K Number
K014174
Device Name
HARVEY PV DRY STERILIZER, MODEL # ST127325; ST127320; ST127320-33
Date Cleared
2002-02-22

(64 days)

Product Code
Regulation Number
880.6880
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
The Harvey® PV Dry is a prevacuum and post vacuum drying sterilizer intended to provide sterilization of wrapped or unwrapped medical instruments, surgical devices and other heat stable devices in pouches, surgical packs, dental handpieces, instruments, surgical devices and outer water is required for operation of the Harvey® PV Dry. The sterilizer has four standard cycles: Unwrapped Instruments (135C for 3 minutes); Wrapped Instruments (121C for 30 minute); Special, for dental handpieces (5 minutes (155c for 11ve manator), I 2015 (15) 10 11 (15 ) 10 routine testing of the steam penetration capability into packs. The parameters are fixed for each cycle.
Device Description
The Harvey® PV Dry is a pre/post-vacuum table top steam sterilizer. Its enclosure is made of plastic. The outside dimensions are 15.5"Hx19"Wx23"D, it weighs 115lbs and the chamber is 10" in diameter. The chamber is made from stainless steel at 45 it weighs 11910s and the enamour is 10 - an Mechanical Engineers (ASME) Pressure Vessel Code.
More Information

Not Found

No
The summary describes a standard steam sterilizer with fixed cycle parameters and does not mention any AI or ML capabilities.

No.
The device is a sterilizer, which is used to sterilize medical instruments and devices, not to treat a disease or condition in a patient.

No

Explanation: The device is described as a "pre/post-vacuum table top steam sterilizer" intended to sterilize medical instruments. Its purpose is to prepare instruments for use, not to diagnose medical conditions.

No

The device description clearly indicates it is a physical steam sterilizer with specific dimensions, weight, and materials, including a stainless steel chamber. This is a hardware device, not software-only.

Based on the provided text, the Harvey® PV Dry is not an In Vitro Diagnostic (IVD) device.

Here's why:

  • Intended Use: The intended use clearly states that the device is a sterilizer for medical instruments, surgical devices, and other heat-stable items. This is a sterilization process, not a diagnostic test performed on biological samples in vitro (outside the body).
  • Device Description: The description details a steam sterilizer with a chamber and pressure vessel, consistent with sterilization equipment.
  • Lack of IVD Characteristics: There is no mention of analyzing biological samples, detecting analytes, or providing diagnostic information.

Therefore, the Harvey® PV Dry falls under the category of medical devices used for sterilization, not IVD devices.

N/A

Intended Use / Indications for Use

The Harvey® PV Dry is a prevacuum and post vacuum drying sterilizer intended to provide sterilization of wrapped or unwrapped instruments, surgical devices and other heat stable devices in pouches, surgical packs, dental handpieces, instruments, surgical devices and other small articles. Water is required for operation of the Harvey® PV Dry.

The sterilizer has four standard cycles: Unwrapped Instruments (135C for 3 minutes); Wrapped Instruments (121C for 30 minute); Special, for dental handpieces (5 minutes (135c for 11 minutes)), and Bowie-Dick, for routine testing of the steam penetration capability into packs. The parameters are fixed for each cycle.

Product codes

FLE

Device Description

The Harvey® PV Dry is a pre/post-vacuum table top steam sterilizer. Its enclosure is made of plastic. The outside dimensions are 15.5"Hx19"Wx23"D, it weighs 115lbs and the chamber is 10" in diameter. The chamber is made from stainless steel (type 316L) and is rated and stamped by the American Society of Mechanical Engineers (ASME) Pressure Vessel Code.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

The Harvey® PV Dry is designed for use in medical offices, hospitals, clinics, and other facilities where a variety of instruments and other heat stable medical materials require sterilization.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Non-clinical Testing: Validation studies were conducted by SPS Medical Supply Corporation located at 6789 West Henrietta Road, Rush, NY 14543. Validation testing is in accordance with ANSI/AAMI ST55, ANSI/AAMI ST37. Since neither standard specifically addresses handpiece cycle validation, additional handpiece cycle validation testing was performed with the protocol being written in accordance with FDA Draft Guidance Document on Dental Handpieces, 1995, and the Guidance on Premarket Notification Submissions for Sterilizers Intended for Use in Health Care Facilities, 1993.

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Successful sterilization was achieved in all three validations and the declarations of conformity to the consensus standards are located in Section III. Testing results and raw data are contained in the product's device master record.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s)

K924955

Reference Device(s)

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information

Not Found

§ 880.6880 Steam sterilizer.

(a)
Identification. A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.(b)
Classification. Class II (performance standards).

0

510(k) SUMMARY

K014174

of the roll and be A. Submitter's Name and Address Barnstead/Thermolyne Corp. P.O. Box 797 Dubuque, IA 52004

FEB 2 2 2002

  • B. Contact Person Mia M. Ware Regulatory Affairs Specialist 563-556-2241 Ext. 485 Fax: 563-557-0612
    C. Establishment Registration Number of Submitter

1950043

Hoch Warry Con

  • D. Contract Manufacturing Facility
    Not Applicable

  • E. Device Name
    Proprietary Name: Harvey® PV Dry Common Name: Steam Sterilizer Classification Name: Steam Sterilizer

  • F. Device Classification
    Class II §880.6880

  • G. Action Taken to Comply with Section 514 of the Act
    The Agency has recognized the ANSVAAMI ST55, Tabletop Steam Sterilizers, and the FDA Guidance on Premarket Notification Submissions for Sterilizers Intended for Use in Health Care Facilities, March 1993, and its addendum, dated September, 1995. Conformance or variance with these standards is described on the following pages.

  • H. Reason for 510(k) Submission

    • Initial Product Introduction
    • New Model for Product-line Extension

Initial Import into the USA

  • Other (Include in Part IV an explanation referenced to Part I. H.)
  • Predicate Device: Harvey® MC10 Steam Sterilizer, K924955 ﻟﺴﻨﺔ

1

ANAMALA DE or with the Describe Description of the Device: The Harvey® PV Dry is a pre/post-vacuum table top steam sterilizer. Its enclosure Description of the Device. The Harvey - 1 - Dis of pastic. The outside dimensions are 15.5"Hx19"Wx23"D, it weighs 115lbs and the chamber is 10" in diameter. The chamber is made from stainless steel at 45 it weighs 11910s and the enamour is 10 - an Mechanical Engineers (ASME) Pressure Vessel Code.

Intended Use of the Device: The Harvey® PV Dry is a pre/post-vacuum table top steam sterilizer that is micined Ose of the Device. The Harry offices, hospitals, clinics, and other facilities where a variety of designed for use in mealed for sterilization of wrapped or unwrapped instruments, dental handpieces, and linen packs.

Technological Characteristics: The Harvey® PV Dry has very similar technological characteristics as the recimented characteriation . In Steam Sterilizer. The Harvey® PV Dry is a table-top autoclave steam predicate uction for the roy - Norto over and sterile post-vacuum drying, similar to larger hospital type units.

The Harvey® PV Dry provides superior performance for dental handpieces and difficult to penetrate fabric packs and superior and rand drying for all loads. With sterile vacuum drying, the Harvey® PV Dry is packs and superfor and rapid of the surfere packs, bagged instruments, and wrapped instrument sets, such as surgical kits are stored for later usage. The integral sterile drying in a 30-40 minute cycle, gives sam as salged this and tions of the instruments at the time of delivery to the operatory. Sterile vacuum drying eliminates the need for extended drying cycles using non-sterile air.

Some features unique to the Harvey® PV Dry also demonstrate safety as well as the efficiency of the unit. The interlocking device on the door prevents manual opening until the chamber pressure decreases to near I he more corroo on the econologies which help in the convenience and ease of operation of the Harvey® PV Dry include indicator lights to tell the user when the waste tank needs to be emptied, and controls preventing the running of a cycle until it is drained, indicator lights telling the user when the water supply needs to be replenished, and cycle parameter display. In the predicate device comparison matrix in Section II, there are detailed differences and similarities between the Harvey® PV Dry and its predicate device, Harvey MC10 Steam Sterilizer.

Non-clinical Testing: Validation studies were conducted by SPS Medical Supply Corporation located at 6789 West Henrietta Road, Rush, NY 14543. Validation testing is in accordance with ANSI/AAMI ST55, ANSVAAMI ST37. Since neither standard specifically addresses handpiece cycle validation, additional handpiece cycle validation testing was performed with the protocol being written in accordance with FDA Draft Guidance Document on Dental Handpieces, 1995, and the Guidance on Premarket Notification Submissions for Sterilizers Intended for Use in Health Care Facilities, 1993. Successful sterilization was achieved in all three validations and the declarations of conformity to the consensus standards are located in Section III. Testing results and raw data are contained in the product's device master record.

Conclusion: It is Barnstead/Thermolyne's conclusion that the Harvey® PV Dry is substantially equivalent to its predicate device, the Harvey® MC10 Steam Sterilizer. Based on the validation results and information submitted, the Harvey® PV Dry provides effective sterilization of the indicated medical materials.

2

Public Health Service

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 2 2 2002

Barnstead/Thermolyne Corporation C/O Mr. Reiner Krumme Responsible Third Party TUV Rheinland of North America, Inc 12 Commerce Road Newton, Connecticut 06470

Re: K014174

Trade/Device Name: Harvey PV Dry Sterilizer, Model # ST127325; ST127320 Regulation Number: 880.6880 Regulation Name: Steam Sterilizer Regulatory Class: II Product Code: FLE Dated: February 12, 2002 Received: February 13, 2002

Dear Ms. Krumme:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting (1 vir 1) it incan be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements

3

Page 2 – Mr. Krumme

of the Act or any Federal statutes and regulations administered by other Federal agencies. of the Act of ally I odelar sates and enginements, including, but not limited to: registration You must comply with an alle Here requires (21 CFR Part 801); good manufacturing practice and listing (21 CFR Part 067), as times (QS) regulation (21 CFR Part 820); and if requirences as set forum in als quality ion control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section This letter will anow you to began marketing your substantial equivalence of your device to 310(k) premailed nourvailed with a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 II you desire specific ad not 10. your for in vitro diagnostic devices), please contact the and additionally 21 OF IC F 21 S 1 4 CL 1 S 94-4618. Additionally, for questions on the FR 1 6 Complia Office of Comphance are, please contact the Office of Compliance at (301) 594-4639. advertising or your are in " Anotification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and the Act may be obtained from and Brumber (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

4

Image /page/4/Picture/0 description: The image shows handwritten text that reads "Attachment" and "02.11.2002". The text is written in black ink on a white background. The word "Attachment" is written on the top left, and the date is written below it.

INDICATIONS FOR USE STATEMENT

510(k) Number : 014174

Device Name: Harvey® PV Dry

Indications for Use:

The Harvey® PV Dry is a prevacuum and post vacuum drying sterilizer intended to provide sterilization I he Harvey " V Dry is a prevacuall and post ratedial artistication of wrapped or unwrapped or unwrapped meal instruments, surgical devices and other heat stable devices in pouches, surgical packs, dental handpieces, instruments, surgical devices and outer water is required for operation of the Harvey® PV Dry.

The sterilizer has four standard cycles: Unwrapped Instruments (135C for 3 minutes); Wrapped I the stellized has four samoure eyeles. One of a (121C for 30 minute); Special, for dental handpieces (5 minutes (155c for 11ve manator), I 2015 (15) 10 11 (15 ) 10 routine testing of the steam penetration capability into packs. The parameters are fixed for each cycle.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) ______________________________________________________________________________________________________________________________________________________________________________ Concurrence of CDRH, Office of Device Evaluation (ODE)

(Optonal Format 3-10-98)

("Tivision Sign-Off) El Vision of Dental, Infection Control, Ceneral Hospital Devices -. .. Privaber _______________________________________________________________________________________________________________________________________________________________