(115 days)
The Medcomp Ash Split-Cath XL is indicated for use in attaining long-term vascular access for hemodialysis and apheresis. It may be inserted percutaneously vand is ideally placed in the internal jugular vein of an adult patient. Alternate insertion sites include the subclavian vein. Catheters greater than 40cm are intended for femoral vein insertion.
The Medcomp Ash Split-Cath-XL is a polyurethane, double lumen catheter used to remove and return blood through two-segregated lumen passages. Both lumens are "D" shaped, tapered at the distal tip, with six side holes. The distal venous lumen extends beyond the arterial lumen to reduce recirculation. The fixed polyester cuff allows for tissue ingrowth for long term placement. A polyurethane sleeve is positioned between the hub to cuff location. The lumens are connected to the extensions via a soft pliable hub with suture wing. Red and blue luer connectors and clamps identify the arterial and venous extensions. The clamps incorporate I.D. Rings which indicate priming volume and site care information.
The provided text describes a 510(k) premarket notification for a medical device, the Medcomp Ash Split-Cath-XL hemodialysis catheter. This submission focuses on comparing the proposed device to a predicate device and outlining performance data. The document does not contain the detailed information necessary to fully answer all aspects of your request, particularly regarding specific acceptance criteria, comprehensive study designs, expert involvement, or AI-related metrics.
Here's an attempt to answer your questions based on the available information, with clear indications of where the information is missing:
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria (Stated or Implied) | Reported Device Performance |
|---|---|
| Material & Design Equivalence: Identical in design and materials to predicate device. | "The proposed device is a product line extension to the legally marketed device, and is identical in design and materials." |
| Functionality (Flow vs. Pressure): Adequate flow characteristics under pressure relevant to hemodialysis. | "In Vitro performance data for the proposed device includes: • Flow vs. Pressure." (No specific quantitative performance values or acceptance criteria are provided in the document.) |
| Safety (Implicit from predicate): Implied to meet safety standards of predicate device regarding force at break, leakage, recirculation, mechanical hemolysis, chemical resistance, packaging, sterilization, and biocompatibility. | "Since the design and materials remain unchanged the following testing or documentation is not deemed necessary and is not included in this submission: • Force at break • Leakage • Recirculation • Mechanical hemolysis • Chemical resistance • Packaging data • Sterilization data • Biocompatibility data." |
| Intended Use: Suitable for long-term vascular access for hemodialysis and apheresis, with specific indication for femoral vein insertion for catheters > 40cm. | The device's intended use matches the stated indications. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size for Test Set: Not specified. The document only mentions "In Vitro performance data" for "Flow vs. Pressure."
- Data Provenance: The study appears to be in vitro testing. Country of origin and whether it's retrospective or prospective are not mentioned, but given it's a 510(k) for a medical device in the US, the testing would typically be expected to meet US regulatory standards.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not Applicable. This device is a physical hemodialysis catheter, and the performance data described (
Flow vs. Pressure) does not involve human expert interpretation for "ground truth." The evaluation seems to be based on engineering and physical performance metrics.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not Applicable. The testing is in vitro and does not involve adjudication by experts as would be required for image interpretation or clinical outcomes where subjectivity might be a factor.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- No. This document describes a physical medical device (hemodialysis catheter), not an AI diagnostic or assistive tool. Therefore, an MRMC study and AI-related metrics are irrelevant and not included.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- No. This is a physical medical device, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- The "ground truth" for the in vitro testing is based on physical measurements (e.g., flow rates, pressure differentials) under controlled experimental conditions. It's not a ground truth derived from expert consensus, pathology, or clinical outcomes data in the usual sense for diagnostic algorithms.
8. The sample size for the training set
- Not Applicable. This is a physical medical device. There is no concept of a "training set" as would be used for machine learning. The device's design is based on previous validated designs (predicate devices) and engineering principles.
9. How the ground truth for the training set was established
- Not Applicable. As there is no training set for a machine learning model, this question does not apply.
{0}------------------------------------------------
K20315
Page 1 of 2
MAY 2 4 2002
510(k) SUMMARY
Submitter Information: A.
Submitter:
Contact: Date Prepared:
- B. Trade Name: Common Name: Classification: C.F.R. Section:
- Predicate Device: C.
MEDCOMP® 1499 Delp Drive Harleysville, PA 19438 (215) 256-4201 Telephone (215) 256-1787 Fax Jeanne M. Cush January 16, 2002
Medcomp Ash Split-Cath-XL Hemodialysis Catheter, Implanted 78 MSD 876.5540
K013162 Medcomp Ash Split-Cath W/Strain Relief K012562 Medcomp 14.5F Double Lumen Hemodialysis Catheter
Device Description: D.
The Medcomp Ash Split-Cath-XL is a polyurethane, double lumen catheter used to remove and return blood through two-segregated lumen passages. Both lumens are "D" shaped, tapered at the distal tip, with six side holes. The distal venous lumen extends beyond the arterial lumen to reduce recirculation. The fixed polyester cuff allows for tissue ingrowth for long term placement. A polyurethane sleeve is positioned between the hub to cuff location.
The lumens are connected to the extensions via a soft pliable hub with suture wing. Red and blue luer connectors and clamps identify the arterial and venous extensions. The clamps incorporate I.D. Rings which indicate priming volume and site care information.
Intended Use: E.
The Medcomp Ash Split-Cath-XL is indicated for use in attaining long-term vascular access for hemodialysis and apheresis. Catheters greater than 40cm are intended for femoral vein insertion.
Page 1 of 2 Medcomp Ash Split-Cath-XL Summary
01 1
{1}------------------------------------------------
F. Comparison to Predicate Device:
The proposed device is a product line extension to the legally marketed device, and is identical in design and materials.
1020315
The modifications include:
- . 55cm length
- Femoral insertion instructions for use .
G. Performance Data:
In Vitro performance data for the proposed device includes:
-
Flow vs. Pressure .
Since the design and materials remain unchanged the following testing or Sinoo the documentation is not deemed necessary and is not included in this submission: -
Force at break ●
-
Leakage ●
-
Recirculation .
-
Mechanical hemolysis ●
-
Chemical resistance ●
-
Packaging data .
-
Sterilization data .
-
Biocompatibility data .
02
1
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is a stylized image of an eagle.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 2 4 2002
Ms. Florence A. Caikoski Regulatory Affairs Associate MEDCOMP® 1499 Delp Drive HARLEYSVILLE PA 19438
Re: K020315
Trade/Device Name: Ash Split-Cath XL, 14F x 55cm, Model ASPC55-XL Regulation Number: 21 CFR 876.5540 Regulation Name: Blood access device and accessories Regulatory Class: III Product Code: 78 MSD Dated: May 1, 2002 Received: May 2, 2002
Dear Ms. Caikoski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Florence Caikoski
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration (21 CFR Part 807); listing (21 CFR Part 807), labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on the labeling regulation (21 CFR Part 801 and additionally Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4616. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely vours,
for
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
KO20315
INDICATIONS FOR USE
K020315 510(k) Number:
Device Name: Medcomp Ash Split-Cath XL
Indications for use:
The Medcomp Ash Split-Cath XL is indicated for use in attaining long-term vascular access for hemodialysis and apheresis. It may be inserted percutaneously vand is ideally placed in the internal jugular vein of an adult patient. Alternate insertion sites include the subclavian vein. Catheters greater than 40cm are intended for femoral vein insertion.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
OR
Over-The-Counter
(Per 21 CFR 801.109)
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices K020315
510(k) Number
(Optional Format 1-2-96)
§ 876.5540 Blood access device and accessories.
(a)
Identification. A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device.
(2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle).
(3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).
(b)
Classification. (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided.
(ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated.
(B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided.
(C) Priming volumes must be established.
(D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(E) Air leakage testing and liquid leakage testing must be conducted.
(F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage.
(G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern.
(H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established.
(iii) Performance data must demonstrate the sterility of the device.
(iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing.
(v) Labeling of implanted blood access devices for hemodialysis must include the following:
(A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated.
(B) Labeling must specify the forward and reverse recirculation rates.
(C) Labeling must provide the arterial and venous priming volumes.
(D) Labeling must specify an expiration date.
(E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device.
(F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient.
(G) Labeling must include a patient implant card.
(H) The labeling must contain comprehensive instructions for the following:
(
1 ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;(
2 ) Proper care and maintenance of the device and device exit site;(
3 ) Removal of the device;(
4 ) Anticoagulation;(
5 ) Management of obstruction and thrombus formation; and(
6 ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following:
(A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port.
(B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following:
(A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied.
(B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties.
(C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens.
(D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness.
(viii) The following must be included for A-V shunt cannulae (with vessel tips):
(A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply.
(B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(2) Class II (performance standards) for the nonimplanted blood access device.
(3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section.
(4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.