K Number
K012726
Device Name
PROSAFE SAFETY SYRINGE
Manufacturer
Date Cleared
2001-10-18

(64 days)

Product Code
Regulation Number
880.5860
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
A properly placed syringe provides access to a vein or tissue for fluid sampling or infusion. The safety syringe is designed for single use and has a needlestick protection feature. The risk of accidental needlesticks is reduced by a needle retraction system which is activated after the infusion procedure is completed.
Device Description
The SafePro* Safety Syringe consists of a syringe assembly and a needle assembly. The device has a built-in safety feature to reduce the risk of accidental needlestick injuries.
More Information

B-D Luer Lok syringe, Retractable Technology, Inc. VanishPoint syringe

Not Found

No
The description focuses on a mechanical safety feature (needle retraction) and does not mention any computational or data-driven components indicative of AI/ML.

No
A therapeutic device is used to treat or alleviate a medical condition. This device is a safety syringe designed for "fluid sampling or infusion," which are diagnostic or procedural uses, and its primary function is to prevent needlestick injuries, not to treat a condition itself.

No
The device is described as providing access for "fluid sampling or infusion" and is designed to deliver or withdraw substances, not to diagnose a condition. Its primary function is in fluid management, with a safety feature for needlestick prevention, not diagnostic analysis.

No

The device description clearly outlines a physical syringe assembly and needle assembly with a built-in safety feature, indicating it is a hardware device.

Based on the provided information, this device is not an IVD (In Vitro Diagnostic).

Here's why:

  • Intended Use: The intended use clearly states the device is for "fluid sampling or infusion" by accessing a vein or tissue. This describes a procedure performed on the patient, not a test performed on a sample taken from the patient.
  • Device Description: The description focuses on the physical components of a syringe and its safety features for preventing needlestick injuries during the procedure.
  • Lack of IVD Characteristics: There is no mention of:
    • Analyzing a sample (blood, urine, etc.)
    • Detecting or measuring substances in a sample
    • Providing diagnostic information based on sample analysis
    • Reagents or test kits

The device is a medical device used for administering or collecting fluids directly from a patient, which falls under a different regulatory category than IVDs.

N/A

Intended Use / Indications for Use

The SafePro* Safety Syringe is designed to gain access to a patient's vascular or intramuscular system for short term fluid sampling or infusion. The device is designed for single use and has a needlestick protection feature.

A properly placed syringe provides access to a vein or tissue for fluid sampling or infusion. The safety syringe is designed for single use and has a needlestick protection feature. The risk of accidental needlesticks is reduced by a needle retraction system which is activated after the infusion procedure is completed.

Product codes (comma separated list FDA assigned to the subject device)

MEG

Device Description

The SafePro* Safety Syringe consists of a syringe assembly and a needle assembly. The device has a built-in safety feature to reduce the risk of accidental needlestick injuries.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

patient's vascular or intramuscular system

Indicated Patient Age Range

Not Found

Intended User / Care Setting

Not Found

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

A total of 500 SafePro* Safety Syringes were evaluated by 50 participants.

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

Summary of Material Testing: The SafePro* Safety Syringe was tested for material safety and biocompatibility. Test results indicated that the SafePro* Safety Syringe meets ISO 10993-1 and US FDA G-95 requirements.

Summary of Simulated Use Study: A total of 500 SafePro* Safety Syringes were evaluated by 50 participants. No sharps injuries or failure of the safety mechanism occurred. Successful completion of the study supports the claim that SafePro* Safety Syringe can reduce the risk of accidental needlestick injuries. The positive responses from the Evaluators regarding functional and performance aspects also indicated that the SafePro* Safety Syringe meets customer requirements.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

Not Found

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

B-D Luer Lok syringe, Retractable Technology, Inc. VanishPoint syringe

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

§ 880.5860 Piston syringe.

(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).

0

Ko12726

OCT 1 8 2001

510(K) - SafePro* Safety Syringe

SECTION VIII

510(K) SUMMARY OF SAFETY AND EFFECTIVENESS for SafePro* Safety Syringe

1. REGULATORY AUTHORITY

Safe Medical Device Act of 1990, CFR 807.92.

2. CONTACT PERSON

Joseph J. Chang, Ph.D., PE HOU-SE MEDICAL, INC. U.S. Liaison Office P. O. Box 630374 Irving, TX 75063-0374

NAME OF MEDICAL DEVICE 3.

Classification Name: Syringe, Antistick Classification Code: 177 MEG Common/Usual Name: Syringe Proprietary Name: SafePro* Safety Syringe

DEVICE CLASSIFICATION 4.

The General Hospital Panel has classified Antistick Syringes (21CFR880.5600) into Class II, Special Controls under section 513 of the Act.

ડ. STATEMENT OF SUBSTANTIAL EQUIVALENCE

The SafePro* Safety Syringe is substantially equivalent to B-D Luer Lok syringe in terms of normal syringe function for fluid transport, and Retractable Technology, Inc. VanishPoint syringe for needlestick protection feature. See also the attached table.

1

510(K) - SafePro* Safety Syringe

6. INTENDED USE

/

The SafePro* Safety Syringe is designed to gain access to a patient's vascular or intramuscular system for short term fluid sampling or infusion. The device is designed for single use and has a needlestick protection feature.

DESCRIPTION OF DEVICE 7.

The SafePro* Safety Syringe consists of a syringe assembly and a needle assembly. The device has a built-in safety feature to reduce the risk of accidental needlestick injuries.

8. SUMMARY OF MATERIAL TESTING

The SafePro* Safety Syringe was tested for material safety and biocompatibility. Test results indicated that the SafePro* Safety Syringe meets ISO 10993-1 and US FDA G-95 requirements.

9. SUMMARY OF SIMULATED USE STUDY

A total of 500 SafePro* Safety Syringes were evaluated by 50 participants. No sharps injuries or failure of the safety mechanism occurred. Successful completion of the study supports the claim that SafePro* Safety Syringe can reduce the risk of accidental needlestick injuries. The positive responses from the Evaluators regarding functional and performance aspects also indicated that the SafePro* Safety Syringe meets customer requirements.

CONCLUSION 10.

The results obtained from bench testing, material safety, and simulated use tests indicate that the SafePro* Safety Syringe is safe and effective for it's intended use.

2

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles three overlapping human profiles or abstract shapes. The logo is in black and white.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Joseph J. Chang, PHD, P.E. Chief Technology Officer Hou-se Medical, Incorporated

P.O. Box 630374 Irving, Texas 75063-0374

Re: K012726

Trade/Device Name: SafePro™ Safety Syringe Regulation Number: 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: MEG Dated: August 10, 2001 Received: August 15, 2001

Dear Dr. Chang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

OCT 1 8 2001

3

Page 2 - Dr. Chang

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 21 CFR Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

Timothy Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

4

INDICATIONS FOR USE

Applicant

HOU-SE MEDICAL, INC. Address in Taiwan Headquarter Office: 295, 7/F, Sec. 2, Ho-Ping E. Rd., Taipei, Taiwan, R. O. C. Zip Code: 106

Address in U.S. Liaison Office: P.O. Box 630374 Irving, TX 75063-0374

510(K) Number K Ol 2726

This is a new submission and FDA has not assigned a number at this time.

Device Name

SafePro* Safety Syringe

Indication for Use

A properly placed syringe provides access to a vein or tissue for fluid sampling or infusion. The safety syringe is designed for single use and has a needlestick protection feature. The risk of accidental needlesticks is reduced by a needle retraction system which is activated after the infusion procedure is completed.

(Please don not write below this line-Continue on another page if needed) Concurrence of CDRH Office of Device Evaluation (ODE)

Prescription Use Per 21 CFR 801.109 (Optional Format 1-2-96)

OR Over-the-Counter __

Sene Naveau for lXC

(Division Sign-Off) Division of Dental, Infection Control, and General Hospital D 510(k) Number

*Trademark