K Number
K021776

Validate with FDA (Live)

Date Cleared
2002-08-15

(78 days)

Product Code
Regulation Number
876.1500
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The KSEA Quadro Switch, in conjunction with its associated handpiece, functions as a suction/irrigation control switch device to control the rinsing and removal of carbon deposits, blood clots, or excised tissue from operative sites during surgical procedures.

Device Description

The Quadro Switch is a microprocessor controlled device connected at the interface between a suction/irrigation handpiece and a legally marketed irrigation/suction pump via two separate pieces of silicone irrigation and suction tubing.

AI/ML Overview

The provided text is a 510(k) summary for the KSEA Quadro Switch, a device intended to control suction/irrigation during surgical procedures. This document does not contain information about acceptance criteria or a study proving the device meets those criteria, as requested.

The text focuses on confirming substantial equivalence to predicate devices, providing administrative details, and describing the device's intended use and function. It explicitly states: "The minor differences between the KSEA Quadro Switch and the predicate devices raise no new questions of safety or effectiveness." This implies that no new performance studies were required or performed to demonstrate safety and effectiveness beyond establishing equivalence to existing technologies.

Therefore, I cannot extract the requested information regarding acceptance criteria and performance studies from this document.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of an eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

JUL 2 7 2015

Karl Storz Endoscopy James A. Lee Senior Regulatory Affairs Specialist 600 Corporate Pointe Drive Culver City, CA 90230-7600

Re: K021776

Trade/Device Name: KSEA Quadro Switch Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: OCX Dated (Date on orig SE Itr): May 28, 2002 Received (Date on orig SE Itr): May 29, 2002

Dear Mr. Lee,

This letter corrects our substantially equivalent letter of August 15, 2002.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be

{1}------------------------------------------------

Page 2 -

found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

K021776

510(k) Number (if known):

Device Name: KSEA Quadro Switch

Indication for Use: The KSEA Quadro Switch, in conjunction with its associated handpiece, functions as a suction/irrigation control switch device to control the rinsing and removal of carbon deposits, blood clots, or excised tissue from operative sites during surgical procedures.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use: (Per 21 CFR 801.109)

(Optional Format 1-2-96)

OR Over-the-Counter Use:

Stypt Swody

(Division Sign-Of (Division of General, Restorative Division ological Devices

KOU776 510(k) Number -

002

{3}------------------------------------------------

AUG 1 5 2002

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS KOQ 2 / 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of KSEA's knowledge.

Application:Karl Storz Endoscopy - America, Inc.600 Corporate Pointe DriveCulver City, CA 90230(310) 338-8100
-----------------------------------------------------------------------------------------------------------------------------

Contact:

James A. Lee, Ph.D. Senior Regulatory Affairs Specialist

Device Identification:

Suction/Irrigation Switch

Common Name:

Trade Name: (optional) KSEA Quadro Switch

Indication: The KSEA Quadro Switch, in conjunction with its associated handpiece, functions as a suction/irrigation control switch device to control the rinsing and removal of carbon deposits, blood clots, or excised tissue from operative sites during surgical procedures.

Device Description: The Quadro Switch is a microprocessor controlled device connected at the interface between a suction/irrigation handpiece and a legally marketed irrigation/suction pump via two separate pieces of silicone irrigation and suction tubing.

Substantial Equivalence: The KSEA Quadro Switch is substantially equivalent to the predicate devices since the basic features and intended uses are the similar. The minor differences between the KSEA Quadro Switch and the predicate devices raise no new questions of safety or effectiveness.

Signed:

signature

James A. Lee, Ph.D. Senior Regulatory Affairs Specialist

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.