K Number
K021350

Validate with FDA (Live)

Date Cleared
2002-05-14

(15 days)

Product Code
Regulation Number
880.5725
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Indicated for use in parenteral, enteral and epidural therapies and the administration of whole blood and blood products.

Device Description

The Abbott Plum A+3 infusion pump consists of three independent Plum A+ pumping units each identical to each other, housed in a single enclosure. It is an electromechanical infusion pump which uses a stepper motor in conjunction with an n-line cassette to meter IV fluids through a dedicated intravenous administration set that is also manufactured and distributed by Abbott Laboratories. The user interface of the infusion pump allows the user to program fluid delivery through a variety of weight and medication based units such as micrograms/kg/hour, grams/hr and other delivery specifications. An individual pump can accurately control the delivery of fluid from two input sources out to a single infusion line. The Abbott Plum A+3 infusion pump allows for six lines in and three lines out. While identical, each pump operates independently from the others and must be programmed separately. Each pump is uniquely labeled and numbered and differentiated by color of the center unit. The proposed device includes a nurse call function that provides for remote monitoring in the case of alarm activation. A computer data port is provided that allows for an exchange of data from the pump to host computers in the healthcare facility. The display on each pump provides visible indication of several functions including active pump operations, alarm and program status and the parameters of fluid flow. Each pump can be used for standard, piggyback, or concurrent fluid delivery.

AI/ML Overview

The Abbott Plum A+3 Multichannel Infusion Pump is a Class II electromechanical infusion pump. It consists of three independent Plum A+ pumping units, each capable of controlling fluid delivery from two input sources to a single infusion line. The device is intended for administering fluids, whole blood, and blood products in parenteral, enteral, and epidural therapies.

The provided document, K021350, is a 510(k) summary for a Special 510(k) submission, indicating a modification to an already legally marketed predicate device (Abbott Plum A+™ Single Channel Infusion Pump). The modification involves housing three identical Plum A+ pumping units into a single enclosure, utilizing a single bar code wand, and a single AC power source.

1. Table of Acceptance Criteria and Reported Device Performance:

The 510(k) summary states: "Data regarding the functional performance of the proposed Abbott Plum A+3 Multichannel Infusion Pump has been generated and reviewed. The results of testing conducted to verify and validate the design modifications demonstrate acceptable performance of the device."

However, the specific acceptance criteria (e.g., flow rate accuracy tolerance, alarm response times, occlusion pressure thresholds, battery life, electromagnetic compatibility limits, etc.) and the detailed quantitative reported device performance against these criteria are not provided in the excerpt. The document only offers a general statement of acceptable performance. For a medical device of this nature, industry standards (e.g., IEC 60601 series, ISO 14971) and FDA guidance documents would typically define the performance criteria.

Acceptance Criteria CategorySpecific Criteria (Example - Not provided in document)Reported Device Performance (Example - Not provided in document)
Flow Rate Accuracye.g., ±5% for flow rates > 1 mL/hre.g., Achieved ±3% for flow rates > 1 mL/hr
Occlusion Detection Pressuree.g., Detect occlusion within X psie.g., Detected occlusion within Y psi
Alarm Response Timee.g., Audible/visual alarm within Z secondse.g., Alarm activated within W seconds
Battery Operation Duratione.g., Operate for A hours on full chargee.g., Operated for B hours on full charge
EMC Compliancee.g., Meets limits of IEC 60601-1-2e.g., Compliant with IEC 60601-1-2
Software Validatione.g., No critical defects identifiede.g., All software tests passed with no critical defects

2. Sample Size Used for the Test Set and Data Provenance:

The document does not specify the sample size used for the test set or the data provenance (e.g., country of origin, retrospective or prospective). It simply states that "Data regarding the functional performance... has been generated and reviewed." For infusion pump testing, this would typically involve various types of units tested under a range of operating conditions (e.g., different flow rates, fluid viscosities, temperatures, pressures, and alarm conditions) to ensure robust performance.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

This information is not applicable to this type of device submission. The Abbott Plum A+3 Multichannel Infusion Pump is a hardware device where "ground truth" for performance is established through objective engineering tests against predefined specifications and regulatory standards, not through clinical expert consensus on complex diagnostic interpretations.

4. Adjudication Method for the Test Set:

This information is not applicable as it pertains to expert consensus on interpretation, which is not relevant for the functional performance testing of an infusion pump.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

An MRMC comparative effectiveness study is not applicable for this device. This type of study is typically used for diagnostic or screening devices (e.g., AI algorithms for medical imaging) where human readers (e.g., radiologists) interpret cases, and their performance with and without AI assistance is compared. The Abbott Plum A+3 is an infusion pump, and its effectiveness is measured by its mechanical and software performance in fluid delivery, not by human interpretation.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

This concept is not applicable to the Abbott Plum A+3 Multichannel Infusion Pump. "Standalone" performance usually refers to the performance of an AI algorithm without human intervention. While the pump has embedded software and "intelligence," its primary function is mechanical fluid delivery guided by programmed parameters. Its "performance" is inherently a standalone measurement against engineering specifications, but it's not an AI-driven interpretive algorithm in the typical sense of standalone AI studies. The "functional performance" mentioned in the document would represent this standalone performance.

7. Type of Ground Truth Used:

The ground truth for the functional performance testing of the Abbott Plum A+3 Multichannel Infusion Pump would be based on objective engineering measurements and predefined technical specifications derived from regulatory standards (e.g., IEC 60601 series for medical electrical equipment, specific infusion pump standards). For example:

  • Flow Rate: Measured actual fluid volume delivered compared to programmed volume using calibrated flow meters or gravimetric methods.
  • Occlusion Pressure: Measurement of pressure required to trigger an occlusion alarm using calibrated pressure sensors.
  • Alarm Functionality: Verification that alarms (audible, visual) activate correctly under specified fault conditions (e.g., air in line, low battery, completion of infusion).
  • Electrical Safety: Measurements against electrical safety standards.
  • EMC: Measurements against electromagnetic compatibility standards.

8. Sample Size for the Training Set:

This information is not applicable as the Abbott Plum A+3 Multichannel Infusion Pump is a hardware device with embedded software for control, not a machine learning model that requires a "training set" of data in the AI sense to learn patterns. The software development lifecycle would involve verification and validation activities, but not "training" data.

9. How Ground Truth for the Training Set Was Established:

This information is not applicable for the reasons stated in point 8.

{0}------------------------------------------------

K.021350

Special 510(k): Device Modification Abbott Plum A+3 Multichannel Infusion Pump Page of

510(k) SUMMARY Abott Plum A+3 ™ Multichannel Infusion Pump

Submitted by:

Frank Pokrop Associate Director, Regulatory Affairs Abbott Laboratories D-389. Bldg. J-45 Abbott Park, IL 60064

Date Prepared:

April 22, 2002

Name/Classification of Device:

Infusion Pump, Class II 80 FRN - 21 CFR Part 880.5725

Proposed Device:

Abbott Plum A+3TM Multichannel Infusion Pump

Predicate Device:

Abbott Plum A+ TM Single Channel Infusion Pump

Proposed Device Description:

The Abbott Plum A+3 infusion pump consists of three independent Plum A+ pumping units each identical to each other, housed in a single enclosure. It is an electromechanical infusion pump which uses a stepper motor in conjunction with an n-line cassette to meter IV fluids through a dedicated intravenous administration set that is also manufactured and distributed by Abbott Laboratories. The user interface of the infusion pump allows the user to program fluid delivery through a variety of weight and

{1}------------------------------------------------

Proposed Device Description: (cont'd)

medication based units such as micrograms/kg/hour, grams/hr and other delivery specifications.

An individual pump can accurately control the delivery of fluid from two input sources out to a single infusion line. The Abbott Plum A+3 infusion pump allows for six lines in and three lines out. While identical, each pump operates independently from the others and must be programmed separately. Each pump is uniquely labeled and numbered and differentiated by color of the center unit.

The proposed device includes a nurse call function that provides for remote monitoring in the case of alarm activation. A computer data port is provided that allows for an exchange of data from the pump to host computers in the healthcare facility. The display on each pump provides visible indication of several functions including active pump operations, alarm and program status and the parameters of fluid flow.

Each pump can be used for standard, piggyback, or concurrent fluid delivery.

Statement of Intended Use:

The Plum A+3 Multichannel Infusion Pump will be used to administer fluids into a patient's vascular system. The pump is intended for use in parenteral, enteral and epidural therapies and the administration of whole blood and blood products.

The pump must be used with sterile, dedicated, intravenous Plum® administration sets.

Summary of Technological Characteristics of New Device Compared to Predicate Device

The proposed pump has similar design, materials of construction, components, labeling and manufacturing process as the currently marketed Abbott Plum A+ Infusion Pump.

The subject and predicate devices have identical indications for use. Abbott proposes to modify the predicate device by placing three identical pumps into a single enclosure. The proposed device will utilize a single bar code wand and a single AC power source as opposed to three individual components for each pump. These differences do not raise new issues of safety and effectiveness.

{2}------------------------------------------------

Discussion and Conclusions from Nonclinical Tests:

Data regarding the functional performance of the proposed Abbott Plum A+3 Multichannel Infusion Pump has been generated and reviewed. The results of testing conducted to verify and validate the design modifications demonstrate acceptable performance of the device.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are black and are set against a white background. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.

MAY 1 4 2002

Food and Drug Administratio 9200 Corporate Boulevard Rockville MD 20850

Mr. Frank Pokrop Abbott Laboratories 200 Abbott Park Road Dept. 0389, Bldg. J-45 Abbott Park, Illinois 60064-6133

Re: K021350

Trade/Device Name: Abbott Plum A+3™ Multichannel Infusion Pump Regulation Number: 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: FRN Dated: April 26, 2002 Received: April 29, 2002

Dear Mr. Pokrop

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{4}------------------------------------------------

Page 2 - Mr. Pokrop

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Susan Ranner

Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Attachment 3

Indications for Use Statement

510(k) Number (if known)

02 1350

Device Name:

Indications For Use:

Abbott Plum A+3 ™ Multichannel Infusion Pump

Abbott Plum A+3 ™ Multichannel Infusion Pump has the following indications for use:

Indicated for use in parenteral, enteral and epidural therapies and the administration of whole blood and blood products.

PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (per 21 CFR 801.109)

OR Over-The_Counter Use

14

Patricia Cusent

(Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices 510(k) Number _ A 0 0

§ 880.5725 Infusion pump.

(a)
Identification. An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.(b)
Classification. Class II (performance standards).