K Number
K020479

Validate with FDA (Live)

Manufacturer
Date Cleared
2002-03-26

(41 days)

Product Code
Regulation Number
870.2700
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Datex-Ohmeda S/5 TM Nellcor Compatible Saturation Module, M-NSAT is intended for use with the Datex-Ohmeda modular multiparameter patient monitors for monitoring arterial oxygen saturation of hospitalized patients.

Indication for use:

The Datex-Ohmeda S/5 TM Nellcor Compatible Saturation Module, M-NSAT, and accessories are indicated for monitoring arterial oxygen saturation of hospitalized patients. The device is indicated for use by qualified medical personnel only.

Device Description

The device description of the S/5 Nellcor Compatible Saturation Module M-NSAT is as follows:

[ a single width plug-in module of the S/5 multiparameter monitor

□ measuring noninvasive arterial oxygen saturation and pulse rate

□ sensor consists two light wavelenghts LEDs and photodetector

[ Nellcor OEM SpO2 measurement board MP 404 and accessories

[ an interface board for connecting the measurement board to the monitor

The modifications to the device are:

  1. The Nellcor measurement board MP 203/204 is changed to Nellcor measurement board MP 404.

  2. The interface board was modified to incorporate an interface for the Nellcor MP 404 board.

  3. The interface software of the interface board was modified to incorporate an interface for the Nellcor MP 404 software.

  4. The D-9 patient connector of the module is changed to the new pre-amplifier board circuit including the new patient connector model. These changes are determinate by Nellcor as part of the MP 404 specification.

  5. Mechanics of the module bezel is redesigned for the new Nellcor patient connector and the pre-amplifier board.

  6. The accessories of the new module are extended to the whole Nellcor sensor line as part of the MP 404 technology.

  7. The new Nellcor trunk cable MC-10 is used.

The Datex-Ohmeda S/5 Nellcor Compatible Saturation Module, M-NSAT and accessories (later referred to as M-NSAT) is a module used to monitor arterial oxygen saturation. The user interface has been implemented in the main software of the

AS/3 Anesthesia Monitor

or AS/3 Compact Monitor using S-STDxx or S-ARKxx software

or CS/3 Monitors using S-ICUxx software

or S/5 Anesthesia Monitor

or S/5 Critical Care monitor

or S/5 Compact Anesthesia Monitor

or S/5 Compact Critical Care Monitor

The M-NSAT module provides continuos non-invasive measurement of the pulse rate and oxygen saturation. Visual and auditory alarms are given for high/low pulse rate and high/low saturation values. The user can adjust alarm limits.

The M-NSAT module is designed using the Nellcor pulse oximetry technology. The user interface of the new M-NSAT module revision -04 is equivalent to the M-NSAT Nellcor Compatible Saturation Module (K943456).

AI/ML Overview

The provided text is a 510(k) Premarket Notification summary for the Datex-Ohmeda S/5 Nellcor Compatible Saturation Module, M-NSAT. This document focuses on demonstrating substantial equivalence to a predicate device (AS/3 Nellcor Compatible Saturation Module, M-NSAT, K943456) rather than presenting a detailed study with specific performance acceptance criteria and results in the format requested.

Therefore, many of the requested fields cannot be directly extracted or are not applicable to a substantial equivalence submission of this nature, which primarily relies on comparing technological characteristics and adherence to safety standards.

Here's an analysis of the requested information based on the provided text:

1. A table of acceptance criteria and the reported device performance:

The document does not provide a table with specific performance acceptance criteria (e.g., accuracy ranges for SpO2 or pulse rate) and corresponding reported device performance values. Instead, it states that the device "complies with the safety standards below and is therefore safe and effective for the intended use."

Acceptance Criteria (Implicit - compliance with standards and equivalence to predicate)Reported Device Performance
Compliance with safety standards (e.g., IEC60601-1, UL 2601-1)The device has been "thoroughly tested" and "complies with the safety standards below."
Substantial equivalence to predicate device (K943456) in terms of safety and effectiveness"no new questions of safety and effectiveness" compared to the predicate device.
Identical intended use and indications for use as predicateConfirmed as identical.
Identical fundamental scientific technology as predicateConfirmed as identical.
Same operating principle as predicateConfirmed as the same.
Same user interface and alarms as predicateConfirmed as the same.

2. Sample sized used for the test set and the data provenance:

The document does not specify a "test set" in the context of clinical performance evaluation (e.g., a set of patient data used to evaluate accuracy). The testing mentioned is primarily non-clinical, focusing on compliance with safety standards, electrical safety, electromagnetic compatibility, mechanical, environmental tolerance, and software validation. Therefore, information on sample size and data provenance for a clinical test set is not provided.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

Not applicable. The submission is not about a clinical efficacy study requiring expert-established ground truth.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

Not applicable for the reasons mentioned above.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This device is a pulse oximeter, not an AI-assisted diagnostic tool for human readers.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

The device is a standalone module for monitoring SpO2 and pulse rate. Its performance is inherent in its design and adherence to standards. The document doesn't detail a specific "standalone performance study" in the way it might for an AI algorithm. Its performance is demonstrated through its compliance with technical specifications and safety standards, as well as its substantial equivalence to the predicate device.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc):

For a pulse oximeter, the "ground truth" for SpO2 accuracy typically comes from co-oximetry (blood gas analysis, a gold standard for blood oxygen saturation). However, the document does not detail specific accuracy studies that would use this type of ground truth. The focus is on the device using a "Nellcor pulse oximetry technology" and having "no new questions of safety and effectiveness" compared to a legally marketed predicate device, which would have undergone such validation.

8. The sample size for the training set:

Not applicable. This is not an AI/machine learning device that requires a training set.

9. How the ground truth for the training set was established:

Not applicable for the reasons mentioned above.

In summary:

This 510(k) submission primarily leverages the concept of substantial equivalence. The acceptance criteria are implicitly met by demonstrating that the modified device has identical intended use, fundamental scientific technology, operating principle, and safety/effectiveness characteristics compared to a previously cleared predicate device (K943456), and by showing compliance with relevant electrical, mechanical, and software safety standards. The submission does not detail specific clinical performance studies with acceptance criteria for accuracy, sensitivity, or specificity, nor does it involve AI or human reader studies. The primary "study" is the non-clinical testing and comparison to the predicate device per 510(k) requirements.

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows a black circle with a white letter D inside. The letter D is positioned in the center of the circle. There is a small black rectangle in the upper right corner of the image.

Premarket Notification 510(k) Summary As required by section 807.92 Datex-Ohmeda S/5 Nellcor Compatible Saturation Module, M-NSAT and accessories

GENERAL COMPANY INFORMATION as required by 807.92(a){1}

COMPANY NAME/ADDRESS/PHONE/FAX:

Datex-Ohmeda 86 Pilgrim Road Needham, MA 02492 USA Tel: 781-449-8685 Fax: 781-433-1344

NAME OF CONTACT:

Mr. Joel Kent

DATE: February 9, 2002

DEVICE NAME as required by 807.92(a)(2)

TRADE NAME:

Datex-Ohmeda S/5 Nellcor Compatible Saturation Module, M-NSAT and accessories

COMMON NAME:

Pulse oximeter

CLASSIFICATION NAME:

The following Class II classification appears applicable:

DQAOximeter870.2700
DPZEar Oximeter870.2710

NAME OF LEGALLY MARKETED DEVICE FOR WHICH A CLAIM OF SUBSTANTIAL EQUIVALENCE IS MADE as required by 807.92(a)(3)

The new Datex-Ohmeda S/5 Nellcor Compatible Saturation Module, M-NSAT is substantially equivalent in safety and effectiveness to the legally marketed (predicate) AS/3 Nellcor Compatible Saturation Module, M-NSAT (K943456).

{1}------------------------------------------------

K020479

DEVICE DESCRIPTION as required by 807.92(a)(4)

The intended and indications for use for the modified device is the same as the predicate (K943456).

There has been no change to the fundamental scientific technology from the predicate.

The device description of the S/5 Nellcor Compatible Saturation Module M-NSAT is as follows:

[ a single width plug-in module of the S/5 multiparameter monitor

□ measuring noninvasive arterial oxygen saturation and pulse rate

□ sensor consists two light wavelenghts LEDs and photodetector

[ Nellcor OEM SpO2 measurement board MP 404 and accessories

[ an interface board for connecting the measurement board to the monitor

The modifications to the device are:

  1. The Nellcor measurement board MP 203/204 is changed to Nellcor measurement board MP 404.

  2. The interface board was modified to incorporate an interface for the Nellcor MP 404 board.

  3. The interface software of the interface board was modified to incorporate an interface for the Nellcor MP 404 software.

  4. The D-9 patient connector of the module is changed to the new pre-amplifier board circuit including the new patient connector model. These changes are determinate by Nellcor as part of the MP 404 specification.

  5. Mechanics of the module bezel is redesigned for the new Nellcor patient connector and the pre-amplifier board.

  6. The accessories of the new module are extended to the whole Nellcor sensor line as part of the MP 404 technology.

  7. The new Nellcor trunk cable MC-10 is used.

The Datex-Ohmeda S/5 Nellcor Compatible Saturation Module, M-NSAT and accessories (later referred to as M-NSAT) is a module used to monitor arterial oxygen saturation. The user interface has been implemented in the main software of the

AS/3 Anesthesia Monitor

or AS/3 Compact Monitor using S-STDxx or S-ARKxx software

or CS/3 Monitors using S-ICUxx software

or S/5 Anesthesia Monitor

or S/5 Critical Care monitor

or S/5 Compact Anesthesia Monitor

or S/5 Compact Critical Care Monitor

The M-NSAT module provides continuos non-invasive measurement of the pulse rate and oxygen saturation. Visual and auditory alarms are given for high/low pulse rate and high/low saturation values. The user can adjust alarm limits.

The M-NSAT module is designed using the Nellcor pulse oximetry technology. The user interface of the new M-NSAT module revision -04 is equivalent to the M-NSAT Nellcor Compatible Saturation Module (K943456).

{2}------------------------------------------------

5020479

INTENDED USE as required by 807.92(a)(5)

The Datex-Ohmeda S/5 TM Nellcor Compatible Saturation Module, M-NSAT is intended for use with the Datex-Ohmeda modular multiparameter patient monitors for monitoring arterial oxygen saturation of hospitalized patients.

Indication for use:

The Datex-Ohmeda S/5 TM Nellcor Compatible Saturation Module, M-NSAT, and accessories are indicated for monitoring arterial oxygen saturation of hospitalized patients. The device is indicated for use by qualified medical personnel only.

SUMMARY OF TECHNOLOGICAL CHARACTERITICS OF DEVICE COMPARED TO THE PREDICATE DEVICE as required by 807.92(a)(6)

The new Datex-Ohmeda S/5 Nellcor Compatible Saturation Module, M-NSAT is substantially equivalent in safety and effectiveness to the legally marketed (predicate) AS/3 Nellcor Compatible Saturation Module, M-NSAT (K943456).

The redesigned M-NSAT module has the following similarities compared to the M-NSAT predicate device.

D identical intended use and indications for use

[] identical fundamental scientific technology

T have the same indicated use,

O use the same operating principle

[ have the same safety and effectiveness

□ have the same user interface and alarms

□ are manufactured using the same processes

The main differences between the new M-NSAT and the predicate is primarily due to fact that the new M-NSAT uses

□ the new measurement board MP 404

[] the modified interface board

[] the modified interface software

[] the new pre-amplifier board and the patient connector

[] the new bezel

□ the whole Nellcor´s sensor line accessories

□ the trunk cable MC-10

The new Datex-Ohmeda S/5 Nellcor Compatible Saturation Module, M-NSAT and AS/3 Nellcor Compatible Saturation Module, M-NSAT (K943456) have the same intended use and Indication for use. The new M-NSAT and the M-NSAT (K943456) are intended for use with the Datex-Ohmeda modular multiparameter patient monitors for monitoring arterial oxygen saturation of hospitalized patients. Both are indicated for use by qualified medical personnel only. The new Datex-Ohmeda S/5 Nellcor Compatible Saturation Module, M-NSAT and AS/3 Nellcor Compatible Saturation Module, M-NSAT (K943456) uses the same user interface and alarm logic of the monitor. The data of both modules are receiving and showing on the monitor screen similar. Alarms of both modules have the same alarm menus and alarms can be adjusted equally.

In summary, the Nellcor Compatible Saturation Module, M-NSAT, described in this submission is substantially equivalent to the predicate device.

{3}------------------------------------------------

SUMMARY OF NONCLINICAL TESTING FOR THE DEVICE and CONCLUSIONS as required by 807.92(b)(1)(3)

The Datex-Ohmeda S/5 Nellcor Compatible Saturation Module, M-NSAT and accessories complies with the safety standards below and is therefore safe and effective for the intended use. The device has been thoroughly tested including electrical safety, electromagnetic compatibility, mechanical and environmental tolerance, software validation and verification of specifications. Verification of compliance with the following mandatory and voluntary standards has been made:

  • 21 CFR 898.12 Performance Standard for Electrode Lead Wires and Patient Cables ●
  • ANSI/AAMI ES1-1993 ●
  • IEC60601-1:1988, Amendment 1: 1991, Amendment 2: 1995 .
  • EN 60601-1:1990 +A1:1993 +A2:1995 +A13:1996 .
  • UL 2601-1 : 1997 .
  • IEC 60601-1-2 .
  • IEC 60601-1-4 .

Conclusion:

The summary above shows that there are no new questions of safety and effectiveness for the Datex-Ohmeda S/5 Nellcor Compatible Saturation Module, M-NSAT and accessories as compared to the predicate device.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular emblem with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the top half of the circle. Inside the circle is a stylized image of an eagle or bird with outstretched wings, depicted in a simple, black and white design.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 2 6 2002

Mr. Joel C. Kent Datex-Ohmeda 86 Pilgrim Road Needham, MA 02492

Re: K020479

Datex-Ohmeda S/5™ Nellcor compatible Saturation Module, M-NSAT and accessories Regulation Number: 870.2700, 870.2710 Regulation Name: Oximeter, Ear Oximeter Regulatory Class: II (two) Product Code: 74 DQA, 74 DPZ Dated: March 13, 2002 Received: March 14, 2002

Dear Mr. Kent:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Mr. Joel C. Kent

Please be advised that FDA's issuance of a substantial equivalence determination does not mean I that FDA has made a determination that your device complies with other requirements of the Act that I Drinas intass and regulations administered by other Federal agencies. You must or any I cacial statutes and regarments, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CFN in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic roral in and quind brovisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) I mis lotter will and w yourse FDA finding of substantial equivalence of your device to a legally promation hotification. The a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 11 you desire specific ad 1809.10 for in vitro diagnostic devices), please contact the Office of additionally 21 Of It For at 4646. Additionally, for questions on the promotion and advertising of Compliance at (301) 594-461 the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). regulation on the responsibilities under the Act may be obtained from the Outler general international and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Elias Mallia

Bram D. Zuckerman, M.D. Acting Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: Datex-Ohmeda S/5™ Nellcor Compatible Saturation Module, M-NSAT, and accessories

The Datex-Ohmeda S/5™ Nellcor Compatible Saturation Module, M-NSAT, and accessories are indicated for monitoring arterial oxygen saturation of hospitalized patients. The device is indicated for use by qualified medical personnel only.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use_ (Per 21 CFR 801.109)

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________

(Optional Format 1-2-96)

Division of Cardiovascular & Respiratory Devic KUZ 510(k) Number

OR

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).