K Number
K020641
Date Cleared
2002-10-29

(244 days)

Product Code
Regulation Number
868.5905
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Spiritus™ Respiratory System is a patient interface accessory intended for use with devices that administer CPAP (continuous positive airway pressure) and bi-level airway pressure in treating adult patients.

The Spiritus™ Respiratory System is a single-patient interface accessory intended for use by patients who have been prescribed positive airway pressure treatment (PAP) including CPAP and BiPAP.

Device Description

The device is a single-patient-use, patient interface accessory for patients receiving positive airway pressure treatment including CPAP and BiPAP and comprises the nasal/tubing assembly and associated headgear.

The nasal cannula protuberances fit into the patient's nares effecting a substantially artight seal, facilitating the delivery of the positive airway pressure. Tubing is fitted to the nasal cannula via elbow connectors which may be used to secure the detachable headgear. The tubing fits to a "Y" connector to effect connection to the flow generator output tube.

The nasal cannula, tubing and connectors may be cleaned with mild soap and water for use by the same patient.

AI/ML Overview

The provided text describes a medical device, the Spiritus™ Respiratory System, a patient interface accessory for positive airway pressure treatment. However, the document is a 510(k) summary and approval letter, which focuses on demonstrating substantial equivalence to a predicate device rather than detailing comprehensive performance studies with acceptance criteria in the manner typically seen for novel AI/ML devices.

Therefore, much of the requested information (like sample size for test/training set, number/qualifications of experts, adjudication methods, MRMC studies, standalone performance, and detailed ground truth establishment methods) is not available in this document. The document describes traditional engineering performance testing, not studies of diagnostic accuracy that would involve expert readers or AI algorithms.

Here's a breakdown of what can be extracted based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of acceptance criteria with corresponding performance metrics in the way that would be expected for a diagnostic AI device. Instead, it lists the types of performance tests conducted and asserts that these tests confirmed the device's performance. The implied acceptance criterion for these tests is "meets performance expectations for a substantially equivalent device."

Performance TestAcceptance Criteria (Implied)Reported Device Performance
Flow ResistanceWithin acceptable limits for patient interface accessories, comparable to predicate device.Performance confirmed through execution of tests.
Enclosed VolumeWithin acceptable limits for patient interface accessories, comparable to predicate device.Performance confirmed through execution of tests.
Intentional Leak RateWithin acceptable limits for patient interface accessories, comparable to predicate device.Performance confirmed through execution of tests.

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: Not specified. Standard engineering-style performance testing was performed, not studies on patient data.
  • Data Provenance: Not applicable in the context of device performance testing described. These are likely bench or laboratory tests.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

  • Number of Experts: Not applicable. These were engineering performance tests, not human-read diagnostic studies.
  • Qualifications of Experts: Not applicable.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done

  • MRMC Study: No. The device is a patient interface accessory, not a diagnostic imaging device or an AI algorithm that would typically undergo MRMC studies.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

  • Standalone Performance: Not applicable. This is a physical medical device, not an algorithm.

7. The Type of Ground Truth Used

  • Type of Ground Truth: Engineering standards, specifications, and comparison to the predicate device's known performance for flow resistance, enclosed volume, and intentional leak rate.

8. The Sample Size for the Training Set

  • Training Set Sample Size: Not applicable. This is a physical medical device, not an AI/ML algorithm that requires a training set.

9. How the Ground Truth for the Training Set was Established

  • Ground Truth Establishment for Training Set: Not applicable.

{0}------------------------------------------------

K02 06 41

OCT 2 9 2002

Non-confidential 510(k) Summary of Safety and Effectiveness

1. Applicant Name & Address

Worldwide Medical Technologies Inc. 3178 Keen Drive Waycross, Georgia 31503.

2. Trade Name of Device

Spiritus™ Respiratory System

3. Common Name(s) of Device

Patient Interface Accessory

4. Classification Name

Accessory to the CPAP/BiPAP device [VENTILATOR, NON-CONTINUOUS (RESPIRATOR)]

5. Classification & Classification Code

21 CFR 868.5905,73 BZD Class II.

6. Device Description

The device is a single-patient-use, patient interface accessory for patients receiving positive airway pressure treatment including CPAP and BiPAP and comprises the nasal/tubing assembly and associated headgear.

The nasal cannula protuberances fit into the patient's nares effecting a substantially artight seal, facilitating the delivery of the positive airway pressure. Tubing is fitted to the nasal cannula via elbow connectors which may be used to secure the detachable headgear. The tubing fits to a "Y" connector to effect connection to the flow generator output tube.

The nasal cannula, tubing and connectors may be cleaned with mild soap and water for use by the same patient.

7. Intended Use

The Spiritus™ Respiratory System is a patient interface accessory intended for use with devices that administer CPAP (continuous positive airway pressure) and bi-level airway pressure in treating adult patients.

8. Predicate Device

Substantial equivalence is claimed to the Puritan-Bennett Airway Delivery & Management (ADAM) Nasal Ventilation, subject of 510(k) K900164.

9. Summary of Performance Testing

Performance was confirmed through the execution of flow resistance, enclosed volume, and intentional leak rate tests.

Conclusion 10.

Based on the above, we conclude the Spiritus™ Respiratory System is substantially equivalent to predicate device, and is safe and effective for its intended use.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized image of a human figure with three overlapping profiles. The profiles are facing to the right and are connected by a series of curved lines.

OCT 2 9 2002

od and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Worldwide Medical Technologies, Incorporated C/O Mr. Jonathan Lee JSL Consulting Incorporated 138 Spartina Avenue St. Augustine, Florida 32080

Re: K020641

Trade/Device Name: Spiritus™ Respiratory System Regulation Number: 868.5905 Regulation Name: Noncontinuous Ventilator, Accessory Regulatory Class: II Product Code: BZD Dated: October 7, 2002 Received: October 9, 2002

Dear Mr. Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{2}------------------------------------------------

Page 2 - Mr. Lee

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.qov/cdrh/dsma/dsmamain.html

Sincerely yours,

Timothy A. Ulatowski Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

- U.S. Food and Drug Administration - Center for Devices and Radiological Health و

Page 1 of 1

510(k) Number (if known): Ho2064)

Device Name: SpiritusTM Respiratory System

Indications for Use:

The Spiritus™ Respiratory System is a single-patient interface accessory intended for use by patients who have been prescribed positive airway pressure treatment (PAP) including CPAP and BiPAP.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription use ✓

le7Wutshamn

(Optional Format 3-10-98)

(Division Sign-Off)
Division of Anesthesiology, General Hospital,
Infection Control, Dental Devices

510(k) Number: K02064

§ 868.5905 Noncontinuous ventilator (IPPB).

(a)
Identification. A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.(b)
Classification. Class II (performance standards).