K Number
K014278
Date Cleared
2002-05-14

(138 days)

Product Code
Regulation Number
878.4460
Panel
HO
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Surgeon's glove (powdered) is a sterile medical device made of natural latex intended to be worn by operating room personnel to protect a surgical wound from contamination and is powdered with absorbable dusting powder USP corn starch as the donning lubricant.

Device Description

Class I Powdered Surgical Glove 79KGO that meets all the requirements of ASTM D3577.

AI/ML Overview

This is an analysis of a 510(k) summary for a medical device, specifically powdered surgeon's gloves. The document primarily focuses on demonstrating substantial equivalence to a predicate device through meeting established performance standards.

Here's the breakdown of acceptance criteria and device performance based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

CharacteristicAcceptance Criteria (ASTM D3577 & other standards)Reported Device Performance (Brightway Gloves Pvt. Ltd.)Meets Criteria?
Dimensions
Length (5 ½)245 mm minimum270-272 mmYes
Length (6 to 9)265 mm minimum270-272 mmYes
Width (5 ½)70 +/- 6 mm68 mmYes (Within range)
Width (6)76 +/- 6 mm73 mmYes (Within range)
Width (6 ½)83 +/- 6 mm79 mmYes (Within range)
Width (7)99 +/- 6 mm87 mmYes (Within range)
Width (7 ½)95 +/- 6 mm92 mmYes (Within range)
Width (8)105 +/- 6 mm103 mmYes (Within range)
Width (8 ½)108 +/- 6 mm106 mmYes (Within range)
Width (9)114 +/- 6 mm112 mmYes (Within range)
Thickness (Cuff, Palm, Fingertip)0.1 mm minimum0.12 mm (cuff), 0.16 mm (palm), 0.19 mm (fingertip)Yes
Physical Properties
Tensile Strength (Before Ageing)24 mpa27 mpaYes
Elongation at break % (Before Ageing)750%850%Yes
Modulus at 500 % elongation (Before Ageing)5.5 mpa (max)3 mpaYes
Tensile Strength (After Ageing)18 mpa min20 mpaYes
Elongation at break % (After Ageing)560% min750%Yes
Performance Requirements
SterilityAs per USPAs per IP (Indian Pharmacopeia)Yes (Assumed equivalent in practice)
Freedom from Holes (AQL)1.51.5 (S4 inspection level)Yes
Dimension (AQL)44 (S2 inspection level)Yes
Physical Property (AQL)44 (S2 inspection level)Yes
Other Properties
Powder Content150 mg/glove max (on Medium size)120 +/- 20 mg on size 8 (Implicitly meets for medium if met for size 8)Yes
Protein Content200 ppm (max) (FDA Requirement)80 +/- 20 ppmYes
Moisture ContentNo value fixed (FDA)0.8% maxN/A (Device provides stricter internal limit)
BiocompatibilityBiologically Compatible (FDA)Biologically CompatibleYes

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the sample sizes (e.g., number of gloves) used for each specific test. It indicates "BGPL Value" for measured parameters, suggesting the company conducted these tests. The provenance of the data is from Brightway Gloves Pvt. Ltd., located in Tamil Nadu, India. The tests appear to be a retrospective comparison against established standards (ASTM D3577 and FDA requirements) rather than a prospective clinical study.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This information is not provided. The "ground truth" for this device is based on compliance with standardized measurements and performance specifications (ASTM D3577, USP, FDA requirements). These standards are developed by expert committees, but the submission does not detail the involvement of specific experts in establishing the ground truth for this particular test set.

4. Adjudication Method for the Test Set

This is not applicable as the compliance is based on objective, standardized measurements against published technical specifications rather than subjective expert assessment requiring adjudication.

5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

A MRMC comparative effectiveness study was not done. This type of study is typically used for diagnostic or interpretive devices where human interpretation of images or data is involved. Surgeon's gloves are a physical product, and their performance is evaluated through material science and physical property tests, not through human reader interpretation.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

This is not applicable. The device is a physical product (surgeon's glove), not an algorithm or AI system. Its performance is inherent to the product itself, not mediated by an AI.

7. Type of Ground Truth Used

The ground truth used is based on established industry standards and regulatory requirements, specifically:

  • ASTM D3577: Standard Specification for Rubber Surgical Gloves.
  • USP (United States Pharmacopeia) / IP (Indian Pharmacopeia): For sterility testing.
  • FDA Requirements: For protein content, moisture content (though no fixed value), and inherent biocompatibility.

These are specifications for physical and chemical properties, not clinical outcomes or expert consensus on a diagnostic interpretation.

8. Sample Size for the Training Set

This is not applicable. As a physical medical device (surgeon's gloves) being assessed for compliance with established performance standards, there is no "training set" in the context of machine learning or AI models. The gloves are manufactured and then tested to ensure they meet specifications.

9. How the Ground Truth for the Training Set Was Established

This is not applicable for the same reasons as point 8. There is no concept of a "training set" or "ground truth for a training set" for this type of device submission.

§ 878.4460 Non-powdered surgeon's glove.

(a)
Identification. A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls).