AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Dual Port Wizard™ replacement gastrostomy device is indicated for percutaneous placement of a low-profile gastrostomy feeding and decompression device into an established, appropriately sized stoma. The Dual Port Wizard™ Feeding/Decompression Tube (REF 00220W) is the only device which should be used for feeding, gastric decompression or medication administration with the Dual Port Wizard replacement gastrostomy device.

Device Description

The Wizard™ is a low profile, balloon-type, feeding device designed for percutaneous insertion through an established, appropriately sized stoma tract. The device is held in place by the distal balloon, which remains in situ within the stomach, and an external bolster. Within the bolster are two ports. One side port is used to inflate the retention balloon and the second, surface port mates with the accompanying feeding/decompression tube to administer enteral nutritional fluid or expel gas or gastric contents.

AI/ML Overview

The provided text describes modifications to an existing gastrostomy device and states that "Bench testing, biocompatibility, and post-risk analysis assessment of the modifications to the Wizard™ demonstrate that design verification/validation activities meet the predetermined acceptance criteria for a safe and effective device."

However, the document does not explicitly list the specific acceptance criteria, nor does it provide a detailed study report with performance metrics against those criteria. It only makes a general statement that the criteria were met.

Therefore, for categories where specific information is requested (like sample size, number of experts, adjudication methods, details of comparative or standalone studies, and ground truth establishment), the information is not present in the provided 510(k) summary. I can only report what is stated directly in the text.

Here's a breakdown of the available information based on your request:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria for Device Performance:

CriterionDescriptionReported Device Performance
Safety and EffectivenessModifications to the device should maintain or improve safety and effectiveness."Bench testing, biocompatibility, and post-risk analysis assessment of the modifications to the Wizard™ demonstrate that design verification/validation activities meet the predetermined acceptance criteria for a safe and effective device."
Material ClassificationAll correlating material changes must be within the same classification as other cleared feeding tubes."All correlating material changes are within the same classification as other cleared feeding tubes."
Mating with Feeding/Decompression TubeThe device must properly mate with the interlocking feeding/decompression tube."The only dimensional modification to the Wizard™ Low Profile Replacement Gastrostomy Device is related to the mating of an interlocking feeding/decompression tube." (Implies successful mating was demonstrated)

Note: The document does not provide quantitative or specific acceptance criteria (e.g., "burst pressure must exceed X psi") or detailed numerical performance results from the bench testing.

2. Sample Size Used for the Test Set and Data Provenance

The document states "Bench testing" was performed, but does not specify the sample size used for these tests. It also does not mention data provenance (e.g., country of origin, retrospective or prospective data), as the tests were on the device itself, not patient data in the typical sense of AI/clinical studies.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

This information is not provided in the document. The "ground truth" here would refer to the standards by which the bench tests were judged successful or unsuccessful, but the document does not detail how these standards were established or by whom.

4. Adjudication Method for the Test Set

Adjudication methods (like 2+1, 3+1) are typically used in clinical studies involving human interpretation of data. Since the document describes "bench testing" of a physical device, this concept does not apply in the usual sense for this submission, and no such method is mentioned.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study was not done. This type of study is relevant for imaging devices or AI algorithms that assist human interpretation, which is not the nature of this product (a gastrostomy device).

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

No, a standalone performance study (in the context of an algorithm) was not done. The device is a physical medical device, not an algorithm. Bench testing would be analogous to standalone testing for a physical product, but not in the AI sense.

7. The Type of Ground Truth Used

For the "bench testing," the "ground truth" would be defined by engineering specifications, design requirements, and possibly industry standards for gastrostomy devices. The document implies these predetermined acceptance criteria were the ground truth, but does not explicitly detail them (e.g., specific forces, flow rates, durability metrics).

8. The Sample Size for the Training Set

This information is not applicable and not provided. The device is a physical product, not an AI algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable and not provided for the same reasons as point 8.

{0}------------------------------------------------

OCT 1 9 2001

BARD ENDOSCOPIC TECHNOLOGIES

BARD

KO13144 1042

C.R. Bard, Inc. 129 Concord Road P.O. Box 7031 Billerica, MA 01821-7031 978-663-8989

VI. 510(k) SUMMARY SAFETY AND EFFECTIVENESS INFORMATION

As required by the Safe Medical Devices Act of 1990, codified under Section 513, Part (i)(3)(A) of the Food Drug and Cosmetic Act, a summary of the safety and effectiveness information upon which substantial equivalence determination is based follows.

  • A. Submitter Information
Submitter's Name:Bard Endoscopic TechnologiesC.R. Bard, Inc.
Address:129 Concord Road, Bldg. #3Billerica, MA 01821
Phone:978 - 262 - 4867
Fax:978 - 262 - 4878
Contact Person:Marion Gordon, R.A.C.
Date of Preparation:September 17, 2001
B. Device Name
Trade Name:Dual Port Wizard™ Low Profile Replacement
Common/Usual Name:Gastrostomy Device & Feeding/Decompression TubeFeeding Device
Classification Name:Gastrointestinal tube and accessories
C. Predicate Device NameTrade Name:Wizard™ Low Profile Replacement Gastrostomy Device

D. Device Description:

The Wizard™ is a low profile, balloon-type, feeding device designed for percutaneous insertion through an established, appropriately sized stoma tract. The device is held in place by the distal balloon, which remains in situ within the stomach, and an external bolster. Within the bolster are two ports. One side port is used to inflate the retention balloon and the second, surface port mates with the accompanying feeding/decompression tube to administer enteral nutritional fluid or expel gas or gastric contents.

  • E. Intended Use:
    The Dual Port Wizard™ replacement gastrostomy device is indicated for percutaneous placement of a low-profile gastrostomy feeding and decompression device into an established, appropriately sized stoma. The Dual Port Wizard™ Feeding/Decompression Tube (REF 00220W) is the only device which should be

{1}------------------------------------------------

used for feeding, gastric decompression or medication administration with the Dual Port Wizard replacement gastrostomy device.

  • F. Technological Characteristics Summary:
    The fundamental scientific technology of the modified device has not changed. The only dimensional modification to the Wizard™ Low Profile Replacement Gastrostomy Device is related to the mating of an interlocking feeding/decompression tube. The feeding/decompression tube will consist of a proximal dual port adapter, which is friction fitted into a shorter length of the same tubing. The distal 90° adapter is one piece and includes a central stem, which inserts into the feeding port of the modified Wizard™, and a locking lip. All correlating material changes are within the same classification as other cleared feeding tubes.

  • G. Performance Data
    Bench testing, biocompatability, and post risk analysis assessment of the modifications to the Wizard™ demonstrate that design verification/validation activities meet the predetermined acceptance criteria for a safe and effective device.

{2}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, facing to the right.

OCT 1 9 2001

Marion Gordon, R.A.C. Senior Regulatory Affairs Coordinator C. R. Bard, Inc. Bard Endoscopic Technologies 129 Concord Road P.O. Box 7031 BILLERICA MA 01821-7031

Re: K013144

Trade/Device Name: The Wizard™ Low Profile Replacement Gastrostomy · Device Regulation Number: 21 CFR §876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: 78 KNT Dated: September 17, 2001 Received: September 20, 2001

Dear Ms. Gordon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration (21 CFR Part 807); listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Ms. Marion Gordon

This letter will allow you to begin marketing your device as described in your Section 510(k) I mis lotter will and in yourse of substantial equivalence of your device to a legally promatic liculicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on the labeling don'to to proceed to the most and additionally Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4616. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at promotion and as returning or note the regulation entitled, "Misbranding by reference to (2017) 594 4059. Plast from 2019.97). Other general information on your responsibilities promation is and may be obtained from the Division of Small Manufacturers, International and ander the Fee its toll free number (800) 638-2041 or (301) 443-6597, or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

D. Indications for Use

510(k) Number (if known):TBD K013144
Device Name:Dual Port Wizard™ Low Profile Replacement Gastrostomy Device

Indications For Use: The Dual Port Wizard™ replacement gastrostomy device is
indicated for percutaneous placement of a low-profile gastrostomy
feeding and decompression device into an established,
appropriately sized stoma.

The Dual Port Wizard™ Feeding/Decompression Tube (REF
00220W) is the only device which should be used for feeding,
gastric decompression or medication administration with the Dual

Port Wizard replacement gastrostomy device.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use(Per 21 CFR 801.109)OROver-The-Counter Use(Optional Format 1-2-96)
------------------------------------------------------------------------------------------------------------------------------------
(Division Sign-Off)Nancy Brogdon
Division of Reproductive, Abdominal,and Radiological Devices
510(k) NumberK013144

§ 876.5980 Gastrointestinal tube and accessories.

(a)
Identification. A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).(b)
Classification. (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.