(85 days)
A latex examination glove is a disposable device intended for medical purpose that is worn on examiner's hand or finger to prevent contamination between patient and examiners
A sterile latex powderfree examination glove is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiners.
Latex Powder free Sterile Examination Gloves
The provided text describes the acceptance criteria and a study for "Latex Powder free Sterile Examination Gloves."
Here's the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
| Characteristic | Acceptance Criteria (Performance Requirement) | Reported Device Performance |
|---|---|---|
| Sterility | Fails sterility (Related Defects) | Meets the sterility assurance level. |
| Freedom from holes | Holes (Related Defects) | Meets FDA pinhole requirement. |
| Dimensions & Thickness | Width, Length (Related Defects) | Meets ASTM D 3578-01 Standard. (Implied) |
| Physical Properties (after ageing) | Before and After (Related Defects) | Meets ASTM D 3578-01 Standard. (Implied)Tensile Strength: 14 Mpa (before & after ageing)Ultimate Elongation: 650% (min, before ageing), 600% (min, after ageing) |
| Powder Free Residue | Exceeds Maximum Limit | Meets labeling claim. (Implied, if labeling claims "powder free")Exceeds Recommended Maximum Limit (Powder Amount, separate) |
| Protein Content | Exceeds Recommended Maximum Limit | Not explicitly stated, but implied to meet claims. |
| Powder Amount | Exceeds Recommended Maximum Limit | Meets labeling claim. (Implied) |
Note: The document states "Performance data is the same as mentioned immediately above," referring to the table of "Technological characteristic of the gloves" and "Performance requirement." The "Non-clinical data" section then certifies that these requirements are met.
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly state the sample sizes for each test set. However, it does provide Inspection Levels and AQL values for some characteristics, which would imply specific sample sizes based on sampling plans. For instance:
- Freedom from holes: Inspection Level 1, AQL 2.5
- Dimensions & Thickness: Inspection Level S-2, AQL 4
- Physical Properties (after ageing): Inspection Level S-2, AQL 4
- Powder Free Residue: N=5
- Protein Content: N=3
- Powder Amount: N=2
The data provenance is not explicitly stated in terms of country of origin but is from the manufacturer PT. Maja Agung Latexindo in North Sumatra, Indonesia. The study appears to be retrospective as it presents data for a device seeking 510(k) clearance based on established standards.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts
This information is not provided in the document. The acceptance criteria are based on ASTM standards (ASTM D 3578-01), FDA pinhole requirements, and generally accepted sterility assurance levels. The compliance is certified by the manufacturer. There is no mention of external experts establishing ground truth for individual test cases.
4. Adjudication Method for the Test Set
This information is not provided. The compliance is self-certified by the manufacturer against established standards.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance
This is not applicable. The device is a physical product (latex gloves), not an AI diagnostic tool. No MRMC study or AI assistance is mentioned.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
This is not applicable. The device is a physical product (latex gloves), not an AI diagnostic tool.
7. The Type of Ground Truth Used
The ground truth is based on established industry standards and regulatory requirements, specifically:
- ASTM D 3578-01 Standard (for dimensions, physical properties)
- FDA pinhole requirement
- Sterility assurance level
- Labeling claims (for powder free, powder amount)
These standards define the objective measures for performance.
8. The Sample Size for the Training Set
This is not applicable as the document describes the testing of a physical product, not the development of a machine learning algorithm that requires a training set.
9. How the Ground Truth for the Training Set Was Established
This is not applicable as there is no training set mentioned for this product.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a stylized logo or graphic. The letters "Ma" are prominently displayed in a bold, sans-serif font. The letters are overlaid on a background of horizontal lines, creating a textured effect. The overall design is simple and modern, with a focus on the typography.
A AGUNG LAT MANUFACTURE OF LATEX GLOVE
Jln. Utama No. 98 PUJI MULYO SUNGGAL - DELI SERDANG SUMATERA UTARA - INDONESIA
Telp. 62-61 - 8459160 62-61 - 8459170 Fax. 62-61 - 8459180
Page Numbers 1 of 2
13996
FEB 2 6 2002 "510 (K)" SUMMARY
| (1) Name of applicant | : Mr. Hansen Laurence |
|---|---|
| Address | : PT. MAJA AGUNG LATEXINDOJl. Utama No. 98 Puji MulioSunggal -Deli SerdangNorth Sumatra - Indonesia |
| Phone No. | : 62-61-845-9170 |
| Fax No. | : 62-61-845-9180 |
| The contact persons within the firm as well as in U.S.A are given below: | |
| Contact person in firm | : Mr. Hansen Laurence |
| Fax No. | : 62-61-845-9180 |
| Contact person in U.S.A | : Emmy Tjoeng |
| Fax No. | : 909-591-8878 |
| (2) Device detailsTrade Name | : Latex Powder free Sterile Examination Gloves |
| Classification Name | : Latex Powder free Sterile Examination Gloves |
| (3) Product Code | : Latex 80 LYY |
| (4) Equivalent device legallymarketed | : Class I Latex Examination Gloves 80 LYYPre - Powdered meeting ASTM D 3578 – 01Sterilized by gamma radiation |
| (5) Intended use | : A latex examination glove is a disposable device intendedfor medical purpose that is worn on examiner's hand orfinger to prevent contamination between patient andexaminers |
{1}------------------------------------------------
196
Jin. Utama No. 98 PUJI MULYO SUNGGAL - DELI SERDANG SUMATERA UTARA - INDONESIA Telp. 62-61 - 8459160 62-61 - 8459170 Fax. 62-61 - 8459180
Page Numbers 2 of 2
(6) Technological characteristic of the gloves.
| DimensionsSizes | XS | S | M | L | XL |
|---|---|---|---|---|---|
| Length | 240 mm | 240 mm | 240 mm | 240 mm | 240 mm |
| Width | $75\pm5$ mm | $80\pm10$ mm | $00\pm10$ mm | $105\pm10$ mm | $115\pm5$ mm |
| THICKNESS1. Cuff (min) | 0.10 mm | 0.10 mm | 0.10 mm | 0.10 mm | 0.10 mm |
| 2. Palm (min) | 0.10 mm | 0.10 mm | 0.10 mm | 0.10 mm | 0.10 mm |
| 3. Finger Tip (min) | 0.10 mm | 0.10 mm | 0.10 mm | 0.10 mm | 0.10 mm |
- b. Physical Properties
a.
| Before ageing | After ageingat 70°C 168 hrs | |
|---|---|---|
| Tensile Strength : | 14 Mpa | 14 Mpa |
| Ultimate Elongation : | 650 % (min.) | 600 % (min.) |
c. Performance requirement
| Characteristic | Related Defects | InspectionLevel | AQL |
|---|---|---|---|
| Sterility | Fails sterility | A | N/A |
| Freedom from holes | Holes | 1 | 2.5 |
| Dimensions& Thickness | Width Length | S-2 | 4 |
| Physical Propertiesafter ageing | Before and | S-2 | 4 |
| Powder Free Residue | Exceeds Maximum Limit | N=5 | N/A |
| Protein Content | Exceeds RecommendedMaximum Limit | N=3 | N/A |
| Powder Amount | Exceeds RecommendedMaximum Limit | N=2 | N/A |
(7) Performance data is the same as mentioned immediately above.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image is a black and white logo. The logo has a large letter "L" on the left side. The letter "Ma" is written in cursive on the right side of the logo. The background of the logo has horizontal lines.
MAJA AGUNG LATE
MANUFACTURE OF LATEX GLO
Jin. Utama No. 98 PUJI MULYO SUNGGAL - DELI SERDANG SUMATERA UTARA - INDONESIA
Telp. 62-61 - 8459160 62-61 - 8459170 62-61 - 8459180 Fax.
- (8) Clinical date is not needed for gloves or for most devices cleared by the 510 (K) process.
- (9) Non-clinical data
We certify that the gloves meet or exceed the ASTM D 3578 - 01 Standard. Meets FDA pinhole requirement. Meets labeling claim. Meets the sterility assurance level.
{3}------------------------------------------------
Image /page/3/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, with three curved lines representing the wings and body.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 2 6 2002
PT. Maja Agung Latexindo C/O Ms. Emmy T. Joeng Shamrock Marketing Company, Incorporated 5445 Daniel Street Chino, California 91710
Re: K013996
Trade/Device Name: Latex Powderfree Sterile Examiantion Gloves Regulation Number: 880.6250 Regulation Name: Patient Examination Gloves Regulatory Class: I Product Code: LYY Dated: November 29, 2001 Received: December 3, 2001
Dear Ms. Joeng:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave roviewed your becament mined the device is substantially equivalent (for the releveloca above and in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device mendments, or to devices that have been reclassified in accordance with the provisions of Antichements, or to do rood alle metic Act (Act) that do not require approval of a premarket the rederal 1 000; Drag, Lrag, therefore, market the device, subject to the general approvin application (the Act. The general controls provisions of the Act include controls providents or a registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting (1 will), it inch be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In your device ear. or rouncements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{4}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618 . Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.qov/cdrh/dsma/dsmamain.html
Sincerely yours,
Susan Kane
Timothy A. Ulatowski Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for PT. MAJA AGUNG LATEXINDO, a manufacturer of latex gloves. The logo features a stylized graphic to the left of the company name in bold, sans-serif font. Above the company name is the number 013996.
Jin. Utama No. 98 PUJI MULYO SUNGGAL - DELI SERDANG SUMATERA UTARA - INDONESIA
Telp. 62-61 - 8459160 62-61 - 8459170 62-61 - 8459180 Fax.
ANNEXURE II
INDICATION FOR USE
| Applicant | : Mr. Hansen Laurence |
|---|---|
| Device Name | : Latex Powderfree Sterile Examination Gloves |
| Indication for use | : |
A sterile latex powderfree examination glove is a disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiners.
Sisan Pani
(Tivision Sign-Off)
Tivision Sign-Off)
Dontal ("Tivision Sign-Off)
Division of Dental, Infection Control,
Division of Dental, Infection Devienti Division of General Hospital D r 10(k) Number -
§ 880.6250 Non-powdered patient examination glove.
(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.