(78 days)
The ACON BAR One Step Barbiturates Test Strip and ACON BAR One Step Barbiturates Test Device are rapid chromatographic immunoassays for the qualitative detection of barbiturates in urine at a cut-off concentration of 300 ng/mL. They are intended for use by healthcare professionals including professionals at point-of-care sites.
The ACON BAR One Step Barbiturates Test Strip and ACON BAR One Step Barbiturates Test Device are competitive binding, lateral flow immunochromatographic assays for the qualitative screening of barbiturates in a urine sample. The test is based on the principle of antigen-antibody immunochemistry. It utilizes the antibody to selectively detect elevated levels of barbiturates in urine specimens at a designated cut-off concentration of 300 ng/mL. These tests can be performed without the use of an instrument. A drug-positive urine specimen will not generate a colored-line in the designated test region, while a negative urine specimen or a urine specimen containing barbiturates at the concentration below the cut-off level will generate a colored-line in the test region. To serve as a procedural control, a colored-line will always appear at the control region, indicating that proper volume of specimen has been added and membrane wicking has occurred.
Here's a breakdown of the acceptance criteria and study details for the ACON® BAR One Step Barbiturates Test Strip and Test Device, based on the provided text:
Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Device Performance (ACON BAR One Step Barbiturates Test Strip) | Device Performance (ACON BAR One Step Barbiturates Test Device) |
|---|---|---|
| Comparison to Predicate Device (MCC Instant Screen Card BAR Test): | ||
| Positive Agreement: High agreement with predicate device | 100% (97% - 100%* CI) | 100% (97% - 100%* CI) |
| Negative Agreement: High agreement with predicate device | 99% (97% - 100%* CI) | 99% (97% - 100%* CI) |
| Overall Agreement: High agreement with predicate device | 99% (98% - 100%* CI) | 99% (98% - 100%* CI) |
| Comparison to GC/MS (at 300 ng/ml secobarbital cut-off): | ||
| Positive agreement with GC/MS: > 90% (implied for substantial equivalence) | 92% (86% - 96%* CI) | 92% (86% - 96%* CI) |
| Negative agreement with GC/MS: > 90% (implied for substantial equivalence) | 98% (94% - 99%* CI) | 98% (94% - 99%* CI) |
| Total agreement with GC/MS: > 90% (implied for substantial equivalence) | 95% (92% - 97%* CI) | 95% (92% - 97%* CI) |
| Clinical Performance: Safe and effective in qualitatively detecting barbiturates in urine specimens at a secobarbital cut-off concentration of 300 ng/mL | Demonstrated and determined to be substantially equivalent to predicate. | Demonstrated and determined to be substantially equivalent to predicate. |
| Suitability for Point-of-Care Use: Demonstrated suitability for healthcare professionals, including point-of-care settings. | Demonstrated through Point-of-Care (POL) study (details not provided beyond mention) | Demonstrated through Point-of-Care (POL) study (details not provided beyond mention) |
| *CI: 95% Confidence Intervals |
Study Details
-
Sample Size Used for the Test Set and Data Provenance:
- Sample Size: 292 clinical urine specimens.
- Composition: Included 10% of samples with barbiturate concentrations at -25% to +25% of the cut-off range (300 ng/mL secobarbital).
- Data Provenance: Not explicitly stated, but it's a "clinical evaluation" using "clinical urine specimens," implying real-world human samples. No country of origin is specified, nor is it explicitly stated as retrospective or prospective; however, given it's a clinical evaluation for regulatory submission, it would typically be a prospective collection or a specifically curated retrospective collection for the validation study.
-
Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
- The document refers to Gas Chromatography/Mass Spectrometry (GC/MS) analysis as the ground truth. GC/MS is an analytical chemistry technique, not a human expert. Therefore, the concept of "number of experts" and "qualifications of experts" does not directly apply in the same way it would for image interpretation or diagnosis. The "experts" would be the laboratory personnel performing and interpreting the GC/MS results, who are presumed to be qualified in analytical chemistry methods.
-
Adjudication Method for the Test Set:
- Not applicable as the ground truth is established by a definitive analytical method (GC/MS), not by human interpretation requiring adjudication. Agreement was measured by comparing the device's qualitative results to the quantitative GC/MS results against the 300 ng/mL cut-off.
-
If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC comparative effectiveness study was not done. This device is an immunoassay for chemical detection, not an AI-assisted diagnostic tool that would typically involve human readers interpreting images or complex data with and without AI assistance.
-
If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Yes, the performance reported for the ACON BAR One Step Barbiturates Test Strip and Test Device is standalone performance. It measures how the device itself performs against the predicate device and the GC/MS ground truth, without human-in-the-loop performance influencing the result-generating mechanism of the immunoassay itself. The "human-in-the-loop" would be the healthcare professional interpreting the visual line on the strip, but the performance metrics are for the device's ability to produce that correct visual indication.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- The primary ground truth used was Gas Chromatography/Mass Spectrometry (GC/MS) analysis at the secobarbital cut-off concentration of 300 ng/ml. This is considered a highly accurate and definitive analytical method for drug detection.
-
The sample size for the training set:
- The document describes a "clinical evaluation" and does not specify a separate "training set" in the context of machine learning algorithms. For in-vitro diagnostic devices like immunoassays, the development process involves internal verification and validation, but not a distinct "training set" in the AI sense. The 292 clinical urine specimens constitute the independent validation set against which the device's performance was formally assessed for regulatory submission.
-
How the ground truth for the training set was established:
- As there is no explicitly defined "training set" in the AI sense, this question is not directly applicable. The device's underlying immunoassay principle is based on antigen-antibody binding, which is a chemical process, not a learned one. The "ground truth" for developing such a device would reside in the chemical properties of the antibodies and antigens and their interaction with known concentrations of the target analyte (barbiturates).
{0}------------------------------------------------
NOV 0 9 2001
SUMMARY OF 510(k) SAFETY AND EFFECTIVENESS 8.
This summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The Assigned 510(k) number is K012824.
Submitter:
ACON Laboratories, Inc. 4108 Sorrento Valley Boulevard San Diego, California 92121
Tel.: 858-535-2030 Fax: 858-535-2038
Date:
August 17, 2001
Contact Person:
Edward Tung, Ph.D.
Product Names:
ACON® BAR One Step Barbiturates Test Strip
ACON® BAR One Step Barbiturates Test Device
Common Name:
Immunochromatographic test for the qualitative detection of barbiturates in urine
Device Classification:
The ACON BAR One Step Barbiturates Test Strip and ACON BAR One Step Barbiturates Test Device are similar to other FDA-cleared devices for the qualitative detection of barbiturates in urine specimens. These tests are used to provide a preliminary analytical result (21 CFR 862.3150). This type of barbiturates test systems (Product Code: DIS) have been classified as Class II devices with moderate complexity.
{1}------------------------------------------------
Classification Name:
Barbiturates test system
Intended Use:
The ACON® BAR One Step Barbiturates Test Strip and ACON BAR One Step Barbiturates Test Device are rapid chromatographic immunoassays for the qualitative detection of barbiturates in urine at a cut-off concentration of 300 ng/mL. They are intended for use by healthcare professionals including professionals at point-of-care sites.
Description:
The ACON BAR One Step Barbiturates Test Strip and ACON BAR One Step Device are competitive binding. lateral flow Test Barbiturates immunochromatographic assays for the qualitative screening of barbiturates in a urine sample. The test is based on the principle of antigen-antibody immunochemistry. It utilizes the antibody to selectively detect elevated levels of barbiturates in urine specimens at a designated cut-off concentration of 300 ng/mL. These tests can be performed without the use of an instrument.
A drug-positive urine specimen will not generate a colored-line in the designated test region, while a negative urine specimen or a urine specimen containing barbiturates at the concentration below the cut-off level will generate a colored-line in the test region. To serve as a procedural control, a colored-line will always appear at the control region, indicating that proper volume of specimen has been added and membrane wicking has occurred.
Predicate Device:
MCC Instant Screen Card BAR Test
Distributed by Medical & Clinical Consortium (MCC), City of Industry, CA 91746-2048
510(k) Number: K002337
Comparison to a Predicate Device:
A comparison of the features of the ACON BAR One Step Barbiturates Test Strip and ACON BAR One Step Barbiturates Test Device versus the MCC Instant Screen Card BAR Test is shown below:
{2}------------------------------------------------
- Both tests are assays intended for the qualitative detection of barbiturates in urine ● samples.
- Both tests are intended as a screening method that provides a preliminary analytical ● test result.
- Both tests are immunochromatographic, lateral flow assays for the rapid detection of ● barbiturates with a visual, qualitative end result.
- Both tests utilize the same basic immunoassay principles that rely on antigen/ ● antibody interactions to indicate a positive or negative result.
- Both tests have a cut-off concentration of 300 ng/mL for secobarbital. .
Safety and Effectiveness Data:
Accuracy
A clinical evaluation was conducted using 292 clinical urine specimens including 10% of the samples with barbiturate concentrations at -25% cut-off to +25% cut-off range. This evaluation compared the test results between ACON® BAR One Step Barbiturates Test Strip and Test Device with MCC Instant Screen Card BAR Test; as well as against data obtained from the customary Gas Chromatography/Mass Spectrometry analysis. The comparisons of data obtained from this study yielded the following results:
ACON BAR One Step Barbiturates Test Strip versus MCC Instant Screen Card BAR Test:
Positive Agreement: 126 / 126 = 100% (97% -100%) Negative Agreement: 165 / 166 = 99% (97% -100%) Overall Agreement: 291 / 292 = 99% (98 % -100%*)
- 95% Confidence Intervals
ACON BAR One Step Barbiturates Test Device versus the MCC Instant Screen Card BAR Test:
Positive Agreement: 126 / 126 = 100% (97% -100%) Negative Agreement: 165 / 166 = 99% (97% -100%) Overall Agreement: 291 / 292 = 99% (98 % -100%*)
- 95% Confidence Intervals
ACON BAR One Step Barbiturates Test Strip versus GC/MS at the secobarbital cut-off concentration of 300 ng/ml:
Positive agreement with GC/MS: 122 / 132 = 92% (86% -96%) Negative agreement with GC/MS: 156 / 160 = 98% (94% -99%) Total agreement with GC/MS: 278 / 292 = 95% (92% -97%*)
- 95% confidence intervals
{3}------------------------------------------------
ACON BAR One-Step Barbiturates Test Device versus GC/MS at the secobarbital cut-off concentration of 300 ng/ml:
Positive agreement with GC/MS: 122 / 132 = 92% (86% -96%) Negative agreement with GC/MS: 156 / 160 = 98% (94% -99%) Total agreement with GC/MS: 278 / 292 = 95% (92% -97%*)
- 95% confidence intervals
Conclusion:
These clinical studies demonstrate the substantial equivalency between the ACON BAR One Step Barbiturates Test Strip, ACON BAR One Step Barbiturates Test Device and the MCC Instant Screen Card BAR Test, which has already being marketed in the United States. It is also demonstrated in the performance studies that the ACON BAR One Step Barbiturates Test Strip, ACON BAR One Step Barbiturates Test Device are safe and effective in qualitatively detecting barbiturates in urine specimens at a secobarbital cut-off concentration of 300 ng/mL. The POL study demonstrated that these tests are suitable for use by healthcare professionals including professionals at point-of-care sites.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
NOV 0 9 2001
Edward Tung. Ph.D. Director of Regulatory Affair ACON Laboratories, Inc. 4108 Sorrento Valley BLvd. San Diego, CA 92121
Re: K012824
Trade/Device Name: ACON® BAR One Step Barbiturates Test Strip and ACON® BAR One Step Barbiturates Test Strip Device Regulation Number: 21 CFR 862.3150 Regulation Name: Barbiturate test system Regulatory Class: Class II Product Code: DIS Dated: August 21, 2001 Received: August 23, 2001
Dear Dr. Tung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 1 additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrb/dsma/dsmamain.html".
Sincerely yours,
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory-Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
NOV 0 9 2001
INDICATIONS FOR USE 10.
510(k) Number: K012824
Device Name: ACON® BAR One Step Barbiturates Test Strip
ACON® BAR One Step Barbiturates Test Device
The ACON BAR One Step Barbiturates Test Strip and ACON BAR Indications for Use: One Step Barbiturates Test Device are rapid chromatographic immunoassays for the qualitative detection of barbiturates in human urine at a level relative to the secobarbital cut-off concentration of 300 ng/mL. They are intended for use by healthcare professionals including professionals at point of care sites.
Rate a charter for Jean Cooper
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number_**_ and the list of the 510(k) number.
(Please do not write below this point)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
Or
Over-The-Counter Use
§ 862.3150 Barbiturate test system.
(a)
Identification. A barbiturate test system is a device intended to measure barbiturates, a class of hypnotic and sedative drugs, in serum, urine, and gastric contents. Measurements obtained by this device are used in the diagnosis and treatment of barbiturate use or overdose and in monitoring levels of barbiturate to ensure appropriate therapy.(b)
Classification. Class II (special controls). A barbiturate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).