(386 days)
Not Found
Not Found
No
The algorithm uses linear discriminant analysis and logistic regression, which are traditional statistical methods, not typically classified as AI/ML in the context of modern medical device regulation.
No.
The CADScor System is a diagnostic device that records heart sounds and vibrations to calculate a CAD risk score, aiding in diagnosis rather than providing therapy.
Yes
The intended use explicitly states "as an aid in cardiac analysis and diagnosis." Additionally, the device calculates a patient-specific score indicating the risk of coronary stenosis, which directly contributes to a diagnostic process.
No
The device description explicitly details hardware components including a docking station, acoustic recording sensor (with display, microphone, and on/off button), and a patch with an RFID tag. While software is used for calculation, the system relies on these physical components for data acquisition and display.
Based on the provided information, the CADScor System is likely an In Vitro Diagnostic (IVD) device, although it doesn't fit the most common definition of an IVD which typically involves testing samples outside the body (in vitro).
Here's why:
- Intended Use: The intended use is to provide a patient-specific score "indicating the risk of presence of coronary stenosis, as an aid in cardiac analysis and diagnosis." This clearly falls under the realm of providing diagnostic information.
- Device Description: The device records acoustic noise and uses computational processing to determine a CAD risk score and classification. This process of analyzing biological signals (heart sounds and vibrations) to generate diagnostic information aligns with the purpose of IVDs.
- Diagnostic Aid: The device is explicitly described as an "aid in cardiac analysis and diagnosis" and helps clinicians determine if "further testing is needed for a CAD diagnosis." This function is characteristic of diagnostic devices.
- Performance Metrics: The provided performance metrics (sensitivity, specificity, PPV, NPV) are standard for evaluating the performance of diagnostic tests.
Why it might not fit the traditional IVD definition perfectly:
The traditional definition of an IVD often focuses on testing biological samples like blood, urine, or tissue in vitro (in glass, or outside the body). The CADScor System records signals directly from the patient's body in vivo.
However, regulatory bodies often broaden the definition of IVD to include devices that analyze biological signals or data from the body to provide diagnostic information. The key is that the device is used to provide information about a patient's health status or condition, which is then used for diagnosis or to guide further diagnostic procedures.
Conclusion:
Given its intended use as an aid in diagnosis, its analysis of biological signals (acoustic noise from the body), and the reporting of diagnostic-like metrics (risk score and classification), the CADScor System functions as a diagnostic tool. While it doesn't test samples in vitro in the traditional sense, it is highly likely to be regulated as an IVD or a similar type of diagnostic medical device due to its role in providing information for diagnosis.
To be absolutely certain, one would need to consult the specific regulatory definitions of IVD in the relevant jurisdiction (e.g., FDA in the US, CE marking in Europe). However, based on the provided description, it strongly aligns with the functional definition of a diagnostic device.
N/A
Intended Use / Indications for Use
The intended use of the CADScor System is to record heart sounds, murmurs and vibration for calculation of a patient specific score, indicating the risk of presence of coronary stenosis, as an aid in cardiac analysis and diagnosis.
Product codes
OMW
Device Description
The CADScor System is a digital stethoscope for recording and quantifying acoustic noise from micro-turbulence originating from stenosis of the coronary arteries. Computational processing of a recording obtained from the chest surface of a patient is used to determine a coronary artery disease (CAD) risk score and associated risk classification.
Three components make up the CADScor System: the Docking station, the Acoustic Recording Sensor and the Patch. The docking station is responsible for charging and qualifying the sensor. These functionalities only happen when the Acoustic Recording Sensor is in direct contact with the docking station. An LED light indicates the status of the sensor. The Acoustic Recording Sensor contains a display, a microphone and an on/off button. The heart sound is recorded through a microphone and once the CAD risk score is calculated, it appears on the display screen. The patch is directly in contact with the patient's chest at the fourth left inter costal space (IC4-L) region. The patch provides a tight seal between the chest surface and the microphone to reduce the external noise that is picked up by the sensor. The patch is for single use only and is discarded after a CAD-score is calculated. A radiofrequency identification (RFID) tag in the patch is used to prevent the patch from being used more than once.
An algorithm is used to calculate a coronary artery disease (CAD) risk score. The algorithm analyzes 8 features of the acoustic recording and three clinical features that are manually inputted into the device by the clinician. The clinical inputs are age, gender and presence of hypertension. The eight acoustic features are put through a linear discriminate analysis and then put through a logistic regression with the clinical features. The risk score and associated categorization color are displayed on the acoustic sensor screen. There are two risk categories: low risk and elevated risk (Figure 2). Clinicians can use the score and categorization to aid in a clinician's determination of whether further testing is needed for a CAD diagnosis.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
chest surface of a patient, fourth left inter costal space (IC4-L) region
Indicated Patient Age Range
over the age of years
Intended User / Care Setting
clinician's office
Description of the training set, sample size, data source, and annotation protocol
The device was tested on 100 patients, who were independent from the algorithm training set.
Description of the test set, sample size, data source, and annotation protocol
A prospective, multicenter, randomized clinical study was conducted to support the safety and effectiveness of the CADScor System for the proposed indication for use. Subjects were over the age of years and were suspected of ischemic heart disease. To determine the subject's diagnosis, which was compared to the result of the CADScor System, the patient underwent coronary computed tomography angiography (CCTA). Subjects who are diagnosed via CCTA to have CAD are defined to have at least one significant stenosis, which has a minimum 11% diameter reduction (DS) of a coronary vessel. Subjects who are classified as Non-CAD are defined as having CT calcium score of zero and no CT stenosis. The primary objective of this study is to determine specificity and sensitivity of the CAD-score in patients with lowintermediate risk for ischemic heart disease.
The device was tested on 100 patients.
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
A prospective, multicenter, randomized clinical study was conducted to support the safety and effectiveness of the CADScor System for the proposed indication for use. The device was tested on 100 patients, who were independent from the algorithm training set. The patient population had a 10.7% prevalence of coronary artery disease. The results show that the sensitivity, specificity and area under the curve for the testing group are 87.5%, 37.5% and 71.5%, respectively, for the threshold score of 20. The positive value (PPV) and negative predictive value (NPV) were reported to be 14.4% and 96.2%, respectively, With the NPV of 96.2%, the Agency has determined that this device could be beneficial for clinicians to use as a "rule-out" device for patients over the age of thears who are presenting with symptoms of ischemic heart disease.
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
sensitivity: 87.5%
specificity: 37.5%
area under the curve: 71.5%
positive value (PPV): 14.4%
negative predictive value (NPV): 96.2%
Predicate Device(s)
Not Found
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 870.1420 Coronary artery disease risk indicator using acoustic heart signals.
(a)
Identification. A coronary artery disease risk indicator using acoustic heart signals is a device that records heart sounds including murmurs and vibrations to calculate a patient-specific risk of presence of coronary artery disease, as an aid in cardiac analysis and diagnosis.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Clinical performance testing must fulfill the following:
(i) Testing must include a discussion of the patient population and any statistical techniques used for analyzing the data; and
(ii) Testing must be representative of the intended use population for the device. Any selection criteria or sample limitations must be fully described and justified.
(2) Acoustic performance testing must evaluate microphone sensitivity, sound acquisition bandwidth, and amplitude accuracy. The acoustic sensor specifications and mechanism used to capture heart sounds must be described.
(3) A scientific justification for the validity of the algorithm(s) must be provided. This justification must fulfill the following:
(i) All inputs and outputs of the algorithm must be fully described;
(ii) The procedure for segmenting, characterizing, and classifying the acoustic signal must be fully described; and
(iii) This justification must include verification of the algorithm calculations and validation using an independent data set.
(4) The patient-contacting components of the device must be demonstrated to be biocompatible.
(5) Software verification, validation, and hazard analysis must be performed.
(6) Human factors/usability testing must demonstrate that the user can correctly use the device, including device placement, based solely on reading the directions for use.
(7) Performance data must demonstrate the electromagnetic compatibility and electrical safety of the device.
(8) Labeling must include the following:
(i) A description of what the device measures and outputs to the user;
(ii) Instructions for proper placement of the device;
(iii) Instructions on care and cleaning of the device;
(iv) Warnings identifying sensor acquisition factors that may impact measurement results and instructions for mitigating these factors; and
(v) The expected performance of the device for all intended use populations and environments.
0
DE NOVO CLASSIFICATION REQUEST FOR CADSCOR SYSTEM
REGULATORY INFORMATION
FDA identifies this generic type of device as:
Coronary artery disease risk indicator using acoustic heart signals. A coronary artery disease risk indicator using acoustic heart signals is a device that records heart sounds including murmurs and vibrations to calculate a patient-specific risk of presence of coronary artery disease, as an aid in cardiac analysis and diagnosis.
NEW REGULATION NUMBER: 21 CFR 870.1420
CLASSIFICATION: Class II
PRODUCT CODE: OMW
BACKGROUND
DEVICE NAME: CADScor System
SUBMISSION NUMBER: DEN190047
DATE DE NOVO RECEIVED: November 4, 2019
SPONSOR INFORMATION:
Acarix A/S Mette Munch Ryvangs Alle 81-83 DK-2900 Hellerup Denmark
INDICATIONS FOR USE
The CADScor System is indicated as follows:
The intended use of the CADScor System is to record heart sounds, murmurs and vibration for calculation of a patient specific score, indicating the risk of presence of coronary stenosis, as an aid in cardiac analysis and diagnosis.
LIMITATIONS
The sale, distribution, and use of the CADScor System are restricted to prescription use in accordance with 21 CFR 801.109.
1
PLEASE REFER TO THE LABELING FOR A COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.
DEVICE DESCRIPTION
The CADScor System is a digital stethoscope for recording and quantifying acoustic noise from micro-turbulence originating from stenosis of the coronary arteries. Computational processing of a recording obtained from the chest surface of a patient is used to determine a coronary artery disease (CAD) risk score and associated risk classification.
Three components make up the CADScor System: the Docking station, the Acoustic Recording Sensor and the Patch. The docking station is responsible for charging and qualifying the sensor. These functionalities only happen when the Acoustic Recording Sensor is in direct contact with the docking station. An LED light indicates the status of the sensor. The Acoustic Recording Sensor contains a display, a microphone and an on/off button. The heart sound is recorded through a microphone and once the CAD risk score is calculated, it appears on the display screen. The patch is directly in contact with the patient's chest at the fourth left inter costal space (IC4-L) region. The patch provides a tight seal between the chest surface and the microphone to reduce the external noise that is picked up by the sensor. The patch is for single use only and is discarded after a CAD-score is calculated. A radiofrequency identification (RFID) tag in the patch is used to prevent the patch from being used more than once.
Image /page/1/Picture/4 description: The image shows a gray, rectangular object with a white label on top, sitting on a silver platform. Next to it is a white, circular object with a small handle. The label on the rectangular object appears to have some text on it, but it is not clear enough to read. The silver platform has a slight incline, and the white circular object is slightly out of focus.
Figure 1. The three components of the CADScor System: Dacking station, Acoustic recording station, and Patch
An algorithm is used to calculate a coronary artery disease (CAD) risk score. The algorithm analyzes 8 features of the acoustic recording and three clinical features that are manually inputted into the device by the clinician. The clinical inputs are age, gender and presence of hypertension. The eight acoustic features are put through a linear discriminate analysis and then put through a logistic regression with the clinical features. The risk score and associated categorization color are displayed on the acoustic sensor screen. There are two risk categories: low risk and elevated risk (Figure 2). Clinicians can use the score and categorization to aid in a clinician's determination of whether further testing is needed for a CAD diagnosis.
CAD-score ≤20 | CAD-score >20 |
---|---|
Low risk | Elevated risk |
Figure 2. CAD score categorization | |||||
---|---|---|---|---|---|
-- | -- | -- | -- | ------------------------------------ | -- |
2
SUMMARY OF NONCLINICAL/BENCH STUDIES
BIOCOMPATIBILITY/MATERIALS
Biocompatibility testing was performed according to FDA Guidance "Use of International Standard ISO 10993-1, Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process." The sponsor tested cytotoxicity, skin sensitization and skin irritation of the patch. The patch is intended to be on the patient for no longer than 45 minutes; therefore, the sponsor tested according to the guidelines for short-term, intact skin contact.
Test | Purpose | Method | Acceptance Criteria | Results | |
---|---|---|---|---|---|
Cytotoxicity | To identify any | ||||
cytotoxic effects to the | |||||
patient | ISO 10993-1 | No evidence of | |||
toxicity or cell lysis | PASS | ||||
Sensitization | To identify tissue | ||||
reactions from the patch | ISO 10993-1 | No evidence of | |||
tissue reaction | PASS | ||||
Irritation | To identify tissue | ||||
irritation | ISO 10993-1 | No evidence of | |||
tissue irritation | PASS |
ELECTROMAGNETIC COMPATIBILITY & ELECTRICAL SAFETY
The CADScor System conforms to the FDA-recognized standards for basic safety and essential performance of Medical Electrical Equipment. These standards include:
- IEC 60601-1-2:2014- Medical Electrical Equipment, Part 1-2: General . requirements for safety- Collateral Standard: Electromagnetic compatibility-Requirements and tests
- IEC 60601-1:2006 + A1:2013- Medical Electrical Equipment, Part 1: General . requirements for basic safety and essential performance
SOFTWARE
Software documentation was provided in accordance with the FDA Guidance Document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," (issued May 11, 2005) for a Moderate Level of Concern (LOC). A Moderate LOC is deemed appropriate as malfunction of the device software or a latent design flaw in the device software may lead to an erroneous diagnosis or a delay in the delivery of appropriate medical care, which would likely result in minor injury but would likely not result in serious injury or death due to the availability of patient history, and vital signs to aid in the clinician's diagnosis.
3
Verification and validation testing were performed to confirm that the device software met the software requirements.
Algorithm unit testing was performed to demonstrate that the device software met the software requirements. This testing was performed using privately collected patient data. Additional details are provided in the "Summary of Clinical Information" section.
PERFORMANCE TESTING - BENCH
Acoustic Performance Testing
Acoustic Performance testing was performed to determine that both the heart and ambient noise microphones function as expected and can correctly record the frequencies that are clinically meaningful for this device.
Test | Purpose | Method | Acceptance Criteria | Results |
---|---|---|---|---|
Free Field Acoustic | ||||
Testing | Ensure the | |||
manufactured parts of | ||||
the CADScor systems | ||||
are performing | ||||
according to test | ||||
specifications | (b) samples were | |||
measured at (b)(4) | ||||
up to | ||||
(b)(4) Hz using a | ||||
loud speaker as a | ||||
source transmitter | Heart and ambient | |||
microphone | ||||
sensitive within 5dB | ||||
throughout the | ||||
range of frequencies | ||||
tested | PASS | |||
Microphone Sensitivity | ||||
Testing | Test microphone | |||
sensitivity before and | ||||
after production | Set-up: Functional | |||
generator for output | ||||
control, loud | ||||
speaker and B&K | ||||
(b)(4) sound level | ||||
meter with receiving | ||||
microphone | Sensitivity of the | |||
heart and ambient | ||||
microphones are | ||||
stable at 5, 10, 20, | ||||
140 Hz before and | ||||
after the plotting | ||||
process | PASS | |||
Acoustic Verification | ||||
Testing | Ensure the | |||
manufactured CADScor | ||||
Sensor functionality | ||||
and performance | The calibration | |||
frequency of (b) | ||||
(b)(4) | ||||
(b)(4) and (b)(4) Hz | ||||
were tested to | ||||
determine if the | ||||
sensor and ambient | ||||
noise detectors | ||||
respond. | ||||
Additionally, broad | ||||
band white noise | ||||
source signal was | ||||
tested. | The sensor and | |||
ambient noise | ||||
detectors respond | ||||
uniformly to all | ||||
seven frequencies. | PASS |
Mechanical Testing
Test | Purpose | Method | Acceptance Criteria | Results |
---|---|---|---|---|
Mechanical Testing- | ||||
Design Verification | Verify the specified | |||
geometric proportions | ||||
and mechanical lifetime | Force Tests. | |||
Dimensions of each | Manufactured to | |||
correct dimension | ||||
specifications | PASS |
4
| | for the following
components of the
CADScor system:
mechanical couplers,
electrical contacts,
sensor push button and
dimensions and weight | component of the
device | Force to remove
sensor from docking
station is 5N.
Qualification
successful before
and after 1000
insertions of the
sensor onto the
docking station | |
-- | -------------------------------------------------------------------------------------------------------------------------------------------------------------- | ---------------------------- | ----------------------------------------------------------------------------------------------------------------------------------------------------------------------------- | -- |
---|
USABILITY TESTING
Usability Testing was provided in accordance with the FDA Guidance Document, "Applying Human Factors and Usability Engineering to Medical Devices- Guidance for Industry and Food and Drug Administration Staff" (issued February 02, 2016).
IDE C subjects belonging to a user group of medical doctors/physicians with valid medical authorization, subjects belonging to a user group of registered nurses with valid authorization and @ @04) subjects belonging to a user group of service technicians (non-clinical user group) participated in the usability testing. The testing was performed in a simulated- use environment using a sequence of tasks including but not limited to correct patch positioning of the device on the patient's chest. entering clinical features and other inputs correctly, and reading the display text and symbols/icons.
SUMMARY OF CLINICAL INFORMATION
A prospective, multicenter, randomized clinical study was conducted to support the safety and effectiveness of the CADScor System for the proposed indication for use. Subjects were over the age of years and were suspected of ischemic heart disease. To determine the subject's diagnosis, which was compared to the result of the CADScor System, the patient underwent coronary computed tomography angiography (CCTA). Subjects who are diagnosed via CCTA to have CAD are defined to have at least one significant stenosis, which has a minimum 11% diameter reduction (DS) of a coronary vessel. Subjects who are classified as Non-CAD are defined as having CT calcium score of zero and no CT stenosis. The primary objective of this study is to determine specificity and sensitivity of the CAD-score in patients with lowintermediate risk for ischemic heart disease.
The device was tested on 100 patients, who were independent from the algorithm training set. The patient population had a 10.7% prevalence of coronary artery disease. The results show that the sensitivity, specificity and area under the curve for the testing group are 87.5%, 37.5% and 71.5%, respectively, for the threshold score of 20. The positive value (PPV) and negative predictive value (NPV) were reported to be 14.4% and 96.2%, respectively, With the NPV of 96.2%, the Agency has determined that this device could be beneficial for clinicians to use as a "rule-out" device for patients over the age of thears who are presenting with symptoms of ischemic heart disease.
5
Pediatric Extrapolation
In this De Novo request, existing clinical data were not leveraged to support the use of the device in a pediatric patient population.
LABELING
The labeling of the device satisfies the special controls listed below:
- A description of what the device measures and outputs to the user; .
- Instructions for proper placement of the device; .
- Instructions on care and cleaning of the device; .
- . Warnings identifying sensor acquisition factors that may impact measurement results and instructions for mitigating these factors
- . The expected performance of the device for all intended use populations and environments.
RISKS TO HEALTH
The table below identifies the risks to health that may be associated with use of the coronary artery disease risk indicator using acoustic heart signals and the measures necessary to mitigate these risks.
Identified Risks to Health | Mitigation Measures |
---|---|
Adverse tissue reaction | Biocompatibility evaluation, |
Labeling, and | |
Usability testing | |
Skin burn/irritation | Electrical safety testing, and |
Electromagnetic compatibility testing | |
False positive leading to | |
unnecessary medical | |
procedures | Software verification, validation, and hazard analysis; |
Usability testing; | |
Acoustic performance testing | |
Clinical performance testing; and | |
Labeling | |
False negative leading to failure | |
to detect coronary artery | |
disease | Software verification, validation, and hazard analysis; |
Usability testing; | |
Acoustic performance testing; | |
Clinical performance testing; and | |
Labeling | |
Delay in calculation due to | |
device failure resulting in a | |
delay of treatment | Software verification, validation, and hazard analysis; |
Clinical performance testing; | |
Usability testing; | |
Acoustic performance testing; and | |
Labeling |
6
SPECIAL CONTROLS
In combination with the general controls of the FD&C Act. the coronary artery disease risk indicator using acoustic heart signals is subject to the following special controls:
- (1) Clinical performance testing must fulfill the following:
- a. Testing must include a discussion of the patient population and any statistical techniques used for analyzing the data; and
- b. Testing must be representative of the intended use population for the device. Any selection criteria or sample limitations must be fully described and justified.
- (2) Acoustic performance testing must evaluate microphone sensitivity, sound acquisition bandwidth, and amplitude accuracy. The acoustic sensor specifications and mechanism used to capture heart sounds must be described.
- (3) A scientific justification for the validity of the algorithm(s) must be provided. This justification must fulfill the following:
- a. All inputs and outputs of the algorithm must be fully described;
- b. The procedure for segmenting, characterizing and classifying the acoustic signal must be fully described; and
- c. This justification must include verification of the algorithm calculations and validation using an independent data set.
- (4) The patient-contacting components of the device must be demonstrated to be biocompatible.
- (5) Software verification, validation, and hazard analysis must be performed.
- (6) Human factors/usability testing must demonstrate that the user can correctly use the device, including device placement, based solely on reading the directions for use.
- (7) Performance data must demonstrate the electromagnetic compatibility (EMC) and electrical safety of the device.
- (8) Labeling must include the following:
- a. A description of what the device measures and outputs to the user;
- b. Instructions for proper placement of the device:
- c. Instructions on care and cleaning of the device;
- d. Warnings identifying sensor acquisition factors that may impact measurement results and instructions for mitigating these factors; and
- e. The expected performance of the device for all intended use populations and environments.
BENEFIT-RISK DETERMINATION
The risks associated with use of the device are based on data collected in a clinical study described above. One risk to a patient is myocardial infarction. This poses a risk to patients when they receive a false negative result where they are classified as "low risk" but may have needed to go through further testing to determine that their symptoms are from CAD. The probability of this occurring is low because the negative predictive value (NPV) is 96.2%. Additionally, the clinician is not intended to use this device as a sole diagnostic decision maker and are instructed in the labeling to recommend to patients categorized in the low risk group to seek medical attention if their symptoms persist or worsen.
7
A risk to a patient categorized in the "elevated risk" group can be excessive testing but are found to be "low risk" after further testing. With a relatively low specificity, it is expected that a false positive result could lead to unnecessary testing. However, without this device, it is likely that the entire patient population would undergo other tests, i.e. stress tests.
Additionally, tissue irritation could occur from the patch: however, the probability of this occurring is very low because no adverse events occurred during the biocompatibility testing. and labeling instructs the clinician to not leave the patient's chest for longer than 45 minutes.
The probable benefits of the device are also based on data collected in a clinical study as described above. By using the CADScor system to evaluate the patient's risk of Coronary Artery Disease, a "low risk" score can help the clinician identify patients who are at low risk of CAD and avoid unnecessary additional testing, which come with their own respective risks. Additionally, this device can be used in the clinician's office to provide results in a timely manner. The probability of correctly identifying a patient who is low risk for CAD is high based on the NPV of 96.2%.
Patient Perspectives
This submission did not include specific information on patient perspectives for this device.
Benefit/Risk Conclusion
In conclusion, given the available information above, for the following indication statement:
The intended use of the CADScor System is to record heart sounds, murmurs and vibration for calculation of a patient specific score, indicating the risk of presence of coronary stenosis, as an aid in cardiac analysis and diagnosis.
The probable benefits outweigh the probable risks for the CADScor System. The device provides benefits and the risks can be mitigated by the use of general controls and the identified special controls.
CONCLUSION
The De Novo request for the CADScor System is granted and the device is classified as follows:
Product Code: OMW Device Type: Coronary artery disease risk indicator using acoustic heart signals Regulation Number: 21 CFR 870.1420 Class: II