(28 days)
Not Found
No
The device description and performance studies focus on mechanical properties and traditional testing methods, with no mention of AI/ML terms or data-driven performance metrics.
Yes
Explanation: The device is intended for "Stabilization of open and/or unstable fractures," which is a therapeutic intervention.
No
Explanation: The device is described as a carbon fiber rod for stabilizing open and/or unstable fractures, and its intended use is purely mechanical support rather than diagnosis or detection of a condition. There is no mention of diagnostic functions or capabilities.
No
The device description clearly outlines physical components made of carbon fiber, aluminum, and stainless steel, indicating it is a hardware device for fracture stabilization.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use is the "Stabilization of open and/or unstable fractures." This describes a device used on the body to treat a physical condition (fractures).
- Device Description: The description details a mechanical device made of carbon fiber, aluminum, and stainless steel, designed for external fixation of fractures. This is consistent with a surgical or orthopedic device, not a diagnostic test.
- Lack of IVD Characteristics: There is no mention of analyzing biological samples (blood, urine, tissue, etc.), detecting analytes, or providing diagnostic information about a patient's health status.
IVD devices are used in vitro (outside the body) to examine specimens from the human body to provide information for diagnostic, monitoring, or compatibility purposes. This device clearly operates in vivo (on the body) for therapeutic purposes.
N/A
Intended Use / Indications for Use
Stabilization of open and/or unstable fractures and where soft tissue injury may preclude the use of other fracture treatments such as IM Rodding, casting, and other means of internal fixation.
Product codes (comma separated list FDA assigned to the subject device)
JDW
Device Description
A carbon fiber rod has an aluminum cap threaded onto one end. That cap slides axially in another bar that is partially externally threaded. There is a slot that runs the length of the external thread that accepts a pin that is part of the aluminum cap. There are two circular nuts that tighten on the pin which allow the device to be used in either compression or distraction mode. The partially threaded rod is made of aluminum and the circular nuts of made of stainless steel.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Non-Clinical Testing: The following tests were used to determine substantial equivalence: Cantilever Testing, Torsional Testing, and Sawbone Lab Testing.
Clinical Testing: None provided as a basis for substantial equivalence.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 888.3040 Smooth or threaded metallic bone fixation fastener.
(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.
0
MAY 13 2002
CORPORAT ARTER
SUMMARY OF SAFETY AND EFFECTIVENESS
| Sponsor: | Biomet, Inc.
56 East Bell Drive
P.O. Box 587
Warsaw, IN 46581-0587 |
|-------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Tracy J. Bickel
(574) 267-6639 |
| Proprietary Name: | Radiolucent Colles Compression/Distraction Bar |
| Common Name: | External Fixator |
| Classification Name: | Pin, Fixation, Threaded (888.3040) |
| Substantially Equivalent Devices: Biomet® Radiolucent Colles Fracture Kit (K001760) | |
| Device Description: | A carbon fiber rod has an aluminum cap threaded onto one end. That cap slides
axially in another bar that is partially externally threaded. There is a slot that runs the
length of the external thread that accepts a pin that is part of the aluminum cap. There
are two circular nuts that tighten on the pin which allow the device to be used in
either compression or distraction mode. The partially threaded rod is made of
aluminum and the circular nuts of made of stainless steel. |
Intended Use: Stabilization of open and/or unstable fractures and where soft tissue injury may preclude the use of other fracture treatments such as IM Rodding, casting, and other means of internal fixation.
Summary of Technologies: The device's technological characteristics (materials, design, sizing, and indications) are similar to or identical to the predicate device.
Non-Clinical Testing: The following tests were used to determine substantial equivalence: Cantilever Testing, Torsional Testing, and Sawbone Lab Testing.
Clinical Testing: None provided as a basis for substantial equivalence.
All trademarks are property of Biomet, Inc.
MAILING ADDRESS P.O. Box 587 Warsaw, IN 46581-0587
문
SHIPPING ADDRESS 000046 56 E. Bell Drive Warsaw, IN 46582
1
OFFICE 219.267.6639
FAX 219.267.8137
E-MAIL biomet@biomet.com
1
Image /page/1/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 1 3 2002
Ms. Tracy J. Bickel Regulatory Specialist Biomet. Inc. 56 East Bell Drive P.O. Box 587 Warsaw, IN 46581-0587
Re: K021182
Trade/Device Name: Radiolucent Colles Compression/Distraction Bar Regulation Number: 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: II Product Code: JDW Dated: April 11, 2002 Received: April 15, 2002
Dear Ms. Bickel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice Ms. Tracy J. Bickel requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
2
Page 2 - Ms. Tracy J. Bickel
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Sincerely yours,
Mark A. Milikman
Celia Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
3
Page of
510(k) Number (if known): Device Name: Radiolucent Colles Compression/Distraction Bar Indications for Use:
Stabilization of open and/or unstable fractures and where soft tissue injury may preclude the use of other fracture treatments such as IM Rodding, casting, and other means of internal fixation.
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
OR
Mark N. Milk
(Division Sign-Off Division of General, Restorative and Neurological Devices
510(k) Number K021182