(133 days)
The Super C™ Hemodialysis/Apheresis Chronic Catheter is indicated for use in attaining short and long term vascular access for hemodialysis. hemoperfusion or apheresis therapy via the jugular or subclavian vein. The catheter is intended for implantation dwell time of greater than 30 days.
The Super C™ Hemodialysis/Apheresis Chronic Catheter is composed of a radiopague polyurethane catheter shaft with two lumens (arterial and venous) in a Crescent (Circle "C") configuration. The lumens are distinguishable by color-coded clamps on clear silicone extensions. The fixed retention cuff on the shaft provides an anchoring site for tissue ingrowth during long-term placement. The Super C™ Hemodialysis/Apheresis Chronic Catheter offered in straight versions are available in 19 cm, 23 cm, and 28 cm cuff to tip insertion lengths. Additionally, the Super C™ Hemodialysis/Apheresis Chronic Catheter is available with a J-Cannula version in 19 cm and 23 cm cuff to tip insertion lengths.
The provided text describes a 510(k) premarket notification for the Super C™ Hemodialysis/Apheresis Chronic Catheter. This type of submission is a regulatory pathway for medical devices that are substantially equivalent to legally marketed predicate devices, not a study designed to prove the device meets specific performance criteria through clinical trials or detailed algorithm performance evaluation.
Therefore, many of the requested categories for device acceptance criteria and studies are not applicable to this submission as it focuses on demonstrating substantial equivalence through comparison of technological characteristics and performance testing data rather than a detailed clinical study with established ground truth, expert adjudication, or MRMC studies.
Here's a breakdown of what can be extracted from the provided text, and where information is not available:
1. Table of Acceptance Criteria and Reported Device Performance
The text does not specify quantitative acceptance criteria. Instead, it states that "Performance testing was conducted data for Super C™ Hemodialysis/Apheresis Chronic Catheter and compared to the predicate devices identified in this 510(k). Test results demonstrated that the Super C™ Hemodialysis/Apheresis Chronic Catheter is substantially equivalent to these predicate devices commercially in distribution for the same intended use."
This implies the acceptance criterion was "substantial equivalence" to the specified predicate devices, meaning its performance characteristics (material type, intended use, operating principle, number of lumens) were similar enough and its performance test results did not raise new questions of safety or effectiveness.
| Acceptance Criterion | Reported Device Performance |
|---|---|
| Substantial Equivalence to Predicate Devices | Performance testing conducted; "Test results demonstrated that the Super C™ Hemodialysis/Apheresis Chronic Catheter is substantially equivalent to these predicate devices" |
| Material Type | Radiopaque polyurethane with silicone extensions, considered similar to predicate devices |
| Intended Use | Short and long-term vascular access for hemodialysis, hemoperfusion, or apheresis therapy; similar to predicate devices |
| Operating Principle | Dual lumens (arterial and venous) in a Crescent configuration; similar to predicate devices |
| Number of Lumens | Two lumens; similar to predicate devices |
| Long-term placement | Fixed retention cuff for tissue ingrowth; similar to predicate devices |
| Insertion Lengths | Available in 19 cm, 23 cm, and 28 cm cuff to tip (straight versions), and 19 cm and 23 cm (J-Cannula version); similar to predicate devices |
2. Sample size used for the test set and the data provenance
- Sample Size: Not specified. The document states "Performance testing was conducted data," indicating in-vitro or bench testing, but does not specify the number of units or tests performed.
- Data Provenance: The nature of 510(k) submissions generally means this would be internal company testing (Horizon Medical Products, Inc.). The data would be prospective in the sense that the company specifically conducted tests for this submission, but it's not a clinical study involving patients or patient data.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not Applicable. This device is a physical medical device (catheter), not one that generates diagnostic output requiring expert interpretation for ground truth. The "ground truth" for its performance would be established by direct physical measurements, material properties testing, and functional assessments, not expert consensus on interpretations.
4. Adjudication method for the test set
- Not Applicable. As no expert interpretation or diagnostic output is involved, there is no need for an adjudication method.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable. This is a physical medical device (catheter), not an AI-powered diagnostic tool. Therefore, MRMC studies and assessment of human reader improvement with AI assistance are irrelevant.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not Applicable. This is a physical medical device, not an algorithm or software.
7. The type of ground truth used
- Not Applicable in the sense of a diagnostic outcome. The "ground truth" for a physical device like this is its physical and functional performance, verified through engineering tests, material science analysis, and potentially some animal or cadaver studies (though not explicitly mentioned here, implied by performance testing). For example, tensile strength would be measured directly, or flow rates assessed under simulated conditions.
8. The sample size for the training set
- Not Applicable. This is a physical medical device, not a machine learning algorithm that requires a training set.
9. How the ground truth for the training set was established
- Not Applicable. (See point 8)
{0}------------------------------------------------
K011716 Page / o
OCT 31 2001
Proprietary to Horizon Medical Products, Inc.
510(k) Summary of Safety and Effectiveness Super C™ Hemodialysis/Apheresis Chronic Catheter
| Date SummaryWas Prepared: | June 15, 2001 |
|---|---|
| Submitter'sInformation: | Horizon Medical Products, Inc.One Horizon WayManchester, Georgia 31816 |
| Telephone Number: | 706-846-3126 |
| Fax Number: | 706-846-3180 |
| Contact Person: | Penny Northcutt, Director of Regulatory Affairs |
| Device Trade Name: | Super C™ Hemodialysis/Apheresis Chronic Catheter |
| Device Common Name: | Long Term Implanted Hemodialysis Catheter |
| Classification Name: | Catheter, Hemodialysis, Implanted |
| Classification Panel: | Gastroenterology |
Legally Marketed Devices To Which Substantial Equivalence Is Claimed To: The Super C™ Hemodialysis/Apheresis Chronic Catheter is substantially equivalent to
- . Horizon Medical Products'13.5 French Chronic Dual Lumen Dialysis Catheter (Akcess Cath)
- Bard Vas Cath Opti-Flow 14.5 French Dual Lumen Catheter ●
- MedComp Ash Split 14 French, 28 cm Chronic Double Lumen Catheter. .
{1}------------------------------------------------
Koi/1916 Page 2
Proprietary to Horizon Medical Products, Inc.
Device Description: The Super C™ Hemodialysis/Apheresis Chronic Catheter is composed of a radiopague polyurethane catheter shaft with two lumens (arterial and venous) in a Crescent (Circle "C") configuration. The lumens are distinguishable by color-coded clamps on clear silicone extensions.
The fixed retention cuff on the shaft provides an anchoring site for tissue ingrowth during long-term placement.
The Super C™ Hemodialysis/Apheresis Chronic Catheter offered in straight versions are available in 19 cm, 23 cm, and 28 cm cuff to tip insertion lengths. Additionally, the Super C™ Hemodialysis/Apheresis Chronic Catheter is available with a J-Cannula version in 19 cm and 23 cm cuff to tip insertion lengths.
Intended Use: The Super C™ Hemodialysis/Apheresis Chronic Catheter is indicated for use in attaining short and long term vascular access for hemodialysis. hemoperfusion or apheresis therapy via the jugular or subclavian vein. The catheter is intended for implantation dwell time of greater than 30 days.
Technological Characteristics: The technological characteristics of the Super C™ Hemodialysis/Apheresis Chronic Catheter, including material type, intended use, operating principle, number of lumens and performance characteristics are similar to the Bard's Vas-Cath Opti-Flow Dual Lumen Catheter and MedComp's Ash Split Catheter.
Performance Data: Performance testing was conducted data for Super C™ Hemodialysis/Apheresis Chronic Catheter and compared to the predicate devices identified in this 510(k). Test results demonstrated that the Super C™ Hemodialysis/Apheresis Chronic Catheter is substantially equivalent to these predicate devices commercially in distribution for the same intended use.
Horizon Medical Products, Inc. Traditional 510(k) for Super C™ Hemodialysis/Apheresis Chronic Catheter Predicate 510(k)s: Akcess Cath Permanent Silicone Catheter - Horizon Medical Products, Inc. Chronic Dual Lumen Catheter (K925819); Vascath Opti-Flow (K981994 & K010567) and Ash Split Cath (K972207) June 18, 2001
0297
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle, indicating the department's name and national affiliation.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 31 2001
Ms. Patricia D. Jones Regulatory Affairs Specialist Horizon Medical Products, Inc. One Horizon Way MANCHESTER GA 31816
Re: K011916
Trade/Device Name: Super C™ Hemodialysis/Apheresis Chronic Catheter Regulation Number: 21 CFR §876.5540 Regulation Name: Blood access device and accessories Regulatory Class: III Product Code: 78 MSD Dated: September 21, 2001 Received: September 24, 2001
Dear Ms. Jones:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Patricia Jones
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration (21 CFR Part 807); listing (21 CFR Part 807), labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR §1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on the labeling regulation (21 CFR Part 801 and additionally Part 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4616. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
PREMARKET NOTIFICATION
INDICATION FOR USE STATEMENT
-Unassigned K01 1911/ 510(k) Number:
Super C™ Hemodialysis/Apheresis Chronic Catheter Device Name:
Indication for Use:
The Super C™ Hemodialysis/Apheresis Chronic Cathere is indicated for use The Super C ™ Hemodialysis/Apricrosis Shremotialysis, hemoperfusion in attailing short and long term vasonan vein. The catheter is intended for aprovels and well time of greater than 30 days.
| Concurrence of CDRH, Office of Device Evaluation (ODE) |
|---|
| -------------------------------------------------------- |
| Prescription Use(Per 21 CFR 801.109) | ✓ |
|---|---|
| ------------------------------------------ | --- |
or
| Over-the-Counter Use | |
|---|---|
| ---------------------- | -- |
Nancy C. Brogdon
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K011916
§ 876.5540 Blood access device and accessories.
(a)
Identification. A blood access device and accessories is a device intended to provide access to a patient's blood for hemodialysis or other chronic uses. When used in hemodialysis, it is part of an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and provides access to a patient's blood for hemodialysis. The device includes implanted blood access devices, nonimplanted blood access devices, and accessories for both the implanted and nonimplanted blood access devices.(1) The implanted blood access device is a prescription device and consists of various flexible or rigid tubes, such as catheters, or cannulae, which are surgically implanted in appropriate blood vessels, may come through the skin, and are intended to remain in the body for 30 days or more. This generic type of device includes various catheters, shunts, and connectors specifically designed to provide access to blood. Examples include single and double lumen catheters with cuff(s), fully subcutaneous port-catheter systems, and A-V shunt cannulae (with vessel tips). The implanted blood access device may also contain coatings or additives which may provide additional functionality to the device.
(2) The nonimplanted blood access device consists of various flexible or rigid tubes, such as catheters, cannulae or hollow needles, which are inserted into appropriate blood vessels or a vascular graft prosthesis (§§ 870.3450 and 870.3460), and are intended to remain in the body for less than 30 days. This generic type of device includes fistula needles, the single needle dialysis set (coaxial flow needle), and the single needle dialysis set (alternating flow needle).
(3) Accessories common to either type include the shunt adaptor, cannula clamp, shunt connector, shunt stabilizer, vessel dilator, disconnect forceps, shunt guard, crimp plier, tube plier, crimp ring, joint ring, fistula adaptor, and declotting tray (including contents).
(b)
Classification. (1) Class II (special controls) for the implanted blood access device. The special controls for this device are:(i) Components of the device that come into human contact must be demonstrated to be biocompatible. Material names and specific designation numbers must be provided.
(ii) Performance data must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(A) Pressure versus flow rates for both arterial and venous lumens, from the minimum flow rate to the maximum flow rate in 100 milliliter per minute increments, must be established. The fluid and its viscosity used during testing must be stated.
(B) Recirculation rates for both forward and reverse flow configurations must be established, along with the protocol used to perform the assay, which must be provided.
(C) Priming volumes must be established.
(D) Tensile testing of joints and materials must be conducted. The minimum acceptance criteria must be adequate for its intended use.
(E) Air leakage testing and liquid leakage testing must be conducted.
(F) Testing of the repeated clamping of the extensions of the catheter that simulates use over the life of the device must be conducted, and retested for leakage.
(G) Mechanical hemolysis testing must be conducted for new or altered device designs that affect the blood flow pattern.
(H) Chemical tolerance of the device to repeated exposure to commonly used disinfection agents must be established.
(iii) Performance data must demonstrate the sterility of the device.
(iv) Performance data must support the shelf life of the device for continued sterility, package integrity, and functionality over the requested shelf life that must include tensile, repeated clamping, and leakage testing.
(v) Labeling of implanted blood access devices for hemodialysis must include the following:
(A) Labeling must provide arterial and venous pressure versus flow rates, either in tabular or graphical format. The fluid and its viscosity used during testing must be stated.
(B) Labeling must specify the forward and reverse recirculation rates.
(C) Labeling must provide the arterial and venous priming volumes.
(D) Labeling must specify an expiration date.
(E) Labeling must identify any disinfecting agents that cannot be used to clean any components of the device.
(F) Any contraindicated disinfecting agents due to material incompatibility must be identified by printing a warning on the catheter. Alternatively, contraindicated disinfecting agents must be identified by a label affixed to the patient's medical record and with written instructions provided directly to the patient.
(G) Labeling must include a patient implant card.
(H) The labeling must contain comprehensive instructions for the following:
(
1 ) Preparation and insertion of the device, including recommended site of insertion, method of insertion, and a reference on the proper location for tip placement;(
2 ) Proper care and maintenance of the device and device exit site;(
3 ) Removal of the device;(
4 ) Anticoagulation;(
5 ) Management of obstruction and thrombus formation; and(
6 ) Qualifications for clinical providers performing the insertion, maintenance, and removal of the devices.(vi) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices that include subcutaneous ports must include the following:
(A) Labeling must include the recommended type of needle for access as well as detailed instructions for care and maintenance of the port, subcutaneous pocket, and skin overlying the port.
(B) Performance testing must include results on repeated use of the ports that simulates use over the intended life of the device.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(vii) In addition to Special Controls in paragraphs (b)(1)(i) through (v) of this section, implanted blood access devices with coatings or additives must include the following:
(A) A description and material characterization of the coating or additive material, the purpose of the coating or additive, duration of effectiveness, and how and where the coating is applied.
(B) An identification in the labeling of any coatings or additives and a summary of the results of performance testing for any coating or material with special characteristics, such as decreased thrombus formation or antimicrobial properties.
(C) A Warning Statement in the labeling for potential allergic reactions including anaphylaxis if the coating or additive contains known allergens.
(D) Performance data must demonstrate efficacy of the coating or additive and the duration of effectiveness.
(viii) The following must be included for A-V shunt cannulae (with vessel tips):
(A) The device must comply with Special Controls in paragraphs (b)(1)(i) through (v) of this section with the exception of paragraphs (b)(1)(ii)(B), (b)(1)(ii)(C), (b)(1)(v)(B), and (b)(1)(v)(C), which do not apply.
(B) Labeling must include Warning Statements to address the potential for vascular access steal syndrome, arterial stenosis, arterial thrombosis, and hemorrhage including exsanguination given that the device accesses the arterial circulation.
(C) Clinical performance testing must demonstrate safe and effective use and capture any adverse events observed during clinical use.
(2) Class II (performance standards) for the nonimplanted blood access device.
(3) Class II (performance standards) for accessories for both the implanted and the nonimplanted blood access devices not listed in paragraph (b)(4) of this section.
(4) Class I for the cannula clamp, disconnect forceps, crimp plier, tube plier, crimp ring, and joint ring, accessories for both the implanted and nonimplanted blood access device. The devices subject to this paragraph (b)(4) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.