(336 days)
Not Found
K01, G01
No
The 510(k) summary describes a blood tubing set for hemodialysis, a purely mechanical device, and contains no mention of AI, ML, image processing, or any other technology typically associated with AI/ML.
Yes
The device is used as part of hemodialysis therapy for patients with renal disease, directly participating in the treatment process.
No.
The device's intended use is to transport blood during hemodialysis therapy; it does not mention diagnosing any condition.
No
The device description clearly states it is a "HEMODIALYSIS BLOOD TUBING SET," which is a physical, hardware component used in hemodialysis therapy.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use clearly states that the device is a "conduit through which blood is taken from the patient, delivered to the hemodialyzer, and returned to the patient during hemodialysis therapy." This describes a device that is used in contact with the patient's blood for a therapeutic procedure (hemodialysis).
- IVD Definition: In vitro diagnostics (IVDs) are devices that are used to examine specimens (like blood, urine, or tissue) outside of the body to provide information for diagnosis, monitoring, or screening.
The RENAX (SUNDER) HEMODIALYSIS BLOOD TUBING SET is a medical device used in vivo (within the body) during a medical procedure, not for testing specimens in vitro.
N/A
Intended Use / Indications for Use
The RENAX (SUNDER) HEMODIALYSIS BLOOD TUBING SET (model: SD-K01, SD-G01) is intended to serve as a conduit through which blood is taken from the patient, delivered to the hemodialyzer, and returned to the patient during hemodialysis therapy. The device is intended to be used for patients suffering from renal disease only.
Product codes
78 FJK
Device Description
Not Found
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Not Found
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Not Found
Predicate Device(s)
Not Found
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 876.5820 Hemodialysis system and accessories.
(a)
Identification. A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient.
(2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860).
(3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860).
(4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.
(b)
Classification. (1) Class II (performance standards) for hemodialysis systems and all accessories directly associated with the extracorporeal blood system and the dialysate delivery system.(2) Class I for other accessories of the hemodialysis system remote from the extracorporeal blood system and the dialysate delivery system, such as the unpowered dialysis chair, hemodialysis start/stop tray, dialyzer holder set, and dialysis tie gun and ties. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
0
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or a bird in flight, composed of three stylized human profiles facing to the right.
Public Health Service
NOV 1 8 2002
Food and Drug Administratio 9200 Corporate Boulevard Rockville MD 20850
Mr. Tony Hung President Sunder Biomedical Tech. Co., Ltd. 10F-1. 1-67. Wu-Chuan Road TaiChung City TAIWAN
Re: K014140 Trade/Device Name: RENAX (SUNDER) Hemodialysis Blood Tubing Set (Models SD-K01, SD-G01) Regulation Number: 21 CFR §876.5820 Regulation Name: Hemodialysis system and accessories Regulatory Class: II Product Code: 78 FJK Dated: August 12, 2002 Received: August 20, 2002
Dear Mr. Hung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
1
Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
8xx.1xxx | (301) 594-4591 |
---|---|
876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 |
884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 |
Other | (301) 594-4692 |
Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Hogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
2
510(k) NUMBER (IF KNOWN): ______k014140 DEVICE NAME: RENAX (SUNDER) HEMODIALYSIS BLOOD TUBING SET (model: SD-K01, SD-G01) __________________________________________________________________________________________________________________________________________________ INDICATIONS FOR USE:
The RENAX (SUNDER) HEMODIALYSIS BLOOD TUBING SET (model: SD-K01, SD-G01) is intended to serve as a conduit through which blood is taken from the patient, delivered to the hemodialyzer, and returned to the patient during hemodialysis therapy. The device is intended to be used for patients suffering from renal disease only.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David G. deyson
on Sian-Off ion of Reproducti and Radiological Devi 510(k) Number
Prescription Use
(Optional Format 3-10-98)