenance is not explicitly stated beyond the fact that the company is WRP Asia Pacific Sdn Bhd, based in Malaysia. The testing would have been done by or for this manufacturer. The data is retrospective relative to the submission date.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

This information is not provided. The tests are based on established ASTM and FDA standards for medical device performance, which define the testing methodologies and criteria rather than relying on expert consensus for each test. For example, a "freedom from pinholes" test would involve a standardized water leak test, not an expert visual assessment.

4. Adjudication Method for the Test Set:

Not applicable. The tests are objective measurements against defined standards, not subjective assessments requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:

No, an MRMC comparative effectiveness study was not done. This type of study is typically associated with AI/diagnostic imaging devices where human readers are interpreting data. This document describes a physical medical device (surgical gloves), and its performance is evaluated against physical and chemical standards, not reader interpretation.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

Not applicable. This device is a physical medical device (surgical gloves), not an algorithm or software. Its performance is inherent to its physical properties, not a standalone algorithm.

7. The Type of Ground Truth Used:

The ground truth used for these tests are the established, objective criteria defined within the referenced ASTM standards (e.g., ASTM D 3577-01E1, ASTM D 6124-01, ASTM D 5712-95) and FDA regulations (21 CFR 800.20) for the physical, chemical, and biological properties of surgical gloves. These standards specify how each characteristic should be measured and what constitutes an acceptable result.

8. The Sample Size for the Training Set:

Not applicable. This is not a machine learning or AI device, so there is no "training set." The device performance is based on manufacturing specifications and testing against established standards.

9. How the Ground Truth for the Training Set was Established:

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

WRP Asia Pacific Sdn Bhd

JUL 2 4 2002

Lot 1, Jalan 3, Kawasan Perusahaan Bandar Baru Salak Tinggi, 43900 Sepang, Selangor Darul Ehsan, MALAYSIA

510(k) SUMMARY

K021784

+60-3-8706 1486 TEL +60-3-8706 1485 FAX cyw wrp@wrpworld.com www.wrpworld.com

1.0 Submitter:

Name:WRP Asia Pacific Sdn Bhd
Address:Lot 1, Jalan 3, Kawasan Perusahaan Bandar Baru SalakTinggi, 43900 Sepang, Selangor Darul Ehsan, MALAYSIA
Phone No.:+60 3 8706 1486
Fax No.:+60 3 8706 1485

27 MAY 2002 Date of Summary Prepared:

2.0 Contact Person:

Name:Mr. Yue Wah, CHOW
Phone No.:+60 3 8706 1486
Fax No.:+60 3 8706 1485

3.0 Name of the device:

Trade Name:1. Profeel2. Multiple or Customer's Trade Name
Device Name:Powder Free Polymer Coated Brown Latex Surgical GlovesSterile (Protein Labeling Claim)
Common Name:Surgical Gloves
Classification Name:Surgeon's Gloves (per 21 CFR 878.4460)

4.0 Identification of The Legally Marketed Device:

Class I Powder Free natural rubber latex Surgeon's gloves, 79KGO, that meets all the requirements of ASTM standard D 3577 - 0181 Type 1 and FDA 21 CFR 800.20.

5.0 Description of The Device:

The Powder Free Polymer Coated Brown Latex Surgical Gloves, Sterile (Protein Labeling Claim) meets all the requirements of ASTM standard D 3577-0151 and FDA 21 CFR 800.20.

Com

{1}------------------------------------------------

WRP Asia Pacific Sdn Bhd 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1 / 1

Intended Use of the Device: 6.0

The Powder Free Polymer Coated Brown Latex Surgical Glove, Sterile (Protein Labeling Claim) is a device made of natural rubber latex intended to be worn by surgeons and/or operating room personnel to protect a surgical wound from contamination,

Summary of The Technological Characteristics of The Device: 7.0

The Powder Free Polymer Coated Brown Latex Surgical Gloves, Sterile (Protein Labeling Claim) are summarized with the following technological characteristics compared to ASTM or equivalent standards.

CHARACTERISTICSSTANDARDSDEVICEPERFORMANCE
DimensionsASTM D 3577 - 01E1Meets
Physical PropertiesASTM D 3577 - 01E1Meets
Freedom from pinholesASTM D 3577 - 01E1FDA 21 CFR 800.20Meets
Powder-FreeASTM D 6124 - 01Meets2 mg/glove maximum
Protein LevelASTM D 5712 – 95< 50 µg/g
BiocompatabilityPrimary Skin Irritation inRabbitsDermal SensitizationPasses(Not a primary skin irritant)Passes(Not a contact sensitizer)

Page 2 of 3

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for WRP Asia Pacific Sdn Bhd. The logo consists of the letters "WRP" in a bold, sans-serif font, enclosed in a black hexagon. To the right of the logo is the text "WRP Asia Pacific Sdn Bhd" in a similar font. The text is black and appears to be printed on a white background.

Substantial Equivalent Based on Assessment of Non-Clinical Performance Data 8.0

The performance test data of the non-clinical tests that support a determination of substantial equivalence is the same as mentioned immediately above.

9.0 Substantial Equivalent Based on Assessment of Clinical Performance Data

Clinical data is not needed for gloves or for most devices cleared by the 510(k) process.

10.0 Conclusion

It can be concluded that the Powder Free Polymer Coated Brown Latex Surgical Gloves, Sterile (Protein Labeling Claim) will perform according to the glove performance standards referenced in section 7 above and meet ASTM standards, and FDA requiremnts for water leak test on pinhole AQL. Consequently, this device is substantially equivalent to currently marketed devices.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for WRP Asia Pacific Sdn Bhd. The logo consists of the letters "WRP" in a stylized font, followed by the words "WRP Asia Pacific Sdn Bhd" in a more standard font. Below the company name is the number 147817V.

INDICATIONS FOR USE

Applicant:

WRP Asia Pacific Sdn Bhd

510(k) Number (if known):

K021784

Device Name:

POWDER FREE POLYMER COATED BROWN LATEX SURGICAL GLOVES, STERILE (PROTEIN LABELING CLAIM)

Indications For Use:

The surgeon's glove is a device made of natural rubber latex intended to be worn by surgeons and/or operating room personnel to protect a surgical wound from contamination.

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use OR Over-The-Counter ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... .. (Per 21 CFR 801.109)

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring a stylized representation of three human profiles facing right, with three parallel lines above them.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 2 4 2002

Mr. Yue Wah Chow Head of Department QA/RA Lot 1, Jalan 3, Kawasan Perusahaan Bandar Baru Salak Tinggi, 43900 Sepang, Selangor Darul Ehsan, MALAYSIA

Re: K021784

Trade/Device Name: Powder Free Polymer Coated Brown Latex Surgical Gloves, Sterile (Protein Labeling Claim) Contains 50 Micrograms or Less of Total Water Extractable Protein Per Gram Regulation Number: 878.4460 Regulation Name: Surgeon's Gloves Regulatory Class: I Product Code: KGO Dated: May 27, 2002 Received: May 30, 2002

Dear Mr. Chow:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Mr. Chow

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Timoth A. Ulatowski Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for WRP Asia Pacific Sdn Bhd. The logo consists of the letters "WRP" in a bold, sans-serif font, with the letters arranged in a slightly overlapping fashion. To the right of the letters is the text "WRP Asia Pacific Sdn Bhd" in a similar font. Below the text is the number "1 4 7 8 1 7 V".

INDICATIONS FOR USE

Applicant:WRP Asia Pacific Sdn Bhd
510(k) Number (if known):K021784
Device Name:POWDER FREE POLYMER COATED BROWNLATEX SURGICAL GLOVES, STERILE(PROTEIN LABELING CLAIM) Contains 50mcgror less of Total Water Extractable proteinper gram.
Indications For Use:

The surgeon's glove is a device made of natural rubber latex intended to be worn by surgeons and/or operating room personnel to protect a surgical wound from contamination.

Concurrence of CDRH, Office of Device Evaluation (ODE)

OR

Prescription Use (Per 21 CFR 801.109)

..............................................................................................................................................................................

Over-The-Counter _____________________________________________________________________________________________________________________________________________________________

Saruta by Clin

(Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices
510(k) Number ___________________________________________________________________________________________________________________________________

Page 1 of 1

Vour Dartnar In Drotoction™

§ 878.4460 Non-powdered surgeon's glove.

(a)
Identification. A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls).