K Number
DEN190055

Validate with FDA (Live)

Manufacturer
Date Cleared
2021-10-01

(653 days)

Product Code
Regulation Number
874.4450
Type
Direct
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The iotaSOFT™ Insertion System is intended to aid the surgeon in placement of cochlear implant electrode arrays into a radiographically normal cochlea by controlling the speed of implant insertion. The iotaSOFT Insertion System is intended for use in cochlear implant patients ages 12 years and older during cochlear implant procedures using either a round window or cochleostomy approach.

Device Description

The iotaSOFT™ Insertion System is designed to assist the surgeon during cochlear implantation by controlling electrode array insertion speed. The system consists of a single-use, sterile drive unit (iotaSOFT™ DRIVE Unit) connected to a reusable, non-sterile touch screen control console and footpedal (iotaSOFT Controller and Accessories) (Figure 1).

AI/ML Overview

Device Acceptance Criteria and Performance Study: iotaSOFT™ Insertion System

The iotaSOFT™ Insertion System is a powered insertion device designed to aid surgeons in placing cochlear implant electrode arrays. The acceptance criteria for this device and the studies conducted to prove it meets these criteria are detailed below, synthesizing information from the provided document.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for the iotaSOFT™ Insertion System are primarily outlined in the "SPECIAL CONTROLS" section and implicitly through the "SUMMARY OF NONCLINICAL/BENCH STUDIES" and "SUMMARY OF CLINICAL TESTING" sections. The reported performance is extracted from the respective study summaries.

Acceptance Criteria CategorySpecific CriteriaReported Device Performance
Clinical PerformancePerforms as intended under anticipated conditions of use, including evaluation of all adverse events.- No clinically significant patient safety or outcomes differences based on surgeon experience. - Electrode array impedance measures within normal limits in all subjects at 1-month follow-up. - Neural response telemetry (NRT) measurements present in all subjects with normal cochlear anatomy (n=20/21) at post-operative follow-up. - Adverse events generally as expected for cochlear implant procedures (1 CSF leak noted, but not attributable to device use per instructions). - Insertion time ranged from 57 sec to 6 min, 1 sec (mean 3 min 15 sec). - Post-insertion cochlea view x-ray imaging satisfactory on all insertions with normal cochlear anatomy.
Non-Clinical Performance(i) Verification of CI attachment force, release force, and insertion speed.- Insertion speed verified to be within 20% of both 0.1 and 1 mm/s. - Maximum insertion force (with array) of <300mN met at both 0.1mm/s and 1mm/s. - Slip force specification of >80mN met at both 1mm/s and 0.1mm/s. - Maximum drive wheel pinch force of <5.6N. - Drive head removal force (when decoupling array) of <100mN.
(ii) Device does not damage or degrade the CI.- Not explicitly stated as a separate acceptance criterion, but implied by successful insertion and normal impedance/NRT. Bench testing on insertion force parameters (pinch force, removal force) suggests no damage.
(iii) Comparison testing with manual insertion to evaluate:
(A) Differences in CI array insertion force.- Average of maximum insertion force for iotaSOFT assisted insertions were lower than manual (51% reduction at 0.1mm/s, 32% reduction at 1mm/s). - Average insertion force variation for iotaSOFT assisted insertions was 78% lower than manual at 0.1mm/s and 70% lower at 1mm/s.
(B) Intracochlear placement of CI array.- Cadaveric comparison: iotaSOFT showed similar results to manual insertion across all surgeons and arrays in terms of:      - Scala Tympani Insertion: Manual 44% (7/16), iotaSOFT 50% (6/12)*      - Scala Media Translocation: Manual 43% (7/16), iotaSOFT 42% (5/12)*      - Scala Vestibuli Translocation: Manual 13% (2/16), iotaSOFT 8% (1/12)*      - Tip Fold-Over: Manual 19% (3/16), iotaSOFT 8% (1/12)*      - Insertion Angle: Manual 309.98 ± 126.6°, iotaSOFT 307.4 ± 95.9° *Four iotaSOFT samples excluded post-insertion.
Usability Testing(i) Successful use to aid in placement of electrode array.- Data show that surgeons with nurse staff support can successfully use iotaSOFT in the hospital environment.
(ii) Harms caused by use errors observed.- Harms associated with use errors deemed minor: prolonged operative/anesthesia time due to delay during surgery and setup. - Mitigated with design changes to screws and screwdriver, and training update. - For all tasks, use errors deemed not to lead to serious patient harm.
CI Compatibility ValidationChanges in CI compatibility are determined to significantly affect safety/effectiveness and must be validated.- Supported with Advanced Bionics HiFocus SlimJ, Cochlear Slim Straight, MED-EL Flex 24/28 (rationale provided for Flex 28 leveraging Flex 24 data).
BiocompatibilityPatient-contacting components demonstrated to be biocompatible.- Patient-contacting components (right/left drive housings, stage base, arm heat shrink, bone screws, silicone) passed all biocompatibility requirements for an external communicating device in limited contact (<24hr). - Testing performed per ISO 10993-1:2009 and FDA Guidance.
EMC/Electrical SafetyPerformance testing must demonstrate EMC, electrical safety, and thermal safety.- Passed all tests for Electrical Safety (compliant with AAMI ES60601-1:2012). - Passed all tests for EMC (compliant with IEC 60601-1-2:2014) including conducted/radiated emissions, ESD immunity, radiated/conducted RF immunity, fast transient/burst immunity, surge immunity, and voltage dips/interruptions/variations.
Sterility/Non-PyrogenicityPatient-contacting components sterile and non-pyrogenic.- Single-use components (iotaSOFT Drive Unit) provided sterile via EO sterilization (SAL 10-6). - EO/ECH residuals comply with ISO 10993-7. - Endotoxin testing performed (LAL kinetic turbidimetric method), all samples met acceptance criteria (< 2.15 EU/device).
Shelf LifePerformance testing supports shelf life (sterility, package, functionality).- Labeled 6-month shelf life. - Packaging validation: Samples subjected to EO, environmental conditioning, transportation simulation; met acceptance criteria for seal strength (≥1 lb/in) and gross leak (no bubble leak). - Shelf life validation: Samples subjected to EO, environmental conditioning, accelerated aging; met acceptance criteria for seal strength (≥1 lb/in) and gross leak (no bubble leak).
Software V&VSoftware V&V and hazard analysis for any software components.- Software and firmware described, verified, and validated (Control Console Software v1.0.3, Drive Unit PCB Firmware v0.1.1). - Documentation provided: SRS, Device Hazard Analysis, Traceability Analysis, V&V, Revision Level History. - Cybersecurity addressed per FDA guidances, threat model analysis (STRIDE) performed.

2. Sample Sizes and Data Provenance

The document describes several studies with varying sample sizes and data provenance:

  • Bench Performance Testing (Device Characteristics):

    • Sample Size: "All samples tested." (Specific number of devices/arrays not quantified, but implied to be sufficient for statistical confidence based on sterilization/accelerated aging protocols).
    • Data Provenance: Non-clinical, bench testing.
  • Synthetic Cochlea Insertion Force Comparison (Surgeon Bench Comparison Testing):

    • Sample Size: Devices "having undergone sterilization." (Specific number not quantified). Involves "multiple surgeons" (number not specified).
    • Data Provenance: Non-clinical, bench testing using synthetic cochlea models.
  • Cadaveric Comparison Testing:

    • Sample Size: 16 cadaver heads (1 cadaver per surgeon).
    • Data Provenance: Non-clinical, cadaveric study.
    • Country of Origin: Not specified but common for such studies to be conducted in the US.
    • Retrospective/Prospective: Prospective, as it involves active experimentation.
  • Usability Testing:

    • Sample Size: 16 Surgeon/Nurse teams, 16 cadavers (1 cadaver per team).
    • Data Provenance: Non-clinical, simulated use environment in cadavers.
    • Country of Origin: Not specified.
    • Retrospective/Prospective: Prospective.
  • Clinical Testing:

    • Sample Size: 21 subjects.
    • Data Provenance: Clinical study.
    • Country of Origin: Not specified, but generally US-based for FDA de novo.
    • Retrospective/Prospective: Prospective.

3. Number of Experts and Qualifications for Ground Truth (Test Set)

For the cadaveric comparison testing and usability testing, the "experts" are the surgeons involved.

  • Cadaveric Comparison Testing:

    • Number of Experts: 16 surgeons.
    • Qualifications: Not explicitly stated for each of the 16, but generally implied to be experienced CI surgeons. The usability testing section mentions "Surgeon CI experience level ranged from 1 to 30+ years: approximately equal numbers of surgeons with less than 10 years, 10 to 20 years, and greater than 20 years. Between 10 and 180 implantations were performed manually in the year preceding this study, across all surgeons." This provides a strong indication of their expertise.
    • Ground Truth Establishment: Array intracochlear position determined via 3D x-ray. This serves as objective measurement for array placement.
  • Usability Testing:

    • Number of Experts: 16 Surgeon/Nurse teams.
    • Qualifications: Same as above for surgeons. ("Surgeon CI experience level ranged from 1 to 30+ years: approximately equal numbers of surgeons with less than 10 years, 10 to 20 years, and greater than 20 years. Between 10 and 180 implantations were performed manually in the year preceding this study, across all surgeons.")
    • Ground Truth Establishment: "Was task successful? Y/N responses," use errors noted, user comments, Likert scale responses. This is a subjective assessment of usability and safety by the users themselves.

For the clinical testing, the assessment is based on clinical outcomes relevant to device safety and function.

  • Number of Experts: 3 surgeons.
  • Qualifications: "The principal investigator (PI) of the study... a neurotologist with nearly 40 years experience who implanted ~75 CIs manually in the year prior to this study." The other two surgeons "implanted 25 to 30 CIs each in their careers prior to commencement of this study." This establishes high levels of expertise.
  • Ground Truth Establishment: Objective clinical measurements (electrode impedance, NRT measurements, x-ray imaging) and observed adverse events.

For bench performance testing, the ground truth is established by objective engineering measurements (e.g., speed, force values) using calibrated equipment.

4. Adjudication Method for the Test Set

  • Cadaveric Comparison Testing: The method for 3D x-ray determination of array position is not detailed in terms of adjudication. It appears to be an objective measurement, not requiring expert consensus adjudication in the typical sense of radiological review.
  • Usability Testing: Adjudication is implicitly through observation of tasks, noting of use errors, and direct feedback from the surgeon/nurse teams. No explicit multi-rater adjudication process is described for this feedback.
  • Clinical Testing: Clinical outcomes are observed and reported. While medical staff would monitor these, a formal adjudication panel (e.g., for adverse events) is not explicitly detailed but standard practice for clinical trials. The outcome (normal impedance, NRT presence, satisfactory x-ray, expected AEs) suggests general medical consensus for interpreting these results.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No explicit MRMC comparative effectiveness study, as typically seen in AI diagnostic imaging (comparing human readers with vs. without AI assistance), was performed. The studies are more focused on validating the device's performance compared to manual insertion or evaluating its safe and effective use.

  • Cadaveric Comparison: This study compares device-assisted insertions to manual insertions, which is a form of comparative effectiveness, but it's human + device vs. human, not human reader with AI vs. human reader without AI assistance in interpretation. The outcomes are about insertion quality, not diagnostic accuracy.
    • Effect Size: Not applicable in the context of an MRMC diagnostic study. The "effect" here is the device's impact on insertion parameters and placement. The device assisted in achieving similar array placement outcomes as manual insertion, while showing lower maximum insertion force and lower insertion force variation.

6. Standalone Performance

The "Bench Performance Testing" section details the standalone performance of the algorithm/device (meaning, the device's mechanical and software functions) without human-in-the-loop performance influencing the intrinsic numerical output of speed, force, etc. This shows whether the device itself can meet its operational specifications.

  • Insertion speed verification within 20% of set rates (0.1 and 1 mm/s).
  • Maximum insertion force <300mN.
  • Slip force >80mN.
  • Maximum drive wheel pinch force <5.6N.
  • Drive head removal force <100mN.

These tests evaluate the device's inherent mechanical and control capabilities.

7. Type of Ground Truth Used

Different types of ground truth were used depending on the study:

  • Bench Performance Testing: Objective engineering measurements (e.g., physical force transducers, speed sensors) and device specifications.
  • Synthetic Cochlea Insertion Force Comparison: Objective measurements of force (mN) and force variation (mN/sec) from the synthetic models.
  • Cadaveric Comparison Testing: Objective anatomical imaging (3D x-ray) to determine intracochlear position, translocations, tip fold-over, and insertion angle.
  • Usability Testing: User-reported success/failure, observed use errors, and subjective Likert scale responses.
  • Clinical Testing: Objective clinical outcomes (electrode impedance, neural response telemetry (NRT), post-insertion x-ray imaging findings) and observed adverse events.

8. Sample Size for the Training Set

The document does not describe a separate "training set" in the context of machine learning. The iotaSOFT Insertion System is a robotic/mechanical device with sophisticated software, but there is no indication that it uses machine learning algorithms that require a specific "training set" in the way a diagnostic AI model would. The software and firmware are verified and validated, which is a different process from training a machine learning model.

9. How the Ground Truth for the Training Set was Established

Since there is no explicit machine learning training set mentioned, this question is not applicable. The device's operational parameters are based on engineering design and validated through bench, cadaveric, and clinical testing, not "training" in the ML sense.

{0}------------------------------------------------

DE NOVO CLASSIFICATION REQUEST FOR IOTASOFTTM INSERTION SYSTEM - DRIVE UNIT, CONTROLLER AND ACCESSORIES

REGULATORY INFORMATION

FDA identifies this generic type of device as:

Powered insertion system for a cochlear implant electrode array. A powered insertion system for a cochlear implant electrode array is a prescription device used to assist in placing an electrode array into the cochlea.

NEW REGULATION NUMBER: 21 CFR 874.4450

CLASSIFICATION: Class II

PRODUCT CODE: QQH

BACKGROUND

DEVICE NAME: iotaSOFT Insertion System - Drive Unit, Controller and Accessories

SUBMISSION NUMBER: DEN190055

DATE DE NOVO RECEIVED: December 18, 2019

SPONSOR INFORMATION:

IotaMotion, Inc. 14 ½ S. Clinton St. Iowa City, Iowa 52240

INDICATIONS FOR USE

The iotaSOFT™ Insertion System is intended to aid the surgeon in placement of cochlear implant electrode arrays into a radiographically normal cochlea by controlling the speed of implant insertion. The iotaSOFT Insertion System is intended for use in cochlear implant patients ages 12 years and older during cochlear implant procedures using either a round window or cochleostomy approach.

LIMITATIONS

The iotaSOFT™ Insertion System is not indicated for use in patients with craniofacial abnormalities, temporal squamosal skull thickness less than 3mm, major cochlear lesions (e.g. fibrosis, fracture, or ossification), or cochlear malformations.

The iotaSOFT™ Insertion System is only intended for use on individuals 12 years of age or older.

{1}------------------------------------------------

The iotaSOFT™ Insertion System is indicated for use only (a) with the Advanced Bionics HiFocus SlimJ, Cochlear Slim Straight, and MED-EL Flex 24 & 28 cochlear implant electrode arrays; and (b) after giving due consideration to all relevant facts and circumstances, including the surgeon's clinical judgment and the array's instructions for use, contraindications, limitations for use, and any accompanying surgical guides.

The sale, distribution, and use of the iotaSOFT Insertion System are restricted to prescription use in accordance with 21 CFR 801.109.

PLEASE REFER TO THE LABELING FOR A COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.

DEVICE DESCRIPTION

The iotaSOFT™ Insertion System is designed to assist the surgeon during cochlear implantation by controlling electrode array insertion speed. The system consists of a single-use, sterile drive unit (iotaSOFT™ DRIVE Unit) connected to a reusable, non-sterile touch screen control console and footpedal (iotaSOFT Controller and Accessories) (Figure 1).

Image /page/1/Figure/5 description: The image shows a medical device system with several components. The components include a control console with a screen displaying "Press Right Footpedal to Insert Electrode" and an insertion rate of 0.5 mm/s, a sterile disposable drive unit, and a foot pedal. The text indicates that the control console and foot pedal are non-sterile and reusable, while the drive unit is sterile and disposable.

Figure 1: iotaSOFTTM Insertion System Overview: The iotaSOFTTM Insertion System is made up of the three main components shown: User Control Console, Drive Unit (screwdriver not shown), and surgeon footpedal controller. Components within Blue dashed region represent components used within the sterile surgical field. Green dash region represents components used outside the sterile surgical field.

{2}------------------------------------------------

System Components

iotaSOFT Drive Unit (sterile, disposable)Drive (Motor) Unit
Drive Head
Unit Base
Self-Drilling Screws x2
Screw Driver
iotaSOFT Controller and Accessories (non-sterile, reusable)Touch Screen
Graphical User Interface
Rolling Stand Assembly
Foot pedal
Accessory Control Console Power Supply
Accessory USB extender cables x2

SYSTEM COMPONENT DESCRIPTION: iotaSOFT™ DRIVE Unit

Image /page/2/Figure/3 description: This image shows a medical device with several labeled parts. The device includes a drive unit, a cable, a drive head, and an arm. It also has a locking tab, a unit base, self-drilling bone screws, a hinge, and wheels. Additionally, there is a loading notch visible in the image.

Figure 2: iotaSOFTTM | DRIVE Unit Sub-Components

Drive (Motor) Unit: The Drive Unit is supplied sterile in a tray with Tyvek lid with the screwdriver packaged within. The Drive unit (Figure 2) couples to the electrode lead via the drive head (A) which is linked to a drive cable housed within the semi-rigid, "stay-put" gooseneck arm (B) of the main drive unit (C). The drive unit is held within the unit base (D) that is secured to the patient via self-drilling bone screws (G). The power/ communications cable (E) is passed off the sterile field and connected to the main control console. The drive unit is designed to be adjustable by sliding forward or backwards within the base or removable for electrode loading then replacing and securing into the base with the side locking tabs (F). Based on patient anatomy, the drive head trajectory and angle may also be adjusted by bending the flexible, semi-rigid arm and/or rotating drive head (Figure 3).

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows six different diagrams of a medical device. The diagrams show the device in different positions and with different functions. The first diagram shows the device in a closed, neutral position. The second diagram shows the device with a bendable neck. The third diagram shows the device with a rotatable head. The fourth diagram shows the device with a clasping action. The fifth diagram shows the device with a drive unit slide. The sixth diagram shows the device with a toggle lock.

Figure 3: iotaSOFTTM | DRIVE Unit adjustability features to manually manipulate drive head position and electrode array insertion trajectory.

Drive Head: The drive head consists of two halves which clamp to the CI electrode lead via spring loaded hinge. The drive wheels are houses within the drive head with the wheel geometries designed to cup around the electrode lead and facilitate loading. The electrode engagement mechanism is compatible with the electrode arrays listed above having multiple and varying lead nominal diameters ranging 0.8 to 1.2mm. The drive head tips are additionally designed with grooved, keyhole inner geometry or track to accommodates for "winged" electrodes to maintain the proper electrode array contact orientation toward the modiolus during insertion (Figure 4) The drive head is design with a spring clamp and engagement gap, (~0.38 mm) to prevent over-compressing the electrode lead. A forceps loading notch is incorporated which increases ease of electrode coupling to the drive head. The loading notch allows both closure of the drive head halves around the electrode while being held with forceps and enables subsequent release of the forceps without re-opening the drive head tip is tapered for improved operative views and made of a semi-transparent polycarbonate plastic which allows visualization of the electrode within the drive tips and operative light transmission.

Image /page/3/Picture/3 description: The image shows a close-up of a mechanical device. The device appears to be made of a light-colored material, possibly plastic or metal, and has several components, including wheels and fasteners. The image also includes text labels pointing to specific parts of the device, such as the "Drive Wheel" and "Idler Wheel."

Figure 4: iotaSOFT | Drive Head representation of drive wheels (right) and clasping hinge coupling mechanism to electrode lead (left).

{4}------------------------------------------------

Unit Base: The unit base provides an adjustable yet stable and secure platform for the drive unit to sit within. The base holds the main drive unit and enables smooth, sliding adjustments with dual locking side toggle tabs. The base inner well "cupped" design allows the drive unit to securely slide within the base yet enables one-handed installation or removal of the unit. Similarly, the dual locking side tabs add additional stabilization once the desired unit positioning has been achieved. Additionally, the base "foot" region holds the pre-loaded, semi-captive screws. The geometry of the base and contact angle of the "foot region" is designed to match curvature of the skull and sit over skin flap between base and skull bone (Figure 5).

Image /page/4/Figure/1 description: This image shows a diagram of a medical device with measurements. The device is 77mm in length. The diagram also shows a screw with measurements of 1mm, 3mm, 7mm, and 3mm, as well as a vertical measurement of 30mm.

Figure 5: iotaSOFTTM | DRIVE Unit Dimensions: Approximate dimensions in mm.

Screwdriver & Self-Drilling Screws: Screwdriver is provided sterile to screw the self-drilling screws x2 into the skull to secure and stabilize the unit base. The bone screws are self-drilling titanium bone screws with nominal outer diameter of 1.5mm and nominal length of 7mm (Figure 5). The bone screws are preloaded and semi-captive in unit base by a retaining silicone plug. When used as intended in combination within the unit base the screw functional length into skull is limited to maximum depth of 3mm.

{5}------------------------------------------------

SYSTEM COMPONENT DESCRIPTION: iotaSOFT™ Controller & Accessories

Touch screen control console with Graphical User Interface (GUI): The Control console (Figure 6) provides power and communication across iotaSOFT™ system components including the Drive unit and surgeon foot pedal. The console consists of an all-in-one computer with high definition Projected Capacitive (PCAP) touchscreen screen and USB connections for the foot pedal and drive unit cables. The iotaSOFT™ software is directly loaded at startup to operate the DRIVE Unit.

Control Footpedal: The surgeon has handsfree control of the Drive unit and insertion rate setting (0.1mm/sec to 1.0mm/sec) via the non-sterile foot pedal (Figure 6). The surgeon controls the CI advancement and retraction using the foot pedal. With left, yellow pedal reversing electrode and right, blue pedal advancing per standard convention. The center, black button can be pressed to access insertion rate control menu and/or select to confirm and close menus.

Accessories: Control console power supply and 2 USB extender cables: The Control Console power supply plugs into the mains power outlet. Two USB 10ft extension cables are secured and connected to the console with excess length coiled on the rolling stand basket (Figure 6). The system is designed such that the foot pedal and drive unit cable ends can be connected to either console USB cable pigtails for normal operation.

Image /page/5/Picture/4 description: The image shows a medical device on a rolling stand. The device has a screen that displays the word "iotamotion" and a button that says "Start". There is a basket attached to the stand, and a foot pedal is on the floor. The device is likely used for physical therapy or rehabilitation.

Figure 6: iotaSOFT | Control console, footpedal, rolling stand assembly with power supply and USB cable extenders.

Rolling Stand Assembly - The rolling stand (Figure 6) allows the surgeons to position the Control Console at a convenient location within 3 feet of surgical table for visibility and accessibility during the procedure. The rolling stand assembly includes a basket and a separate control console power supply mounting bracket. (NOTE: console, rolling stand, cable connections, and foot pedal are intended for operation outside of the sterile field).

SUMMARY OF NONCLINICAL/BENCH STUDIES

The non-clinical/bench studies conducted on the iotaSOFT System to demonstrate a reasonable assurance of safety and effectiveness of the device are summarized in the sections below.

BIOCOMPATIBILITY

The iotaSOFT™ Drive Unit components and type of contact are described below. The patient-contacting components of the iotaSOFTTM Drive Unit passed all biocompatibility requirements for an external communicating device in limited contact with tissue/bone/dentin with an exposure duration of limited (<24hr). The following

{6}------------------------------------------------

biocompatibility testing was performed to ensure the safety of the iotaSOFT Insertion System:

  • . Cytotoxicity
  • . Irritation
  • Sensitization
  • Acute Systemic Toxicity ●
  • Material Mediated Pyrogenicity ●
  • Bacterial Endotoxin ●
  • . Ethylene Oxide Residuals

The components that come into direct body contact with the patient are the right and left drive housings (in contact with tissue/bone/CSF), the stage base, arm (gooseneck) heat shrink, bone screws, and silicone in contact with tissue/bone. The rest of the disposable device and the reusable components (Control Console and Footpedal) do not come into any direct contact with the patient.

Image /page/6/Figure/9 description: The image shows a diagram of the iotaSOFT Drive System Assembled. The diagram labels the different parts of the system, including the USB cable, housing, stage base, gooseneck, and drive head. The image also shows the preshrunk heat shrink, bone screws and silicone, left drive housing, and right drive housing. The image is labeled as Figure 1 and Figure 2.

Figure 2. iotaSOFT DRIVE Biocompatible Components

The biocompatibility testing is performed according to the ISO 10993-1:2009 Biological evaluation of medical devices and the FDA Guidance Document on Use of International Standard ISO 10993-1. FDA guidance includes a modified table from ISO 10993-1 with additional recommended tests.

SHELF LIFE/STERILITY

Sterilization

The reusable components (control console, power supply, footpedal, extension cable and rolling stand) do not contact the patient tissue and thus do not need to be sterilized.

{7}------------------------------------------------

These reusable components are provided non-sterile to the user. Cleaning/disinfection instructions were provided in the Instructions for Use.

The single use components (including the iotaSOFT Drive Unit) are provided sterile to the user. The sterilization method is ethylene oxide (EO). The sterilization validation of the iotaSOFT Drive Unit was performed via adoption per AAMI TIR28. It was demonstrated that the iotaSOFT Drive Unit can be sterilized effectively to a minimum sterilization assurance level (SAL) of 10-6.

The EO and ECH residuals were measured using the iotaSOFT Drive Unit demonstrating that the residual levels comply to the limits specified in ISO 10993-7.

Packaging and Shelf Life

Packaging validation

A representative number of product samples (95% confidence/90% reliability) were subjected to two full cycles of EO, environmental conditioning (per ASTM D4332) and transportation simulation (per ASTM D4169, DC13, AL1). For the package integrity, seal strength (per ASTM F88) and gross leak (per ASTM F2096) testing were performed. All samples met the acceptance criteria (≥1 1b/in and no bubble leak).

Shelf life validation

The iotaSOFT Insertion System is labeled with a 6-month shelf life. Product samples (95% confidence/90% reliability) were subjected to two full cycles of EO, environmental conditioning (per ASTM D4332) and accelerated aging (per ASTM F1980, 55°C ± 2°C for 19 days). For the package integrity, seal strength (per ASTM F88) and gross leak (per ASTM F2096) testing were performed. All samples met the acceptance criteria (≥1 1b/in and no bubble leak).

Non-Pyrogenicity

Endotoxin testing was performed using the LAL kinetic turbidimetric method (AAMI ST72, USP <85> and USP <161>) with the sterilized iotaSOFT Drive Unit. All samples met the acceptance criteria (< 2.15 EU/device).

ELECTROMAGNETIC CAPABILITY & ELECTROMAGNETIC SAFETY

The iotaSOFT Insertion System was tested for Electrical Safety and Electromagnetic Compatibility using final production units and the following software and firmware versions: Control Console Software: v1.0.3. PCB Firmware: v0.1.1.

The iotaSOFT Insertion System passed all tests for Electrical Safety and was found compliant with AAMI ES60601-1:2005/(R)2012+A1:2012. All tests and compliance investigations were completed with no alternative risk control measures or test methods employed.

{8}------------------------------------------------

The iotaSOFT Insertion System passed all tests for EMC and was found compliant with IEC 60601-1-2:2014. All tests and compliance investigations were completed with no alternative risk control measures or test methods employed. The table below provides a summary of the tests completed.

SpecificationTest DescriptionCompliance
CISPR 11: 2009 + A1: 2010Conducted Emissions-Voltage-Class A, Group 1Compliant
CISPR 11: 2009 + A1: 2010Radiated Emissions - Class A, Group 1Compliant
IEC 61000-4-2: 2008Electrostatic Discharge ImmunityCompliant
IEC 61000-4-3:2006 +Radiated Electromagnetic Field ImmunityCompliant
A1:2007 + A2:2010Immunity to proximity fields from RF wirelesscommunications equipmentCompliant
IEC 61000-4-4: 2012Electrical Fast Transient/Burst ImmunityCompliant
IEC 61000-4-5: 2005Surge ImmunityCompliant
IEC 61000-4-6: 2013Conducted Radio-Frequency ImmunityCompliant
IEC 61000-4-11: 2004Voltage Dips, Interruptions and VariationsCompliant

SOFTWARE

Software/Firmware

Device software and firmware was described, verified and validated. Considerations were taken based on the FDA Guidance Document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". The iotaSOFT™ Insertion System contains two components housing software. the control console (software) and the Drive Unit (firmware). Device communication (1-way) between the control console and the Foot Pedal is via a wired connection. The control console also conducts (2-way) communication with the Drive unit via a wired connection. The device is in no way connected to the internet or IT Network. No patient data or other sensitive information is held or communicated by the device.

The following software and firmware versions were used for software verification:

  • Control Console Software: v1.0.3
  • Drive Unit PCB Firmware: v0.1.1 .

Software documentation, including the following, was provided:

  • Software Description/Summary of Functional Requirements from SRS ●
  • Device Hazard Analysis
  • Traceability Analysis .
  • Verification and Validation Documentation ●
  • . Revision Level History

Cybersecurity

Cybersecurity was addressed during the design and development of the iotaSOFT Insertion System. Cybersecurity considerations were taken based on the FDA Guidance Documents " Content of Premarket Submissions for Management of Cybersecurity in Medical Devices", and "Postmarket Management of Cybersecurity in Medical Devices."

{9}------------------------------------------------

(b) (4)

Figure 1: iotaSOFT System Diagram for Threat Model Analysis

A cybersecurity threat model analysis was performed for the iotaSOFT Insertion System using the STRIDE method. As a result of the threat analysis, the iotaSOFT Insertion System Hazard Analyses identify the hazards, mitigations, and design considerations pertaining to intentional or unintentional cybersecurity risks associated with the device.

Mechanical testing - unit base attachment screws

Bone screw and screwdriver testing (bench testing): The bone screws used in the iotaSOFT Insertion System to secure the unit base to the skull were characterized per ASTM F543-17 Annex A1 for maximum torque. Bone screws were tested to withstand a maximum torque ≥0.2Nm without stalling, stripping, or breaking. The rotational breaking angle, vield torque, vield strength and fracture location were also evaluated at failure, and these parameters exceeded the acceptance criteria with respect to the intended use.

Summary of iotaSOFT performance data (non-clinical and clinical) supporting compatibility with labeled cochlear implants

Four sets of non-clinical performance testing (comprising bench performance, synthetic cochlea insertion force comparison (bench testing), cadaveric comparison, usability/human factors) and one set of clinical performance testing are summarized below. As discussed below, cochlear implant (electrode array) models tested include the following straight arrays:

  • MED-EL Flex 24 .
  • . MED-EL Flex 28
  • . Cochlear Slim Straight
  • Advanced Bionics (AB) HiFocus SlimJ .

{10}------------------------------------------------

Data was provided to support the safe and effective use of iotaSOFT to assist in the insertion of the above cochlear implant electrode arrays (i.e., cochlear implant compatibility validation).

As listed below, iotaMotion conducted four nonclinical validation tests using the MED-EL Flex 24; iotaMotion provided rationale to leverage this testing to support iotaSOFT use with the MED-EL Flex 28 array. iotaMotion asserts the primary difference between the electrode arrays is the additional length of the array portion (4 mm) between the MED-EL Flex 24 and Flex 28, which does not affect compatibility or use with the iotaSOFT. All other dimensions and materials between the MED-EL Flex 24 and Flex 28 arrays/leads are reported by iotaMotion to be identical.

PERFORMANCE TESTING - BENCH

Bench performance testing (verification of device characteristics). Devices having undergone sterilization, accelerated aging and simulated transportation/distribution were tested with the following arrays: Advanced Bionics HiFocus SlimJ, Cochlear Slim Straight, and MedEl Flex 24. For all samples tested:

  • insertion speed was verified to be within 20% of both 0.1 and 1 mm/s (lowest and O highest) insertion rates
  • maximum insertion force (with array) of <300mN was met at both the 0.1mm/s о and 1mm/s insertion rates
  • slip force specification of >80mN was met at both the 1mm/s and 0.1mm/s о insertion rates
  • maximum drive wheel pinch force of <5.6N o
  • O drive head removal force (when decoupling array) of <100mN

Synthetic cochlea insertion force comparison (surgeon bench comparison testing): Devices having undergone sterilization were tested with the following arrays: Advanced Bionics HiFocus SlimJ, Cochlear Slim Straight, and MedEl Flex 24. Both maximum insertion force and insertion force variation are lower on average for iotaSOFT versus manual insertion across multiple surgeons:

  • о The average of maximum insertion force for iotaSOFT assisted electrode array insertions were lower than average of maximum insertion force for manually inserted electrodes for both insertion rates, a 51% and 32% reduction at 0.1mm/s and 1mm/s.respectively.
  • The average insertion force variation for iotaSOFT assisted electrode array o insertions was 78% lower than the average insertion force for manually inserted electrode arrays at 0.1mm/s and 70% lower at 1mm/s.

Representative insertion force versus time profiles and summary tables of maximum insertion force and insertion force variation follow:

{11}------------------------------------------------

Image /page/11/Figure/0 description: The image contains two line graphs comparing insertion force profiles. The graph on the left shows a 'Representative Device-Assisted Insertion Force Profile' with two lines, one for 'Device 1.0 mm/s' and another for 'Device 0.1 mm/s'. The graph on the right shows a 'Representative Manual Insertion Force Profile' with two lines, one for 'Manual 1.0 mm/s' and another for 'Manual 0.1 mm/s'.

Maximum insertion force

InsertionRate(mm/sec)InsertionMethodMean (mN)
0.1iotaSOFT(b) (4)
Manual
1.0iotaSOFT
Manual

Insertion force variation

InsertionRate(mm/sec)InsertionMethodMean (mN/sec)
0.1iotaSOFT(b) (4)
Manual
1.0iotaSOFT
Manual

PERFORMANCE TESTING - CADAVER

Cadaveric comparison testing: iotaSOFT utilized and manual insertions by same surgeon in cadaver were compared. 16 surgeons implanted CIs in 16 cadaver heads. Each surgeon implanted an electrode array with one head using iotaSOFT on one side and manually contralaterally (same CI model, one device per side). MED-EL Flex 24 (n=6), AB HiFocus SlimJ (n=5), Cochlear Slim Straight (n=5) CI arrays were randomly assigned

{12}------------------------------------------------

and implanted per side by the 16 surgeons. During insertion, any array resistance. buckling, or incomplete insertion was noted. 3D x-ray was utilized to determine array intracochlear position. Histology was not assessed. Results across all surgeons and arrays in terms of array positioning summarized in table below:

ScalaTympaniInsertionScala MediaTranslocationScala VestibuliTranslocationTip Fold-OverInsertion Angle(°)
Manual7/16 (44%)7/16 (43%)2/16 (13%)3/16 (19%)309.98 ± 126.6(N=16)
iotaSOFT6/12* (50%)5/12* (42%)1/12* (8%)1/12* (8%)307.4 ± 95.9(N=12)*

*Four samples were excluded post insertion from the comparative analysis due to the electrode inadvertently being removed from the cochlea following insertion either during electrical functional testing or tissue processing prior to scanning.

Results across all surgeons and arrays in terms of array positioning in the cochlea showed similar array placement in the cochlea (including cochlear translocations), tip foldover, and insertion angle.

Post insertion CI device impedance measurements when compared between manual insertion and iotaSOFT appeared unchanged and within normal limits.

Usability testing: Surgeon/Nurse teams (16 teams) performed cadaveric implantations (in 16 cadavers) with iotaSOFT in a simulated use environment (1 cadaver implanted by 1 team). MED-EL Flex24. AB HiFocus SlimJ. and the Cochlear Slim Straight devices were implanted, one device implanted by each team. Data collected: "was task successful?" Y/N responses (e.g., to "lock drive unit into base"), use errors, user comments. Likertscale response to "able to secure base unit to skull." Surgeons were freshly trained with the device prior to study participation and training decay time was included. Surgeon CI experience level ranged from 1 to 30+ years: approximately equal numbers of surgeons with less than 10 years, 10 to 20 years, and greater than 20 years. Between 10 and 180 implantations were performed manually in the year preceding this study, across all surgeons.

  • o Harms associated with use errors were deemed minor: prolonged operative/anesthesia time due to delay during surgery and prolonged operative/anesthesia time due to delay during set up. Mitigated with design changes to final De Novo device involving screws and screwdriver (revalidated) & training update.
  • o For all tasks, use errors deemed not to lead to serious patient harm. Any further mitigations deemed not to reduce the overall or individual risks.
  • o Data show that surgeons with nurse staff support can successfully use iotaSOFT in the hospital environment to aid in placement of the electrode array into the cochlea.

SUMMARY OF CLINICAL TESTING

{13}------------------------------------------------

Clinical testing in 21 subjects was performed by 3 surgeons with varying experience levels. CI models used: Cochlear Slim Straight (n=13), MED-EL Flex 28 (n=6), MED-EL Flex 24 (n=1), 1 AB HiFocus SlimJ. The principal investigator (PI) of the study was involved in the early device design phase and usability testing and completed 15 (of the 21) clinical cases; the other two study investigators were new users of the device and completed 2 cases and 4 cases respectively. The PI is a neurotologist with nearly 40 years experience who implanted ~75 CIs manually in the year prior to this study. The latter two surgeons implanted 25 to 30 Cls each in their careers prior to commencement of this study. Study subjects were aged 25 and older, and 21 iotaSOFT assisted implantations (in 21 subjects) were conducted.

As the IotaSOFT device is intended for cochlear implant array insertion, outcomes related to arrav insertion were collected and outcomes related to the cochlear implant system such as residual hearing measures, vestibular outcomes, and speech recognition metrics were not reported in the clinical study.

Summary of Results:

  • . There were no clinically significant patient safety or outcomes differences based on surgeon years of practice experience nor cochlear implantation experience.
  • Electrode array impedance measures were within normal limits in all subjects as ● of 1 month follow up. Neural response telemetry (NRT) measurements were present in all subjects with normal cochlear anatomy (n=20/21) at post-operative follow up.
  • Adverse events following the proposed surgical protocol were generally as . expected for cochlear implant procedures. (CSF leak occurred in one case when otologic drill was used for pilot hole, drill use is not according to iotaSOFT's instructions for use).
  • Insertion time of the electrode array ranged from 57 sec to 6 min, 1 sec (mean 3 min 15 sec).
  • Post-insertion cochlea view x-ray imaging of the electrode array was satisfactory on all insertions with normal cochlear anatomy when utilizing the iotaSOFT Insertion System.

Pediatric Extrapolation

In this De Novo request, the clinical data summarized above was leveraged to support the benefit versus risk profile of the proposed device use in the pediatric sub-population of children 12 to 21 years old. This skull thickness and mastoid size are considered sufficiently developed for the iotaSOFT Insertion System to be used with patients aged 12 and older. The clinical data are thus considered leverageable to this pediatric sub-population.

1 For one subject, NRT measurements were not obtained due to un-identified temporal bone fracture through the internal auditory canal.

{14}------------------------------------------------

LABELING

A surgical instructions for use is a necessary part of the labeling. This surgical instructions for use includes a list of compatible validated cochlear implants and a summary of supporting nonclinical and clinical performance data.

The contraindications identified in the iotaSOFT labeling contribute to the defined indications for use for the iotaSOFT. In accordance with 21 CFR 807.81(a)(3), FDA has determined that removal or modification of any of the contraindications will require submission of a premarket notification [510(k)], which includes non-clinical performance testing to demonstrate that intended users can use the device as intended with compatible cochlear implants, or rationale for omission of any testing (e.g., applicable data from a comparable model of cochlear implant).

RISKS TO HEALTH

The table below identifies the risks to health that may be associated with use of a powered insertion system for a cochlear implant electrode array and the measures necessary to mitigate these risks.

Identified Risks to HealthMitigation Measures
Risks to health relating to device interface with patient anatomy, including:Damage to skull tissue Damage to dura mater Bone damage Cerebrospinal fluid leak Damage to cochlea; hearing loss, tinnitus, vertigoClinical performance testingUsability testingNon-clinical performance testingLabeling
Cochlear implant insertion failure leading to:Trauma to cochlear structures resulting in residual hearing loss or nerve degeneration Suboptimal array placement (including array rotation) leading to poor hearing performance Failure to disengage from cochlear implant at end of procedure, leading to manual correction and insertionClinical performance testingNon-clinical performance testingUsability testingCochlear implant compatibility validationSoftware verification, validation, and hazard analysisLabeling
Damage to cochlear implant during insertion leading to poor cochlear implant performance and/or compromised implant reliabilityNon-clinical performance testingUsability testingCochlear implant compatibility validationShelf life testingSoftware verification, validation, and hazard analysisLabeling

{15}------------------------------------------------

Adverse tissue reaction, including irritation /inflammation of surgical siteBiocompatibility evaluation
Electromagnetic interference, thermal injury,or electric shockElectrical safety testingElectromagnetic compatibility (EMC) testingLabeling
InfectionSterilization validationShelf life testingLabeling
Excessive operation time leading to increasedexposure to anesthesiaClinical performance testingUsability testingLabeling

SPECIAL CONTROLS

In combination with the general controls of the FD&C Act, the powered insertion system for a cochlear implant electrode array is subject to the following special controls:

  • Clinical performance testing must demonstrate that the device performs as intended (1) under anticipated conditions of use, including evaluation of all adverse events.
  • Non-clinical performance testing must demonstrate that the device performs as (2) intended under anticipated conditions of use. Testing must include:
    • Verification of cochlear implant attachment force, release force and insertion (i) speed;
    • (ii) Testing to demonstrate the device does not damage or degrade the cochlear implant (including the lead and array portions of the cochlear implant);
    • (iii) Comparison testing with manual insertion to evaluate:
      • Differences in cochlear implant array insertion force associated with use of (A) the device; and
      • Intracochlear placement of the cochlear implant array (intended scala (B) placement and array insertion depth, together with minimal array tip foldover and cochlear scala translocation).
  • Usability testing in a simulated hospital environment with an anatomically relevant (3) model (e.g., cadaver testing) that evaluates the following:
    • Successful use to aid in placement of the electrode array into the cochlea; and (i)
    • Harms caused by use errors observed. (ii)
  • (4) Changes in cochlear implant compatibility are determined to significantly affect the safety or effectiveness of the device and must be validated through performance testing or a rationale for omission of any testing.
  • The patient-contacting components of the device must be demonstrated to be (ર) biocompatible.
  • (6) Performance testing must demonstrate the electromagnetic compatibility (EMC), electrical safety, and thermal safety of the device.
  • (7) The patient-contacting components of the device must be demonstrated to be sterile and non-pyrogenic.

{16}------------------------------------------------

  • (8) Performance testing must support the shelf life of device components provided sterile by demonstrating continued sterility, package integrity, and device functionality over the labeled shelf life.
  • Software verification, validation, and hazard analysis must be performed for any (9) software components of the device.
  • (10) Labeling must include:
    • The recommended training for the safe use of the device; (i)
    • (ii) Summary of the relevant clinical and non-clinical testing pertinent to use of the device with compatible electrode arrays; and
    • A shelf life. (iii)

BENEFIT-RISK DETERMINATION

CI manufacturer surgical guides recommendations vary. They may include discussion of speed of insertion and insertion forces and recommentation. Insertion utilizing a powered insertion device incorporates a hybrid approach. Utilization of the iotaSOFT device should be performed in consideration of the CI surgical instructions. Stabilized, steady, and slow insertions are recognized in the cochlear implant industry recommended surgical practices and academic literature recognizes these insertions as surgical approaches which decrease the risk of additional hearing loss or vestibular symptoms from inherently uncontrolled, variable, and excessively forceful insertions. 1 ,2,3,4,5.6,7

The iotaSOFT Insertion System stabilizes the electrode and permits slower constant motion (down to 0.1mm/sec insertion speed) and provides a more controlled insertion than is possible with manual, human capabilities. A study by Kesler et al. found that the lower limit of a manual, constant forward insertion motion by a CI surgeon to be a speed of 0.87 mm/sec. Y Lower maximum insertion force and lower insertion force variation were demonstrated for iotaSOFT versus manual insertion in synthetic cochlear model testing. Additionally, iotaSOFT holds and stabilizes the electrode during insertion to free up the surgeon's hand to guide a cochlear implant electrode array with manual tools. This design aspect of iotaSOFT attempts to reduce the risk of unwanted movement of the electrode array such as tremor or inadvertent manual surgeon movements. However, iotaSOFT versus manual insertions were similar per comparative cadaveric assessment in terms of array placement in the cochlea (including cochlear translocations), tip foldover, and insertion angle. Therefore, the benefits that were observed preclinically for iotaSOFT were restricted to more controlled insertion and lower insertion forces (and insertion force variation) than manual insertion.

Considering the clinical performance testing of 21 iotaSOFT implantations by 3 physicians, as of 1-month post activation, electrode impedances and neural telemetry measures were considered normal. Adverse events following the proposed surgical were generally as expected for cochlear implant surgeries2 and as expected when considered relative to manual cochlear implantation, e.g., there were no vestibular related adverse events outside of the mild and

2 CSF leak occurred in one case when otologic drill was used for pilot hole (drill use is not according to iotaSOFT's instructions for use).

{17}------------------------------------------------

temporary symptoms as reported in cochlear implantations, in general. The following limitations are associated with the clinical study:

  • . Direct comparison data to manual insertions was not included in the scope of the study.
  • Preservation of residual hearing was not within the scope of this clinical confirmatory . study and 1-month postoperative hearing threshold data were not reported.
  • Clinical study endpoint interval was at 1-month, postoperative; therefore, there was no ● long-term data provided.

Direct clinical benefit was not observed in the clinical and cadaveric studies of powered insertion as compared to manual insertion, and therefore, uncertainty exists in terms of the actual clinical benefit of powered insertion. However, the clinical confirmatory study measures of safety and effectiveness did indicate patient clinical performance using iotaSOFT for electrode arrav insertion was as expected and consistent with that seen at 1-month followup with manual insertions.

Risks associated with the use of iotaSOFT can be considered mitigated with surgeon training and labeling for indicated patients (12 years and older) for the 4 tested and validated cochlear implant arrays: Cochlear Slim Straight, MED-EL Flex 28, MED-EL Flex 24, Advanced Bionics HiFocus SlimJ. When used as intended, the risks due to use of iotaSOFT appeared to be no greater in severity or prevalence than risks associated with manual insertion.

In conclusion, the benefit/risk profile for the iotaSOFT Insertion System is considered favorable for the following reasons:

  • . Regarding clinical risk, adverse events/unmitigated risks related to the IotaSOFT insertion device use were not observed in the clinical and usability studies.
  • As evidenced by benchtop performance characterizations, the iotaSOFT permits ● slower, more controlled insertions with a reduction in both maximum insertion force and insertion force variation of cochlear implant arrays versus manual insertion.
  • The benefit of the device design related to slow, controlled insertion and the potential for reduced intracochlear trauma is supported by cochlear implant industry recommended surgical practices and academic literature.

References:

  • Lehnhardt E. Intracochlear placement of cochlear implant electrodes in soft surgery 1. technique. HNO. 1993;41:356-359.
  • Rajan GP, Kontorinis G. Kuthubutheen J. The Effects of Insertion Speed on Inner Ear 2. Function during Cochlear Implantation: A Comparison Study. Audiol Neurotol. 2013;18(1):17-22.
    1. Mirsalehi M, Rau TS, Harbach L, et al. Insertion forces and intracochlear trauma in temporal bone specimens implanted with a straight atraumatic electrode array. Eur Arch Oto-Rhino-Larvngology, 2017:274(5):2131-2140.

{18}------------------------------------------------

    1. De Seta D, Torres R, Russo FY, et al. Damage to inner ear structure during cochlear implantation: Correlation between insertion force and radio-histological findings in temporal bone specimens. Hear Res. 2017;344:90-97. doi:https://doi.org/10.1016/j.heares.2016.11.002
  • న. Eshraghi AA, Van De Water TR. Cochlear implantation trauma and noise-induced hearing loss: Apoptosis and therapeutic strategies. Anat Rec Part A Discov Mol Cell Evol Biol. 2006;288A(4):473-481. doi:10.1002/ar.a.20305
  • Kontorinis G, Lenarz T, Stöver T, Paasche G. Impact of the Insertion Speed of Cochlear 6. Implant Electrodes on the Insertion Forces. Otol Neurotol. 2011;32(4).
    1. Roland PS, Wright CG. Surgical Aspects of Cochlear Implantation: Mechanisms of Insertional Trauma. In: Advances in Oto-Rhino-Laryngology. Vol 64. ; 2006:11-30. doi:10.1159/000094642
  • Kesler K, Dillon NP, Fichera L, Labadie RF. Human Kinematics of Cochlear Implant 8. Surgery: An Investigation of Insertion Micro-Motions and Speed Limitations. Otolaryngol Neck Surg. 2017:157(3):493-498. doi:10.1177/0194599817704391

Patient Perspectives

This submission did not include specific information on patient perspectives for this device.

Benefit/Risk Conclusion

In conclusion, given the available information above, for the following indication statement:

The iotaSOFT™ Insertion System is intended to aid the surgeon in placement of cochlear implant electrode arrays into a radiographically normal cochlea by controlling the speed of implant insertion. The iotaSOFT Insertion System is intended for use in cochlear implant patients ages 12 years and older during cochlear implant procedures using either a round window or cochleostomy approach.

The probable benefits outweigh the probable risks for the iotaSOFT Insertion System. The device provides benefits and the risks can be mitigated by the use of general controls and the identified special controls.

CONCLUSION

The De Novo request for the iotaSOFT Insertion System is granted and the device is classified as follows:

Product Code: OOH Device Type: Powered insertion system for a cochlear implant electrode array Regulation Number: 21 CFR 874.4450 Class: II

§ 874.4450 Powered insertion system for a cochlear implant electrode array.

(a)
Identification. A powered insertion system for a cochlear implant electrode array is a prescription device used to assist in placing an electrode array into the cochlea.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including evaluation of all adverse events.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Testing must include:
(i) Verification of cochlear implant attachment force, release force, and insertion speed;
(ii) Testing to demonstrate the device does not damage or degrade the cochlear implant (including the lead and array portions of the cochlear implant); and
(iii) Comparison testing with manual insertion to evaluate:
(A) Differences in cochlear implant array insertion force associated with use of the device; and
(B) Intracochlear placement of the cochlear implant array (intended scala placement and array insertion depth, together with minimal array tip foldover and cochlear scala translocation).
(3) Usability testing in a simulated hospital environment with an anatomically relevant model (
e.g., cadaver testing) that evaluates the following:(i) Successful use to aid in placement of the electrode array into the cochlea; and
(ii) Harms caused by use errors observed.
(4) Changes in cochlear implant compatibility are determined to significantly affect the safety or effectiveness of the device and must be validated through performance testing or a rationale for omission of any testing.
(5) The patient-contacting components of the device must be demonstrated to be biocompatible.
(6) Performance testing must demonstrate the electromagnetic compatibility, electrical safety, and thermal safety of the device.
(7) The patient-contacting components of the device must be demonstrated to be sterile and non-pyrogenic.
(8) Performance testing must support the shelf life of device components provided sterile by demonstrating continued sterility, package integrity, and device functionality over the labeled shelf life.
(9) Software verification, validation, and hazard analysis must be performed for any software components of the device.
(10) Labeling must include:
(i) The recommended training for the safe use of the device;
(ii) Summary of the relevant clinical and non-clinical testing pertinent to use of the device with compatible electrode arrays; and
(iii) A shelf life.