K Number
K010576
Manufacturer
Date Cleared
2001-05-23

(85 days)

Product Code
Regulation Number
888.3070
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Monarch Spine System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scollosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).

The Monarch Spine System is also indicated for pedicle screw fixation for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.

The Monarch Spine System is also a hook and sacral/iliac screw fixation system of the noncervical spine indicated for degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, lordosis and/or kyphosis), tumor, and previous failed fusion (pseudarthrosis).

The Monarch Spine System Dual Rod Connectors can be used to connect Monarch Spine System Rods to rods of other DePuy AcroMed 4.75mm, 5.5mm, and 6.35mm diameter rod systems.

Device Description

The Monarch Spine System is both a rod-based and plate-based system designed to interface with various spinal anatomies. The plate-based system consists of pedicle screws, spine plates, ISOLA Modular Cross Connector (MCC) transverse connectors, J-hooks, washers, and nuts. The rod-based system consists of spinal rods, pedicle screws, polyaxial screws, caps, various slotted connectors, open and closed hooks, various rod-to-rod connections, ISOLA Modular Cross Connector (MCC) transverse connectors, and ISOLA EZ-X transverse connectors.

The Monarch Spine System is manufactured from implant grade titanium alloy that conforms to ASTM standard F-136.

AI/ML Overview

This document is a 510(k) premarket notification for the Monarch Spine System. It focuses on establishing substantial equivalence to previously cleared devices rather than providing detailed acceptance criteria or a dedicated study report for performance. Therefore, most of the requested information regarding acceptance criteria and a study proving device performance is not present in the provided text.

Specifically:

  1. A table of acceptance criteria and the reported device performance: This information is not provided. The document states that substantial equivalence is based on "equivalence in design, materials, manufacturing methods, intended use, and relative indications and contraindications" to predicate devices, but no specific performance criteria or their achievement are reported.

  2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective): Not applicable, as no pre-market clinical or performance study beyond substantial equivalence comparison is described.

  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience): Not applicable, as no independent expert ground-truthing for a test set is described.

  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable, as no test set requiring adjudication is described.

  5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a spinal fixation system, not an AI-driven diagnostic or assistive device that would involve human readers.

  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. The Monarch Spine System is a physical implant, not an algorithm.

  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable, as no ground truth for performance evaluation is described.

  8. The sample size for the training set: Not applicable. The device is a physical implant, not a machine learning model requiring a training set.

  9. How the ground truth for the training set was established: Not applicable.

Summary of what is available in the document:

  • Device Name: Monarch Spine System
  • Intended Use: Immobilization and stabilization of spinal segments as an adjunct to fusion for various spinal conditions (degenerative spondylolisthesis, fracture, dislocation, scoliosis, kyphosis, spinal tumor, failed previous fusion, severe spondylolisthesis (Grades 3 and 4) of L5-S1, degenerative disc disease, trauma, spinal stenosis, deformities).
  • Basis for Clearance: Substantial equivalence to predicate devices (TiMX Low Back System, Moss Miami Spinal System, ISOLA System) based on equivalence in design, materials, manufacturing methods, intended use, and indications/contraindications.
  • Material: Implant grade titanium alloy conforming to ASTM standard F-136.

The provided text is a 510(k) summary, which is a pre-market notification to the FDA. Its purpose is to demonstrate that a new device is "substantially equivalent" to a legally marketed predicate device, rather than to present detailed de novo performance study results against specific acceptance criteria.

{0}------------------------------------------------

MAY 2 3 2001

K010576

510(k) Summary

Proprietary name: Monarch Spine System

Common Name: Spinal Fixation System

Classification Name and Reference:

Spinal interlaminal fixation orthosis. §888,3050 Pedicle screw spinal fixation §888.3070

Proposed Regulatory Class: Class II

Device Product Code: 87/KWP, 87/MNH, 87/MNI

The Monarch Spine System is both a rod-based and plate-based system designed to interface with various spinal anatomies. The plate-based system consists of pedicle screws, spine plates, ISOLA Modular Cross Connector (MCC) transverse connectors, J-hooks, washers, and nuts. The rod-based system consists of spinal rods, pedicle screws, polyaxial screws, caps, various slotted connectors, open and closed hooks, various rod-to-rod connections, ISOLA Modular Cross Connector (MCC) transverse connectors, and ISOLA EZ-X transverse connectors.

The Monarch Spine System is manufactured from implant grade titanium alloy that conforms to ASTM standard F-136.

The Monarch Spine System is substantially equivalent to the TiMX Low Back System previously cleared in K981274 and K981714, the Moss Miami Spinal System previously cleared in 510(k) K955348 and K983583, and the ISOLA System previously cleared in 510(k) K980485.The substantial equivalence is based upon an equivalence in design, materials, manufacturing methods, intended use, and relative indications and contraindications.

The Monarch Spine System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scollosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis).

{1}------------------------------------------------

The Monarch Spine System is also indicated for pedicle screw fixation for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.

The Monarch Spine System is also a hook and sacral/iliac screw fixation system of the noncervical spine indicated for degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, lordosis and/or kyphosis), tumor, and previous failed fusion (pseudarthrosis).

The Monarch Spine System Dual Rod Connectors can be used to connect Monarch Spine System Rods to rods of other DePuy AcroMed 4.75mm, 5.5mm, and 6.35mm diameter rod systems.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a symbol composed of three stylized human profiles facing to the right, with flowing lines suggesting movement or connection.

MAY 2 3 2001

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Frank Maas Manager, Regulatory Affairs DePuv AcroMed 325 Paramount Drive Raynham, Massachusetts 02767-0350

Re: K010576

Trade Name: Monarch Spine System Regulation Number: 888.3050 and 888.3070 Regulatory Class: II Product Codes: MNI, KWP and MNH Dated: February 26, 2001 Received: February 27, 2001

Dear Mr. Maas:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

{3}------------------------------------------------

Page 2 - Mr. Frank Maas

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Mark N Melkum

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K010576

Device Name:

Indications for Use:

The Monarch Spine System is a pedicle screw system intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral degenerative spondylolisthesis with objective evidence of spine: neurological impairment, fracture, dislocation, scoliosis, spinal tumor, and failed previous fusion (pseudarthrosis).

The Monarch Spine System is also indicated for pedicle screw fixation for the treatment of severe spondylolisthesis (Grades 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to sacrum) with removal of the implants after the attainment of a solid fusion.

The Monarch Spine System is also a hook and sacral/iliac screw fixation system of the noncervical spine indicated for degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (fracture and/or dislocation), spinal stenosis, deformities (scoliosis, lordosis and/or kyphosis), tumor, and previous failed fusion (pseudarthrosis).

The Monarch Spine System Dual Rod Connectors can be used to connect Monarch Spine System Rods to rods of other DePuy AcroMed 4.75mm, 5.5mm, and 6.35mm diameter rod systems.

PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(per 21 CRF 801.109)

OR Over-The-Counter Use

(Division Sign-Off) (Optional Format 1-2-96)
Division of General, Restorative
and Neurological Devices

K010576

510(k) Number

§ 888.3070 Thoracolumbosacral pedicle screw system.

(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.