K Number
K003816

Validate with FDA (Live)

Manufacturer
Date Cleared
2001-03-09

(88 days)

Product Code
Regulation Number
888.3040
Age Range
All
Reference & Predicate Devices
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Titanium Fixation Device is intended for arthroscopic and limited mini-open surgeries requiring soft tissue fixation to bone for repair of shoulder, hand/wrist, foot/ankle, knee, elbow, and pelvis.

The Titanium Corkscrew is intended for fixation of suture to bone. This product is intended for the following indications:

  • Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps Shoulder: Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
  • Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Foot/Ankle: Hallux Valgus Reconstruction, Midfoot Reconstruction, Metatarsal Ligament Repair
  • Medial Collateral Ligament Repair, Lateral Collateral Ligament Knee: Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, and Illiotibial Band Tenodesis
  • Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction
  • Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Elbow: Radial Collateral Ligament Reconstruction
  • Bladder Neck Suspension for female urinary incontinence due to Pelvis: urethral hyper mobility or intrinsic sphincter deficiency
Device Description

The Arthrex, Inc. Titanium Fixation Device is a manually operated surgical device intended for suture fixation in the repair of tendons and ligaments. The implant is constructed of titanium and the suture is composed of #2 braided polyester. These components are used in a wide variety of medical devices and prior approved implants of this type.

AI/ML Overview

The provided text is a 510(k) summary for the Arthrex, Inc. Titanium Fixation Device. It is a premarket notification for a medical device and not a study report. Therefore, it does not contain information about acceptance criteria, device performance results from a study, sample sizes, expert ground truth establishment, or any details related to comparative effectiveness studies (MRMC or standalone).

This document mainly focuses on:

  • Device Identification: Trade name, classification, contact person.
  • Intended Use: Specifies the surgical applications for the device (e.g., soft tissue fixation to bone for repair in various body parts).
  • Description: Materials (titanium, braided polyester suture).
  • Substantial Equivalence: Compares the device to predicate devices, asserting that it is "substantially equivalent" in basic features and intended uses, implying no new questions of safety or effectiveness.
  • FDA Response: A letter from the FDA confirming receipt of the 510(k) and stating that the device is substantially equivalent to legally marketed predicate devices, thus allowing it to be marketed.

Therefore, I cannot provide the requested information regarding acceptance criteria, study findings, sample sizes, or expert involvement because this document is a regulatory submission, not a study report or clinical trial result.

To answer your questions, one would need access to the actual design validation and verification testing reports that would have been submitted to the FDA as part of the 510(k) application, but which are not included in this public summary document.

{0}------------------------------------------------

510(k) Summary

510(k) Number: Contact Person: Date Prepared:

100 38 1 Vernon C. Brown, Manager of Regulatory Affairs November 22, 2000

Trade/Proprietary Name: Titanium Fixation System

Fastener, fixation, non-degradable, soft tissue Classification Name: ANCHORLOK™ Soft Tissue Anchor System Wright Predicate Devices: Medical, Mitek Knotless Anchor Mitek Surgical Products, Straight-In Orthopedic fixation system, Influence, Inc., KSEA Flipptack Karl Storz Endoscopy-America, and Acumed Suture Anchor by Acumed, Inc.

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

Intended Use:

The Titanium Fixation Device is intended for arthroscopic and limited mini-open surgeries requiring soft tissue fixation to bone for repair of shoulder, hand/wrist, foot/ankle, knee, elbow, and pelvis.

Description:

The Arthrex, Inc. Titanium Fixation Device is a manually operated surgical device intended for suture fixation in the repair of tendons and ligaments. The implant is constructed of titanium and the suture is composed of #2 braided polyester. These components are used in a wide variety of medical devices and prior approved implants of this type.

Substantial Equivalence:

The Arthrex, Inc. Titanium Fixation Device is substantially equivalent to predicate devices where the basic features and intended uses are the same. Minor differences between the Titanium Fixation Device and predicate devices do not raise any questions concerning safety and effectiveness and have no apparent effect on the performance, function, or intended use of this device.

00 0080

{1}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image is a black and white circular logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which is a stylized representation of a human figure embracing a globe. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper half of the circle.

Public Health Service

MAR - 9 2001

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Mr. Vernon C. Brown Manager, Regulatory Affairs Arthrex. Inc. 2885 South Horseshoe Drive Naples, Florida 34104

Re: K003816

Trade Name: Arthrex Titanium Corkscrew, 3.5mm, 5.0mm, 6.5mm Regulatory Class: II Product Code: HWC and MBI Dated: December 8, 2000 Received: December 11, 2000

Dear Mr. Vernon:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations ( recting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

{2}------------------------------------------------

Page 2 - Mr. Vernon C. Brown

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597, or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

le Mark n Millkusa

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

The Titanium Corkscrew is intended for fixation of suture to bone. This product is intended for the following indications:

  • Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps Shoulder: Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction
  • Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Foot/Ankle: Hallux Valgus Reconstruction, Midfoot Reconstruction, Metatarsal Ligament Repair
  • Medial Collateral Ligament Repair, Lateral Collateral Ligament Knee: Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, and Illiotibial Band Tenodesis
  • Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction
  • Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Elbow: Radial Collateral Ligament Reconstruction
  • Bladder Neck Suspension for female urinary incontinence due to Pelvis: urethral hyper mobility or intrinsic sphincter deficiency

Mark N. Milheusen

Division Sign-Off) Division of General, Restorative and Neurological Devices K003816

510(k) Number_

000004

N/A