Intended Use

SafeBreak® Vascular is intended to separate when excessive tension is exerted across a peripheral IV administration set. When SafeBreak® Vascular separates, fluid flow is stopped from the infusion pump and blood flow is stopped from the patient's IV catheter.

SafeBreak® Vascular is intended to aid in reduction of peripheral IV mechanical complications requiring IV replacement.

SafeBreak® Vascular is intended to be used on peripheral IV catheters in adults and adolescent populations eighteen (18) years of age and older receiving intermittent or continuous infusions with an electronic pump.

Device Description

SafeBreak® Vascular is used with peripheral intravenous (IV) catheters and installed for in-line use with an IV administration set. It connects between the IV administration tubing and the patient's peripheral IV access device. When excessive force is applied to the IV administration set, SafeBreak® Vascular separates at a force lower than the force required to dislodge the peripheral IV catheter or disrupt the IV securement dressing (4 ± 1 1bf). When SafeBreak® Vascular separates, fluid flow is stopped from the infusion pump, and blood flow is stopped from the patient's IV catheter. SafeBreak® Vascular may be used for a maximum of seven days.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the studies that prove the SafeBreak® Vascular device meets those criteria, based on the provided text:

1. Acceptance Criteria and Reported Device Performance

Acceptance Criteria / Performance CharacteristicDescription / Acceptance CriteriaReported Device Performance
Primary Clinical Endpoint: Delays of TherapyTo evaluate if use of the SafeBreak® Vascular device resulted in a delay of therapy that was non-inferior to the delays in the Control Group by comparing the total delay of therapy per 24 hours per subject between groups (Control and SafeBreak®). Non-inferiority margin: 19 minutes.Met. The upper confidence limit (UCL) of the two-sided 95% confidence interval (CI) of the treatment difference (SafeBreak® minus Control) was within the non-inferiority margin (19 minutes). This was true for both the Intention to Treat (ITT) and Per Protocol (PP) populations, demonstrating that the total delay of subjects in the SafeBreak® Group was non-inferior to the Control Group. The median total delay of therapy time per 24 hours was 0 minutes in both groups for the primary analysis. When considering only subjects with events, the median total delay was lower in the SafeBreak® Group ((b)(4) minutes) compared to the Control Group ((b)(4) minutes), although these sub-analyses were underpowered for non-inferiority testing.
Secondary Clinical Endpoint: Reduction in Mechanical ComplicationsTo compare the number of peripheral IV mechanical complications (dislodgement, infiltration, occlusion, phlebitis) and need for IV catheter replacements between the Control and SafeBreak® groups.Met. There were fewer peripheral IV mechanical complications requiring IV catheter replacement in the SafeBreak® Group (13 mechanical complications in 70 subjects) compared to the Control Group (25 mechanical complications in 69 subjects). This was deemed a clinically significant benefit. Specifically:
  • Dislodgement: SafeBreak® Group: 2 (2.9% subjects, 2.5% catheters); Control Group: 4 (5.8% subjects, 4.3% catheters).
  • Infiltration: SafeBreak® Group: 8 (11.4% subjects, 9.9% catheters); Control Group: 10 (14.5% subjects, 10.8% catheters).
  • Occlusion: SafeBreak® Group: 2 (2.9% subjects, 2.5% catheters); Control Group: 1 (1.5% subjects, 1.1% catheters).
  • Phlebitis (VIP Score ≥2): SafeBreak® Group: 1 (1.4% subjects, 1.2% catheters); Control Group: 10 (14.5% subjects, 10.8% catheters).
  • Total MCs: SafeBreak® Group: 13 (18.6% subjects); Control Group: 25 (36.2% subjects). |
    | Separation Force Testing | The device separates at the specified force, 4 ± 1 lbf. (This specification was derived from a characterization study of peripheral IV securement devices to understand when the catheter would move/dislodge/securement device disrupt on swine skin). | Met. Non-clinical performance testing must demonstrate that the device performs as intended, including separation force testing. (The text states this was done and implied conformance without giving a specific value for the product itself, but rather stating the acceptance criteria was derived from this study. Therefore, the device is presumed to have met its own derived specification.) |
    | Validation of Anti-Reconnect Feature | After the separation of the two sub-assemblies, the device cannot be reconnected by hand. | Met. Non-clinical performance testing demonstrated this. |
    | Air Leak Testing | Air leakage on aged non-separated and separated device meets specified acceptance criteria: Non-separated (intact) (b) (4); Patient side (male) (b) (4); Administration set side (female) (b) (4). | Met. Non-clinical performance testing demonstrated this. |
    | Burst Pressure (check valve leakage) per ISO 8536-12 | No signs of air or water leakage when the IV tubing line is under pressure. | Met. Non-clinical performance testing demonstrated this. |
    | Counterflow (check valve leakage) per ISO 8536-12 A.1 | The check valve from the IV tubing line withstands a pressure of 200 kPa in the counter flow direction, while it is subjected to a water excess pressure at 40°C for fifteen (15) minutes. | Met. Non-clinical performance testing demonstrated this. |
    | Duckbill Valve Testing | The valve on the patient side withstands (b) (4) back pressure. | Met. Non-clinical performance testing demonstrated this. |
    | Luer Connection Testing (ISO 80369-7, ISO 80369-20) | Device passes tests for Falling Drop Positive Pressure Liquid Leakage, Sub-atmospheric pressure air leakage, Stress Cracking, Resistance to Separation from Axial load, Resistance to Separation from Unscrewing, Resistance to Overriding. | Met. Non-clinical performance testing demonstrated this. |
    | Pump Occlusion Alarm Testing | The check valve on the IV tubing side (separated) withstands (b) (4) of back pressure at various flow rates and triggers the occlusion alarm. | Met. Non-clinical performance testing demonstrated this. |
    | Flow Inhibition Testing | The device does not impact the flowrate accuracy of commonly used infusion pumps. | Met. Non-clinical performance testing demonstrated this. |
    | Crush Testing | The snap fit fingers are not broken or dislodged when the device is subjected to force of 200 lbf, which simulates a patient lying on the device. | Met. Non-clinical performance testing demonstrated this. |
    | Feature Verification: Transparency | The air-water interface is visually detectable, when the device is partially filled with distilled water, per ISO 8536-9, A.1. | Met. Non-clinical performance testing demonstrated this. |
    | Biocompatibility | Evaluated in accordance with FDA's Guidance Document titled Use of International Standard ISO 10993-1. Prolonged contact duration (>24hr to 30 days). Includes Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Hemocompatibility, Pyrogenicity, Subacute/Subchronic Toxicity, Particulates. | Met. Biocompatibility testing demonstrated that the device is biocompatible for its intended use across all specified endpoints. |
    | Shelf-Life/Sterility (1 year) | Shelf-life established at 1 year, based on testing non-aged devices after simulated shipping, 1-year accelerated aged devices, and 1-year real-time aged devices. Passed dye penetration and burst strength testing. Validated ethylene oxide sterilization to SAL of 10-6 per ISO 11135:2014. Sterilant residuals assessed per ISO 10993-7:2008. Bacterial endotoxin testing per ANSI/AAMI ST72:2011 ( (b)(4) | Met. Non-clinical performance testing demonstrated this. |

2. Sample Size and Data Provenance for the Clinical Test Set

  • Sample Size (Test Set):
    • Intention to Treat (ITT) population: (b)(4) subjects (divided into roughly equal SafeBreak® and Control groups). Specific numbers: (b)(4) SafeBreak® and (b)(4) Control.
    • Per Protocol (PP) population: 139 subjects (70 SafeBreak®, 69 Control).
  • Data Provenance:
    • Country of Origin: Not explicitly stated, but the study was performed in (b)(4), indicating a specific location rather than multiple. The fact it's a De Novo submission to the FDA implies the study was conducted to US regulatory standards, likely in the US or a country with comparable ethical and clinical trial standards.
    • Retrospective or Prospective: Prospective, randomized, controlled trial.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • The clinical study did not use "experts" in the sense of adjudicators for imaging or similar interpretations. Instead, it was a randomized controlled trial where patient outcomes and events were directly measured and recorded by clinical staff.
  • The "ground truth" for clinical endpoints like delays of therapy and mechanical complications was based on the direct observation and documentation of events in the study subjects during the clinical trial.
  • The qualifications of the clinicians involved (e.g., nurses, physicians) are implied by their roles in a clinical trial setting but not explicitly detailed (e.g., "radiologist with 10 years of experience"). However, human factors testing did involve 15 registered nurses (RNs) as participants.

4. Adjudication Method for the Test Set

  • For the clinical trial, there's no mention of an external adjudication committee or specific adjudication method (like 2+1 or 3+1).
  • Events like mechanical complications and delays of therapy were recorded as part of the trial protocol. The data appears to have been collected directly by the study sites and analyzed as part of the trial results.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, an MRMC comparative effectiveness study was not done. This type of study is typically used for diagnostic or screening devices where different readers interpret cases (e.g., radiologists interpreting images).
  • The SafeBreak® Vascular is a mechanical device, and its effectiveness was measured by clinical outcomes (delays of therapy, mechanical complications) rather than interpretive performance.

6. Standalone (Algorithm Only) Performance

  • Yes, standalone performance was extensively done, but for bench testing, not clinical performance.
  • The "Summary of Nonclinical/Bench Studies" section details numerous standalone tests performed on the device itself (e.g., Separation Force Testing, Anti-Reconnect Feature, Air Leak Testing, Luer Connection Testing, Flow Inhibition, Crush Testing, Biocompatibility, Shelf-Life/Sterility, Microbial Ingress, Transparency). These tests evaluate the device's inherent physical and functional characteristics without human interaction in a clinical setting.
  • "Algorithm only" performance typically refers to AI/software, which is not applicable here as SafeBreak® Vascular is a mechanical device.

7. Type of Ground Truth Used

  • For the clinical study (test set): The ground truth was clinical outcomes data obtained directly from the randomized controlled trial. This includes:
    • Direct measurements of "delay of therapy" (minutes per 24 hours per subject).
    • Observed and documented "peripheral IV mechanical complications" (dislodgement, infiltration, occlusion, phlebitis).
    • Observed and documented "SafeBreak® Vascular separations."
    • Reported "Adverse Events."
  • For bench testing: The ground truth was established by engineering specifications, regulatory standards (e.g., ISO, ASTM, USP), and physiological models (e.g., swine skin for separation force characterization).

8. Sample Size for the Training Set

  • Not applicable in the conventional sense for AI/machine learning. The SafeBreak® Vascular is a mechanical device, not typically trained using data sets.
  • Instead of a "training set," the device underwent extensive design, development, and iterative testing to refine its performance characteristics (e.g., separation force). The "Separation Force characterization study" is an example of such a process, where a specification was defined based on understanding peripheral IV securement devices on swine skin. This could be considered analogous to a "training" or "development" phase for the device's physical design and specifications.

9. How the Ground Truth for the Training Set Was Established

  • As mentioned above, for a mechanical device, the concept of a "training set ground truth" is not directly applicable.
  • Ground truth for defining device specifications (which could be likened to "training"):
    • Bench Testing Standards: Established through validated engineering principles and national/international standards (e.g., ISO, ASTM).
    • Clinical Relevance Studies: The "Separation Force characterization study" used a clinically relevant model (swine skin) to understand critical forces on peripheral IVs. This empirical data informed the acceptable force range for the SafeBreak® Vascular's separation. This data acted as the "ground truth" to establish the device's target performance specification.

§ 880.5220 Intravenous catheter force-activated separation device.

(a)
Identification. An intravenous catheter force-activated separation device is placed in-line with an intravenous (IV) catheter and an intravascular administration set, including any administration set accessories. It separates into two parts when a specified force is applied. The device is intended to reduce the risk of IV catheter failure(s) requiring IV catheter replacement.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Performance data must be provided to demonstrate clinically acceptable performance for the intended use of the device.
(2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. The following performance characteristics must be tested:
(i) Separation force testing;
(ii) Validation of anti-reconnect features;
(iii) Air and liquid leakage testing, both before and after separation;
(iv) Luer connection testing;
(v) Flow rate testing;
(vi) Particulate testing; and
(vii) Microbial ingress testing.
(3) The device must be demonstrated to be biocompatible.
(4) Performance testing must demonstrate that the device is sterile and non-pyrogenic.
(5) Performance testing must support the shelf life of the device by demonstrating continued sterility and device functionality over the identified shelf life.
(6) Device labeling must include:
(i) Instructions for use; and
(ii) A discussion of catheter dressings intended to be used with the device.