K Number
K020769
Date Cleared
2002-04-03

(27 days)

Product Code
Regulation Number
862.3200
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Single Analyte Urine Drugs of Abuse (DAU) Calibrators are intended for in vitro diagnostic use for the calibration of their respective DAU enzyme immunoassays to detect phencyclidine, opiates, cocaine metabolite, or amphetamines in human urine.

The Single Analyte Urine Drugs of Abuse (DAU) Controls are intended for in vitro diagnostic use for the validation of their respective DAU enzyme immunoassays to detect phencyclidine, opiates, cocaine metabolite, or amphetamines in human urine.

Device Description

All of the Single Analyte Urine DAU Calibrators and Controls are human urine-based liquid, and ready to use. These Calibrators and Controls do not have any especially unique technical characteristics. When applicable, they contain known concentration of a specific drug analyte.

The Negative DAU Calibrator is a processed, drug-free human urine matrix. The Low, Cutoff, Intermediate, and High Calibrators, as well as the 2 levels of Controls are prepared by spiking known concentrations of drug analyte into the Negative DAU Calibrator matrix. The various concentrations of each drug analyte in their corresponding calibrators and controls are summarized as follows:

Phencyclidine EIAOpiate EIACocaine Metabolite EIAAmphetamines EIA
Reference MaterialPhencyclidineMorphineBenzoylecgonined-Methamphetamine
Low Calibrator12.5 ng/mL150 ng/mL150 ng/mL500 ng/mL
Cutoff Calibrator25 ng/mL300 ng/mL300 ng/mL1000 ng/mL
Intermediate Calibrator50 ng/mL600 ng/mL1000 ng/mL1500 ng/mL
High Calibrator100 ng/mL1000 ng/mL3000 ng/mL2000 ng/mL
Control Level 118 ng/mL225 ng/mL225 ng/mL750 ng/mL
Control Level 232 ng/mL375 ng/mL375 ng/mL1250 ng/mL

The nominal concentrations of the analyte in the calibrators and controls are determined and confirmed by GC/MS.

AI/ML Overview

The provided document describes a 510(k) premarket notification for "Single Analyte Urine Drugs of Abuse Calibrators and Controls". This device is a set of calibrators and controls used for in vitro diagnostic tests, rather than an AI-powered diagnostic device. As such, many of the requested categories in the prompt (e.g., AI performance, expert adjudication, MRMC studies, training set details) are not applicable.

Here's an analysis based on the information available:

1. Table of Acceptance Criteria and Reported Device Performance:

The document does not explicitly state "acceptance criteria" for the LZI device. Instead, it demonstrates substantial equivalence to a predicate device (DRI's Drugs of Abuse Urine Calibrators and Controls) by highlighting similarities in intended use, matrix, and performance characteristics. The performance characteristics mentioned as similar are precision, accuracy, and stability. However, specific numerical targets or benchmarks for these characteristics for the LZI device are not provided in this summary.

The table below summarizes the device's characteristics and how they compare to the predicate, which implicitly defines the performance expectations for equivalence.

CharacteristicPredicate Device (DRI's DAU Urine Cals/Controls)LZI's Single Analyte Urine DAU Cals/ControlsAcceptance Criteria / Performance
Intended UseCalibration and validation of DAU immunoassays for phencyclidine, opiates, cocaine metabolite, or amphetamines in human urine.Calibration and validation of DAU immunoassays for phencyclidine, opiates, cocaine metabolite, or amphetamines in human urine.Similar (Intended use is the same)
MatrixUrine-based liquidsHuman urine-based liquidsSimilar (Urine-based)
Cutoff ConcentrationSame as SAMHSA recommendationsSame as SAMHSA recommendationsSimilar (Matches SAMHSA guidelines)
Control Levels2 levels, ± 25% of cutoff per SAMHSA guideline2 levels, concentrations (18/32 ng/mL for PCP, 225/375 ng/mL for Opiate, 225/375 ng/mL for Cocaine, 750/1250 ng/mL for Amphetamines) set around cutoffSimilar (2 levels, concentrations set around cutoff per SAMHSA guideline implicitly met)
Analyte Concentration ConfirmationConfirmed by GC/MSConfirmed by GC/MSSimilar (Uses GC/MS for nominal concentration confirmation)
Storage Condition2℃ to 8℃2℃ to 8℃Similar (Same storage conditions)
Performance Characteristics (General)Precision, accuracy, stability are similar to LZI's productPrecision, accuracy, stability are similar to DRI's productSimilar (General statement of similar precision, accuracy, and stability. Specific numerical data not provided in the summary.)

2. Sample Size Used for the Test Set and Data Provenance:

The document does not provide details on sample size or data provenance for any "test set" in the context of device performance testing. The evaluation focused on demonstrating substantial equivalence to a predicate device based on characteristics and general performance attributes, rather than presenting a detailed clinical study with a test set. The nominal concentrations of analytes in the calibrators and controls are "determined and confirmed by GC/MS," which implies internal analytical testing, but specifics are not given.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

Not applicable. This device is a set of calibrators and controls for laboratory assays, not an AI or diagnostic device that requires expert ground truthing of clinical data for performance evaluation. The "ground truth" for the calibrators and controls is the precisely known concentration of drug analytes, which is confirmed by analytical methods like GC/MS.

4. Adjudication Method:

Not applicable for the reasons stated above.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This is not an AI-powered device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

Not applicable. This is not an AI-powered device with an algorithm for standalone performance. The device itself (calibrators and controls) is a component used in an assay.

7. The Type of Ground Truth Used:

The ground truth for the device's characteristics is the precisely known (nominal) concentrations of drug analytes spiked into the urine matrix, which are confirmed by Gas Chromatography/Mass Spectrometry (GC/MS).

8. The Sample Size for the Training Set:

Not applicable. This is not an AI-powered device that requires a "training set." The development of calibrators and controls relies on precise chemical formulations and analytical validation.

9. How the Ground Truth for the Training Set was Established:

Not applicable. As there is no training set for this type of device, this question is not relevant. The "ground truth" for the product itself (known analyte concentrations) is established through analytical chemistry methods like GC/MS.

{0}------------------------------------------------

510(k) Summary of Safety and Effectiveness

APR 0 3 2002

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Introduction

According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence.

Submitter name, Address, and Contact

Lin-Zhi International, Inc. 2391 Zanker Road. Suite 340 San Jose, CA 95131 Phone: (408) 944-0360 (408) 944-0359 Fax:

Contact: Chiu Chin Chang, Ph.D. VP, R&D

Device Name and Classification

(a) Classification Name:Calibrators, Drug Specific;Class II, DLJ (91 Toxicology), 21 CFR 862.3200
Common/Usual Name:Phencyclidine Calibrators,Opiate Calibrators,Cocaine Calibrators, andAmphetamine Calibrators
Proprietary Name:None
(b) Classification Name:Single (Specified) Analyte Controls (Assayed and Unassayed);Class I, LAS (91 Toxicology), 21 CFR 862.3280
Common/Usual Name:Phencyclidine Controls,Opiate Controls,Cocaine Controls, andAmphetamines Controls
Proprietary Name:None

Legally Marketed Predicate Device(s)

Lin-Zhi International, Inc.' Single Analyte Urine Drugs of Abuse Calibrators and Controls are substantially equivalent to the Drugs of Abuse Urine Calibrators and Controls (Diagnostic Reagents, Inc., now Microgenics Corporation), cleared under premarket notification K983159.

{1}------------------------------------------------

Device Description

All of the Single Analyte Urine DAU Calibrators and Controls are human urine-based liquid, and ready to use. These Calibrators and Controls do not have any especially unique technical characteristics. When applicable, they contain known concentration of a specific drug analyte.

The Negative DAU Calibrator is a processed, drug-free human urine matrix. The Low, Cutoff, Intermediate, and High Calibrators, as well as the 2 levels of Controls are prepared by spiking known concentrations of drug analyte into the Negative DAU Calibrator matrix. The various concentrations of each drug analyte in their corresponding calibrators and controls are summarized as follows:

Phencyclidine EIAOpiate EIACocaine Metabolite EIAAmphetamines EIA
Reference MaterialPhencyclidineMorphineBenzoylecgonined-Methamphetamine
Low Calibrator12.5 ng/mL150 ng/mL150 ng/mL500 ng/mL
Cutoff Calibrator25 ng/mL300 ng/mL300 ng/mL1000 ng/mL
Intermediate Calibrator50 ng/mL600 ng/mL1000 ng/mL1500 ng/mL
High Calibrator100 ng/mL1000 ng/mL3000 ng/mL2000 ng/mL
Control Level 118 ng/mL225 ng/mL225 ng/mL750 ng/mL
Control Level 232 ng/mL375 ng/mL375 ng/mL1250 ng/mL

The nominal concentrations of the analyte in the calibrators and controls are determined and confirmed by GC/MS.

Intended Use

The Single Analyte Urine Drugs of Abuse (DAU) Calibrators are intended for in vitro diagnostic use for the calibration of their respective DAU enzyme immunoassays to detect phencyclidine, opiates, cocaine metabolite, or amphetamines in human urine.

The Single Analyte Urine Drugs of Abuse (DAU) Controls are intended for in vitro diagnostic use for the validation of their respective DAU enzyme immunoassays to detect phencyclidine, opiates, cocaine metabolite, or amphetamines in human urine.

Comparison to Predicate Device

LZI's Single Analyte Urine DAU Calibrators and Controls are similar in intended use, matrix, and performance to the DRI's Drugs of Abuse Urine Calibrators and Controls.

Similarities:

  • Both are for the calibration and validation of DAU enzyme immunoassays to detect the . same commonly abused drug analytes (phencyclidine, opiates, cocaine metabolite, or amphetamines) in human urine.

{2}------------------------------------------------

  • The cutoff concentration for each analyte is the same per recommendations of The . Substance Abuse and Mental Health Services Administration (SAMHSA).
  • Both use 2 levels of Controls, and the concentrations are set at equal to or approximately . ± 25% of the cutoff concentration for each analyte according to the SAMHSA guideline.
  • The nominal concentrations of the analyte in the calibrators and controls are determined . and confirmed by GC/MS.
  • . Both are urine-based liquids.
  • . Storage condition is the same, at 2℃ to 8℃.
  • Performance characteristics on precision, accuracy and stability are similar. �
CharacteristicsDRI's DAU UrineCalibrators and ControlsLZI's Single Analyte UrineDAU Calibrators and Controls
No. of Analytes in eachCalibrator or ControlAll 4 drugs in each Calibrator or ControlSingle drug only in eachCalibrator or Control.
No. of Calibrators3 levels* including theNegative Calibrator.5 levels including the NegativeCalibrator
Nomenclature/Labelingof CalibratorsNegative, Low (= Cutoff),and HighNegative, Low, Cutoff,Intermediate, and High
Concentration ofAnalyte(s)Phencyclidine Controls: 20and 35 ng/mLCocaine High Calibrator:1000 ng/mLPhencyclidine Controls: 18 and32 ng/mL,Cocaine High Calibrator: 3000ng/mL

Differences:

  • Additional calibrators are now available. Currently 5 levels of calibrators (Cal 0, 1, 2, 3, and 4) are available from DRI/Microgenics Corp. (under "Multi-drug Urine Calibrators and Controls" product name).

Conclusion

The information provided in the premarket notification demonstrates that the LZI's Single Analyte Urine Drugs of Abuse Calibrators and Controls are substantially equivalent to previously approved predicate devices, notably the DRI's DAU Urine Calibrators and Controls, and safe and effective for its intended use.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows a logo for the Department of Health & Human Services U.S.A. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES U.S.A." arranged around the circumference. Inside the circle is a stylized emblem featuring three abstract shapes resembling human profiles or figures, stacked on top of each other.

Public Health Service

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Chiu Chin Chang, Ph.D. VP, R&D Lin-Zhi International, Inc. 2391 Zanker Road - Suite 340 San Jose, CA 95131

APR 0 3 2002

K020769 Re:

Trade/Device Name: Single Analyte Urine Drugs of Abuse Calibrators and Controls Regulation Number: 21 CFR 862.3280; 21 CFR 862.3200 Regulation Name: Clinical toxicology control material; Clinical toxicology calibrator Regulatory Class: Class I; Class II Product Code: LAS; DLJ Dated: February 28, 2002 Received: March 7, 2002

Dear Dr. Chang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 -

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Putman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Premarket Notification

Indications for Use Statement

510(k) Number (if known): K020769

Device Name: Single Analyte Urine Drugs of Abuse Calibrators and Controls

Indications for Use:

The Single Analyte Urine Drugs of Abuse (DAU) Calibrators are intended for in vitro diagnostic use for the calibration of their respective DAU enzyme immunoassays to detect phencyclidine, opiates, cocaine metabolite, or amphetamines in human urine.

The Single Analyte Urine Drugs of Abuse (DAU) Controls are intended for in vitro diagnostic use for the validation of their respective DAU enzyme immunoassays to detect phencyclidine, opiates, cocaine metabolite, or amphetamines in human urine.

P. Bernhardt for J. Cooper

ision of Clinical Laboratory Devices 510(k) Number LO2 O

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use

(Per 21 CFR 801.109)

OR

Over-The-Counter Use

(Optional Format 1-2-96)

§ 862.3200 Clinical toxicology calibrator.

(a)
Identification. A clinical toxicology calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. A clinical toxicology calibrator can be a mixture of drugs or a specific material for a particular drug (e.g., ethanol, lidocaine, etc.). (See also § 862.2 in this part.)(b)
Classification. Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.