K Number
DEN200005
Date Cleared
2021-12-03

(668 days)

Product Code
Regulation Number
878.4880
Type
Direct
Panel
SU
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Klox Biophotonic LumiHeal™ System is indicated to provide blue light and fluorescent light energy for use on post-surgical incisions for scar management. The System is intended to be used in FST I-IV female patients 22 years and over.

Device Description

The Klox Biophotonic LumiHeal System is a device which consists of a blue light emitting Multi-LED Light device (KT-L Lamp) with emitted wavelengths of 440-460nm and a topical photoconverter gel (LumiHeal Gel). When the gel is illuminated by the LED device, it will emit fluorescence with blue, green, yellow, and orange wavelengths between 400mm to 625mm. The fluorescence mixed with the excitation blue light is used for scar management (i.e., reducing the appearance of acute post-surgical incisions).

The KT-L Lamp (Figure 1) has a timer and a distance sensor with which the system set the illumination time at 5 min and distance between the lamp and the wound area at 5cm. For the KT-L Lamp interface, press the Distance Verification Button to measure the distance. Display Screen should read approximately 50mm. Press the Time Display Button to return display to Timer Mode. During the illumination, it is possible to toggle between the Timer Mode or Measure Mode by pressing either the Time Display Button or the Distance Verification Button. Time remaining during illumination will be displayed in minutes (min). The lamp comes equipped with a power cable, 2 pairs of protective eyeglasses, and a user manual. The device specification is shown in the Table 1.

The LumiHeal Gel is provided as a two-component gel (Figure 2), specifically Jar A (25g) and Jar B (2.5g). The two components of the LumiHeal Gel are intended to be mixed immediately prior to use. Jar A is the carrier gel and Jar B is the chromophore gel. The LumiHeal Gel is applied just before use and is intended to remain in contact with the surface of wound for 5 minutes during the light exposure and application. After the 5-minute application period, exposure to the KT-L Lamp is discontinued and the LumiHeal Gel is removed.

The purpose of the LumiHeal Gel is to facilitate conversion of the non-coherent blue light wavelength from the KT-L Lamp into blue, green, yellow and orange wavelength light between 400nm to 625mm at the skin surface.

AI/ML Overview

Acceptance Criteria and Device Performance for Klox Biophotonic LumiHeal System

This response synthesizes the acceptance criteria and study findings for the Klox Biophotonic LumiHeal System, as described in the provided regulatory document.

1. Table of Acceptance Criteria and Reported Device Performance

The provided document does not explicitly state "acceptance criteria" for effectiveness in a quantitative manner (e.g., "device must achieve X% improvement in Y score"). Instead, it describes general expectations for safety and performance (mitigation of risks) and then reports the study's effectiveness findings. Thus, the acceptance criteria are inferred from the stated benefits and the demonstrated performance.

Acceptance Criteria (Inferred from Risks & Benefits)Reported Device Performance
Safety:Safety:
1. Biocompatibility: Patient-contacting components must be biocompatible (prevent adverse tissue reactions).Biocompatibility assessment was adequate for both light-exposed and non-exposed photoconverter gel. Conclusion: "risk of a clinically significant biocompatibility concern is low." (Section: BIOCOMPATIBILITY/MATERIALS)
2. Sterility: Patient-contacting components must be sterile to prevent infection.LumiHeal Gel Jar B is terminally sterilized via autoclaving. Preservative effectiveness testing per USP for real-time aged Jar A and packaging integrity testing were performed. "The test gels met the acceptance criteria for each test." (Section: SHELF LIFE/STERILITY)
3. Shelf Life: Photoconverter gel must maintain sterility and functional performance over its shelf life.Shelf life testing performed (preservative effectiveness, packaging integrity); "The test gels met the acceptance criteria for each test." (Section: SHELF LIFE/STERILITY) Performance after simulated 5 years of service life met specifications. (Section: PERFORMANCE TESTING - BENCH, last bullet point)
4. Electrical & Thermal Safety: Device must be electrically/thermally safe and electromagnetically compatible.The KT-L lamp passed all relevant portions of IEC 60601-1 and IEC 60601-1-2 testing. (Section: ELECTROMAGNETIC CAPABILITY & ELECTROMAGNETIC SAFETY) Evaluation of heat dissipation showed skin temperatures remained "*within a safe temperature range (

N/A