K Number
K012460
Manufacturer
Date Cleared
2001-08-14

(13 days)

Product Code
Regulation Number
870.3680
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Model 6495 Bipolar Temporary Myocardial Pacing Lead is designed for temporary atrial and ventricular pacing and sensing for a contemplated implant duration of 7 days or less. The device is supplied sterile and intended for single use only.

Device Description

The Model 6495 Bipolar Temporary Myocardial Pacing Lead consists of an insulated multi-filament lead, which contains a distal, discrete, ring electrode, a discrete, tip electrode, and a coaxial conductor lead body. Each discrete electrode is crimped onto a conductor and terminates in an atraumatic myocardial curved needle. A blue monofilament coil provides fixation while the lead is implanted in myocardial tissue. An atraumatic chest needle at the proximal end of the conductor wire permits exiting the pacing lead through the chest wall. Terminated on the back of the chest needle are two breakaway connector pins. To remove the pacing lead, gentle traction should be applied. No part of the lead remains in the body.

AI/ML Overview

This document does not contain acceptance criteria or a study proving the device meets acceptance criteria in the manner requested. The submission is a Special 510(k) Premarket Notification for a modified medical device. This type of submission focuses on demonstrating substantial equivalence to a previously cleared predicate device, rather than proving performance against specific acceptance criteria through a clinical study.

Here's why the requested information cannot be fully extracted and what the document does provide:

  • No specific acceptance criteria table for device performance: The document does not present a table of quantitative acceptance criteria (e.g., "sensitivity > X%", "accuracy > Y%") alongside reported performance metrics for the device.
  • No clinical study demonstrating device performance: The submission outlines "system compatibility testing" which includes environmental conditioning, visual verification, dimensional testing, electrical testing, and mechanical testing. These are bench tests, not clinical studies involving patients or human readers.
  • Comparison to a predicate device: The core of a 510(k) submission is to show that the new device is "substantially equivalent" to a legally marketed predicate device, implying that its safety and effectiveness are comparable.

Here's the information that can be extracted or inferred from the provided text, structured to address your questions where possible:


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategoryReported Device Performance
Environmental Conditioning"meets the specified requirements"
Visual Verification"meets the specified requirements"
Dimensional Testing"meets the specified requirements"
Electrical Testing"meets the specified requirements"
Mechanical Testing"meets the specified requirements"
Sterilization Validation"Processes appropriate for sterilizing the devices were validated."

Note: The document only states that the device "meets the specified requirements" for each test. It does not provide the specific quantitative requirements or the exact performance values achieved.

2. Sample size used for the test set and the data provenance

  • Sample Size: Not explicitly stated. The "Summary of Studies" refers to "system compatibility testing" which typically involves a defined number of units, but this number is not provided in the document.
  • Data Provenance: Not explicitly stated, but these are bench tests and likely conducted internally by Medtronic, Inc. There is no indication of country of origin of data or whether it was retrospective or prospective in the context of human data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not applicable. This document describes engineering and product safety testing (e.g., electrical, mechanical, dimensional, sterilization), not a study requiring expert clinical ground truth assessment.

4. Adjudication method for the test set

  • Not applicable. As the tests are technical and functional, not clinical assessments requiring human interpretation or consensus.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No. This is not an AI/software device or an imaging device, and therefore, an MRMC study is not relevant to this submission.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

  • Not applicable. This is a physical medical device (pacing lead), not an algorithm or AI.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Not applicable in the conventional sense. The "ground truth" for the tests performed (e.g., electrical testing, dimensional testing) would be the established engineering specifications and industry standards for safety and performance of such a device.

8. The sample size for the training set

  • Not applicable. There is no "training set" as this is not an AI/machine learning device. The device itself is manufactured, and its performance is verified against specifications.

9. How the ground truth for the training set was established

  • Not applicable. See point 8.

In summary: This 510(k) submission for the Medtronic Model 6495 Bipolar Temporary Myocardial Pacing Lead focuses on establishing substantial equivalence to an existing predicate device through a series of bench and validation tests (environmental, visual, dimensional, electrical, mechanical, and sterilization). It does not involve a clinical study with human subjects, AI performance, or expert evaluation of clinical cases to determine acceptance criteria for diagnostic or prognostic accuracy. The acceptance criteria for the tests performed are implicitly meeting "specified requirements," which would be Medtronic's internal engineering and quality standards for the device.

{0}------------------------------------------------

K012460 p. 1/3

MEDTRONIC CONFIDENTIAL

Special 510(k) Premarket Notification Model 6495 Bipolar Temporary Myocardial Pacing Lead Attachment D: 510(k) Summary of Substantial Equivalence

AUG 1 4 2001

ATTACHMENT D 510(K) SUMMARY OF SUBSTANTIAL EQUIVALENCE

Submitter

Medtronic, Inc. 7000 Central Avenue N.E. Minneapolis, MN 55432

Contact: Tina Benoit, Associate Product Regulation Manager Telephone: (763) 514-4112 Fax: (763) 514-6424 E-Mail: tina.benoit@medtronic.com

Date Prepared: July 31, 2001

Name of Device

Trade Name: Temporary Pacemaker Electrode, 74 LDF Common Name: Temporary Pacing Lead Classification: Class II

{1}------------------------------------------------

K012460 p. 2/3

MEDTRONIC CONFIDENTIAL

Special 510(k) Premarket Notification Model 6495 Bipolar Temporary Myocardial Pacing Lead Model 6495 Dipolar \ Summary of Substantial Equivalence

Predicate Devices

The predicate device for the Model 6495 Bipolar Temporary Myocardial Pacing The production and the currently market released Model 6495 Bipolar Temporary Myocardial Pacing Lead.

Device Description

The Model 6495 Bipolar Temporary Myocardial Pacing Lead consists of an insulated multi-filament lead, which contains a distal, discrete, ring electrode, a discrete, tip electrode, and a coaxial conductor lead body. Each discrete electrode is crimped onto a conductor and terminates in an atraumatic myocardial curved needle. A blue monofilament coil provides fixation while the lead is implanted in myocardial tissue. An atraumatic chest needle at the proximal end of the conductor wire permits exiting the pacing lead through the chest wall. Terminated on the back of the chest needle are two breakaway connector pins. To remove the pacing lead, gentle traction should be applied. No part of the lead remains in the body.

Packaging

The sterile packaging for the Model 6495 Bipolar Temporary Myocardial Pacing Lead consists of a double pouch configuration. The inner pouch (or package liner) and outer pouch materials are transparent Tyvek - polyester/polyethylene laminate. The pouches are heat-sealed.

Intended Use

The Model 6495 Bipolar Temporary Myocardial Pacing Lead is designed for temporary atrial and ventricular pacing and sensing for a contemplated implant duration of 7 day or less. The device is supplied sterile and intended for single use only.

Technological Characteristics

The technology used with the Model 6495 Bipolar Temporary Myocardial Pacing Lead is the same technological characteristics as the predicate device.

{2}------------------------------------------------

K012460 p.3/3

MEDTRONIC CONFIDENTIAL

Special 510(k) Premarket Notification Model 6495 Bipolar Temporary Myocardial Pacing Lead Model 0435 Dipolar > Substantial Equivalence

Summary of Studies

Medtronic, Inc. performed system compatibility testing to support that the Model Mcdirolie, the perfective syocardial Pacing lead is equivalent to the predicate device. Device testing included:

  • Environmental Conditioning .
  • Visual Verification .
  • Dimensional Testing .
  • Electrical Testing .
  • Mechanical Testing .

All system compatibility tests performed have demonstrated that the modified Model 6495 heartwire meets the specified requirements.

Sterilization Validation

The Model 6495 Bipolar Temporary Myocardial Pacing Lead is sterilized using a The Model 6495 Dipolar Tomporilization process. Processes appropriate for sterilizing the devices were validated.

Conclusion

Through data and information presented, numerous similarities support a I hrough data and information pivalence and show the device modifications do not determination of bactuation equive or alter the fundamental scientific technology of the device. Market clearance of the Model 6495 Bipolar Temporary Myocardial Pacing Lead is supported through this Special 510(k) Premarket Notification.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/3/Picture/2 description: The image is a logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an image of an eagle with its wings spread.

AUG 1 4 2001

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Tina L. Benoit Medtronic, Inc. Cardiac Rhythm Management 7000 Central Avenue NE Minneapolis, MN 55432-3576

Re: K012460

Trade Name: Model 6495 Bipolar Temporary Myocardial Pacing Lead Regulation Number: 21 CFR 870.3680 Regulatory Class: Class II (two) Product Code: 74 LDF Dated: July 31, 2001 Received: August 1, 2001

Dear Ms. Benoit:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this

{4}------------------------------------------------

Page 2 - Ms. Tina L. Benoit

response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed nredicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4646. Additionally, for questions on the promotion and advertising of your device, (1) be se contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Oakteh
James E. Dillard III

Director Division of Cardiovascular and Respiratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Special 510(k) Premarket Notification Model 6495 Bipolar Temporary Myocardial Pacing Lead Indications for Use

INDICATIONS FOR USE

NA KO 12460

510(k) Number (if known):

Device Name:

Medtronic® Model 6495 Bipolar Temporary Myocardial Pacing Lead

Indications For Use:

The Model 6495 Bipolar Temporary Myocardial Pacing Lead is designed for temporary atrial and ventricular pacing and sensing for a contemplated implant duration of 7 days or less. The device is supplied sterile and intended for single use only.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)

OR

Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________

Division of Cardiovascular & Respiratory Devices
510(k) Number K012460

(Optional Format 1-2-96)

§ 870.3680 Cardiovascular permanent or temporary pacemaker electrode.

(a)
Temporary pacemaker electrode —(1)Identification. A temporary pacemaker electrode is a device consisting of flexible insulated electrical conductors with one end connected to anexternal pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator.(2)
Classification. Class II (performance standards).(b)
Permanent pacemaker electrode —(1)Identification. A permanent pacemaker electrode is a device consisting of flexible insulated electrical conductors with one end connected to an implantable pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator.(2)
Classification. Class III (premarket approval).(c)
Date PMA or notice of completion of PDP is required. A PMA or notice of completion of a PDP is required to be filed with the Food and Drug Administration on or before October 4, 2012, for any permanent pacemaker electrode device that was in commercial distribution before May 28, 1976, or that has, on or before October 4, 2012, been found to be substantially equivalent to any permanent pacemaker electrode device that was in commercial distribution before May 28, 1976. Any other pacemaker repair or replacement material device shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.