(491 days)
K/DEN number: Not Found
Not Found
No
The device description and performance studies focus on mechanical and fluid dynamics, with no mention of AI, ML, image processing, or data-driven decision making.
No
The device is described as a surgical wound retraction and protection device that integrates surgical retraction, wound barrier protection, and controlled fluid delivery and removal. Its intended uses are to retract the incision, provide access, irrigate the wound, and aid in preventing wound edge contamination, which are supportive functions for a surgical procedure, not therapeutic in nature.
No
The device description and intended use clearly state that the CleanCision™ System is a surgical retraction and protection device designed to physically retract incisions, provide access, irrigate, and protect the wound edge during abdominal surgery. It does not mention any function related to diagnosing medical conditions.
No
The device description clearly outlines physical components like a flexible, double-walled sheath, rings, and connections for fluid delivery and removal, indicating it is a hardware device with integrated fluid management. The performance studies also focus on physical characteristics and functionality in a surgical setting.
Based on the provided information, this device is not an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use clearly describes a surgical device used during abdominal surgery to retract, protect, and irrigate the wound edge. This is a direct intervention on the patient's body.
- Device Description: The description details a physical device with components for retraction, barrier protection, and fluid management within the surgical site.
- Lack of IVD Characteristics: There is no mention of the device being used to examine specimens (like blood, tissue, or urine) in vitro (outside the body) to provide information for diagnosis, monitoring, or screening.
IVD devices are used to perform tests on samples taken from the body to provide diagnostic or other health-related information. This device is a surgical tool used directly on the patient during a procedure.
N/A
Intended Use / Indications for Use
The CleanCision™ Wound Retraction and Protection System is intended for use by a surgeon during abdominal surgery to: retract the surgical incision, provide access to the abdominal cavity, and irrigate the surgical wound edge. The device may aid in the prevention of wound edge contamination. This device is intended to deliver a sterile irrigant solution and serve as a conduit for fluid removal from the surgical wound edge.
Product codes
PQI
Device Description
The CleanCision™ Wound Retraction and Protection System (CleanCision™ System) is a sterile, single-use surgical wound retraction and protection device that integrates surgical retraction, wound barrier protection, and controlled fluid delivery and removal, as shown in Figure 1, below. The wound barrier protection component is comprised of a flexible, doublewalled sheath with an inner layer that may aid in the prevention of wound edge contamination. The wound retraction component is formed by attaching the sheath to a fixed-diameter ring at the bottom, designed to be inserted into the abdominal cavity, and a radially-adjustable upper retraction ring at the top, designed to remain outside of the body and be actuated to achieve wound retraction. The double-walled sheath also includes integrated fluid delivery and removal. Fluid is delivered via gravitational feed from an external sterile irrigation solution bag into the device and delivered to the wound edges through the permeable outer layer of the sheath. Excess fluid is removed through a separate chamber within the sheath via a connection with the hospital's standard vacuum suction mechanism.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Abdominal cavity / surgical wound edge during abdominal surgery
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Surgeon, during abdominal surgery
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies
Non-clinical/Bench Studies:
- Biocompatibility: Tested for cytotoxicity, sensitization, irritation, acute systemic toxicity, particulate testing, rabbit pyrogen test, Limulus Amebocyte Lysate (LAL) bacterial endotoxin testing, and hemocompatibility. Tests performed on finished devices after sterilization according to ISO 10993 standards. Results showed the device to be non-cytotoxic, non-sensitizing, non-irritating, non-toxic, non-pyrogenic, and non-hemolytic.
- Extractables and Leachables: Chemical characterization studies completed per ISO 10993-18:2005. Extraction performed with polar and non-polar solvents. A toxicological risk assessment was conducted per ISO 10993-17:2002.
- Shelf Life/Sterility: The device is sterile with a sterility assurance level (SAL) of 10-6. Shelf life of two (2) years was validated using accelerated aging, inspection of packaging components, gross leak detection, and seal strength testing.
- Bench Performance Testing: Included actuation force testing (21.1N, Pass). Testing conducted on devices exposed to sterilization, environmental conditioning, simulated distribution, and simulated use. Performance verified after two years of accelerated aging.
Animal Study:
- Study Type: Animal study to demonstrate proper device functioning under simulated use in a porcine large animal model.
- Sample Size: Not explicitly stated, but implies multiple animals ("all animals").
- Key Results: All animals had normal clinical observations. Clinical pathology excursions were mild and clinically insignificant. All test devices were successfully placed and utilized. Fluid flow patency and successful fluid removal were observed. No procedural complications, inflammation, infection, or inadvertent organ aspiration were found. No adverse device-related gross or microscopic lesions were identified. No evidence of hematoma, tissue trapping, denudation, erosion, maceration, adhesions, or bacterial infection. Particulates were not identified in tissues.
Human Factors/Usability Study:
- Study Type: Cadaver study.
- Key Results: Confirmed risk assessments and identified no unforeseen use-related hazardous situations. Device performed at an equivalent level compared to current devices on the market for barrier wound protection and retraction. Surgeons found the force required to deploy the device acceptable (21.1N (Pass)
- Sterility Assurance Level (SAL): 10-6
- Shelf Life: 2 years
Predicate Device(s)
Not Found
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 878.4371 Irrigating wound retractor device.
(a)
Identification. An irrigating wound retractor device is a prescription device intended to be used by a surgeon to retract the surgical incision, to provide access to the surgical wound, to protect and irrigate the surgical wound, and to serve as a conduit for removal of fluid from the surgical wound.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The patient-contacting components of the device must be demonstrated to be biocompatible and evaluated for particulate matter.
(2) Performance data must demonstrate the sterility and pyrogenicity of the patient-contacting components of the device.
(3) Performance data must support shelf life by demonstrating continued functionality and sterility of the device over the identified shelf life.
(4) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Performance testing must:
(i) Characterize the tear resistance, tensile strength, and elongation properties of the barrier material;
(ii) Demonstrate that the liquid barrier material is resistant to penetration by blood, and is non-flammable;
(iii) Characterize the forces required to deploy the device;
(iv) Characterize the device's ranges of operation, including flow rates and maximum suction pressures;
(v) Demonstrate the ability of the device irrigation apparatus to maintain a user defined or preset flow rate to the surgical wound; and
(vi) Demonstrate the ability of the device to maintain user defined or preset removal rates of fluid from the surgical wound.
(5) The labeling must include or state the following information:
(i) Device size or incision length range;
(ii) Method of sterilization;
(iii) Flammability classification;
(iv) Non-pyrogenic;
(v) Shelf life; and
(vi) Maximum flow rate and suction pressure.
0
DE NOVO CLASSIFICATION REQUEST FOR CLEANCISION "11 WOUND RETRACTION AND PROTECTION SYSTEM
REGULATORY INFORMATION
FDA identifies this generic type of device as:
Irrigating Wound Retractor Device: An irrigating wound retractor device is a prescription device intended to be used by a surgeon to retract the surgical incision, to provide access to the surgical wound, to protect and irrigate the surgical wound, and to serve as a conduit for removal of fluid from the surgical wound.
NEW REGULATION NUMBER: 21 CFR 878.4371
CLASSIFICATION: II
PRODUCT CODE: PQI
BACKGROUND
DEVICE NAME: CleanCision™ Wound Retraction and Protection System
SUBMISSION NUMBER: DEN150038
DATE OF DE NOVO: August 13, 2015
Prescient Surgical CONTACT: 1585 Industrial Road San Carlos, CA 94040
INDICATIONS FOR USE
The CleanCision™ Wound Retraction and Protection System is intended for use by a surgeon during abdominal surgery to: retract the surgical incision, provide access to the abdominal cavity, and irrigate the surgical wound edge. The device may aid in the prevention of wound edge contamination. This device is intended to deliver a sterile irrigant solution and serve as a conduit for fluid removal from the surgical wound edge.
LIMITATIONS
The sale, distribution, and use of the device are restricted to prescription use in accordance with 21 CFR §801.109.
1
Contraindications
- Use of the CleanCision™ Wound Retraction and Protection System is ● contraindicated when, in the judgement of the physician, use of this device would be contrary to the best interest of the patient.
Warnings
- . This device was designed and tested for single patient use only. Do not reuse, reprocess, or resterilize this device.
- Avoid contact with potential ignition sources. Do not use in the presence of . flammable anesthesia.
- Device has been tested with a max hospital suction pressure of 370 mmHg. .
PLEASE REFER TO THE LABELING FOR A MORE COMPLETE LIST OF WARNINGS, PRECAUTIONS AND CONTRAINDICATIONS.
DEVICE DESCRIPTION
The CleanCision™ Wound Retraction and Protection System (CleanCision™ System) is a sterile, single-use surgical wound retraction and protection device that integrates surgical retraction, wound barrier protection, and controlled fluid delivery and removal, as shown in Figure 1, below. The wound barrier protection component is comprised of a flexible, doublewalled sheath with an inner layer that may aid in the prevention of wound edge contamination. The wound retraction component is formed by attaching the sheath to a fixed-diameter ring at the bottom, designed to be inserted into the abdominal cavity, and a radially-adjustable upper retraction ring at the top, designed to remain outside of the body and be actuated to achieve wound retraction. The double-walled sheath also includes integrated fluid delivery and removal. Fluid is delivered via gravitational feed from an external sterile irrigation solution bag into the device and delivered to the wound edges through the permeable outer layer of the sheath. Excess fluid is removed through a separate chamber within the sheath via a connection with the hospital's standard vacuum suction mechanism.
Image /page/1/Figure/9 description: The image shows a device with various labeled parts. The labels include numbers 1 through 16, which point to different components of the device. Some of the labeled parts include a rim (1), a base (2), a handle (3), and a sensor unit (16). The device appears to be a type of container or apparatus with additional sensors and attachments.
Figure 1. CleanCision™ Wound Retraction and Protection System
2
Image /page/2/Picture/0 description: The image is a gray rectangle. The rectangle takes up most of the image. In the upper right corner of the image, there is the text "(b) (4)".
Table 1. - CleanCision™ Wound Retraction and Protection System (refer to Figure 1)
SUMMARY OF NONCLINICAL/BENCH STUDIES
The sponsor conducted a series of non-clinical performance testing to demonstrate that the CleanCision™ System would perform as anticipated. Non-clinical testing included: biocompatibility, shelf-life, sterility, package integrity, bench, animal, and usability performance testing.
BIOCOMPATIBILITY/MATERIALS
A. Biocompatibility
The CleanCision™ System is classified as an externally communicating, blood path indirect contact, limited exposure (≤ 24 hours) device. Biocompatibility testing was performed according to ISO 10993-1:2009, "Biological evaluation of medical devices-Part 1: Evaluation and testing within a risk management process." Testing (b) (4) sterilization. As was completed on finished devices, summarized in Table 2, the CleanCision™ System was found to be non-cytotoxic, non-sensitizing, non-toxic, non-irritating, non-pyrogenic and non-hemolytic, mitigating the risk of adverse tissue reaction.
Test | Purpose | Methods | Results |
---|---|---|---|
Cytotoxicity | Determine the potential | ||
biological reactivity of a | |||
mammalian cell culture | |||
(L929) in response to | |||
the test article extract. | ISO 10993-5 – | ||
Biological evaluation | |||
of medical devices – | |||
Part 5: Tests for in | |||
vitro cytotoxicity | Non-cytotoxic | ||
Sensitization | Determine the | ||
allergenic potential or | |||
sensitizing capacity of | |||
the test article after | |||
extraction with a polar | |||
and non-polar solvent. | ISO 10993-10 – | ||
Biological evaluation | |||
of medical devices – | |||
Part 10: Tests for | |||
irritation and skin | |||
sensitization | Non-sensitizer |
Table 2 – Summary of Biocompatibility Testing
3
Test | Purpose | Methods | Results |
---|---|---|---|
Irritation | Determine the potential | ||
irritation effects of the | |||
test article extraction | |||
with a polar and non-polar solvent. | ISO 10993-10 – | ||
Biological evaluation | |||
of medical devices – | |||
Part 10: Tests for | |||
irritation and skin | |||
sensitization | Non - irritant | ||
Acute Systemic | |||
Toxicity | Determine the potential | ||
toxic effects of the test | |||
article extract as a result | |||
of a single-dose | |||
systemic injection in | |||
mice. | ISO 10993-11 – | ||
Biological evaluation | |||
of medical devices – | |||
Part 11: Tests for | |||
systemic toxicity | No acute systemic | ||
toxicity response | |||
was observed from | |||
test article extract. | |||
Particulate Testing | To determine the | ||
number of particulates | |||
(b) (4) | |||
and the | |||
concentration of | |||
particulates from the | |||
test article from water | |||
injection. | USP | ||
Particulate Matters in | |||
Injections | Results met USP | ||
Limits of ≤ 25 | |||
particles per mL for | |||
(b) (4) | |||
particles per mL for | |||
(b) (4) µm for nominal | |||
volumes of | |||
(4) | |||
Rabbit Pyrogen Test | |||
(Material Mediated) | Determine the presence | ||
of chemical pyrogens in | |||
test article extracts. | ISO 10993-11 – | ||
Biological evaluation | |||
of medical devices – | |||
Part 11: Tests for | |||
systemic toxicity | Non-pyrogenic | ||
Limulus Amebocyte | |||
Lysate (LAL) | |||
Bacterial Endotoxin | |||
Testing | In vitro assay for | ||
detection and | |||
quantitation of bacterial | |||
endotoxin in test article | |||
extract. | USP Bacterial | ||
Endotoxin Test | Non-pyrogenic | ||
Hemocompatibility | |||
(Hemolysis) | Determine the potential | ||
hemolytic activity on | |||
rabbit blood in response | |||
to the test article and to | |||
its extract. | ISO 10993-4 – | ||
Biological evaluation | |||
of medical devices – | |||
Part 4: Selection of | |||
tests for interaction | |||
with blood. | Non - hemolytic |
B. Extractables and Leachables
Chemical characterization studies were completed per ISO 10993-18: 2005 "Biological evaluation of medical devices -- Part 18: Chemical characterization of materials" to determine the amounts of extracted inorganic and polymer-related organic substances. The extractions of the CleanCision™ System were performed at & with both polar and non-polar extraction solvents. The chosen conditions represented aggressive worst-case conditions. Results indicate that a total of six chemical compounds from the polar solvent were characterized above the limit of detection. No chemical compounds were observed above the limit of detection in the nonpolar solvent. Subsequently, a toxicological risk assessment was conducted per ISO 10993-17:2002 "Biological evaluation of medical devices -- Part 17: Methods for the establishment of allowable
4
for leachable substances." The margin of safety for all six compounds was Therefore, based on the results of these chemical characterizations, the toxicological risk assessment concluded that the use of the CleanCision™ System would not be expected to result in exposure to chemicals during clinical use at levels that would result in an adverse biological response in patients.
C. Conclusions
In summary, the CleanCision™ System was evaluated to determine the potential for toxicity resulting from contact of the device component materials with the body. The results of this testing demonstrate that the CleanCision™ System is biocompatible when used as intended.
SHELF LIFE/STERILITY
The CleanCision™ System is packaged in a Tyvek / Polyethylene-Nylon pouch and is (b)(4). The validation of the sterilization process sterilized using complies with -- Requirements for the development, validation and routine control of a sterilization process for medical devices." The sterilization method achieves a sterility assurance level (SAL) of 10-6.
The identified shelf life of two (2) years was validated using (b) (4)
(b) (4)
(b) (4)inspection of the packaging components and product labels, gross leak detection (bubble) testing for assessing package integrity, and seal strength (peel) testing for evaluating the pouch seal strength. In addition, microbial ranking testing for the pouch was conducted in accordance with ASTM F1608-00 "Standard Test Method for Microbial Ranking of Porous Packaging Materials."
5
PERFORMANCE TESTING - BENCH
Performance testing included actuation force testing, resistance to synthetic blood, tear and tensile strength, elongation characterization, flammability, fluid delivery and removal flow rate testing, and suction pressure testing. Testing was conducted on devices which had been exposed to sterilization, environmental conditioning, simulated distribution and simulated use, where applicable. Test results confirmed the CleanCision™ System met all device requirements. The device requirements are appropriate for the device's indications for use. Device performance was also verified after two years of accelerated aging by performing the following tests (Table 3) at zero and 24 months of accelerated aging. The tested samples met all requirements supporting a labeled shelf life of 2 years.
Test | Purpose | Method | Acceptance Criteria | Results |
---|---|---|---|---|
Retraction Ring | ||||
Actuation Force | Determine that the | |||
forces required to deploy | ||||
the device clinically do | ||||
not lead to device | ||||
failures. | 15 retraction | |||
assemblies placed in | ||||
a simulated wound | ||||
model; measured | ||||
force required to | ||||
actuate retraction ring | 21.1N | |||
21.1 N is the force required to remove the bag spike from a fluid bag and the tubing connection should be stronger so the bag spike would disconnect from the bag before the tubing would pull out of the pliable membrane | Pass |
1 NFPA = National Fire Protection Association
PERFORMANCE TESTING - ANIMAL
An animal study was conducted to demonstrate proper device functioning under simulated use. The purpose was to evaluate the performance and safety of the CleanCision™ System in a simulated animal study under defined worst case conditions. A 6 hour abdominal surgical procedure simulated use of the device, and evaluated the local, regional, and systemic effects of the device in a porcine large animal model. Baseline and terminal CBC and Serum Chemistry Panel samples were collected and analyzed; terminal peritoneal fluid samples were collected and analyzed; and a complete necropsy was performed, including a gross assessment and procurement of appropriate tissue specimens for histological evaluation.
From the results provided, all animals had normal clinical observations at study enrollment and a normal baseline physical examination. All clinical pathology excursions were mild and clinically insignificant, none of which were believed to be test or control article related. In addition, all test devices were successfully placed and utilized as laid out in the study protocol and each device was successfully evaluated for all the device functions and parameters listed in the study protocol. All devices showed fluid flow patency throughout the procedure and operated as intended. Fluid that was run through the device was successfully removed through the suction port throughout the procedure. There were no procedural complications related to the use of the test device during the procedure. The study surgeon was successfully able to insert, remove and reinsert the device without any complication for the duration of the procedure. There was no evidence of inflammation,
7
infection or inadvertent organ aspiration from the cytology analysis. No important adverse device-related gross or microscopic lesions were identified in the surgical wound margins, abdominal tissues in contact with the device, or in more distant systemic organs. There was no evidence of hematoma at the surgical site, tissue trapping under the device, denudation of the peritoneum or any organ serosa. tissue erosion or maceration, adhesions, or evidence of bacterial infection. At least 14 tissues with serosa in contact with the device were examined and particulates were not identified using bright field and polarized light.
The results of this study, in combination with the bench testing described above and the usability testing described below, established the usability of this device and demonstrated that the benefits of this device outweigh the risks.
HUMAN FACTORS/ USABILITY
The Human Factors / Usability Validation Testing was performed to confirm the risk assessments and to identify any unforeseen use-related hazardous situations with this device. The usability validation testing consisted of a cadaver study to determine whether the CleanCision™ System is successfully able to meet surgeon-user needs with respect to its usability and functionality throughout the entirety of its intended use, and by extension determining whether the technical usability of the device is acceptable for clinical use in patients.
CleanCision™ System performed at an equivalent level compared to current devices on the market for barrier wound protection and retraction. Surgeons found that the force required to deploy the device was acceptable, which was less than 10N. In addition, the effectiveness of the fluid delivery and retrieval mechanism was rated uniformly high by the surgeon-user. The feedback specifically indicated that the device should prevent the majority of fluid from leaking into the abdominal cavity. All users were able to successfully insert, utilize, and remove the device as expected. No unacceptable risks were identified during the usability testing.
The results of this study, in combination with the bench testing and animal study described above, established the usability of this device and demonstrated that the benefits of this device outweigh the risks.
PEDIATRIC EXTRAPOLATION
In this de novo request, existing data were not leveraged to support the use of the device in a pediatric patient population.
LABELING
Labeling for the CleanCision™ System includes Instructions for Use, which includes the intended use, product description, contraindications, warnings and precautions (contraindications and warnings are identified above, under Limitations).
8
The labeling provided is adequate and includes appropriate information regarding specifications, instructions for the surgeon on proper use and removal, as well as an appropriate prescription statement as required by 21 CFR 801.109.
RISKS TO HEALTH
Table 4 below identifies the risks to health that may be associated with use of Irrigating Wound Retractor Device and the measures necessary to mitigate these risks.
Identified Risk | Mitigation Measure |
---|---|
Adverse Tissue Reaction | Biocompatibility Evaluation |
Tissue or Wound Damage | Non-clinical Performance Testing |
Shelf Life Testing | |
Labeling | |
Infection | Sterilization Validation |
Non-clinical Performance Testing | |
Shelf Life Testing | |
Labeling |
Table 4. - Identified Risks to Health and Mitigation Measures
SPECIAL CONTROLS
In combination with the general controls of the FD&C Act, the Irrigating Wound Retractor Device is subject to the following special controls:
- The patient-contacting components of the device must be demonstrated to be 1. biocompatible and evaluated for particulate matter.
- Performance data must demonstrate the sterility and pyrogenicity of the patient-2. contacting components of the device.
-
- Performance data must support shelf life by demonstrating continued functionality and sterility of the device over the identified shelf life.
-
- Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Performance testing must:
- Characterize the tear resistance, tensile strength, and elongation properties of the a. barrier material:
- b. Demonstrate that the liquid barrier material is resistant to penetration by blood, and is non-flammable:
- Characterize the forces required to deploy the device; C.
- Characterize the device's ranges of operation, including flow rates and maximum d. suction pressures:
- Demonstrate the ability of the device irrigation apparatus to maintain a user defined e. or pre-set flow rate to the surgical wound;
- f. Demonstrate the ability of the device to maintain user defined or pre-set removal rates of fluid from the surgical wound.
9
-
- The labeling must include or state the following information:
- a. Device size or incision length range;
- b. Method of sterilization:
- c. Flammability classification;
- d. Non-pyrogenic:
- e. Shelf life;
- f. Maximum flow rate and suction pressure.
BENEFIT/RISK DETERMINATION
The risks of the device are based on biocompatibility studies and non-clinical performance testing. The risks include adverse tissue reaction, tissue or wound damage and infection. There are alternative devices in the market for wound retraction and wound protection. However, the current device is unique in design because it includes a system for delivery and removal of fluid to and from the wound edge.
The probable benefits of the device are also based on non-clinical laboratory, animal studies, and usability testing. The probable benefits for the CleanCision™ Wound Retraction and Protection System include wound retraction and barrier protection to the edge of the surgical incision.
Patient Perspectives
This submission did not include specific information on patient perspectives for this device.
Benefit/Risk Conclusion
In conclusion, given the available information above, the data support that for use of the device by a surgeon during abdominal surgery to retract the surgical incision, provide access to the abdominal cavity, and irrigate the surgical wound edge, the probable benefits outweigh the identified risks to health for the CleanCision™ Wound Retraction and Protection System. The device provides benefits and the risks can be mitigated by the use of general and the identified special controls.
CONCLUSION
The de novo request for the CleanCision™ Wound Retraction System is granted and the device is classified under the following:
Product Code: PQI Device Type: Irrigating Wound Retractor Device Class: II Regulation: 21 CFR 878.4371