K Number
K011882
Date Cleared
2001-07-10

(22 days)

Product Code
Regulation Number
880.6250
Panel
HO
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

A patient examination glove is a disposable device intended for medical purposes that is worn on the hand of healthcare and similar personnel to prevent contamination between healthcare personnel and the patient's body, fluids, waste or environment.

Device Description

Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 880.6250, Powder-Free Vinyl Patient Examination Glove, 80L YZ, and meets all requirement of ASTM Standard D5250-00.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study information based on the provided text, using the requested format.


This document describes the acceptance criteria and studies conducted for the Shijiazhuang Eversharp Plastics Products Co., Ltd. Synthetic Powder Free (Yellow) Vinyl Patient Examination Gloves.

1. Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Physical and Dimensions Testing: Meets ASTM-D-5250-00 requirements, Inspection Level S-2, AQL 4.0.The device meets requirements for Physical and Dimensions Testing conducted on gloves, Inspection Level S-2, AQL 4.0, based on ASTM-D-5250-00.
Pinhole Test (FDA 1000 ml Water Fill Test): Meets requirements for AQL 2.5, Inspection Level I.The FDA 1000 ml Water Fill Test was conducted with samplings of AQL 2.5, Inspection Level I, and the device met these requirements.
Biocompatibility (Primary Skin Irritation and Skin Sensitization): No primary skin irritant or sensitization reactions.Primary Skin irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted, with results showing no primary skin irritant or sensitization reactions.
Residual Powder Test (for "powder-free" claim): Contains no more than 2 mg powder per glove (based on ASTM D6124-97).A Residual Powder Test based on ASTM D6124-97 for Starch at finished inspection was conducted, ensuring the gloves meet the "powder-free" claim (containing no more than 2 mg powder per glove).
General Conformance: Conforms fully to ASTM-D-5250-00 standard and applicable 21 CFR references.The device conforms fully to ASTM-D-5250-00 standard as well as applicable 21 CFR references.
Labeling Claims Conformance: Adherence to labeling claims (no special claims, not hypoallergenic).There are no special labeling claims, and the gloves are not claimed as hypoallergenic. The device meets "labeling claims as shown by data in Section 7."

2. Sample size used for the test set and the data provenance

  • Sample Size for Test Set:
    • Physical and Dimensions Testing: The sample size is implicitly defined by "Inspection Level S-2, AQL 4.0" as per ASTM-D-5250-00, but a specific number is not provided.
    • Pinhole Test: The sample size is implicitly defined by "AQL 2.5, Inspection Level I" as per the FDA 1000 ml Water Fill Test, but a specific number is not provided.
    • Biocompatibility (Skin Irritation/Sensitization): Not specified.
    • Residual Powder Test: Not specified.
  • Data Provenance: Not explicitly stated as "country of origin" or "retrospective/prospective." However, the tests were conducted by the manufacturer, Shijiazhuang Eversharp Plastics Products Co., Ltd., based in Luquan, Hebei Province, P.R. China, implying the data originated from China. The nature of the tests (manufacturing quality control, laboratory testing) suggests they are prospective in nature, performed on the manufactured product.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. The tests performed are objective, quantitative measurements (e.g., physical dimensions, water fill test, chemical analysis for powder, biological response in irritation tests) rather than subjective assessments requiring expert consensus for ground truth.

4. Adjudication method for the test set

Not applicable. The tests are objective measurements against established standards, not requiring multi-reader adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This device is a patient examination glove, not an AI-assisted diagnostic or imaging device.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical medical device, not an algorithm.

7. The type of ground truth used

The ground truth used for performance validation is defined by:

  • Established ASTM Standards: ASTM-D-5250-00 for physical characteristics and dimensions, and ASTM D6124-97 for residual powder.
  • FDA Requirements: Specific FDA requirements for pinhole testing (1000 ml Water Fill Test).
  • Biocompatibility Testing Protocols: Standard methods for primary skin irritation and sensitization testing.
    These are all objective, quantitative, and standardized criteria.

8. The sample size for the training set

Not applicable. This device is a physical product, not a machine learning model, so there is no "training set."

9. How the ground truth for the training set was established

Not applicable. As there is no training set, this question is not relevant.

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.