K Number
DEN230007
Device Name
LightStrike+ (MXSUV1-SL and MXSUV1-FT)
Date Cleared
2023-09-01

(212 days)

Product Code
Regulation Number
880.6510
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdparty
Intended Use
The Xenex LightStrike™+ is a pulsed, broad-spectrum, high-intensity, germicidal UV light system intended to perform microbial reduction on non-porous, non-critical medical device surfaces, free from visual soiling, after manual cleaning and disinfection practices. LightStrike+ is intended for use in unoccupied operating rooms, hospital rooms, and other clinical settings where non-critical medical devices are present as an adjunct to existing manual cleaning and disinfection practices. The system is for over-the-counter (OTC) use.
Device Description
LightStrike+ is a system that emits UV germicidal light to reduce the microbial load present on non-porous, non-critical medical device surfaces in healthcare environments, after manual cleaning and disinfection practices. Designed for microbial reduction cycles, the system will achieve the labeled log reductions through the utilization of Xenon lamp technology for the generation of UV germicidal light in the system's intended use environments. It is designed to be moved to targeted locations within a facility. The graphical user interface allows operators (trained cleaning staff) to select preprogrammed rooms and their associated protocols for cycle execution. The system is for use in unoccupied areas and includes design and redundant safety features to help prevent accidental exposure to the UV light.
More Information

Not Found

Not Found

No
The summary describes a UV light system for disinfection and does not mention any AI or ML components or functionalities.

No
The device is described as a germicidal UV light system intended for microbial reduction on non-porous, non-critical medical device surfaces in healthcare settings. Its purpose is disinfection, not direct treatment or diagnosis of a disease or condition in a patient, which is the definition of a therapeutic device.

No

Explanation: The device is described as a germicidal UV light system intended for microbial reduction on surfaces, not for diagnosing medical conditions.

No

The device description explicitly states it is a "system that emits UV germicidal light" and utilizes "Xenon lamp technology," indicating it is a hardware device with a software component for control and interface.

Based on the provided information, the Xenex LightStrike™+ is not an In Vitro Diagnostic (IVD) device.

Here's why:

  • Intended Use: The intended use is to perform microbial reduction on non-porous, non-critical medical device surfaces in clinical settings. This is an environmental disinfection function, not a diagnostic test performed on biological samples.
  • Device Description: The device emits UV light to reduce microbial load on surfaces. It does not analyze biological samples to provide diagnostic information about a patient's health or condition.
  • Lack of IVD Characteristics: The description does not mention any of the typical characteristics of an IVD, such as:
    • Analyzing biological samples (blood, urine, tissue, etc.)
    • Detecting or measuring analytes (proteins, DNA, antibodies, etc.)
    • Providing information for the diagnosis, monitoring, or treatment of a disease or condition.

The Xenex LightStrike™+ is a device intended for environmental disinfection in healthcare settings.

N/A

Intended Use / Indications for Use

The Xenex LightStrike™+ is a pulsed, broad-spectrum, high-intensity, germicidal UV light system intended to perform microbial reduction on non-porous, non-critical medical device surfaces, free from visual soiling, after manual cleaning and disinfection practices. LightStrike+ is intended for use in unoccupied operating rooms, hospital rooms, and other clinical settings where non-critical medical devices are present as an adjunct to existing manual cleaning and disinfection practices. The system is for over-the-counter (OTC) use.

Product codes (comma separated list FDA assigned to the subject device)

QXJ

Device Description

LightStrike+ is a system that emits UV germicidal light to reduce the microbial load present on non-porous, non-critical medical device surfaces in healthcare environments, after manual cleaning and disinfection practices. Designed for microbial reduction cycles, the system will achieve the labeled log reductions through the utilization of Xenon lamp technology for the generation of UV germicidal light in the system's intended use environments. It is designed to be moved to targeted locations within a facility. The graphical user interface allows operators (trained cleaning staff) to select preprogrammed rooms and their associated protocols for cycle execution. The system is for use in unoccupied areas and includes design and redundant safety features to help prevent accidental exposure to the UV light.

Mentions image processing

Not Found

Mentions AI, DNN, or ML

Not Found

Input Imaging Modality

Not Found

Anatomical Site

Not Found

Indicated Patient Age Range

Not Found

Intended User / Care Setting

LightStrike+ is intended for use in unoccupied operating rooms, hospital rooms, and other clinical settings where non-critical medical devices are present. The system is for over-the-counter (OTC) use.

Description of the training set, sample size, data source, and annotation protocol

Not Found

Description of the test set, sample size, data source, and annotation protocol

Not Found

Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)

The sponsor conducted the following performance tests to support that the device can achieve its intended use:

  • Optimized Kill Curve testing were provided, determining the dosage required for LightStrike+ to achieve a 2-log reduction of select bacterial organisms.
  • Bacteriostasis of Intended In-use Surfaces were provided to verify potential inherent bacteriostatic properties of surfaces intended for future In-use testing.
  • Recovery validation testing were provided to verify expected effectiveness of microbial recovery methods used to verify microbial reduction system results in a simulated use environment.
  • Simulated-use tests were provided, demonstrating microbial reduction on medical device surfaces, using the determined most UV resistant organism, achieving an average of 2-log reduction across all inoculated surfaces.
  • Accelerated UV materials damage test were provided, to evaluate the impact of regular exposure to LightStrike+ with metallic and non-metallic materials commonly present in healthcare environments, including materials commonly used as medical device enclosures.
  • Ozone tests were provided, to ensure that LightStrike+ users are safe, notwithstanding the very low concentration of ozone that is produced by the device as a byproduct of UV-C light emission (to which the user could be exposed), per 21 CFR 801.415 "Maximum acceptable level of ozone". The levels of toxic chemicals (other than ozone) generated during the use of LightStrike+ that would be safe for users/patients following the completion of the UV microbial reduction cycle were also tested, per 29 CFR 1910.1000.
  • Photobiological safety of lamps and lamp system test were provided, to evaluate the safety of the lamp that emits UV light when the robot is in use, conducted using a new lamp, because the light intensity is at its maximum when the lamp is new and that represents the worst-case scenario, per IEC 62471 "Photobiological Safety of Lamps and Lamp Systems.

Key Results:
The device has been appropriately evaluated for performance on the bench, demonstrating that is capable to achieve a microbial reduction of 2-log. Materials damage testing results met ISO 4582, ASTM D256-10, and ASTM A370-22 requirements. Ozone generation levels were below 21 CFR 801.415 acceptance levels of toxic chemicals were below 29 CFR 1910.1000 acceptance levels. Photobiological safety of lamps and lamp system met IEC 62471 criteria.

Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)

2-log reduction

Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.

Not Found

Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.

Not Found

Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).

Not Found

N/A

0

DE NOVO CLASSIFICATION REQUEST FOR LIGHTSTRIKE+

REGULATORY INFORMATION

FDA identifies this generic type of device as:

Whole room microbial reduction device. A whole room microbial reduction device is a medical device to be used to reduce microbial load on medical device surfaces following cleaning and disinfection.

NEW REGULATION NUMBER: 21 CFR 880.6510

CLASSIFICATION: Class II

PRODUCT CODE: QXJ

BACKGROUND

DEVICE NAME: LightStrike+

SUBMISSION NUMBER: DEN230007

DATE DE NOVO RECEIVED: February 01, 2023

SPONSOR INFORMATION:

Xenex Disinfection Services, Inc. 1074 Arion Circle\ Suite 116 San Antonio, TX 78216 USA

INDICATIONS FOR USE

The Xenex LightStrike™+ is a pulsed, broad-spectrum, high-intensity, germicidal UV light system intended to perform microbial reduction on non-porous, non-critical medical device surfaces, free from visual soiling, after manual cleaning and disinfection practices. LightStrike+ is intended for use in unoccupied operating rooms, hospital rooms, and other clinical settings where non-critical medical devices are present as an adjunct to existing manual cleaning and disinfection practices. The system is for over-the-counter (OTC) use.

LIMITATIONS

Limitations on device use are provided in the Instructions for Use Manual:

1

Warnings and Precautions

  • . do not use the device for microbial reduction of the critical or semi-critical surfaces of medical devices.
  • do not use the device for microbial reduction of food or water. .
  • do not use the device for microbial reduction of a human or animal body or . corresponding bodily fluids.
  • . not indicated for use in occupied rooms, areas, and spaces.
  • follow unit setup and positioning instructions. .

PLEASE REFER TO THE LABELING FOR A COMPLETE LIST OF WARNINGS. PRECAUTIONS AND CONTRAINDICATIONS

DEVICE DESCRIPTION

LightStrike+ is a system that emits UV germicidal light to reduce the microbial load present on non-porous, non-critical medical device surfaces in healthcare environments, after manual cleaning and disinfection practices. Designed for microbial reduction cycles, the system will achieve the labeled log reductions through the utilization of Xenon lamp technology for the generation of UV germicidal light in the system's intended use environments. It is designed to be moved to targeted locations within a facility. The graphical user interface allows operators (trained cleaning staff) to select preprogrammed rooms and their associated protocols for cycle execution. The system is for use in unoccupied areas and includes design and redundant safety features to help prevent accidental exposure to the UV light.

Image /page/1/Figure/9 description: The image shows a LightStrike+ robot with several labeled parts. The robot has a dome and xenon lamp on top, a user interface, and a handle. The robot also has an appliance inlet and cable retainer at the bottom, as well as a tethered motion detection cone, remote status/stop cone, and door placard.

2

System ComponentFunction Summary
RobotGenerates UV light for whole room microbial reduction
Xenon LampXenon gas lamp that emits germicidal UV at wavelengths from 240
to 315nm when pulsed by the robot. The lamp is stored inside of the
robot and rises to deliver the UV light
User InterfaceIncludes a room guide to allow the user to select the room where
LightStrike+ will be used
Cycle times are preprogrammed for each room and downloaded to
the robot over wireless communications
When a room is selected, the microbial reduction cycle will start once
the physical button above the user interface screen is pressed and the
robot senses the operator has exited the room
The microbial reduction cycle will automatically stop when the cycle
time countdown expires
Tethered Motion
Detection ConeSafety feature to help prevent inadvertent UV exposure to anyone
entering the room while a cycle is active
Placed inside the room/area where LightStrike+ will be used, facing
the entry door
Detects motion from the entry door to a room during a cycle and
signals the robot to terminate the microbial reduction cycle when
motion is sensed
Remote Status/Stop
ConeFeature to help prevent users from entering an area during a microbial
reduction cycle. The cone serves as an indicator that LightStrike+ is
in use within a room/area.
Located outside the entrance of the room where LightStrike+ is in use
Includes a feature to allow the user to terminate a cycle from outside
the room
Provides a visual countdown to indicate how much time is remaining
in an ongoing microbial reduction cycle
Provides warning to indicate that an individual should not enter the
room during an ongoing cycle
Once countdown expires, it provides a visual cue that it is safe to
enter the room
HandleUsed to wheel and position the robot in the area where LightStrike+
will be used
Can be pulled/pushed from both sides
Door PlacardWarning sign that adheres to the door exterior to indicate that
LightStrike+ is being used in the Room/Area

SUMMARY OF BENCH STUDIES

REPROCESSING, STERILITY AND SHELF-LIFE

The LightStrike+ does not require reprocessing and sterility. The LightStrike+ does not have a labeled shelf life; device performance is not adversely affected by aging or

3

storage. Built-in design features monitor when the lamp is nearing the end of its usable life and prevent use of the device at that point.

BIOCOMPATIBILITY

The LightStrike+ uses a Xenon gas flash lamp to produce pulsed broad-spectrum germicidal UV light that reduces microbial load on non-porous, non-critical medical device surfaces in the whole room, for a period of time, after the room has been manual cleaned and disinfected. The operation occurs only if the room is unoccupied.

Testing was conducted to demonstrate that the operation of LightStrike+ generates Ozone at concentration levels that are below the maximum acceptable level of ozone in medical devices, according to 21 CFR 801.415, and is without an appreciable risk of deleterious effects. Additionally, test results demonstrate that the levels of toxic chemicals (other than ozone) generated during the use of LightStrike+ that would be safe for users/patients following the completion of the UV microbial reduction cycle are either below the safe exposure limit or will dissipate to safe levels when a user enters the room after the completion of the cycle.

Specifically, the concentrations of Carbon monoxide (CO), Carbon dioxide (CO2), Nitric oxide (NO) and Nitrogen dioxide (NO2) are well below the safe exposure limits, per 29 CFR 1910.1000, after the use of the LightStrike+ device. Test results demonstrate that there is a very low probability for the generation of toxic levels of Hydroxyl radical. singlet oxygen, and superoxide anion radicals by the LightStrike+ device in an indoor healthcare environment due to the low energy of the photons compared to the required Bond Dissociation Energy and the lack of high levels of pollutants. This is also clear from the low concentration of gases such as Ozone. NO. and NO2 which are generated from the secondary interaction of these species with other molecules. The small amount of the generated radicals will also decay to safe levels because of the extremely low halflife and high reactivity of the species before a user enters the room after a disinfection cycle.

SOFTWARE

The LightStrike+ Software System is a collection software that executes on the Brazos physical device and cloud-based platforms. The Software System includes both off-theshelf software and software that is developed and maintained by Xenex. The Software System is composed of the Microbial Reduction Management System (MRMS). Device Interface, and Firmware Software Items. The MRMS Software Item runs on cloud-based platforms. The Device Interface and Firmware Software Items run on the LightStrike+ physical device.

The interfaces between the top-level Software Items are indicated in Figure 1. MRMS and the Device Interface communicate remotely through two TCP/IP based protocols. HTTPS and MOTT. The Device Interface and the Firmware communicate through a RS232 link using a custom communication protocol and over TCP/IP sockets. In addition

4

to the interfaces between Software Items in Figure 1, each Software Item includes some form of interface. The LightStrike+ Software System includes two distinct user types, Device Users and MRMS Users. Device Users are those users that are authorized to use the LightStrike+ physical device and interact with the Device Interface and Firmware Software Items. MRMS Users are those authorized to access the user interfaces of the MRMS Software Item. Adequate cybersecurity and software functionality were demonstrated according to FDA guidance "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices".

The LightStrike+ controls the integrity of the software by limiting access to trusted users and ensuring trusted content through software controls as defined within the Cybersecurity System, according to FDA guidance document titled "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" (https://www.fda.gov/media/86174/download). The effectiveness of the software security controls is determined through a threat model - risk assessment. Xenex controls deployment of software updates as per change control and procedures within the Software Development Plan. All potential firmware and software updates are reviewed against the most recent versions of the SRS.

ELECTROMAGNETIC COMPATIBILITY & ELECTRICAL SAFETY

The LightStrike+ electrical safety is in conformance with IEC 60601-1-6. and the electromagnetic compatibility (EMC) is in accordance with IEC 60601-1-2.

PERFORMANCE TESTING - BENCH

The sponsor conducted the following performance tests to support that the device can achieve its intended use:

  • . Optimized Kill Curve testing were provided, determining the dosage required for LightStrike+ to achieve a 2-log reduction of select bacterial organisms.
  • . Bacteriostasis of Intended In-use Surfaces were provided to verify potential inherent bacteriostatic properties of surfaces intended for future In-use testing.
  • . Recovery validation testing were provided to verify expected effectiveness of microbial recovery methods used to verify microbial reduction system results in a simulated use environment.
  • Simulated-use tests were provided, demonstrating microbial reduction on medical . device surfaces, using the determined most UV resistant organism, achieving an average of 2-log reduction across all inoculated surfaces.
  • . Accelerated UV materials damage test were provided, to evaluate the impact of regular exposure to LightStrike+ with metallic and non-metallic materials commonly present in healthcare environments, including materials commonly used as medical device enclosures.
  • . Ozone tests were provided, to ensure that LightStrike+ users are safe, notwithstanding the very low concentration of ozone that is produced by the device as a byproduct of UV-C light emission (to which the user could be exposed), per 21 CFR 801.415 "Maximum acceptable level of ozone". The levels

5

of toxic chemicals (other than ozone) generated during the use of LightStrike+ that would be safe for users/patients following the completion of the UV microbial reduction cycle were also tested, per 29 CFR 1910.1000.

  • Photobiological safety of lamps and lamp system test were provided, to evaluate . the safety of the lamp that emits UV light when the robot is in use, conducted using a new lamp, because the light intensity is at its maximum when the lamp is new and that represents the worst-case scenario, per IEC 62471 "Photobiological Safety of Lamps and Lamp Systems.
    The device has been appropriately evaluated for performance on the bench. demonstrating that is capable to achieve a microbial reduction of 2-log. Materials damage testing results met ISO 4582. ASTM D256-10. and ASTM A370-22 requirements. Ozone generation levels were below 21 CFR 801.415 acceptance levels of toxic chemicals were below 29 CFR 1910.1000 acceptance levels. Photobiological safety of lamps and lamp system met IEC 62471 criteria.

LABELING

The labeling consists of a user manual, instructions for use, and packaging labels. The instructions for use include the indications for use, including the time to achieve a 2-log reduction; a description of the device, contraindications, warnings, precautions, including a door placard with caution signs advising "DO NOT ENTER" to be affixed on entry ways of rooms/areas, warns of possible UV exposure during operation and warns of risk group 3 UV light generated by device; and instructions for the safe use of the device.

Labeling for this device is in accordance with the special controls listed below.

RISKS TO HEALTH

The table below identifies the risks to health that may be associated with use of a whole room microbial reduction device and the measures necessary to mitigate these risks.

Risks to HealthMitigation Measures
Exposure to microbiocidal agent,
leading to skin and eye damageNon-clinical performance testing
Biocompatibility evaluation
Software verification, validation, and hazard analysis
Labeling
Respiratory mucous membrane irritation
and pulmonary edema due to chemical
exposureNon-clinical performance testing
Biocompatibility evaluation
Patient cross-contamination due to
device failure leading to inadequate
microbial reductionNon-clinical performance testing
Labeling
Software verification, validation, and hazard analysis
Electrical shockElectrical safety testing
Non-clinical performance testing

6

Risks to HealthMitigation Measures
Interference with other devicesLabeling
Electromagnetic compatibility testing
Electrical safety testing
Wireless coexistence testing
Labeling

SPECIAL CONTROLS

In combination with the general controls of the FD&C Act, the whole room microbial reduction device is subject to the following special controls:

  • Non-clinical performance testing must demonstrate that the device performs as (1) intended under anticipated conditions of use. The following performance characteristics must be tested:
    • (i) Performance testing must demonstrate microbial log reduction of the demonstrated most resistant microorganism on medical device surfaces commensurate with the intended level of microbial reduction.
    • (ii) Simulated use testing must evaluate device performance under simulated worstcase use conditions (e.g., soiling, room objects and surfaces, distances).
    • In-use testing must evaluate device performance under real-world use conditions. (iii)
    • (iv) Performance testing must demonstrate the photobiological safety of any lamps or lamp systems.
    • Performance testing must evaluate safety features intended to prevent exposure (v) and ensure that device operation can only occur in an unoccupied environment: and
    • Performance testing must characterize the long-term material compatibility of the (vi) microbiocidal agent on clinically relevant surfaces and/or devices.
  • (2) Biocompatibility testing must demonstrate safe residual levels of chemicals on medical devices surfaces and/or gaseous byproducts in air.
  • Software verification, validation, and hazard analysis must be performed for any (3) software components.
  • Performance data must demonstrate the electromagnetic compatibility (EMC) and (4) electrical safety of the device.
  • Labeling must include: (5)
    • Warnings and instructions to ensure the device is operated in an unoccupied (i) environment.
    • (ii) Setup and positioning instructions; and
    • (iii) Information regarding material compatibility.

BENEFIT-RISK DETERMINATION

Risks:

The risks of the device are based on nonclinical laboratory (bench) studies described above.

Subject device is to be used in an unoccupied room. The only risk of injury is skin and eve damage resulting from exposure to dangerous light source, respiratory issues from exposure to

7

ozone, electrical shock and interference with other devices are mitigated with the performance testing, warnings, precautions, and special controls. Risk of device failure leading to microbial growth, increasing the risk of patient-to-patient cross contamination and study limitations, are mitigated with the performance testing, labeling, special controls, and the use as adjunct to manual cleaning and disinfection only.

Benefits:

The probable benefits of the device are based on nonclinical laboratory (bench) studies described above.

The LightStrike+ device demonstrated 2-log average reduction in simulated use testing using worst-case microorganisms. The device will be used as an adjunct to existing manual cleaning practices. LightStrike+ provides additional microbial reduction after manual cleaning and disinfection, strengthening the risk reduction of transmitting pathogens from one patient to the next, or between health care providers and patients. The magnitude of the clinical benefit (i.e., transmission reduction) has not been established based on the non-clinical testing provided by the sponsor. Data was robust for 2-log average reduction on medical device surfaces, in the simulated environment.

Patient Perspectives

This submission did not include specific information on patient perspectives for this device.

Benefit/Risk Conclusion

Based on the performance testing provided (in particular, log reduction test, in a simulated-use test, optimized kill curve determination test, bacteriostasis test, microbial recovering test, biocompatibility test, material degradation test, ozone test, photobiological safety of lamps and lamp systems tests, and labeling), the sponsor has adequately supported that the subject device is a UV-C light microbial reduction device adjunctive to manual cleaning. This feature will provide an additional microbial reduction to surfaces that were previous cleaned and disinfected, helping to prevent cross contamination among patients and healthcare staff.

In conclusion, given the testing provided above, for the following indication statement:

The Xenex LightStrike™+ is a pulsed. broad-spectrum. high-intensity, germicidal UV light system intended to perform microbial reduction on non-porous, non-critical medical device surfaces. free from visual soiling, after manual cleaning and disinfection practices. LightStrike+ is intended for use in unoccupied operating rooms, hospital rooms, and other clinical settings where non-critical medical devices are present as an adjunct to existing manual cleaning and disinfection practices. The system is for over-the-counter (OTC) use.

The probable benefits outweigh the probable risks for the LightStrike+. The device provides benefits and the risks can be mitigated using general controls and the identified special controls.

8

CONCLUSION

The De Novo request for the LightStrike+ is granted and the device is classified as follows:

Product Code: QXJ Device Type: Whole room microbial reduction device Regulation Number: 21 CFR 880.6510 Class: II